K Number
K972462

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-09-22

(83 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This in vitro diagnostic assay is intended to quantitatively measure a-fetoprotein (AFP) in human serum on the Bayer Immuno 1TM system as an aid in the management of nonseminomatous testicular cancer. AFP values obtained using the Bayer Immuno 1™ AFP assay method must be interpreted in conjunction with all other available clinical and laboratory data before a medical decision is determined.

Device Description

The Bayer Immuno ITM AFP Assay utilizes a well-established immunoassay technology in which one monoclonal AFP antibody is conjugated to fluorescein (designated Reagent 1, or R1) and a second monoclonal AFP antibody is conjugated to alkaline phosphatase (Reagent 2, or R2). The R1 and R2 conjugates are reacted with patient sample, calibrator, or control and are incubated at 37°C on the system. An Immuno 1 Magnetic Particle coated with an anti-fluorescein antibody is then added and a second incubation occurs during which the antibody complex is bound. The magnetic particles complexed with the immunological sandwich are then washed to separate unbound molecules, and a colorimetric substrate is added. The rate of conversion of substrate to a compound with absorbance at 405 and 450 nm is measured and the measured rate is proportional to the concentration of AFP antigen in the sample. A cubic-through-zero curve fitting algorithm is used to generate standard curves.

AI/ML Overview

Here's an analysis of the provided text regarding the Bayer Immuno 1™ AFP Assay, structured according to your requirements:


Bayer Immuno 1™ AFP Assay Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

The document primarily focuses on demonstrating substantial equivalence to a predicate device (Abbott IMx AFP Assay) rather than pre-defined acceptance criteria with specific thresholds for all metrics. However, performance metrics are presented in comparison to the predicate and generally found to be "excellent" or "comparable."

Metric / Acceptance Criteria (Implied)Reported Device Performance (Bayer Immuno 1™ AFP Assay)
Nonclinical Performance
Specificity (Interference)Endogenous Interferents: - Triglycerides (900 mg/dL): 99.9% Recovery - IgG (0.053 g/mL): 98.2% Recovery - Hemoglobin (10.0 mg/mL): 102.1% Recovery - Heparin (0.15 mg/mL): 101.5% Recovery - Bilirubin (0.25 mg/mL): 99.7% Recovery - Albumin (0.065 g/mL): 104.0% Recovery - CEA (10,000 ng/mL): 99.9% Recovery Conclusion: None demonstrated significant interfering effects.
Exogenous Interferents (Chemotherapeutic drugs, OTC drugs, Vitamins): - All tested showed % Recovery ranging from 95.2% to 104.8%. Conclusion: None demonstrated significant interfering effects.
Heterophilic Antibodies (Rheumatoid Factor, HAMA): - RF (high titers): 96.4-103.2% Recovery - HAMA (high titers): 93.5-102.4% Recovery Conclusion: Lack of significant heterophilic interference.
Minimum Detectable Concentration (MDC)Range: 0.04 ng/mL to 0.09 ng/mL across three sites. Claim: 0.1 ng/mL. (Deemed acceptable for an assay of this type).
Imprecision (Total CV%)Not greater than 3.6% for all tested concentrations (e.g., Calibrator 2 mean 5.0 ng/mL: Total CV 2.5%; Calibrator 6 mean 398.2 ng/mL: Total CV 2.6%). Conclusion: Highly reproducible.
Linearity (Recovery%)95.7% to 102.8% of expected value over the entire calibration range (0-400 ng/mL). Conclusion: Clearly demonstrates linearity.
Hook EffectNo "hook effect" observed until a concentration of 204,268 ng/mL AFP was reached. Conclusion: Lack of hook effect at AFP assay values ≤ 160,000 ng/mL.
Spiked Recovery (%)96.7% to 99.3% in five patient samples spiked with ~300 ng/mL AFP. Conclusion: Accurate quantitation of spiked and recovered AFP values.
Parallelism (Recovery%)92.7% to 111.0% in diluted patient samples. Linear regression showed no deviation from linearity. Conclusion: Further illustrates linearity and acceptability of 0 ng/mL Calibrator as diluent.
Clinical Performance (Method Comparison with Abbott IMx)
Normal Range Cut-off (97.5th percentile)Immuno 1: 8.9 ng/mL (based on male subjects from DOCRO). Abbott IMx: 8.1 ng/mL (based on male subjects from DOCRO). Conclusion: Similar to predicate device and manufacturer's suggested range (<8.9 ng/ml for 99% of healthy subjects).
Regression Analysis (r, Slope, Intercept)DOCRO (N=545): - 0-400 ng/mL: r=0.993, Slope=1.047 (OLS); r=0.993, Slope=1.109 (Passing-Bablok) - >400 ng/mL: r=0.992, Slope=1.046 (OLS); r=0.992, Slope=0.985 (Passing-Bablok) MSKCC (N=318): - 0-400 ng/mL: r=0.997, Slope=0.931 (OLS); r=0.997, Slope=1.024 (Passing-Bablok) - >400 ng/mL: r=0.999, Slope=0.939 (OLS); r=0.999, Slope=0.931 (Passing-Bablok) Conclusion: Excellent comparison with slopes near 1.0 and r > 0.99.
Clinical Sensitivity (Nonseminomatous Testicular Ca.)Immuno 1: 64.0% Abbott IMx: 64.0% Conclusion: Comparable to predicate and previous studies (50-75% for nonseminomatous and mixed GCT).
Clinical Sensitivity (Mixed Germ Cell Tumor)Immuno 1: 56.5% Abbott IMx: 54.3% Conclusion: Comparable.
Clinical Specificity (Healthy Individuals)Immuno 1: 98.3% Abbott IMx: 98.6% Conclusion: Consistent with predicate and normal population.
Clinical Specificity (Other Benign Diseases)Immuno 1 & Abbott IMx: 88%-100% (for cirrhosis, hepatitis, benign GU, other non-malignant diseases). Conclusion: Consistent with predicate and literature.
Longitudinal Monitoring (Patient Course Tracking)Immuno 1 AFP Assay demonstrated clinical utility for 78.8% (41/52) of all longitudinal patients (testicular and hepatocellular cancer) and 76.6% (36/47) of testicular cancer patients. Conclusion: Identical trends observed with both Immuno 1 and IMx, demonstrating equivalent clinical utility.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Total Samples for Method Comparison (Cross-sectional): 863 serum samples (545 from DOCRO + 318 from MSKCC).
    • DOCRO (Diagnostic Oncology CRO, Stratford, CT): 545 samples from Western States Plasma Company (Fallbrook, CA).
      • Clinical Classifications: Normal (male/female), various testicular cancers, hepatocellular carcinoma, genito-urinary cancer, gastrointestinal/lung/pancreatic cancer, hepatitis, cirrhosis, benign genito-urinary diseases, other non-malignant diseases.
    • MSKCC (Memorial Sloan-Kettering Cancer Center, New York, NY): 318 samples from MSKCC's own in-house specimen collection.
      • Clinical Classifications: Normal, various testicular cancers, prostate cancer, liver cancer, colon cancer, breast cancer, lung cancer, benign GU disease, and samples from longitudinal patients (counted in the 318 for cross-sectional and separately for longitudinal).
  • Longitudinal Monitoring Test Set:
    • 314 longitudinal samples from 56 patients (51 testicular cancer, 5 hepatocellular cancer).
    • All these samples were obtained from MSKCC's in-house specimen collection.
  • Normal Range Determination: 350 healthy subjects (100 males from MSKCC, 150 males from DOCRO, 100 females from DOCRO).
  • Provenance: Data was retrospective, collected from two U.S. clinical trial sites (Diagnostic Oncology CRO and Memorial Sloan-Kettering Cancer Center) and a plasma company (Western States Plasma Company, CA).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth for the test set beyond mentioning that "physician evaluations" were guides. The ground truth for longitudinal studies relied on "medical, diagnostic, and course of therapy information" and "diagnostic testing, therapeutic interventions, and physician evaluations." The evaluation of the Immuno 1 AFP Assay's correspondence with the clinical course was performed by "Bayer scientists." No specific qualifications (e.g., years of experience as a radiologist) are provided for these "Bayer scientists" or the "physicians" whose evaluations were used.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method for discrepancies in the ground truth for the clinical test set. For the longitudinal studies, the Immuno 1 AFP results were "graphically compared to the patient's course of disease using diagnostic testing, therapeutic interventions, and physician evaluations as guides to describe the clinical picture." The judgment of correspondence was made by "Bayer scientists." This implies a single-party evaluation of the clinical course against the assay results, rather than a multi-expert adjudication method like 2+1 or 3+1.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This study is an in vitro diagnostic assay comparing a new device (Bayer Immuno 1™ AFP Assay) to a predicate device (Abbott IMx AFP Assay) directly, not assessing how human readers (e.g., radiologists, pathologists) improve with or without AI assistance. The "AI" in this context refers to the automated immunoassay system, not an independent AI algorithm assisting human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, the study presents the standalone performance of the Bayer Immuno 1™ AFP Assay. The results (e.g., AFP values, sensitivity, specificity, correlation, longitudinal trends) are generated directly by the Immuno 1 system and evaluated against clinical classifications or compared to the predicate device's standalone results. There is no human-in-the-loop component being evaluated or improved upon.

