K Number
K970877

Validate with FDA (Live)

Device Name
IMMULITE CEA
Date Cleared
1997-09-22

(196 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DPC's IMMULITE CEA is intended for use with the IMMULITE Automated Immunoassay Analyzer. The IMMULITE CEA is a solid-phase, chemiluminescent enzyme immunoassay designed for the quantitative detection of carcinoembryonic antigen in human serum. It is intended strictly for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis.

Device Description

IMMULITE CEA is a clinical device for use with the IMMULITE Automated Immunoassay Analyzer. IMMULITE CEA is designed for the quantitative detection of carcinoembryonic antigen in human serum. It is intended strictly for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis. DPC's IMMULITE CEA is an in vitro diagnostic medical device for use with DPC's IMMULITE Automated Immunoassay Analyzer, a random access instrument which performs chemiluminescent immunoassays. The assay is designed for the quantitative measurement of carcinoembryonic antigen (CEA) in serum and is intended as an aid in the management of patients with cancer and in the assessment of prognosis. The IMMULITE CEA assay is a solid-phase, two-site sequential chemiluminescent immunometric assay. The solid phase consists of a polystyrene bead (coated with a monoclonal antibody specific for CEA) which is enclosed within an IMMULITE Test Unit (LCE1) which acts as a reaction vessel. The patient serum sample (or CEA Adjustors, LCEL and LCEH) and reagent (LCEA, buffer/serum matrix, with preservative) are incubated for approximately 30 minutes at 37℃ in the Test Unit. With intermittent agitation, CEA in the sample becomes bound to the surface of the bead. Unbound serum is then removed by a centrifugal wash. A second reagent (LCEB, an alkaline phosphatase-labeled anti-ligand) is introduced and the Test Unit is incubated for an additional 30 minutes. Unbound conjugate is removed by a centrifugal wash, after which a chemiluminescent substrate (LSUB, a phosphate ester of adamantyl dioxetane) is added and the Test Unit is incubated for a further 10 minutes. The substrate undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light. The bound complex (and thus also the photon output as measured by the luminometer) is proportional to the concentration of CEA in the sample. The concentration of CEA in the patient sample is obtained using a stored master calibration curve within the IMMULITE analyzer. The IMMULITE CEA assay has a calibration range up to 550 nanograms of CEA per milliliter.

AI/ML Overview

The provided text describes the IMMULITE® CEA device, a chemiluminescent immunoassay for quantitative detection of carcinoembryonic antigen (CEA) in human serum, intended as an aid in cancer patient management and prognosis assessment. The submission is a 510(k) summary for premarket notification, asserting substantial equivalence to the Abbott IMx® CEA kit.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list "acceptance criteria" in a numerical or categorical format. Instead, the basis for acceptance is stated as "substantial equivalence" to a predicate device. The performance is demonstrated through a method comparison study.

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
Qualitative Equivalence:
Intended UseIMMULITE® CEA: Aid in management of cancer patients and assessment of prognosis. (Matches predicate)
Quantitative detection of CEA in human serumIMMULITE® CEA: Quantitative detection of CEA in human serum. (Matches predicate)
Quantitative Equivalence:
Correlation with predicate device (Abbott IMx® CEA)Linear regression: IMMULITE® = 1.19 * IMx® - 0.67 ng/mL
Correlation coefficient (r) with predicate0.987
Calibration RangeIMMULITE® CEA: Up to 550 nanograms of CEA per milliliter. (Implied to be comparable or better than predicate, as 374 specimens were within the calibration range of both assays.)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: 374 specimens
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). It is simply referred to as "specimens."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

This information is not provided in the document. The "ground truth" for the method comparison study was the results obtained from the predicate device, Abbott IMx® CEA.

4. Adjudication Method for the Test Set:

This information is not applicable/provided. The comparison was between two automated assay systems, not involving human expert adjudication in the traditional sense of diagnostic imaging or clinical assessment. The predicate device's results served as the reference.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an in vitro diagnostic immunoassay, not an AI-assisted diagnostic imaging tool that involves human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, essentially. The study described is a comparison of the IMMULITE® CEA assay (an automated system) against the Abbott IMx® CEA assay (also an automated system). Both systems operate without "human-in-the-loop" performance in the interpretive sense, although human operators perform the tests. The performance metrics (linear regression, correlation coefficient) reflect the standalone analytical performance of the IMMULITE® CEA device compared to an established method.

7. The Type of Ground Truth Used:

The "ground truth" for this comparative study was the measurement of CEA in human serum using the legally marketed predicate device, Abbott IMx® CEA. This is a form of "reference method" comparison.

