K Number
K970965

Validate with FDA (Live)

Date Cleared
1997-04-09

(23 days)

Product Code
Regulation Number
874.4250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Midas Rex III pneumatic motor is designed for used in Ear, Nose and Throat, Otolaryngology and Maxillofacial specifications.

Device Description

Ergonomically designed motor, 62 gram in weight, 15 mm in diameter with an overall length of approximately 10.25 centimeters. The Midas Rex III motor operates at variable speeds from 25 to 70 thousand rpm's. Operating pressures range from 20-150 psi (2-10 bar), with a suggested operating pressure of 100-psi (6.9 bar) and approximately 60 thousand rpm. The Midas Rex III has a standard quick disconnect GS dissecting tool and tube attachment system, which includes a fixed or rotating footed attachment, the telescoping micro systems, and the micro mini drill guide attachment.

AI/ML Overview

The provided document is a 510(k) summary for a medical device submitted in 1996. This type of document is for premarket notification to the FDA for medical devices and primarily focuses on demonstrating substantial equivalence to a predicate device already on the market.

It is highly unlikely that a 510(k) summary from 1996 for a pneumatic cranial drill motor will contain acceptance criteria or a detailed study proving performance in the manner you've requested. The information you're asking for (e.g., sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance of AI, specific ground truth types) is standard for performance evaluations of AI/ML-driven medical devices, which became prevalent much later than 1996.

Therefore, based on the provided text, I cannot extract the information you've requested. The document describes a physical medical device (a surgical drill motor) and its characteristics, not an AI/ML algorithm.

However, I can explain why the information isn't present based on the nature of the document:

  • Acceptance Criteria & Reported Performance Table: For mechanical devices like this, acceptance criteria would typically involve engineering specifications (e.g., RPM range, torque, pressure tolerance, material strength, sterility) and performance would be reported against these in internal testing or design validation, but not usually detailed in a public 510(k) summary in this format. The 510(k) focuses on showing it's "as safe and effective as" a predicate, not necessarily detailed performance metrics.
  • Sample Size for Test Set & Data Provenance: These concepts are relevant for statistical studies involving data, particularly in AI/ML, not for verifying the mechanical operation of a drill motor in the way described here.
  • Number of Experts, Qualifications, Adjudication Method: These are crucial for establishing ground truth in diagnostic or image analysis studies. While experts might be involved in evaluating the usability or safety of a surgical tool, it's not the "ground truth" establishment in the AI sense.
  • MRMC Comparative Effectiveness Study: This is specific to evaluating diagnostic AI algorithms against human readers. It's entirely inapplicable to a surgical drill motor.
  • Standalone Performance: This term is for AI algorithms operating without human intervention. Again, not relevant to a physical surgical tool.
  • Type of Ground Truth Used: For AI, this refers to the definitive labels (e.g., pathology report, clinical outcome). For a drill, "ground truth" would be its engineering specifications and successful (safe, effective) operation in a surgical context, evaluated through validation studies.
  • Sample Size for Training Set & Ground Truth for Training Set: These are concepts exclusively for machine learning model development.

In summary, the provided 510(k) document describes a traditional medical device (a surgical drill motor) and does not contain the information requested about acceptance criteria and study details relevant to AI/ML device performance evaluation.

{0}------------------------------------------------

APR - 9 1997

510 (K) SUMMARY OF SAFETY AND EFFECTIVEN

Date prepared:

Submitter:

Contact Person:

March 10. 1996

Midas Rex. L. P. 3001 Race Street Fort Worth, Texas 76111 (817) 831-4181

Michael Fowler Director Regulatory Affairs

Midas Rex Motor (Midas III)

Pneumatic Cranial Drill Motor 874.4250 77 ERL Class II

Midas Rex III motor, attachments and accessories Midas GS and GS20 Surgical Tool System Midas Drill Guide Attachment Midas Telescoping System Dissecting tools K962483 class II

Ergonomically designed motor, 62 gram in weight, 15 mm in diameter with an overall length of approximately 10.25 centimeters. The Midas Rex III motor operates at variable speeds from 25 to 70 thousand rpm's. Operating pressures range from 20-150 psi (2-10 bar), with a suggested operating pressure of 100-psi (6.9 bar) and approximately 60 thousand rpm. The Midas Rex III has a standard quick disconnect GS dissecting tool and tube attachment system, which includes a fixed or rotating footed attachment, the telescoping micro systems, and the micro mini drill guide attachment.

Trade (proprietary) Name:

Common / Classification Name: Regulation no. Product code Device Classification:

Predicate Device:

Description of the Device:

{1}------------------------------------------------

Statement of Intended Use:

Technological Characteristics:

The Midas Rex III pneumatic motor is designed for used in Ear, Nose and Throat, Otolaryngology and Maxillofacial specifications.

The Midas Rex III motor, attachments and accessories have the same technological characteristics (i.e., design, material, energy source) as the predicate devices.

মা

§ 874.4250 Ear, nose, and throat electric or pneumatic surgical drill.

(a)
Identification. An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.(b)
Classification. Class II.