7. The Type of Ground Truth Used

The ground truth for the clinical studies relied on a combination of:

  • Clinical Classifications: Based on patient diagnoses (e.g., "Normal," "Testicular Cancer," "Hepatocellular Carcinoma," "Hepatitis," "Cirrhosis").
  • Absence/Presence of Disease: For sensitivity/specificity calculations.
  • Clinical Course Information: For longitudinal studies, this included "documented medical, diagnostic, and course of therapy information," as well as "diagnostic testing, therapeutic interventions, and physician evaluations." Essentially, expert clinical judgment and patient outcomes data.

The ground truth for the non-clinical studies (e.g., specificity, imprecision, linearity) was based on known spiked concentrations or established reference materials/methods.

8. The Sample Size for the Training Set

The document does not explicitly state a separate training set size beyond implying that the Immuno 1 system and its assay were developed and optimized internally by Bayer. It describes the characterization of antigens and antibodies and the development of the assay method (e.g., cubic-through-zero curve fitting algorithm). However, there is no mention of a distinct "training set" of patient data used for machine learning or algorithm development in the way one might find for a modern AI device. The study primarily focuses on validation against the predicate and clinical performance.

9. How the Ground Truth for the Training Set Was Established

Since a distinct "training set" for an AI algorithm is not described, the concept of ground truth establishment for such a set is not applicable here. The ground truth for assay development and calibration (implicitly the "training" phase for the assay itself) would have been established using known concentrations of AFP calibrators, controls, and characterized biological samples to define the standard curve and ensure assay performance characteristics like linearity and imprecision.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS K972462 For the Bayer Immuno 1™ AFP Assay

SEP 22 1997

This premarket notification is to add the quantitative measurement of AFP in human serum to the intended use of the Bayer Immuno 1TM Immunoassay System. The performance of the Bayer Immuno 1™ AFP Assay has been established by comparison to a predicate device, the Abbott IMx AFP Assay, in accordance with Section VI. (A) of the "Guidance Document For Submission of Tumor Associated Antigen Premarket Notifications, 510(k), to the FDA." Clinical evaluations of the Bayer Immuno 17M AFP Assay at two US clinical trial sites demonstrated clinical safety and effectiveness and substantial equivalence to the predicate device in accordance with Section VI.(B) of the Guidance Document. The information presented in this Summary of Safety And Effectiveness was derived from nonclinical performance and clinical evaluation studies comparing the performance of the Immuno 1 AFP Assay with that of the FDA-approved Abbott IMx AFP Assay. Clinical studies were conducted at two clinical sites with a suitable sampling of patients to support the diagnostic claims for this device.

INDICATIONS FOR USE

The Bayer Immuno 1™ AFP Assay is an in vitro diagnostic assay intended to quantitatively measure AFP in human serum on the Bayer Immuno 1™ System as an aid in the management of nonseminomatous testicular cancer. AFP values obtained using the Bayer Immuno 1™ AFP Assay must be interpreted in conjunction with all other available clinical and laboratory data before a medical decision is determined. The assay is designed to run on the Bayer Immuno 1™ Immunoassay System, a fully automated random-access analyzer which performs both homogeneous and heterogeneous immunoassays.

Background

Alpha-fetoprotein (AFP), a major serum protein of the developing fetus, is a General. monomeric, heterogeneous, glycoprotein of approximately 70 kilodaltons (kDa) (1). The molecular heterogeneity of AFP, which can be observed either electrophoretically or chromatographically, is thought to be due to either differences in peptide sequence and/or glycosylation among the various isoforms detectable in serum (2-4). AFP consists of a single polypeptide chain and is 4.3% carbohydrate (1). Produced by the fetal yolk sac, liver, and gastrointestinal tract (5), serum levels of AFP peak at about 14 weeks gestation and gradually decline until birth. The AFP concentration is approximately 50 ug/mL at birth and is present in a normal adult at a low basal range of 0.5 to 10 ng/mL (6). AFP is thought to function as a natural ligand for fatty acids and estrogens, and possesses physiochemical properties and an amino acid sequence which are very similar to albumin (reviewed in 4). AFP from different species all reversibly bind fatty acids, particularly polyunsaturated fatty acids. It has been

{1}------------------------------------------------

postulated that the transport and cell delivery of these acids could be the major biological role of AFP (7).

AFP Serum Levels in Testicular Cancer. Elevated AFP levels are seen in patients with nonseminomatous testicular cancer. More than 95% of testicular cancers belong to a heterogeneous group called germ-cell turnors because it is widely believed that they arise in primordial germ cells (8). Germ cell turnors (GCTs) are classified either as seminomatous or as nonseminomatous (NSGCTs). The latter can be further classified as embryonal carcinoma, teratoma, or choriocarcinoma. The seminoma histologic subtype can be found in 40% of all GCTs while the nonseminoma histologic subtype can be found in 60% of GCTs (9). The different histologic types of germ-cell tumors may occur singly or in various combinations. Elevated AFP levels have been observed in patients diagnosed as having seminomatous testicular cancer with nonseminomatous elements, but not in patients with pure seminoma (10-ાર).

Both AFP and HCG are measured in testicular cancer. Approximately 40% of patients with NGSCTs have elevation of only one marker (16). During the clinical course of the disease, the levels of the two markers do not always parallel each other.

A direct relationship has been observed between the incidence of elevated AFP levels in nonseminomatous testicular cancer, and the stage of disease (10-12). Elevation of AFP (>10 IU/L or 12.1 ng/mL) occurs in 80% of metastatic and in 57% of stage 1 nonseminomatous germ cell tumors (16). In Clinical Stage 2B or higher, AFP and/or HCG are elevated in 65-80% of the cases with increasing frequency according to the bulk of the disease (18).

The usefulness of AFP measurements in the management of nonseminomatous testicular cancer patients undergoing cancer therapy has been well established (10, 12, 19). Current management of the testicular germ cell tumors (GCTs) relies upon the use of serum turnor markers which can indicate the presence of small foci of active tumor that cannot be detected by currently available imaging techniques (16). Markers augment and complement information obtained from radiographic and other staging procedures (20). Also, the short half-lives of tumor markers facilitate their use in assessing turnor burden during therapy. AFP has a serum half-life of 3.5-6 days (21). AFP and/or HCG levels are elevated before orchiectomy in about 60% of all Clinical Stage I patients but follow a normal decline after the testicle is removed (18). It is not unusual to have AFP levels at 1000 ng/mL at orchiectomy, which means 5-6 weeks before normalization according to expected half-life values.

For patients in clinical remission following treatment, AFP levels generally decrease (12). Post-operative AFP levels which fail to return to normal strongly suggest the presence of residual tumor (10, 12, 22). Following successful resection of primary or metastatic disease, AFP and HCG decline at a rate proportional to their respective half-lives (21). An elevated actual half-life (AHL) of serum markers following orchiectomy or retroperitoneal lymph node dissection (RPLND) may indicate the presence of occult, persistent disease (20). Residual disease is the rule if such a decline does not ensue.

{2}------------------------------------------------

As recently as the 1970s, NSGCTs were often fatal. Due to advances in chemotherapy, most patients are cured, even those with disseminated disease (8). The clinical use of AFP and HCG measurements has been essential to this success. Many patients have a marker surge during the first week of chemotherapy, presumably secondary to turnor lysis. AFP may increase from 20% to 200% over pretreatment levels (20). Chemotherapeutic responses are accompanied by a decline in marker levels. Persistent marker elevation is usually the result of residual malignancy. Rising marker values may occur before or after clinical recurrence and one marker may rise in discordance with the other (21). Tumor recurrence is often accompanied by a rise in serum AFP values prior to clinical evidence of progressive disease (10-11).

AFP Serum Levels in Non-Testicular Malignancies. Elevated serum levels of AFP are also associated with some non-testicular cancers. Increased serum concentrations of AFP were first observed in human subjects with primary hepatocellular carcinoma (17). Subsequently, elevated serum AFP values have been associated with other malignant diseases such as teratocarcinoma (with yolk sac components) of the ovary, endodermal sinus tumors, certain gastrointestinal tumors (with and without liver metastasis), and tumors of other tissues (18-19, 22-26). A study performed at the National Institutes of Health and the Mayo Clinic demonstrated elevated AFP values in patients with pancreatic, gastric, colon, and lung cancer (20). In additional studies, AFP was elevated in 60-80% of patients with hepatocellular cancer, in 23% of patients with gastrointestinal cancer and in 10% of patients with liver metastasis from various turnor types (18). However, a normalization of markers does not mean that all viable tumor has been eliminated (20).

AFP Serum Levels in Healthy Controls and Benign Disease. Several studies have defined the specificity of the AFP assay. In the NIH study (20) , 210 control patients (blood donors) were tested for AFP by RIA. All controls had serum values below 40 ng/mL. AFP assay values for healthy adults, as determined in additional studies employing different assay systems, ranged from 1 to 25 ng/mL (18, 27-28).

