K Number
K972020
Device Name
OSCILLOMATE NON-INVASIVE BLOOD PRESSURE MONITOR
Date Cleared
1998-05-05

(337 days)

Product Code
Regulation Number
870.1270
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Oscillomate 9001D monitor non-invasively measures the blood pressure of adult and pediatric patients primarily in the emergency care environment. The monitor automatically inflates an occluding cuff and, using the Oscillometric measurement technique, determines systolic, diastolic and pulse rate.
Device Description
The 9001D monitor is compact, lightweight, and durable. The device is housed in and ABS enclosure. The device and all of its accessories are further enclosed in a rugged Cordura nylon carry bag. Power is supplied by an internal rechargeable battery. An external battery charger is provided. Information is displayed in an easy to read LED display. Readings may be taken manually, or at preset intervals from 1 to 60 minutes. A message center display provides information and troubleshooting prompts. A history mode displays previous readings and time readings were taken. The Oscillometric technique monitors the changes of pressure caused by the flow of blood through the artery. The monitor will inflate the cuff around the patients arm to a value that occludes the artery. The monitor then deflates the cuff in steps. As the cuff pressure goes down, blood continues to flow through the artery. The increasing blood flow causes the amplitude of the pressure pulses in the cuff to increase. As the monitor steps down further, the pulses will reach a peak amplitude and then start to decrease with additional pressure steps. The rising and falling of the amplitude of these pressure pulses as the cuff pressure is stepped down, creates a curve that is used to find the systolic pressure and diastolic pressure. Counting the pulses over a time period will give a pulse rate. Motion artifact rejection techniques are used to provide accurate results under most operating conditions.
More Information

Not Found

No
The description details a standard oscillometric blood pressure monitor with motion artifact rejection, which is a common signal processing technique, not necessarily AI/ML. There is no mention of AI, ML, or related concepts in the text.

No.
The device is a monitor that measures blood pressure; it does not provide any therapy or treatment.

Yes

The device measures blood pressure (systolic, diastolic, and pulse rate) using the oscillometric technique. These measurements provide information about the patient's physiological state, which can be used by healthcare professionals to assess health, monitor conditions, and guide diagnosis and treatment, thus qualifying it as a diagnostic device.

No

The device description explicitly mentions hardware components such as an ABS enclosure, internal rechargeable battery, external battery charger, LED display, and an occluding cuff, indicating it is a physical medical device, not software-only.

Based on the provided information, the Oscillomate 9001D is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • IVD Definition: In Vitro Diagnostics are devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health. This testing is performed outside of the body (in vitro).
  • Oscillomate 9001D Function: The Oscillomate 9001D measures blood pressure and pulse rate non-invasively by applying a cuff to the patient's arm. This is a measurement taken on the body (in vivo), not on a specimen taken from the body.
  • Intended Use: The intended use clearly states "non-invasively measures the blood pressure of adult and pediatric patients." This aligns with in vivo measurement.
  • Device Description: The description details the mechanical process of cuff inflation and deflation and the analysis of pressure pulses within the cuff, all of which occur externally to the body.

Therefore, the Oscillomate 9001D is a non-invasive physiological monitoring device, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Oscillomate 9001D monitor non-invasively measures the blood pressure of adult and pediatric patients primarily in the emergency care environment. The monitor automatically inflates an occluding cuff and, using the Oscillometric measurement technique, determines systolic, diastolic and pulse rate.

Product codes (comma separated list FDA assigned to the subject device)

DXW

Device Description

The 9001D monitor is compact, lightweight, and durable. The device is housed in and ABS enclosure. The device and all of its accessories are further enclosed in a ruqged Cordura nylon carry bag. Power is supplied by an internal rechargeable battery. An external battery charger is provided. Information is displayed in an easy to read LED display. Readings may be taken manually, or at preset intervals from 1 to 60 minutes. A message center display provides information and troubleshooting prompts. A history mode displays previous readings and time readings were taken.

The Oscillometric technique monitors the changes of pressure caused by the flow of blood through the artery. The monitor will inflate the cuff around the patients arm to a value that occludes the artery. The monitor then deflates the cuff in steps. As the cuff pressure goes down, blood continues to flow through the artery. The increasing blood flow causes the amplitude of the pressure pulses in the cuff to increase. As the monitor steps down further, the pulses will reach a peak amplitude and then start to decrease with additional pressure steps. The rising and falling of the amplitude of these pressure pulses as the cuff pressure is stepped down, creates a curve that is used to find the systolic pressure and diastolic pressure. Counting the pulses over a time period will give a pulse rate. Motion artifact rejection techniques are used to provide accurate results under most operating conditions.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adult and pediatric patients

Intended User / Care Setting

primarily in the emergency care environment.
The Oscillomate 9001D is compact lightweight and portable, allowing it to be easily transported and used in a variety of clinical settings.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Performance Testing:
The Oscillomate 9001D and the predicate device Propaq 106EL were evaluated in a patient transport environment, specifically in an ambulance, to assess performance under patient motion, road noise, and vibration. The report concluded that the 9001D provides equivalent performance under these conditions.
Accuracy comparison using a patient simulation device: The Oscillomate 9001D met the conditions for acceptance and outperformed the predicate device.
Other non-clinical testing performed includes compliance with UL544, IEC 601-1, IEC 601-1-2, IEC 601-2-30, BS EN 1060-1, ISTA packaging shipping testing, and IEC 68 (series) environmental shock and vibration testing. CAS has passed all submitted testing.