8. The Sample Size for the Training Set:

This information is not provided in the document. The summary focuses on the comparative study for substantial equivalence, not the internal development or training of the IMMULITE® CEA assay.

9. How the Ground Truth for the Training Set was Established:

This information is not provided. As above, the document details the comparative study against a predicate device, not the development process of the IMMULITE® CEA assay itself, which would involve its own calibration and validation.

{0}------------------------------------------------

K97087

Diagnostic Products Corporation IMMULITE CEA

SEP 22 1997

510 (k) Summary Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the reguirements of SMDA 1990 and 21 CFR Part 807.92.

Name: Address:

Telephone Number:

Contact Person:

Date of Preparation:

Device Name: Trade: Catalog Number: Common:

Classification:

Manufacturer:

Establishment Registration #:

Substantially Equivalent Predicate Device:

Description of Device:

Intended Use of the Device:

Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045

(213) 776-0180

Edward M. Levine, Ph.D.

March 7, 1997

IMMULITE® CEA LKCE1 (100 tests); LKCE5 (500 tests) Reagent system for the determination of carcinoembryonic antigen in human serum.

Class II device (866.3780)

Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045

DPC's Registration # is 2017183

Abbott Laboratories' IMx® CEA

IMMULITE CEA is a clinical device for use with the IMMULITE Automated Immunoassay Analyzer.

IMMULITE CEA is designed for the quantitative detection of carcinoembryonic antigen in human serum. It is intended strictly for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis.

{1}------------------------------------------------

Summary and Explanation of the Test:

Carcinoembryonic antigen (CEA) comprises a heterogeneous family of glycoproteins ranging in weight from 175,000 to 200,000 daltons due to varying carbohydrate and amino acid content. Its biological function is not well defined, but it may play a role in intercellular recognition, regulation of the immune response, and metastasis of colorectal cancer. The name derives from the previous belief that CEA occurred only in gastrointestinal carcinomas and the fetal digestive tract. Elevated levels have since been detected in a number of malignant and nonmalignant conditions of the gastrointestinal tract and other sites. These conditions include various hepatic diseases: inflammatory lessons, particularly of the gastrointestinal tract: infections, trauma, infarction, collagen vascular disease, renal impairment, and smoking (current and past). CEA also occurs at low levels in normal colon and other tissues. Serum values in healthy adults are generally less than 5 ng/mL. I Though serum values exceeding five times the normal range usually indicate malignancy, values seen in malignant and nonmalignant conditions can overlap considerably, ruling out the use of CEA as a screen for malignancy. The value of CEA measurement lies, instead, in patient prognosis, status assessment, and monitoring.

Until 1980, the only assay method for CEA used in the United States was a radioimmunoassay requiring extraction. Since then, several other nonextraction RIAs and sandwich-type EIAs have been developed for use on serum and plasma samples.

CEA levels at diagnosis of colorectal carcinoma (CRC) correlate with prognosis. Elevated preoperative serum CEA predicts increased risk of recurrence or hepatic metastases. The site of the primary cancer may influence interpretation of elevated CEA: in colon carcinomas, high CEA levels indicate a poor prognosis, but not necessarily so in rectal carcinomas. In addition, monitoring levels during chemotherapy before surgery can be informative, and failure of CEA to fall during preoperative radiotherapy usually indicates the presence of a turnor outside the field of radiation and a poorer prognosis. Preoperative elevated CEA has also been observed to correlate inverselv with estimated mean time to recurrence, and to correlate directly with degree of tumor stage, thickness, and differentiation in CRC patients at diagnosis. CEA is not always elevated, however, in sera of those with poorly differentiated CRCs, which produce less CEA but which are more aggressive than moderately or well-differentiated carcinomas. Rising CEA is deemed the most accurate indicator of recurrence in CRC. Levels decrease to normal in nearly all patients after complete resection of CRCs, usually within 4 to 6 weeks after surgery. Failure to decline after surgery may suggest incomplete resection, and increasing levels can precede recurrent CRCs an average of 4 to 6 months before they are clinically evident in about two-thirds of cases. Rising CEA levels have also been used to assist in selecting candidates for second-look surgery, in conjunction with antibody imaging using radiolabeled antibodies against CEA to provide information on tumor location. Several studies have suggested a correlation between the rate of rise in CEA and the presence or probability of hepatic metastases; more rapid rises are seen with hepatic metastases than with localized recurrences and resectable localized tumors. In some cases, it has been CEA that signaled the asymptomatic recurrences.