Notably however, elevated serum AFP concentrations have also been reported in patients with noncancerous diseases such as ataxia telangiectasia, hereditary tyrosinemia, neonatal hyperbilirubinemia, acute viral hepatitis, chronic active hepatitis, cirrhosis, and other benign hepatic conditions (20, 22, 29-34). AFP is modestly elevated (up to 100 ng/mL) in 20% of patients with non-malignant liver disease (18). In the NIH study, most benign liver diseases were associated with levels below 500 ng/mL (20). Elevated serum AFP concentrations are also associated with pregnancy (35-36). Due to its lack of specificity for malignant conditions, AFP testing is not recommended as a screening procedure to detect cancer in the general population.

DEVICE DESCRIPTION

Indicated Use. The Bayer Immuno 1™ AFP Assay is an in vitro diagnostic assay intended to quantitatively measure AFP in human serum on the Bayer Immuno 1™ System as an aid in the management of nonseminomatous testicular cancer. AFP values obtained using the Bayer

{3}------------------------------------------------

Immuno 1™ AFP Assay must be interpreted in conjunction with all other available clinical and laboratory data before a medical decision is determined. The assay is designed to run on the Bayer Immuno 1™ Immunoassay System, a fully automated random-access analyzer which performs both homogeneous and heterogeneous immunoassays.

Description of the Method. The Bayer Immuno ITM AFP Assay utilizes a well-established immunoassay technology in which one monoclonal AFP antibody is conjugated to fluorescein (designated Reagent 1, or R1) and a second monoclonal AFP antibody is conjugated to alkaline phosphatase (Reagent 2, or R2). The R1 and R2 conjugates are reacted with patient sample, calibrator, or control and are incubated at 37°C on the system. An Immuno 1 Magnetic Particle coated with an anti-fluorescein antibody is then added and a second incubation occurs during which the antibody complex is bound. The magnetic particles complexed with the immunological sandwich are then washed to separate unbound molecules, and a colorimetric substrate is added. The rate of conversion of substrate to a compound with absorbance at 405 and 450 nm is measured and the measured rate is proportional to the concentration of AFP antigen in the sample. A cubic-through-zero curve fitting algorithm is used to generate standard curves. A schematic representation of the magnetic separation sandwich immunoassay technique used by the Bayer Immuno 1™ Analyzer is depicted in Figure 1 below.

The assay has a range of 0.1 to 400 ng/mL. The assay uses six calibrators with AFP concentrations of 0, 5, 25, 100, 200, and 400 ng/mL. A typical standard curve for the assay is presented in Figure 2.

Image /page/3/Figure/4 description: The image shows a diagram of a laboratory process involving magnetic particles, antibodies, and a sample analyte. The process includes steps of incubation and washing. The diagram also shows the different reagents used in the process, including IMP reagent (magnetic particles antifluorescein), reagent 2 (enzyme-labeled monoclonal antibody), and reagent 1 (fluorescein-labeled monoclonal antibody).

Figure.1 Schematic Representation of the Magnetic Separation Sandwich Immunoassay on the Bayer Immuno ITM System.

{4}------------------------------------------------

Image /page/4/Figure/1 description: The image shows a graph and a table. The graph plots absorbance units (AU/min) on the y-axis against concentration in ng/mL on the x-axis, showing a linear relationship. The table lists specific data points, including concentrations of 0, 5, 25, 100, 200, and 400 ng/mL, with corresponding absorbance units of 0.0044, 0.0488, 0.2301, 0.8649, 1.7061, and 3.0557 AU/min, respectively.

Figure 2 Standard Curve for the Bayer Immuno 1TM AFP Assay.

POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

The Immuno 1 AFP Assay is intended for in vitro diagnostic use only. There are no known potential adverse effects on the health of clinically managed patients when this device is used as indicated. It is imperative that the physician use the Immuno 1 AFP test results in conjunction with the patient's overall clinical assessment and other diagnostic tests. False test results could affect physician decisions regarding treatment. If falsely low, treatment may be delayed in cases of recurring testicular cancer. If falsely high, new therapy may be instituted unnecessarily. These false positive and false negative values should not lead to patient mismanagement as it is indicated that AFP assay values be used in conjunction with the results of the patient's overall clinical assessment.

PRECAUTIONS AND WARNINGS

This device is not indicated for testicular cancer screening or as a sole diagnostic tool to confirm the presence or absence of malignant testicular disease. AFP assay values should be used for the management of nonseminomatous testicular cancer patients in conjunction with the information from a complete clinical evaluation including a physical exam and other diagnostic tests.

In some cases, confirmed testicular carcinoma patients may show declining AFP levels while tumor masses continue to enlarge (16). Additionally, patients with certain non-malignant conditions (pregnancy, hepatitis, and cirrhosis) and other benign hepatic conditions (6, 20-24) and patients with certain non-testicular malignancies (hepatocellular carcinoma, teratocarcinoma of the ovary, endodermal sinus tumors, certain gastrointestinal turnors, and turnors of other tissues) can exhibit elevations in AFP assay levels (8, 16-18, 22-25, 31-36). As such, serum AFP assay levels should not be interpreted as absolute evidence of the presence of malignant disease.

{5}------------------------------------------------

The concentration of AFP in a given specimen determined with assays from different manufacturers can vary due to differences in assay methodology and reagent specificity. The results reported by the laboratory to the physician must include the identity of the AFP assay used, Values obtained with different AFP assays cannot be used interchangeably. If in the course of monitoring the patient, the assay method used for determining serial AFP levels is changed, additional sequential testing should be carried out to confirm baseline values. At the same time, a range of normal values for the new assay should be determined based on normal sera.

SUMMARY OF STUDIES

Nonclinical studies were performed to evaluate assay specificity and interfering substances, minimum detectable concentration, imprecision, linear range, hook effect, parallelism, and spiked recovery.

The clinical evaluation of the Bayer Immuno 1TM AFP Assay, as an aid in the management of patients with nonseminomatous testicular cancer, was performed at two clinical trial sites. Cross-sectional sensitivity and specificity, reference range, and longitudinal monitoring were evaluated. Immuno 1 AFP values were compared to Abbott IMx AFP values in a method concordance study using a panel of 863 cross-sectional samples and 56 longitudinal monitored patients.

NONCLINICAL STUDIES

Characterization of the Antigen used in the Immuno 1 AFP Assay is purified from human fetal cord serum. The AFP antigen is manufactured by Scripps Laboratories (San Diego, CA) and is supplied to Bayer. Antigen preparations, analyzed by reducing SDS-PAGE, and immunoblotted with rabbit anti-human alpha-fetoprotein followed by horseradish peroxidase conjugated goat anti-rabbit IgG, showed that the immunoreactive material migrated with a molecular weight of approximately 72 kDa. This electrophoretic analysis demonstrated that the antigen preparation used in the assay is consistent with previous descriptions of AFP. Isoelectric focusing of the AFP antigen resolved a pI of 4.8 which is consistent with literature references which show a pI of 4.75 for AFP (2).

Immunoreactivity of the Antibodies. Two monoclonal antibody preparations are used in the Immuno 1 AFP Assay. Both antibodies are manufactured by Medix and are supplied to Bayer following purification by protein A chromatography.

The monoclonal antibodies were characterized in a series of experiments. Isotype analysis demonstrated that the monoclonal antibodies are both of the murine IgG1 subclass. Relative affinity analysis revealed that both antibodies bound to AFP antigen in a similar and saturable manner. Biochemical analysis of MAb preparations, under conditions of reducing or nonreducing SDS-PAGE and isoelectric focusing, revealed bands characteristic of murine immunoglobulin molecules of the IgG isotype. These results demonstrate that the monoclonal antibodies bind to the AFP antigen quantitatively and display biophysical properties expected of mouse monoclonal antibodies.

{6}------------------------------------------------

Specificity And Interfering Substances. The recovery of AFP values from human serum was studied before and after spiking with the potentially interfering substances listed below. Each potential interferent was tested at a maximum concentration. Several potential interferents were tested at five equally spaced concentration levels (including no interferent and the maximum concentration of interferent). All testing was performed on one Immuno I system, using one lot of assay reagent.

Endogenous Interferents. The Immuno 1 AFP Assay was performed on serum samples or pools of serum to which was added various concentrations of either triglycerides, immunoglobulin, hemoglobin, heparin, bilirubin, albumin, or carcinoembryonic antigen (CEA). The highest concentration of each potential interferent used was greater than that normally observed during routine clinical testing. The highest concentration of each potential endogenous interferent tested and the maximum effect on the observed AFP recovery are summarized in Table 1. None of the potential endogenous interferents demonstrated any significant interfering effects on AFP recovery.

Exogenous Interferents. Because of the possibility that serum AFP measurements might be performed while patients are undergoing a regimen of chemotherapy, AFP values were measured in serum samples after spiking with either an individual drug or a cocktail of drugs commonly used to treat cancer. The identities of the individual chemotherapeutic drugs and their final concentrations in the test are presented in Tables 2 and 3. Because of the possibility that serum AFP measurements might be performed while patients are taking "Over the Counter" (OTC) drugs or dietary supplements, AFP values were also measured in serum samples after spilking with either an individual OTC drug or vitamin. The identities of the individual OTC drugs/dietary supplements and their final concentrations in the test are presented in Table 4. None of the potential exogenous interferents demonstrated any significant interfering effects on AFP recovery.