Clinical Performance Testing:
CAS submitted its NB module to AAMI/ANSI SP10: 1992 clinical accuracy testing using a model 9010 monitor as a test platform. The NB module is unchanged from the testing performed. The clinical testing included a clinical trial protocol, data, and analysis of results. The Oscillomate 9001D meets the clinical performance criteria of AAMI/ANSI SP10: 1992.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K910772

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1270 Intracavitary phonocatheter system.

(a)
Identification. An intracavitary phonocatheter system is a system that includes a catheter with an acoustic transducer and the associated device that processes the signal from the transducer; this device records bioacoustic phenomena from a transducer placed within the heart, blood vessels, or body cavities.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows a handwritten number sequence. The sequence appears to be "K972020". The numbers are written in a casual, slightly uneven style, with varying stroke thicknesses.

MAY 5 1008

510(K) SUMMARY

Name of device:

Oscillomate 9001D Non-invasive Blood Pressure Monitor

Submitted by:

CAS Medical Systems, Inc.

May 28, 1997

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Date prepared:

Contact person:

Ron Jeffrey Quality & Regulatory Affairs Manager

1

Device trade name:Oscillomate® 9001D NIBP Monitor
Common name:Blood Pressure Monitor
Classification name:Non-invasive Blood Pressure Measurement System
(870.1130)

CAS is claiming substantial equivalence to the following legally marketed device:

Propag model 106EL Portable Patient Monitor

DEVICE DESCRIPTION

General information:

The 9001D monitor is compact, lightweight, and durable. The device is housed in and ABS enclosure. The device and all of its accessories are further enclosed in a ruqged Cordura nylon carry bag. Power is supplied by an internal rechargeable battery. An external battery charger is provided. Information is displayed in an easy to read LED display. Readings may be taken manually, or at preset intervals from 1 to 60 minutes. A message center display provides information and troubleshooting prompts. A history mode displays previous readings and time readings were taken.

The Oscillometric technique monitors the changes of pressure caused by the flow of blood through the artery. The monitor will inflate the cuff around the patients arm to a value that occludes the artery. The monitor then deflates the cuff in steps. As the cuff pressure goes down, blood continues to flow through the artery. The increasing blood flow causes the amplitude of the pressure pulses in the cuff to increase. As the monitor steps down further, the pulses will reach a peak amplitude and then start to decrease with additional pressure steps. The rising and falling of the amplitude of these pressure pulses as the cuff pressure is stepped down, creates a curve that is used to find the systolic pressure and diastolic pressure. Counting the pulses over a time period will give a pulse rate. Motion artifact rejection techniques are used to provide accurate results under most operating conditions.

2

INTENDED USE

The 9001D is designed to measure the blood pressure of the adult or pediatric patients primarily in the emergency care environment. The monitor automatically inflates an occluding cuff and, using the Oscillometric measurement technique, determines systolic, diastolic, and pulse rate.

The Oscillomate 9001D is compact lightweight and portable, allowing it to be easily transported and used in a variety of clinical settings.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The tables on the following pages compare characteristics between the CAS Oscillomate 9001D and the predicate device "Propaq 106EL". There are no major differences in blood pressure technological characteristics between the two devices.

NOTE: The Propaq 106EL is a multi-parameter monitor with other functions aside from non-invasive blood pressure. The comparison table does not address the other parameters.

NON-CLINICAL PERFORMANCE TESTING AND SUBSTANTIAL EQUIVALENCE

The Oscillomate 9001D and the predicate device Propaq 106EL were evaluated in a patient transport environment that represents their primary area of use. A comparative study was made to evaluate both monitors in an ambulance. The series of tests measured subjects while the ambulance was in motion.

This test includes the protocol, data and final report. See 510(k) appendix 1.

The report concludes that the 9001D provides equivalent performance when encountering patient motion, road noise and vibration conditions in a vehicular transport environment.

The Oscillomate 9001D and the predicate device Propaq 106EL were compared for accuracy using a patient simulation device over a range of input conditions.

The test data concludes that the Oscillomate 9001D met the conditions for acceptance and outperformed the predicate device. See 510(k) appendix 2.

