K Number
K972535
Device Name
MAGLIFE C
Manufacturer
Date Cleared
1998-01-13

(189 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N Q), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.

Device Description

MAGLIFE C is a multi-parameter patient monitor which has been designed for use in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations. The parameters that can be monitored are electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N D), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO).

The monitor consists of a line-powered console (with optional back-up battery) which is placed outside the magnet bore, and appropriate electrodes, transducers, cables and tubes to allow the patient to be monitored during the examination. The console enclosure is a Faraday cage for the protection of sensitive electronic circuits.

AI/ML Overview

The provided text does not contain detailed acceptance criteria and a study proving the device meets those criteria. Instead, it is a 510(k) summary for the Maglife C patient monitor, primarily focusing on demonstrating substantial equivalence to predicate devices based on non-clinical testing and compliance with various standards.

Therefore, I cannot provide the requested information in the format specified.

However, I can extract the available information regarding testing:

Key Takeaways from the document:

  • Non-clinical tests were conducted to demonstrate compliance with a list of standards (ANSI/AAMI EC13-1992, EC12-1991, SP10, SP9, BP22-1994, BP23-1986, ES21-1993, and IEC 601-1).
  • The device was evaluated for electrical testing based on the November 1993 Reviewer Guidance for Premarket Notification Submissions (Anesthesiology and Respiratory Devices Branch).
  • Tests in an MRI unit were performed to:
    • Determine the maximum field strength the Maglife C can withstand without performance degradation.
    • Verify sufficiently small ferrous material content to prevent movement due to magnetic attraction.
    • Assess the effect of Maglife C on the homogeneity of the magnetic field (with a 1.5 Tesla MR unit, showing it could be placed as close as 60 cm from the magnet without visible image differences).
  • Conclusion from non-clinical testing: "The testing of the Maglife C demonstrates that the performance is substantially equivalent to predicate devices cited above."

Since the document focuses on non-clinical testing and substantial equivalence, and does not describe clinical studies or specific quantitative acceptance criteria for device performance with corresponding results, it's impossible to fill out the table and answer most of the questions as requested.

{0}------------------------------------------------

K912335

JAN 1 3 1998

APPENDIX B

Revised 510(k) Summary

Additional Information - K972535 O.D.A.M. Maglife C

.

... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10/13/97

{1}------------------------------------------------

510(k) Summary for Maglife C

  • Date this summary was prepared: October 13, 1997 1.
  • Submitter's Name and Address 2.

O.D.A.M. 19, avenue de la gare 67162 Wissembourg cedex, France

  • Contact Person 3.
    Mr. Court Goehry Groupe BRUKER O.D.A.M. 19, avenue de la gare 67162 Wissembourg cedex, France

Telephone 011 33 3 88 63 36 00 011 33 3 88 54 36 32 Telefax cg@bruker.fr E-mail

  • Device Name 4.
    Maglife C Trade/Proprietary Name:

MR Safe Patient Monitor Common Name:

patient physiological monitor Classification Names:

  • 、 ડ. Predicate Devices
    The legally marketed devices to which equivalence is being claimed are:

  • MAGLIFE, marketed by O.D.A.M. (K950264) .

{2}------------------------------------------------

  • MRI-Compatible Model 9500 manufactured by Magnetic Resonance Corp. (K954120)
  • Omni Trak 3100 MRI manufactured by In Vivo Research (K864889) .
  • Device Description 6.

MAGLIFE C is a multi-parameter patient monitor which has been designed for use in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations. The parameters that can be monitored are electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N D), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO).

The monitor consists of a line-powered console (with optional back-up battery) which is placed outside the magnet bore, and appropriate electrodes, transducers, cables and tubes to allow the patient to be monitored during the examination. The console enclosure is a Faraday cage for the protection of sensitive electronic circuits.

    1. Intended Use
      MAGLIFE C is intended for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N,O), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) in the immediate vicinity of Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.

{3}------------------------------------------------

  • Comparison of Technological Characteristics 8.
    All four MRI compatible patient monitors employ established measurement methods combined with non-metallic sensors and fiber optic signal transmission to minimize the risk of burns and special shielding and filtering to minimize interference to and from the MRI unit.

  • Nonclinical Tests Used in Determination of Substantial Equivalence 9.
    The design of the Maglife C has been thoroughly validated at the unit and system level. Non-clinical tests were conducted to demonstrate compliance with the following standards:

  • Cardiac Monitors, Heart Rate Meters, and Alarms. . ANSI/AAMI EC13 - 1992

  • American National Standard for Pregelled ECG Disposable . Electrodes. ANSI/AAMI EC12 - 1991

  • Electronic or Automated Sphygmomanometers. . ANSI/AAMI SP10

  • . Nonautomated Sphygmomanometers. ANSI/AAMI SP9

  • . Blood Pressure Transducers. ANSI/AAMI BP22 - 1994

  • . American National Standard for Interchangeability for Resistive Bridge Type Blood Pressure Transducers. ANSI/AAMI BP23 - 1986

  • . Safe Current Limits for Electromedical Apparatus ANSI/AAMI ES21 1993

  • . IEC 601-1

{4}------------------------------------------------

Maglife C has been evaluated with respect to the recommendations for electrical testing included in the November 1993 Reviewer Guidance for Premarket Notification Submissions (Anesthesiology and Respiratory Devices Branch).

Tests were conducted in an MRI unit to determine the maximum field strength that Maglife C can withstand without degradation of performance and to verify that the amount of ferrous material in the Maglife C is sufficiently small so that movement of the unit due to magnetic attraction is not possible.

Tests were conducted to assess the effect of the Maglife C on the Homogeneity of the magnetic field inside the magnet bore. These tests, using a 1.5 Tesla MR unit, showed that the Maglife could be placed as close as 60 cm from the front of the opening from the magnet without causing any visible difference in the images.

  • Conclusions From Nonclinical Testing 10.
    The testing of the Maglife C demonstrates that the performance is substantially equivalent to predicate devices cited above.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 3 1998

0.D.A.M. c/o Mr. James R. Veale Medical Device Consultants, Inc. 49 Plain Street 92760 North Attleboro, MA

Re: K972535 Maglife C Regulatory Class: II (two) Product Code: 74 MHX Dated: October 13, 1997 Received: October 15, 1997

Dear Mr. Veale:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good ........ manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special. Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{6}------------------------------------------------

Page 2 - Mr. James R. Veale

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your dev`ice to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callahan, Ph.D.

Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

(972535 ... 510(k) Number (if known):

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N Q), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

in. Png?

Division of Cardiovascular, Respiratory, and Neurological Device 510k) Nur

Prescription Use
(Per 21 CFR 801.109) ✓

Over-The-Counter Use

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.