{7}------------------------------------------------

InterferentHighest ConcentrationTested% Recovery
Triglycerides900 mg/dL99.9
IgG0.053 g/mL98.2
Hemoglobin10.0 mg/mL102.1
Heparin0.15 mg/mL101.5
Bilirubin0.25 mg/mL99.7
Albumin0.065 g/mL104.0
CEA10,000 ng/mL99.9

Table 1. Endogenous Interference

PoolNumberDrug CocktailHighestConcentrationTested%Recovery
DrugCocktail#1Cyclophosphamide (Cytoxan)Doxorubicin (Adriamycin)Bleomycin800ug/mL51.8 ug/mL0.16 U/mL99.7%
DrugCocktail# 2Cis-platinEtoposideDactinomycin173 ug/mL415.2 ug/mL16 ug/mL98.7%

Table 2. Chemotherapeutic Drugs Used For Interference Testing Analyzed in a Drug Cocktail

DrugConcentrationTested (1X)% Recovery
Mitomycin C73 ug/mL99.2
Vinblastine5.11 ug/mL97.8
Vincristine13.5 ug/mL100.8
5-Fluorouracil1600 ug/mL101.3
Amethopterin (Methotrexate)450 ug/mL95.2

Table 3. Chemotherapeutic Drugs Used For Interference Testing Analyzed Individually

{8}------------------------------------------------

OTC Drug orConcentration Tested% Recovery
Dietary Supplement
Acetaminophen200 ug/mL104.8
Aspirin500 ug/mL99.4
Ibuprofen400 ug/mL102.8
Caffeine100 ug/mL101.1
Vitamin A10 IU/mL101.8
Vitamin B₁ (Thiamin)3 ug/mL102.6
Vitamin B₂ (Riboflavin)3.4 ug/mL100.5
Vitamin B₆4 ug/mL104.8
Vitamin B₁₂12 ng/mL99.8
Vitamin C (Ascorbic Acid)30 ug/mL100.6
Vitamin D₂0.8 IU/mL100.5
Vitamin E0.6 IU/mL101.3
Folic Acid0.8 ug/mL99.9
Niacin40 ug/mL100.6

Table 4. OTC Drugs and Dietary Supplements Used For Interference Testing

Heterophilic Antibodies. To investigate the effectiveness of the assay's reagent formulation in minimizing heterophilic antibody interferences, ten samples with high rheumatoid factor titers and nine samples with high human anti-mouse antibody titers were spiked with approximately 200 ng/mL of AFP antigen. Each sample was assayed in triplicate with one lot of Immuno 1 reagent. The percent recovery range for samples high in rheumatoid factor was 96.4-103.2% while the percent recovery range for samples high in HAMA was 93.5-102.4%. The observed AFP recoveries indicate a lack of significant heterophilic interference in the assay and demonstrate the effectiveness of the reagent formulation in minimizing these interferences.

Minimum Detectable Concentration. Analytical sensitivity of the Immuno 1 AFP assay was evaluated by determination of the Minimum Detectable Concentration (MDC). The MDC is defined as the minimum concentration of AFP which can be statistically distinguished from the concentration of the lowest standard as calculated from a typical standard curve. The estimate of the minimum detectable concentration of the assay is obtained from the imprecision of Level 1 calibrator's reaction rates. Specifically, the MDC of the Immuno 1 AFP assay is the AFP concentration corresponding to an absorbance value two within-run standard deviations above the reaction rate of the zero calibrator.

The minimum detectable concentration (MDC) of the Immuno 1 AFP Assay was based on multiple determinations of the Level 1 calibrator using two lots of reagents and two lots of calibrators. The study was performed on two systems at Tarrytown and on one system each at Memorial Sloan-Kettering Cancer Center and Diagnostic Oncology CRO (DOCRO). The MDC's from each of the sites are shown in Table 5.

{9}------------------------------------------------

SITENLEVEL 1WITHIN-RUN SD(mA)MDC(ng/mL)
Tarrytown3200.230.05
Memorial Sloan-Kettering1600.180.04
DOCRO1440.330.09
TOTAL6240.250.06

Table 5. Minimum Detectable Concentration at all Clinical Trial Sites

The Minimum Detectable Concentrations from three clinical trial sites ranged from 0.04 ng/mL to 0.09 ng/mL. These levels of analytical sensitivities are acceptable for an assay of this type and support a method sheet claim of 0.1 ng/mL.

Imprecision. Within-run and total assay imprecision were evaluated by testing five levels of Immuno 1 AFP Assay Calibrators and BioRad Tumor Marker Controls in a minimum of 18-20 assay qualification runs performed over a minimum period of ten days. The study was performed on two systems at Site 1 and on one system each at Sites 2 and 3. Typical imprecision obtained at a clinical trial site is presented in Table 6. The total imprecision was not greater than 3.6%.

SAMPLEMEANCONCENTRATION(ng/mL)OBSERVATIONSWITHIN-RUNTOTAL
SD(ng/mL)CV(%)SD(ng/mL)CV(%)
BioRad Level131.01600.652.10.882.8
BioRad Level2129.91604.503.54.693.6
BioRad Level3258.01608.173.28.863.4
Calibrator 25.03200.091.70.132.5
Calibrator 325.43200.351.40.532.1
Calibrator 498.33201.511.52.172.2
Calibrator 5199.93206.443.27.263.6
Calibrator 6398.23206.261.610.542.6

Table 6. Imprecision at a Clinical Site

These results demonstrate that the recovery of Immuno 1 AFP assay values are highly reproducible over time.

{10}------------------------------------------------

Linearity. To determine if AFP recoveries are linear over the entire calibration range, three clinical sample pools containing a high concentration of AFP values were diluted with normal serum (low AFP values) to final concentrations of 100% (undiluted), 75%, 50%, 25%, and 0% (low AFP serum only). Each pool was assayed with two lots of Immuno 1 AFP reagent.

The three pools of sera with AFP values of approximately 275-375 ng/mL were diluted. Recoveries of the intermediate dilutions were all between 95.7% and 102.8% of the expected value. These results clearly demonstrate the linearity of AFP recoveries over the entire calibration range.

Hook Effect. Extremely high concentrations of AFP seen in some malignant conditions may cause a "hook effect" in an assay. An excess of analyte saturates both label and capture antibody and causes the reported concentration to "hook" back into the assay range rather than be flagged as above range. AFP antigen, isolated from human fetal cord serum and purchased from Scripps Laboratories (San Diego, CA), was value-assigned by testing on the Immuno 1, and was diluted in Level 1 Calibrator at concentrations of 32.0 ng/mL to 204,268 ng/mL. This collection of samples was tested in the assay using two lots of reagents. The reaction rate did not "hook" back into the assay range (Calibrator Level 6 reaction rate) until a concentration of 204,268 ng/mL AFP was reached. These results clearly demonstrate the lack of a hook effect in the Immuno 1 AFP Assay at AFP assay values ≤ 160,000 ng/mL.

Spiked Recovery. To determine how well AFP antigen, when spiked into a patient sample, is recovered by the Immuno 1 AFP Assay, antigen was spiked into five patient samples at an AFP concentration of approximately 300 ng/mL. Each sample was assayed in triplicate with one lot of Immuno 1 reagent.

Recoveries of assay values for all samples ranged from 96.7 to 99.3%. No significant deviation was noted with regard to expected versus observed assay values. These results demonstrate the accurate quantitation of spiked and recovered AFP values using the Immuno 1 assay.

Parallelism. As a further verification of assay linearity and to determine the acceptability of 0 ng/mL Calibrator as a diluent, four patient serum sample pools containing a high level of AFP values were diluted with Level 1 calibrator to final concentrations of 100% (undiluted), 75%, 50%, 25%, and 0% (Level 1 calibrator only). Each dilution of each sample pool was assayed with two lots of reagents. Two serum pools were tested on one Immuno 1 system while the other two serum pools were tested on a second Immuno 1 system. Linear regression analysis for the determination of deviations from linearity for each of these clinical samples showed no deviation from linearity. The recovery of AFP assay values ranged from 92.7-111.0%. The accurate recovery of AFP values in diluted patient samples further illustrates the linearity of the Immuno 1 AFP Assay throughout the entire calibration range and the acceptability of 0 ng/mL Calibrator as diluent.

{11}------------------------------------------------

CLINICAL STUDIES AND METHOD COMPARISON STUDIES

Introduction. The objective of the method comparison studies was to examine the concordance of sample assay values obtained using the Bayer Immuno 1™ AFP Assay with those obtained using the Abbott IMx AFP Assay. Patient sample AFP values generated by the two methods were compared by correlation analysis, a determination of normal range cutoff, an analysis of clinical sensitivity and specificity, and an analysis of longitudinal sample AFP values.

This retrospective study was performed at two investigational sites. Both the Immuno 1 testing and Abbott IMx analysis on patient specimens was performed in the laboratories at Diagnostic Oncology CRO ("DOCRO") in Stratford, CT and at Memorial Sloan-Kettering Cancer Center ("MSKCC") in New York, NY. Human serum samples from 545 patients obtained from Western States Plasma Company (Fallbrook, CA) were tested at DOCRO and a total of 318 samples from MSKCC's own in-house specimen collection were analyzed for method correlation using the Immuno 1 AFP Assay and the Abbott IMx AFP Assay. The number, source, and clinical classification of the single point patient samples used in this comparison study are summarized in Tables 7 and 8.