3

CLINICAL PERFORMANCE TESTING AND SUBSTANTIAL EQUIVALENCE

CAS submitted its NB module to AAMI/ANSI SP10: 1992 clinical accuracy testing using a model 9010 monitor as a test platform. The 9010 monitor, originally intended to be a dual parameter monitor with a pulse oximeter, was never developed or marketed. The NB module is unchanged from the testing performed in the study to its new use in the 9001D monitor.

The clinical testing includes a clinical trial protocol, data and analysis of results. See 510(k) appendix 10.

The Oscillomate 9001D meets the clinical performance criteria of AAMI/ANSI SP10: 1992. The predicate device makes the claim that it meets AAMI/ANSI SP10: 1987 standard for clinical accuracy with the NIBP portion of their monitor.

CONCLUSIONS DRAWN FROM CLINICAL AND NON-CLINICAL TESTS

CAS has tested and met all aspects of the AAMI/ANSI SP10: 1992 (510k section 5) which includes overall system efficacy, through clinical trial, environmental performance and stability, safety requirements, and performance requirements. For non-invasive blood pressure monitoring this is the standard to meet.

Other non-clinical testing has been performed as well. Included here are UL544, IEC 601-1, IEC 601-1-2, IEC 601-2-30, BS EN 1060-1, ISTA packaging shipping testing, and IEC 68 (series) environmental shock and vibration testing.

CAS has passed all submitted testing.

