(88 days)
The 2.3mm ROC XS Suture Bone Fasteners are intended for the reattachment of soft tissue to bone for the following indications:
ANKLE
- Lateral instability
- Medial instability
- Achilles tendon reconstruction and repair
- Mid-foot reconstructions
HAND
- Ulnar or lateral collateral ligament reconstruction
WRIST
- Scapholunate ligament reconstruction
FOOT
- Hallux valgus reconstruction
ELBOW
- Tennis elbow repair
- Biceps tendon reattachment
- Medial and lateral repairs
The ROC XS suture bone fastener implant tip portion consists of a shear pin, crown and anvil. The crown and anvil are fitted onto the shear pin component such that the crown component is located below the anvil on the shear pin. As the crown is pulled up during the device deployment, the anvil forces it open. The anvil is tapered such that it will fit into the crown component as the deployment action progresses. As the crown is forced open by the anvil, the crown expands to make contact with the surrounding bone. This expanding crown results in the final fixation properties of the device. Once the anvil is completely seated inside the crown, the shearing pin shears from the fastener assembly resulting in the device being fixated into the bone hole.
In addition to the Fastener, a stainless steel drill and drill guide is available to establish the proper hole in the bone for the Fastener along with a deployment handle and hole finder. All of the instrumentation except the Fastener will be offered as reusable devices and can be autoclaved in the sterilization tray provided for this purpose.
The Fastener will be available as a sterile, single use device. Both sutured and sutureless versions will be marketed.
The provided text describes a 510(k) summary for the Innovasive 2.3mm ROC XS Suture Bone Fastener. It details the device, its intended use, and a comparison to predicate devices, along with performance data. However, the information provided does not contain acceptance criteria or detailed study methodologies that would allow for a comprehensive description as requested in your prompt.
Specifically, the document states:
- "Bone model testing: Comparison of the ultimate holding strength in a bone model compared to the predicate devices. The Innovasive 2.3mm ROC XS Suture Bone Fastener holding strength was found to be equivalent to the strength of the predicate devices."
- "Cadaver testing: Comparison of the ultimate holding strength in cadaver compared to the predicate device. Again, the Innovasive 2.3mm ROC XS Suture Bone Fastener holding strength was found to be equivalent to the strength of the predicate devices."
This broadly indicates that the acceptance criteria for these tests were likely based on equivalency to the predicate devices in terms of "ultimate holding strength." However, it does not specify:
- Specific numerical acceptance values or ranges.
- The exact methodology for determining "ultimate holding strength."
- The statistical methods used for "equivalency."
Given the limitations of the provided text, I cannot create the detailed table of acceptance criteria and reported device performance, nor can I provide information on sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.
Therefore, I will outline what is available and explicitly state what information is missing based on your request.
Acceptance Criteria and Study for the Innovasive 2.3mm ROC XS Suture Bone Fastener
The provided 510(k) summary (K972436) for the Innovasive 2.3mm ROC XS Suture Bone Fastener focuses on demonstrating substantial equivalence to predicate devices, primarily through comparison of physical performance characteristics. The core acceptance criterion appears to be equivalency in ultimate holding strength compared to the predicate devices in both bone model and cadaveric testing.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Bone Model Testing: | |
| Ultimate holding strength equivalent to predicate devices (Innovasive 3.5mm ROC XS Suture Bone Fastener, Innovasive 1.9mm/2.3mm ROC Suture Bone Fastener, Mitek Mini Anchor). | Holding strength was found to be equivalent to the strength of the predicate devices. |
| Cadaver Testing: | |
| Ultimate holding strength equivalent to predicate devices (Innovasive 3.5mm ROC XS Suture Bone Fastener, Innovasive 1.9mm/2.3mm ROC Suture Bone Fastener, Mitek Mini Anchor). | Holding strength was found to be equivalent to the strength of the predicate devices. |
Missing Information for this Section:
- Specific numerical thresholds or ranges for "ultimate holding strength."
- The definition of "equivalent" (e.g., within a certain percentage, statistical non-inferiority margin).
- Detailed experimental protocols.
2. Sample Size and Data Provenance
The provided text does not specify the sample size used for either the bone model testing or the cadaver testing.
The data provenance (e.g., country of origin, retrospective/prospective) is not mentioned. These studies would generally be considered prospective laboratory or preclinical studies conducted as part of the device development and regulatory submission process.
3. Number of Experts and Qualifications for Ground Truth
This information is not applicable or not provided as the studies described are biomechanical performance tests comparing holding strength, rather than diagnostic studies requiring expert human readers to establish ground truth for image interpretation or clinical diagnosis. The "ground truth" here is the measured ultimate holding strength of the devices.
4. Adjudication Method for the Test Set
This information is not applicable or not provided for the type of performance testing described. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in diagnostic studies, not for biomechanical strength measurements.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not mentioned and is not applicable to this type of device (suture bone fastener) or the described performance tests. MRMC studies are typically used for evaluating diagnostic imaging devices or AI algorithms that assist human readers.
6. Standalone (Algorithm Only) Performance
Standalone performance, typically referring to an algorithm without human involvement, is not applicable as the device is a physical medical implant, not a software algorithm.
7. Type of Ground Truth Used
The "ground truth" for the described studies was biomechanical measurement of ultimate holding strength. This would involve standardized testing equipment to apply forces and record the point of failure for the device in the bone model or cadaveric bone.