Clinical ClassificationSourceNumber of Samples
Normal (Male)Western States Plasma150
Normal (Female)Western States Plasma100
Testicular Cancer (Mixed Germ Cell Tumor)Western States Plasma5
Testicular Cancer (Seminomatous Germ Cell Tumor)Western States Plasma1
Hepatocellular CarcinomaWestern States Plasma52
Genito-Urinary CancerWestern States Plasma33
Gastrointestinal, Lung & Pancreatic CancerWestern States Plasma48
HepatitisWestern States Plasma50
CirrhosisWestern States Plasma50
Benign Genito-Urinary DiseasesWestern States Plasma19
Other non-malignant DiseasesWestern States Plasma37
TotalWestern States Plasma545

Table 7. Summary of Single Point Patient Samples Used In Comparison Study At DOCRO Site.

{12}------------------------------------------------

ClinicalClassificationSourceNumber of Samples
Normal SerumMemorial Sloan-Kettering100
Testicular Cancer(Nonseminomatous Germ CellTumor)Memorial Sloan-Kettering74
Testicular Cancer (MixedGerm Cell Tumor)Memorial Sloan-Kettering22
Testicular Cancer(Seminomatous Germ CellTumor)Memorial Sloan-Kettering6
Prostate CancerMemorial Sloan-Kettering10
Liver CancerMemorial Sloan-Kettering10
Colon CancerMemorial Sloan-Kettering10
Breast CancerMemorial Sloan-Kettering10
Lung CancerMemorial Sloan-Kettering10
Benign GU DiseaseMemorial Sloan-Kettering10
Randomly-Selected TesticularCancer Longitudinal PatientsMemorial Sloan-Kettering51
Randomly-SelectedHepatocellular CancerLongitudinal PatientsMemorial Sloan-Kettering5
TotalMemorial Sloan-Kettering318

Table 8. Summary of Single Point Patient Samples Used In Comparison Study At MSKCC Site.

Longitudinal patient samples were also tested at MSKCC as part of the method correlation analysis. A total of 314 longitudinal samples from 51 testicular cancer patients and 5 hepatocellular cancer pateints were tested by Immuno 1 and IMx assays and compared. A summary of the longitudinal patients and sample number is presented in Table 9.

Cancer Type# Patients# Samples per Patient
5678
Testicular5131983
Hepatocellular55000
Total5636983
Number of Total Samples180545624

Table 9. Summary of Longitudinal Patient Samples Used In Comparison Study At MSKCC Site

Normal Range. The normal range cut-off for normal specimens was determined by calculation of the 97.50 percentile of AFP values in 150 normal healthy male patients. The distribution of

{13}------------------------------------------------

Immuno 1 and Abbott IMx values for healthy subjects used in this study are shown in Tables 10 and 11 respectively. Male subjects ranged in age from 17 to 60.

The results of the normal range analysis show the mean serum AFP values for all samples analyzed by the Immuno 1 AFP Assay and the Abbott IMx AFP Assay were 3.02 ng/mL and Females were included in this analysis only for comparative 2.61 ng/mL respectively. purposes. To determine the upper range of normal for the Immunol AFP Assay, data derived only from males was used. The greatest 97.50 percentile normal range cut-off found at any one site for the Immuno 1 was 8.9 ng/mL compared to 8.1 ng/mL for Abbott IMx AFP Assay. These results are consistent with the manufacturer's suggested normal range cutoff ranges for the Abbott IMx AFP Assay which states that 99% of the healthy subjects had AFP levels less than 8.9 ng/ml. Overall, the means and normal range cut-off of the AFP assay values calculated for the normal samples using these two methods were similar, and demonstrate the concordance of assay values determined by these methods. As the 97.5" percentile values ranged between 4.6 and 8.9 ng/mL, a cut-off of 8.9 ng/ml was determined to be appropriate for the Immuno 1 AFP Assay.

Normal,Healthy,SubjectsNDistribution of IMMUNO 1 AFP AssayResults (ng/mL)FOR NORMAL SUBJECTSMean(SD)AFPNon-Para-metric97.5th %(ile)
0-8.9(N%)>8.9-15(N%)>15-20(N%)>20-30(N%)>30(N%)ng / mLng / mL
MSKCC(males)100100(100%)0(0%)0(0%)0(0%)0(0%)2.48(1.03)4.6
DOCRO(males)150146(97.3%)2(1.3%)2(1.3%)0(0%)0(0%)3.38(2.46)8.9
DOCRO(females)10098(98%)0(0%)1(1%)1(1%)0(0%)3.02(3.33)8.2
TOTAL350344(98.3%)2(0.6%)3(0.9%)1(0.3%)0(0%)3.02(2.48)8.0

Table 10. Distribution of Immuno1 AFP Assay Results for Normal Subjects

{14}------------------------------------------------

Normal,Healthy,SubjectsNDistribution of ABBOTT IMX 1 AFP AssayResults (ng/mL)FOR NORMAL SUBJECTSMean(SD)AFPNon-Para-metric97.5th%(ile)
0-8.9(N%)>8.9-15(N%)>15-20(N%)>20-30(N%)>30(N%)ng / mLng / mL
MSKCC(males)100100(100%)0(0%)0(0%)0(0%)0(0%)2.30(0.91)4.5
DOCRO(males)150147(98%)2(1.3%)1(0.7%)0(0%)0(0%)2.86(2.27)8.10
DOCRO(females)10098(98%)0(0%)1(1%)1(1%)0(0%)2.53(2.83)7.80
TOTAL350345(98.6%)2(0.6%)2(0.6%)1(0.3%)0(0%)2.61(2.18)7.0

Table 11. Distribution of Abbott IMx AFP Assay Results for Normal Subjects

Method Comparison. In order to compare the values obtained from serum samples analyzed by the Immuno 1 AFP Assay with the Abbott IMx AFP assay values, a correlation study was performed using data collected at two clinical sites. A summary of the patient populations used for these studies at Diagnostic Oncology CRO and Memorial Sloan-Kettering Cancer Center is shown in Tables 12 and 13 respectively. This study included the analysis of 545 single point serum samples at DOCRO and 318 single point serum samples at MSKCC.

Passing-Bablok and ordinary least squares linear regression analyses were performed to compare the Immuno 1 and Abbott IMx AFP assay values. Sample values were obtained in singlicate determinations for both the Immuno 1 and Abbott IMx AFP assays.

The data for 0-400 ng/mL and >400 AFP ranges are summarized in Tables 12 and 13 for both clinical sites. Results include 95% confidence intervals for the slopes and intercepts. An excellent comparison is demonstrated by slopes near 1.0 and correlation coefficients (r) greater than 0.99. Irrespective of range, site of analysis or regression analysis algorithm, the correlation between the two methods is excellent.

ાં ર

{15}------------------------------------------------

Range ofAnalyteConc.ng/mLNrSlope95%ConfidenceInterval (slope)Intercept95% ConfidenceInterval (intercept)RegressionMethod
0-4005160.9931.0471.036 to 1.0580.350-0.040 to 0.741OrdinaryLeastSquares
>400290.9921.0460.992 to 1.099-165.8-1452.8 to 1121.2OrdinaryLeastSquares
0-4005160.9931.1091.075 to 1.1410.2580.183 to 0.341Passing-Bablok
>400290.9920.9850.922 to 1.072-1.843-63.9 to 51.7Passing-Bablok

Table 12. Method Comparison at DOCRO

Range ofAnalyteConc.ng/mLNrSlope95%ConfidenceInterval (slope)Intercept95% ConfidenceInterval (intercept)RegressionMethod
0-4003020.9970.9310.923 to 0.9390.9320.390 to 1.473OrdinaryLeastSquares
>400160.9990.9390.914 to 0.96526.462-129.139 to 182.064OrdinaryLeastSquares
0-4003020.9971.0241.000 to 1.0490.0890.002 to 0.201Passing-Bablok
>400160.9990.9310.878 to 0.96721.238-12.968 to 127.8Passing-Bablok

Table 13. Method Comparison at MSKCC

Clinical Sensitivity and Specificity. The clinical sensitivity of the Immuno 1 and Abbott IMx AFP assays was determined by measurement of AFP values in serum samples from 213 patients with either active testicular or liver cancer. The testicular cancer samples used in this analysis were obtained from patients with either nonseminomatous germ cell turnor (GCT) or mixed germ cell tumor by DOCRO and MSKCC. The clinical specificity of the Immuno 1 AFP Assay and Abbott IMx assays was determined by measurement of AFP values in 641 serum samples from patients without nonseminomatous testicular cancer. These included

{16}------------------------------------------------

000208

healthy individuals, or patients with seminomatous testicular cancer, non-testicular cancers, or non-malignant diseases.

The frequency distribution of AFP values from patients in all categories are presented in Table 14 for Immuno 1 values and in Table 15 for Abbott IMx values. The upper limit of the reference interval that was used was 8.9 ng/mL AFP for both assays.