{2}------------------------------------------------

Summary and Explanation of the Device:

DPC's IMMULITE CEA is an in vitro diagnostic medical device for use with DPC's IMMULITE Automated Immunoassay Analyzer, a random access instrument which performs chemiluminescent immunoassays. The assay is designed for the quantitative measurement of carcinoembryonic antigen (CEA) in serum and is intended as an aid in the management of patients with cancer and in the assessment of prognosis. The IMMULITE CEA assay is a solid-phase, two-site sequential chemiluminescent immunometric assay. The solid phase consists of a polystyrene bead (coated with a monoclonal antibody specific for CEA) which is enclosed within an IMMULITE Test Unit (LCE1) which acts as a reaction vessel. The patient serum sample (or CEA Adjustors, LCEL and LCEH) and reagent (LCEA, buffer/serum matrix, with preservative) are incubated for approximately 30 minutes at 37℃ in the Test Unit. With intermittent agitation, CEA in the sample becomes bound to the surface of the bead. Unbound serum is then removed by a centrifugal wash. A second reagent (LCEB, an alkaline phosphatase-labeled anti-ligand) is introduced and the Test Unit is incubated for an additional 30 minutes. Unbound conjugate is removed by a centrifugal wash, after which a chemiluminescent substrate (LSUB, a phosphate ester of adamantyl dioxetane) is added and the Test Unit is incubated for a further 10 minutes. The substrate undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light. The bound complex (and thus also the photon output as measured by the luminometer) is proportional to the concentration of CEA in the sample. The concentration of CEA in the patient sample is obtained using a stored master calibration curve within the IMMULITE analyzer. The IMMULITE CEA assay has a calibration range up to 550 nanograms of CEA per milliliter.

Performance Equivalence - Technology Comparison:

Diagnostic Products Corporation (DPC) asserts that IMMULITE® CEA is substantially equivalent to the IMx® CEA kit marketed by Abbott Laboratories (Abbott Park, IL).

Each product is designed for the quantitative measurement of carcinoembryonic antigen (CEA) in serum and is intended as an aid in the management of patients with cancer and in the assessment of prognosis.

IMMULITE® CEA is a chemiluminescent enzyme immunoassay and IMx® CEA is a microparticle enzyme immunoassay (MEIA). The technology in DPC's IMMULITE CEA is identical to technology used in previously cleared and commercially marketed IMMULITE® products.

In the IMx CEA assay, the patient sample and Anti-CEA coated microparticles are added to the incubation well of a reaction cell. CEA in the specimen binds to the anti-CEA coated microparticles forming an antibody-antigen complex. An aliquot of the antigen-antibody complex is transferred to the glass fiber matrix. The microparticles bind irreversibly to the glass fiber matrix. The matrix is washed to remove unbound materials. The anti-CEA/alkaline phosphatase conjugate is dispensed onto the matrix and binds to the antigen-antibody complex. Finally, the matrix is washed to remove unbound materials, and the substrate, 4-Methylumbelliferyl Phosphate, is added to the matrix, and the fluorescent product is measured by the optical assembly.

{3}------------------------------------------------

Diagnostic Products Corporation IMMULITE CEA

Performance Equivalence - Method Comparison:

The clinical performance of the IMMULITE CEA procedure was compared to the Abbott IMx® CEA. Linear regression analysis of the 374 specimens that were within the calibration range of both the IMMULITE and IMx® assays yielded:

IMMULITE® = 1.19 * IMx® - 0.67 ng/mL

with a correlation coefficient (r) of 0.987.

Conclusion:

The conclusions drawn from the clinical and nonclinical studies demonstrate that the device is safe, effective, and performs as well as, or better, than the current legally marketed device.

Gd. Vriel H. Kaine

3/7/97

Date

Edward M. Levine, Ph.D. Director of Clinical Affairs

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping lines forming the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 2 2 1997

Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045

Re: K970877 Trade Name: IMMULITE® CEA Requlatory Class: II Product Code: DHX Dated: July 2, 1997 Received: July 3, 1997

Dear Dr. Levine:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): K970877___________________________________________________________________________________________________________________________________________

Device Name: IMMULITE CEA ____________________________________________________________________________________________________________________________________________________

Indications For Use:

t

DPC's IMMULITE CEA is intended for use with the IMMULITE Automated Immunoassay Analyzer. The IMMULITE CEA is a solid-phase, chemiluminescent enzyme immunoassay designed for the quantitative detection of carcinoembryonic antigen in human serum. It is intended strictly for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Clinical Laboratory Devices

510(k) NumberK970877
------------------------
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use
--------------------------------------------------------------------

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.