8. Sample Size for the Training Set
The concept of a "training set" is not applicable to the type of device and performance testing described. This device is a physical implant, not an AI/ML algorithm that requires a training dataset.
9. How Ground Truth for the Training Set was Established
As there is no "training set" for this physical device, this question is not applicable.
{0}------------------------------------------------
SEP 2 6 1997
510(K) SUMMARY
K9 72436
1. SUBMITTER:
Innovasive Devices, Inc. 734 Forest St. Marlborough, MA 01752 Telephone: 508-460-8229
Contact: Stephen M. Page, Manager of Regulatory Affairs Date Prepared: June 27, 1997
2. DEVICE:
Innovasive 2.3mm ROC XS Suture Bone Fastener Classification Name: Single/multiple component bone fixation appliances and accessories. Trade Name: Innovasive Devices ROC XS Suture Bone Fastener
3. PREDICATE DEVICES:
The predicate devices used to determine substantial equivalence for the modified Innovasive Devices 2.3mm ROC XS Suture Bone Fastener were:
- The 3.5mm Innovasive Devices ROC XS Suture Bone Fastener, a. marketed by Innovasive Devices, Marlborough, MA,
- The 1.9mm/2.3mm Innovasive Devices ROC Suture Bone Fastener b. marketed by Innovasive Devices, Marlborough, MA, and
- The 1.8mm Mitek Mini Anchor marketed by Mitek Products, 60 Glacier C. Drive, Westwood, MA.
4. DEVICE DESCRIPTION:
The ROC XS suture bone fastener implant tip portion consists of a shear pin, crown and anvil. The crown and anvil are fitted onto the shear pin component such that the crown component is located below the anvil on the shear pin. As the crown is pulled up during the device deployment, the anvil forces it open. The anvil is tapered such that it will fit into the crown component as the deployment action progresses. As the crown is forced open by the anvil, the crown expands to make contact with the surrounding bone. This expanding crown results in the final fixation properties of the device. Once the anvil is completely seated inside the crown, the shearing pin shears from the fastener assembly resulting in the device being fixated into the bone hole.
.
{1}------------------------------------------------
In addition to the Fastener, a stainless steel drill and drill guide is available to establish the proper hole in the bone for the Fastener along with a deployment handle and hole finder. All of the instrumentation except the Fastener will be offered as reusable devices and can be autoclaved in the sterilization tray provided for this purpose.
The Fastener will be available as a sterile, single use device. Both sutured and sutureless versions will be marketed.
5. INTENDED USE:
t
The 2.3mm ROC XS Suture Bone Fasteners are intended for the reattachment of soft tissue to bone for the following indications:
ANKLE
-
Lateral instability
-
Medial instability
-
Achilles tendon reconstruction and repair
-
Mid-foot reconstructions
HAND
- Ulnar or lateral collateral ligament reconstruction
WRIST
- Scapholunate ligament reconstruction
FOOT
- Hallux valgus reconstruction
ELBOW
-
Tennis elbow repair
-
Biceps tendon reattachment
-
Medial and lateral repairs
6. COMPARISON OF CHARACTERISTICS:
The existing Innovasive Devices 3.5mm ROC XS Suture Bone Fastener components are comprised of acetyl plastic. This device is used to secure a suture in a predrilled hole in bone. It remains fixed in the bone through radial compression as the device is deployed. This remains true for all sizes of the device.
20151
{2}------------------------------------------------
The 2.3mm ROC XS device incorporates the same design as the existing 3.5mm ROC XS. The only differences are the size (2.3mm) and the material composition of the anvil (polysulfone). As with the predicate device, the 2.3mm ROC XS uses a crown and anvil fitted on a shear pin such that the crown component is located below the anvil on the shear pin. As the crown is pulled up during device deployment, the anvil forces it open. The shear pin then shears resulting in the device being fixed into the bone site.
7. PERFORMANCE DATA:
The following performance data was provided in support of the substantial equivalence determination:
-
- Bone model testing: Comparison of the ultimate holding strength in a bone model compared to the predicate devices. The Innovasive 2.3mm ROC XS Suture Bone Fastener holding strength was found to be equivalent to the strength of the predicate devices.
-
- Cadaver testing: Comparison of the ultimate holding strength in cadaver compared to the predicate device. Again, the Innovasive 2.3mm ROC XS Suture Bone Fastener holding strength was found to be equivalent to the strength of the predicate devices.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Stephen M. Page Manager of Regulatory Affairs Innovasive Devices, Inc. 734 Forest Street Marlborough, Massachusetts 01752
EP 26 1997
K972438 Re: Innovasive 2.3mm ROC™ XS Suture Bone Anchor Regulatory Class: II Product Code: MBI June 27, 1997 Dated: Received: June 30, 1997
Dear Mr. Page:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2 - Mr. Stephen M. Page
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
P
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE
ANKLE
Lateral instability Medial instability Medial inistability
Achilles tendon reconstruction and repair Midfoot reconstructions
HAND
HAND
Ulnar or lateral collateral ligament reconstruction
WRIST
WKIST
Scapholunate ligament reconstruction
FOOT Hallux valgus reconstruction
ELBOW Tennis elbow repair Biceps tendon reattachment Medial and lateral repairs
bioll
Division Sign-Off
(Division Sign-Off) Division of General Restorative I 510(k) Number
Prescription Use Prescription -----------------------------------------------------------------------------------------------------------------------------------------------------------------
000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
N/A