BAYER IMMUNO 1 SYSTEM
AFP ASSAY VALUES
(ng/mL)
PATIENTPOPULATIONN0-8.9(%)>8.9-100(%)>100-400(%)>400(%)MEDIAN
HealthySubjects35098.31.70.00.02.5
Testicular Ca.- nonseminomatous10036.040.014.010.020.2
Testicular Ca.-Mixed GermCell Tumor4643.545.76.54.311.6
Testicular Ca.-seminomatous875.025.00.00.02.5
Prostate Ca./Bladder Ca.4097.52.50.00.03.4
Lung Ca.2996.60.00.03.43.8
Colorectal Ca.3889.510.50.00.04.5
Liver Ca.670.031.320.947.8310.9
Breast Ca.1080.020.00.00.04.8
Cirrhosis5088.012.00.00.04.0
Hepatitis5088.012.00.00.03.9
Benign Genito-Urinary Disease2993.16.90.00.03.1
Other Non-malignant37100.00.00.00.02.3

Table 14. Clinical Sensitivity and Specificity with Immuno I AFP Values

{17}------------------------------------------------

ABBOTT IMxAFP ASSAY VALUES(ng/mL)
PATIENTN0-8.9>8.9-100>100-400>400
POPULATION(%)(%)(%)(%)MEDIAN
HealthySubjects35098.61.40.00.02.1
Testicular Ca .-nonseminomatous10036.038.016.010.017.9
Testicular Ca .-Mixed GermCell Tumor4645.741.38.74.310.8
Testicular Ca .-seminomatous875.025.00.00.02.5
Prostate Ca./Bladder Ca.4097.52.50.00.02.8
Lung Ca.2996.60.00.03.42.8
Colorectal Ca.3892.17.90.00.03.4
Liver Ca.670.031.320.947.8306.7
Breast Ca.1080.020.00.00.04.3
Cirrhosis5088.012.00.00.04.0
Hepatitis5080.020.00.00.03.8
Benign Genito-Urinary Disease29100.00.00.00.02.6
Other Non-malignant3797.32.70.00.01.9

Table 15. Clinical Sensitivity and Specificity with IMx AFP Values

The clinical sensitivity of both the Immuno 1 and IMx AFP assays for nonseminomatous testicular cancer was 64.0%. The sensitivity for mixed germ cell tumor testicular cancer (containing both nonseminomatous and seminomatous tumors) was 56.5 % for the Immuno 1 assay and 54.3% for the IMx assays. This is comparable to previous studies reporting 50-75% of AFP values above normal for nonseminomatous and mixed GCT (8-10, 12, 15, 18).

The clinical specificity of the Immuno 1 AFP Assay for healthy individuals was 98.3% while the specificity of the IMx assay was 98.6%. Both are consistent with results generally seen with the normal population (28, 30). Specificity across all other benign diseases (cirrhosis, hepatitis, benign genito-urinary disease, and other non-malignant disease) ranged from 88%-100% for both the Immuno 1 and IMx assays. Elevated Immuno 1 and Abbott IMx AFP Assay results were observed in 12% of cirrhosis and hepatitis patients. These observations are consistent with previous literature reports of AFP elevations in cirrhosis, hepatitis and other benign conditions (20, 28, 30, 33-34).

{18}------------------------------------------------

Elevated Immuno 1 observations were also observed in 2.5% prostate or bladder cancer patients. Of the 40 patients tested in this group, only one patient with bladder cancer possessed an elevated AFP level. Additionally, elevated Immuno 1 AFP values were observed in 3.4% of lung cancer specimens, 10.5 % of colorectal cancer specimens, and 20% of breast cancer specimens. Most notably, virtually identical results were obtained using the IMx assay. These results are consistent with previous published studies which also noted that elevations of AFP is not limited to malignant disease of the testes (8, 16, 22, 26, 28). Interestingly, elevated AFP levels were seen in all patients (N=67) with liver cancer. This observation correlates with the published observations that elevated AFP was first noted in patients with hepatocellular cancer (17, 24) and that elevated AFP values can be found in 60-90% of these patients (18, 21-22, 29-30).

The data presented in Table 15 demonstrate the concordance of AFP assay values determined by the Immuno 1 and IMx assays. Percent sensitivity and percent specificity values are comparable. The data in these tables also confirm the conclusion that AFP is not recommended as a screening procedure to detect cancer in the general population. Not all nonseminomatous testicular cancers produce AFP while at the same time, elevated AFP levels may be found in other conditions, both malignant and non-malignant (29).

Analysis of Longitudinal Sample AFP Values The principal utility of AFP testing is in the evaluation of longitudinal or serial serum samples from testicular cancer patients. The level of AFP can be used to assess the effectiveness of chemotherapy, the prognosis of the disease, and the possible recurrence of the cancer (37-39). Tumor markers for testicular cancer are measured regularly in patients on surveillance for stage I disease, on chemotherapy, and on follow up after surgery (37). The regression of serum tumor marker levels such as AFP help to predict treatment outcome (38). Patients with an appropriate serum tumor marker decline have a longer event-free survival. Decreasing levels indicate regression even though a return to normal levels does not mean the eradication of all tumor cells (39).

In this part of the study, the clinical utility of the Bayer Immuno 1™ AFP Assay as an adjunctive test in the management of nonseminomatous testicular cancer patients was evaluated.

Longitudinal samples from specimen banks at Memorial Sloan-Kettering Cancer Center included serum samples from 47 testicular cancer patients and for comparison purposes only, 5 hepatocellular cancer patients. Each patient had ≥ 5 specimens which met inclusion criteria, spanning the longitudinal profile. Each patient's age, documented medical, diagnostic, and course of therapy information was available. In the course of the evaluation, 4 testicular cancer patients were eliminated from the analysis because there was insufficient accompanying clinical information to allow complete evaluation of their longitudinal AFP profiles.

All patient samples were tested using both the Immuno 1 AFP Assay and the Abbott IMX AFP Assay. Each patient's serial Immuno 1 AFP results were graphically compared to the patient's

{19}------------------------------------------------

course of disease using diagnostic testing, therapeutic interventions, and physician evaluations as guides to describe the clinical picture.

The ability of the Immuno 1 AFP Assay to accurately reflect each patient's course of disease was evaluated by Bayer scientists and the profile was classified into one of the following categories as listed in Tables 16-18.

Table 16 presents the comparison of serial Immuno 1 AFP results with the clinical course of disease for all of the patients evaluated (47 testicular and 5 hepatocellular cancer patients). Table 17 presents the comparison of serial Immuno 1 AFP results with the clinical course of disease for testicular cancer patients and Table 18 presents results for hepatocellular cancer patients.

IMMUNO 1 AFP LONGITUDINAL PATIENT EVALUATION RESULTS
ALL PATIENTS
Correspondence(Parallels Clinical Course)NPercentage
Increasing AFP with Progression23.8
Decreasing AFP with Response1732.7
Increasing and Decreasing AFPwith Progression and Response1936.5
Elevated AFP withActive Disease00
Normal AFP with No Evidenceof Disease11.9
Stable AFP with StableDisease23.8
Total ParallelingClinical Course4178.8
No Correspondence(Does Not Parallel ClinicalCourse)1121.2
Table 16. Evaluation of All Longitudinal Patients

{20}------------------------------------------------

IMMUNO 1 AFP LONGITUDINAL PATIENT EVALUATION RESULTSTESTICULAR CANCER PATIENTS ONLY
Correspondence(Parallels Clinical Course)NPercentage
Increasing AFP with Progression00
Decreasing AFP with Response1736.2
Increasing and Decreasing AFPwith Progression and Response1736.2
Elevated AFP withActive Disease00
Normal AFP with No Evidenceof Disease12.1
Stable AFP with StableDisease12.1
Total ParallelingClinical Course3676.6
No Correspondence(Does Not Parallel ClinicalCourse)1123.4
Total

Table 17. Evaluation of Testicular Cancer Longitudinal Patients

ﮐﺴﺘﺎ

{21}------------------------------------------------

IMMUNO 1 AFP LONGITUDINAL PATIENT EVALUATION RESULTSHEPATOCELLULAR CANCER PATIENTS ONLY
Correspondence(Parallels Clinical Course)NPercentage
Increasing AFP with Progression240
Decreasing AFP with Response00
Increasing and Decreasing AFPwith Progression and Response240
Elevated AFP withActive Disease00
Normal AFP with No Evidenceof Disease00
Stable AFP with StableDisease120
Total ParallelingClinical Course5100
No Correspondence(Does Not Parallel ClinicalCourse)00
Total

Table 18. Evaluation of Hepatocellular Cancer Longitudinal Patients

The percent of patients whose changes in longitudinal AFP values reflected changes in the clinical profile was calculated. The Bayer Immuno 17M AFP Assay demonstrated clinical utility for 78.8% (41/52) of the all patients evaluated longitudinally and 76.6% (36/47) of testicular cancer patients evaluated. Among the patients where Immuno 1 AFP accurately reflected disease course, 2 patients demonstrated increasing AFP concentrations with progressive cancer, 17 patients demonstrated decreasing AFP concentrations in response to therapy and 19 patients demonstrated both increasing and decreasing AFP concentrations corresponding to periods of both disease progression and response to therapy. For 1 patient with no evidence of disease, AFP values were consistently normal throughout the monitoring period. For 1 patient under therapy with stable disease, AFP levels were consistently normal and stable throughout the monitoring period. For 1 hepatocellular cancer patient, AFP concentrations were consistently stable at high concentrations.