4

List of Predicate Device

Monitoring FunctionPredicate DeviceDate of Substantial Equivalence and K Number
Non-Invasive Blood Pressure(NIBP)
using oscillometric methodProtocol Systems, Inc. PROPAQ 100
Series Ultra-Portable Patient MonitorK910772 NIBP Module referenced in K914838 January 10, 1992
FeatureCAS OSCILLOMATEProtocol Propaq 100 seriesDifference and
9001DJustification
l. Intended UsePhysical environmentHospital ER, Clinics, Emergency
Medical ServiceHospital, Air medical/emergency
transport, field medical applicationCAS device is not designed as a
bed-side monitor.
Patient PopulationAdult and PediatricAdult and PediatricNone
2. Monitoring Mode -
using blood pressure
cuffNon-invasive Blood PressureUses single-hose, inflatable cuff to
determine systolic and diastolic.Uses single-hose, inflatable cuff to
determine systolic, diastolic and
mean arterial pressureNone
Pulse RateDetermines pulse rate through cuff
pressure pulsationDetermines pulse rate through cuff
pressure pulsationNone
3. NIBP FunctionsBlood Pressure determining
methodOscillometric, step-wise de-
pressurization of cuffOscillometricNone
Operating modesMANUAL
Automatic cycles of 1, 2, 3, 4, 5,
10, 15, 30, 60, 90 minutesMANUAL
Automatic cycles of 1, 2, 3, 5, 10,
15, 30 and 60 minutes
TURBOCUFF for 5 minutesContinuous (TURBOCUFF)
measurements mode is only a
convenience feature. It does not
provide added benefit to the
patient when health provider is
operating the device next to the
patient in close quarters such as a
field situation.
Typical Determination Time25 seconds15 to 40 secondsSimilar
CalibrationDigital manometer modeOn-screen manometerNone
AlarmsEquipment/Measurement error
indications onlyPatient parameters and
Equipment/Measurement errorsCAS device does not provide
patient parameter alarms for the
simple reason that patients are
closely observed during transport
It is not being used as a bed-side
monitoring device.
4. Blood pressure RangeSystolic25-255 mmHg30-250 mmHgBoth ranges cover majority of
patient population adequately.
FeatureCAS OSCILLOMATE
9001DProtocol Propaq 100 seriesDifference and
Justification
4. CONT.MAP18-255 mmHg25-240 mmHgSame as above
Diastolic10-220 mmHg20-230 mmHgSame as above
5. NIBP System
Efficacy
(Voluntary Performance
Standards)Blood Pressure AccuracyMeets ANSI/AAMI SP-10, 1992.
Mean difference in Systolic(and
Diastolic) pressure comparison: $\pm$ 5mmHg or less with a standard
deviation of 8 mmHg or less.Meets ANSI/AAMI SP-10, 1987.
Mean Systolic(and Diastolic)
pressure comparison: $\pm$ 5mmHg or
less with a standard deviation of 8
mmHg or less.None
6. Pulse Rate RangeDesign specification40-240 beats per minute25-200 bpmBoth ranges cover majority of
patient population adequately.
7. Pulse Rate AccuracyDesign specification$\pm$ 2 bpm or $\pm$ 2%, whichever is
greaterwithin $\pm$ of 6 bpm or 6%,
whichever is greaterNo significant difference
8. Measurements
StorageDesign specificationStores measurements within the
last 5 hours of up to 99 cyclesUp to 128 readings that are less
than 8 hours oldNo significant difference
9. AC Power AdapterInput RequirementOptional 120 VAC or 100-240
VACOptional 100 VAC, 120 VAC or
220-240 VACNo significant difference
10. Battery PowerBattery TypeRechargeable sealed lead acid, 6V
3.2AHRechargeable sealed lead acid, 8V
3AHNo significant difference
Low, or Dead Battery AlertDisplays warning on battery
condition; does not make
measurement in dead battery
conditionDisplays LOW BATT alarmNo significant difference
Charge Time for a depleted
pack12 hours~8 hoursBoth provide overnight charging
method.
Operating Time on a fully
charged pack300 reading at 1-minute automatic
cycles; 7 hour continuous6.5 hours, NIBP every 15 minutesCAS device is dedicated to BP
measurement only. It has more
reserved power to run
electromechanical components
11. Alarm IndicatorsVisualBlinking 8-character message
displayRED LED and on-screen messageNo significant difference
AuralShort Audio beeps for alertBeeperNo significant difference
12. Physical
CharacteristicsMeasurement Data DisplayRED 8-segment LED'sBacklit LCD or EL 276x128 pixel
graphical displayUser preference item. No
graphical display for NIBP
Message DisplayRED 8-character dot-matrix LED
message centersame as aboveSame as above
Unit Size (H x W x D)7.5" x 8" x 5"6.6" x 8.3" x 4.8"No significant difference
FeatureCAS OSCILLOMATE 9001DProtocol Propaq 100 seriesDifference and Justification
12. CONT.Unit Weight4.5 lb.5.8 lb..No significant difference
Unit mounting/carryingA padded, fabric carrying bag is an integral part of the monitor, battery and cuff storage system.Unit with carrying handle, optional carrying bag for transport.CAS device is a more convenient and complete package for the intended environment.
13. EnvironmentalOperating AltitudeTested to -500 to +10,000 ft. Measurement accuracy is referenced to local atmospheric pressure and will not be affected by operating altitude-1,000 to +15,000 ft.No significant difference
Operating Temperature0° to 50° C0° to 50° CNone
Operating Humidity15 to 90%(non-condensing)0 to 97%(non-condensing)No significant difference
Storage Temperature-20° to 70° C-20° to 60° CNo significant difference
Storage Humidity15 to 95%0 to 97%No significant difference
Electromagnetic CompatibilityEN 60601-1-2
IEC 801-1-2; IEC 801-1-3
IEC 801-1-4; IEC 801-1-5
CISPR - 11IEC-801-2, level 4 ESD
VDE 0871 class B EMI
FDA MDS-201-0004(emissions only)CAS elected to use the current International Standards.
Shock & Vibration ResistanceIEC 68-2-27; SHOCK
IEC 68-2-6; Sin. VIBRATION
IEC 68-2-34; RANDOM VIBRATION, wide bandMIL-T-28800CAS elected to use FDA's Performance Standard for Infant Apnea Monitor as guidance for S&V resistant testing
14. Safety
CharacteristicsElectricalUL544, EN 60601-1UL, CSA, TUV and othersNone
Cuff Over-Pressurization limitsSoftware Limit: 290 mm Hg
Redundant switch: 330mmHg
(Meets IEC 601-2-30, AAMI SP-10)Software Limit: 260 mmHg
Redundant switch: 330mmHg
(Meets IEC 601-2-30, AAMI SP-10)No significant difference
Prolonged Cuff Inflation time-outUp to 120 seconds (Meets IEC 601-2-30, AAMI SP-10)Up to 180 seconds (Meets IEC 601-2-30, AAMI SP-10)CAS uses a shorter time for patient comfort

5

--

6

--

7

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

5 1898 MAY

Mr. Ron Jeffrey Quality Manager CAS Medical Systems, Inc. Technology Applied to Medicine 21 Business Park Drive Branford, CT 06405

Re : K972020 OSCILLOMATE 9001D NON-Invasive Blood Pressure Monitor Regulatory Class: II (Two) Product Code: DXW Dated: April 17, 1998 Received: April 28, 1998

Dear Mr. Jeffrey:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any. obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

8

Page 2 - Mr. Ron Jeffrey

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

9

Device Name: OSCILLOMATE 9001D NON-INVASIVE BLOOD PRESSURE MONITOR

Indications For Use:

The Oscillomate 9001D monitor non-invasively measures the blood pressure of adult and I he Oschnoliate 2001D inomer non agency care environment. The monitor automatically peniatic patients primarily in also entrigens in the Oscillometric measurement technique, determines systolic, diastolic and pulse rate.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bete Lumperle
(Division Sign-Off)

Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number _

Prescription Use
(Per 21 CFR 801.109)

OR

Artistics

Over-The-Counter Use

(Optional Format 1-2-96)

.