One testicular cancer patient demonstrated only seminomatous germ cell tumor on histology which usually does not show elevated AFP levels. At the point of increasing AFP levels, metastasis to the liver was discovered. The literature supports increasing AFP levels in seminomatous GCT when metastasis to the liver is present (24). The data from the 5 hepatocellular cancer patients also demonstrates the correspondence of AFP values with the

{22}------------------------------------------------

clinical course of the disease. The AFP values paralleled the clinical course of the disease in all of these patients.

The data presented in Table 17 demonstrate the utility of AFP testing in the management of nonseminomatous testicular cancer. The level of AFP can be used to assess the effectiveness of chemotherapy, the prognosis of the disease, and the possible recurrence of the cancer (37-39), Abbott IMx results were directly compared to Immuno 1 AFP values and the correlation between the two methods was excellent. There were clear rising trends with both methods in correspondence to disease progression and declining trends in response to therapy. The trends agreed in 100% of the graphs when comparing the Immuno 1 and IMx AFP values.

The usefulness of AFP measurements in the management of nonseminomatous testicular cancer in patients undergoing cancer therapy has been well established (10, 12, 19). Current management of the testicular germ cell tumors (GCTs) relies upon the use of serum tumor markers which can indicate the presence of small foci of active tumor that cannot be detected by currently available imaging techniques (16). Markers augment and complement information obtained from radiographic and other staging procedures (20). Also, the short half-lives of tumor markers facilitate their use is assessing turnor burden during therapy. AFP has a serum half-life of 3.5-6 days (21). AFP and/or HCG levels are elevated before orchiectomy in about 60% of all Clinical Stage I patients but follow a normal decline after the testicle is removed (18). It is not unusual to have AFP levels at 1000 ng/mL at orchiectomy, which means 5-6 weeks before normalization according to expected half-life values. At the same time, in patients recently diagnosed with testicular cancer, a highly elevated AFP value indicates a poor prognosis (39).

For patients in clinical remission following treatment, AFP levels generally decrease (12). Post-operative AFP levels which fail to return to normal strongly suggest the presence of residual turnor (10, 12, 22). Following successful resection of primary or metastatic disease, AFP declines at a rate proportional to its respective half-life (21). An elevated actual half-life (AHL) of serum markers following orchiectomy or retroperitoneal lymph node dissection (RPLND) may indicate the presence of occult, persistent disease (20). Residual disease is the rule if such a decline does not ensue. AFP values may also rise before there is other evidence of disease progression (10) and is sometimes used as a first indicator of relapse (16).

The results of this study demonstrates the clinical utility of the Bayer Immuno 1™ AFP Assay as an aid in the management of testicular cancer patients. Identical trends in longitudinal results were observed with both the Immuno 1 and IMx AFP assays. This demonstrates equivalent clinical utility for the two methods.

Conclusions from Clinical Studies and Method Comparison Studies. The results of these clinical studies and comparative analysis of serum AFP assay values in the determination of AFP in normal, malignant, and non-malignant patient serum, clearly demonstrates the concordance of the Immuno 1 AFP Assay with the predicate device, the Abbott IMx AFP Assay. Normal reference ranges, clinical sensitivity and analysis of longitudinal

{23}------------------------------------------------

Correlation statistics showed good concordance serum samples were essentially equivalent. between the Immuno 1 and the IMx assays. Longitudinal monitoring data demonstrated the utility of the Immuno 1 AFP Assay in the management of patients with nonseminomatous testicular cancer.

REAGENT STABILITY TESTING

Reagent Shelf-Life Stability Testing. Reagents were subjected to temperature stress at 25℃, 30°C, and 40°C and tested at selected timepoints. Additional reagent stored at 2-8°C was tested in parallel with the stressed reagents at all timepoints tested. At each timepoint, the sensitivity of the calibrators, as well as the recovery of controls were monitored. Data obtained using three lots of reagents support a shelf life stability claim of 12 months.

Reagent On-System Stability Testing. The On-System stability of Immuno 1 AFP assay reagents was evaluated using three lots of reagent. At all timepoints tested, all control recoveries, calculated from either "week 0" or timepoint calibration curves, were within specifications. On-system stability studies on the three lots of Immuno 1 AFP reagents support an on-system stability recommendation of 21 days.

Reagent Shipping Stabilty Testing. Three lots of reagents were subjected to three freeze/thaw cycles (three days at -80°C, and then three days at 2-8°C) and three heat/chill cycles (three days at 40°C, then three days at 2-8°C). Following these cycles, the reagents were stored at 2-8°C, and performance was evaluated at selected timepoints. Additional reagents were stored at 2-8°C and run alongside the stressed reagents at all timepoints. The sensitivity of the calibrators and the recovery of controls were monitored at each timepoint.

No significant change in control recovery, as compared to the non-stressed control reagent, was noted throughout the duration of testing. Recoveries for the cycled reagent remained well within specification. Shipping, temperature stress, and temperature cycling stability data on three lots of Immuno 1 AFP assay reagents support the requirement of refrigerated (2-8°C) shipping of these reagents.

Labeling. Reagent stability is summarized in the Immuno 1 AFP method insert sheet. Expiration dates are also indicated on the labels of each reagent kit.

CALIBRATOR STABILITY TESTING

Calibrator Shelf-Life Stability Testing. The calibrator lots were stressed at 2-8°C, 30°C and 40℃ and tested at selected timepoints. Additional calibrators stored at 2-8℃ and at <-80℃ were tested in parallel with the stressed calibrators at all timepoints. At each testpoint, recovery of controls was determined. Additionally, the recovery of the temperature stressed calibrators (analyzed as unknowns) using the -80°C stored calibrators for a calibration curve, was also determined.

{24}------------------------------------------------

Three lots of calibrators were tested for shelf-life stability. For all lots tested, recoveries of calibrators stored at 2-8℃ were within specifications. Recoveries of calibrators stored at the elevated temperatures exceeded the specification of ±13%. Recoveries of control materials remained within specifications for the duration of the testing. These results support the conclusion that the Immuno 1 AFP calibrators are stable for 9 months when stored at 2-8℃.

Calibrator Open Vial Stability Testing. After opening, calibrators were stored at 2-8ºC and analyzed at selected weekly timepoints. In addition, at each timepoint a fresh (control) set of calibrators was run. Recovery of the calibrators and controls, run as unknowns at Day 0, 1 week, 2 weeks, 3 weeks, and 52 day timepoints were determined. Analyte values were derived using a calibration curve generated at that timepoint using the calibrators opened and stored at 2-8°C.

Three lots of calibrators were tested for open vial stability. At all timepoints tested, control recoveries remained within specifications. These results support the claim that opened Immuno 1 AFP calibrators are stable for 30 days when stored at 2-8°C.

Shipping stability studies were conducted on three lots of Calibrator Shipping Studies. calibrators. Recoveries were determined for calibrators subjected to three cycles of freezing and thawing using a calibration curve generated with control calibrators stored at -80°C. Based on these studies, it is recommended that assay calibrators be shipped in a refrigerated package (2-8°C).

Calibrator stability is summarized in the Immuno 1 AFP method insert sheet. Labeling. Expiration dates are also indicated on the labels of each reagent kit.

CONCLUSIONS DRAWN FROM ALL THE STUDIES

Valid Scientific Evidence. The conclusions drawn from these studies are based upon valid scientific evidence. Data were gathered following a well designed protocol, in a research laboratory operating under the principles of Good Laboratory Practices.

Immuno 1 AFP Assay nonclinical performance, including analytical Method Performance. sensitivity (minimum detectable concentration), imprecision, parallelism, linear range, hook effect, and spiked recovery met accepted specifications for an assay of this type.

The clinical studies demonstrate the safety and effectiveness of Safety and Effectiveness. measuring AFP in human serum on the Bayer Immuno 1™ system to aid in the management of nonseminomatous testicular cancer. The correlations between the Immuno 1 AFP Assay values and disease course demonstrate that this assay may be used in conjunction with other available clinical and laboratory data to manage nonseminomatous testicular cancer patients.

{25}------------------------------------------------

The method concordance studies confirm the substantial clinical Substantial Equivalence. equivalence of the Immuno 1 AFP Assay with the Abbott IMx AFP Assay predicate device. There is a high degree of correlation between the Immuno 1 and IMx AFP specimen values. Multi-site clinical studies demonstrate cross-sectional and longitudinal specimen results that are Therefore, based upon the analytical and clinical concordance equivalent for the two tests. established in these studies, the Bayer Immuno 1TM AFP Assay and the Abbott IMx AFP Assay are equivalent with respect to method performance, clinical utility, and device safety and effectiveness.

{26}------------------------------------------------

Cited References

  • Ruoslahti E. Seppala M. Studies of Carcino-fetal Proteins: Physical and Chemical 1. Properties of Human Alpha-Fetoprotein. Int. J. Cancer 7: 218-225 (1971).
  • Hirai H. Alpha-Fetoprotein. In Biochemical Markers for Cancer. Ming Chu T 2. (ed.) New York, Dekker. 1982. pages 25-59.
  • Deutsch HF. Chemistry and Biology of Alpha-Fetoprotein. Adv. Cancer Res. 56: 253 3. (1991).
  • Nunez EA. Biological Role of Alpha-Fetoprotein in the Endocrinological Field: Data 4. and Hypothesis. Tumor Biol. 15: 63-72 (1994).
  • Gitlin D, Perricelli A, Gitlin GM. Synthesis of Alpha-Fetoprotein by Liver, Yolk Sac, న. and Gastrointestinal Tract of the Human Conceptus. Cancer Res. 32: 979-982 (1972)..
  • Sundram s., Goldstein PJ, Saravanan M, et al. Alpha-Fetoprotein and Screening 6. Markers of Congenital Disease. Clin. Lab. Med. 12: 481-493 (1992).
  • Esteban C., Trojan A., Mishal Z., et al. Activation of an Alpha-Fetoprotein/Receptor 7. Pathway in Human Normal and Malignant Peripheral Blood Mononuclear Cells. Leukemia. 7: 1807-1816 (1993).
  • Lange PH. Testicular Cancer Markers. in Human Cancer Markers. Sell S and Wahren 8. · B (ed.) Clifton, Humana. 1985. pages 259-273.
  • Small EJ, Torti, FM. Testes. in Clinical Oncology. New York, Churchill Livingstone. 9. 1995. pages 1493-1526.
    1. Kohn J, Orr AH, McElwain TJ, et al. Serum Alpha-Fetoprotein in Patients with Testicular Tumours. Lancet 2: 433-436, (1976).
  • Scardino PT, Cox HD, Waldmann TA, et al. The Value of Serum Tumor Markers in the 11. Staging and Prognosis of Germ Cell Tumors of the Testis. J. Urol. 118: 994 (1977).
    1. Lange PH, McIntire KR, Waldmann TA, et al. Serum Alpha-Fetoprotein and Human Chorionic Gonadotropin in the Diagnosis and Management of Nonseminomatous Germ Cell Testicular Cancer. Medical Intelligence 295: 1237 (1976).
  • Javadpour N, McIntire KR, Waldmann TA. Human Chorionic Gonadotropin 13. (HCG) and Alpha-Fetoprotein (AFP) in Sera and Turnor Cells of Patients with Testicular Seminoma A Prospective Study . Cancer 42: 2768-2772 (1978).

{27}------------------------------------------------

  • Lange PH, Nochomovitz LE, Rosai J, et al. Serum Alpha-Fetoprotein and Human 14. Chorionic Gonadotropin in Patients with Seminoma. J. Urol. 124: 472-478 (1980).
  • Jacobsen GK. Alpha-Fetoprotein (AFP) and Human Chorionic Gonadotropin (HCG) in ાં રે . Testicular Germ Cell Tumors. Acta. Path. Microbiol. Immunol. Scand. 91: 183-190 (1983).
  • Doherty AP, Bower M, Christmas TJ. The Role of Tumour Markers in the Diagnosis 16. and Treatment of Testicular Germ Cell Cancers. Brit. J. Urol. 79: 247-252 (1997).
  • Tatarinov YS. Finding of an Embryonic Alpha Globulin in the Blood Stream in a 17. Patient with Primary Hepatic Cancer. Vopr. Med. Khim. 10: 90 (1964).
  • Klepp O. Serum Tumour Markers in Testicular and Extragonadal Germ Cell 18. Malignancies. Scand. J. Clin. Lab. Invest. Suppl. 51: 28-41 (1991).
    1. Perlin E. Engeler JE, Edson M, et al. The Value of Serial Measurement of Both Human Chorionic Gonadotropin and Alpha-Fetoprotein for Monitoring Germinal Cell Tumors. Cancer 37: 215-219 (1976).
  • Bartlett NL, Freiha FF, Torti FM. Serum Markers in Germ Cell Neoplasms. Hem./Onc. 20. Clinics of N.A. 5: 1245-1260 (1991).
  • Jacobs EL, Haskell CM. Clinical Use of Tumor Markers in Oncology. in Current 21. Problems in Cancer. Littleton, Mosby-Year Book. 1991. Pages 299-359.
    • · Waldmann TA, McIntire KR. The Use of a Radioimmunoassay for Alpha-22. Fetoprotein in the Diagnosis of Malignancy. Cancer 34: 1510-1515 (1974).
  • Silver HKB, Gold P, Feder S, et al. Radioimmunoassay for Human Alpha-23. Fetoprotein. Proc. Nat. Acad. Sci. USA 70: 526-530 (1973).
  • Abelev GI. Alpha-Fetoprotein in Ontogenesis and Its Association With Malignant 24. Tumors. Adv. Cancer Res. 14: 295, 1971.
  • Maeyama M, Tayama C, Inoue S, et al. Serial Serum Determination on Alpha-25. Fetoprotein as a Marker of the Effect of Postoperative Chemotherapy in Ovarian Endodermal Sinus Tumor. Gynecol. 17: 104-116 91984 0.
  • Yasunami R, Hashimoto Z, Ogura T, et al. Primary Lung Cancer Producing 26. Alpha-Fetoprotein: A Case Report. Cancer 47: 926-929, 91981 0.
  • D'Costa M, Feld R, Laxdal V., et al. A Multicenter Evaluation of the Boehringer 27. Mannheim ES 300 Immunoassay System. Clin. Biochem. 26: 51-57 (1993).

{28}------------------------------------------------

  • Cattini R., Cooksey M, Robinson D, et al. Measurement of Alpha-Fetoprotein, 28. Carcinoembryonic Antigen and Prostate-Specific Antigen in Serum and Heparinised Plasma by Enzyme Immunoassay on the Fully Automated Serono SR1™ Analyzer. Eur. J. Clin. Chem. Clin. Biochem. 31: 517-524 (1993).
  • Wepsic HT. Alpha-Fetoprotein: Its Quantitation and Relationship to Neoplastic 29. Disease. in Alpha-Fetoprotein, Laboratory Procedures and Clinical Applications.
  • Chen DS, Sung JL. Relationship of Hepatitis B Surface Antigen to Serum Alpha-30. Fetoprotein in Non-Malignant Diseases of the Liver. Cancer 44: 984-992 (1979).
  • Waldmann TA, McIntire KR. Serum Alpha-Fetoprotein Levels In Patients with 31. Ataxia Telangiectasia. Lancet 2: 1112-1115 (1972).
  • Belanger L. Tyrosinemie Hereditaire et Alpha-Foetoproteine. II. Recherche 32. Tissulaire Comparee de L'Alpha-Foetoproteine dans Deux Cas dr Tyrosinemie Hereditaire. Considerations sur L'Ontogense de la Foetoproteine Humaine. Pathol. Biol. 21: 457-462 (1973
  • Kew MC, Purves LR, Bersohn I. Serum Alpha-Fetoprotein Levels in Acute Viral 33. Hepatitis. Gut 14: 939-942 (1973).
  • Endo Y, Kanai K, Oda T, et al. Clinical Significance of Alpha-Fetoprotein in 34. Hepatitis and Liver Cirrhosis. Ann. NY Acad. Sci. 259: 234-238 (1975).

:

  • Purves LR. Purves M. Serum Alpha-Fetoprotein. VI. The Radioimmunoassay 35. Evidence for the Presence of AFP in the Serum of Normal People and During Pregnancy. S. Afr. Med. J. 46: 1290 (1972).
    1. Seppala M. Ruoslahti E. Alpha-Fetoprotein: Physiology and Pathology During Pregnancy and Application to Antenatal Diagnosis. J. Perinat. Med. 1: 104 (1973).
  • Seckl MJ, Rustin GJS, Bagshawe KD. Frequency of Serum Tumour Marker 37. Monitoring in Patients with Non-Seminomatous Germ Cell Tumours. Br. J. Cancer. 61: 916 (1990).
    1. Murphy BA, Motzer RJ, Mazumdar M, et al. Serum Tumor Marker Decline is an Early Predictor of Treatment Outcome in Germ Cell Tumor Patients Treated with Cisplatin and Ifosfamide Salvage Chemotherapy. Cancer. 73: 2520 (1994).
    1. Motawy MS, Szymendera JJ, Al-Jazzaf H, et al. Serum AFP, hCG and CEA in the Management of Patients with Testicular, Ovarian and Extragonadal Germ Cell Tumors. Intl. J. Biol. Markers. 7: 80 (1992).

{29}------------------------------------------------

Image /page/29/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is arranged horizontally, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or informational context.

Image /page/29/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.

SEP 2 2 1997

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Gabriel J. Muraca, Jr. Manager, Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097

Re: K972462 AFP Assay for the Bayer Immuno 1 System Trade Name: Regulatory Class: II Product Code: LOJ Dated: June 30, 1997 Received: July 1, 1997

Dear Mr. Muraca:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{30}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{31}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K973462

Device Name: Bayer Immuno 1TM AFP Assay

Indications For Use:

This in vitro diagnostic assay is intended to quantitatively measure a-fetoprotein (AFP) in human serum on the Bayer Immuno 1TM system as an aid in the management of nonseminomatous testicular cancer. AFP values obtained using the Bayer Immuno 1™ AFP assay method must be interpreted in conjunction with all other available clinical and laboratory data before a medical decision is determined.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) ………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

Concurrence of CDRH, Office of Device Evaluation (ODE)

tetu E. Mdumi

(Division Sign-Off) Division of Clinical Laboratory 510(k) Number

Prescription Use レ (Per 21 CFR 801.109)

OR

Over-the-counter Use

(Optional Format 1-2-96)

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.