K Number
K970424
Date Cleared
1997-03-06

(30 days)

Product Code
Regulation Number
862.1155
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Excel OneStep Urine hCG Pregnancy DipStick Test is Excel Scientific, Inc.'s name for professional use of solid Ekol of one of the more regioner in vitro diagnostic test kit for the qualitative detection of human phonic Gonadotropin (hCG) in urine.

Device Description

Excel OneStep Urine hCG Pregnancy DipStick Test is Excel Scientific, Inc.'s name for professional use of solid Ekol of one of the more regioner in vitro diagnostic test kit for the qualitative detection of human phonic Gonadotropin (hCG) in urine. Excel OneStep Urine hCG Pregnancy DipStick Test is based on comparative data with Syntron Quikpac EIA, an assay for human Chorionic Gonadotropin currently being marketed.

The Principle of Excel OneStep Urine hCG Pregnancy DipStick Test is as follows:

  • Both the Excel OneStep Urine hCG Pregnancy DipStick and the QuikStrip (or E.Z. Strip) Pregnancy Home 1. tests are a sandwich solid phase dye-conjugated non-enzyme immunoassay. When the Excel OneStep Urine hCG Pregnancy DipStick is placed in a urine test cup containing the urine specimen, the urine will migrate by capillary action. The gold sol-antibody conjugate in the gold sol pad of the Excel OneStep Urine hCG Pregnancy DipStick binds to the hCG molecules in the specimen forming an antibody-antigen complex. This complex binds to the anti-hCG antibody immobilized in the positive reaction zone in the white membrane area and produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL. In the absence of detectable levels of hCG, there is no rose-pink band in the positive reaction zone. The reaction mixture continues to flow through the absorbent device, pass the positive reaction zone, producing a rose-pink color band in the control zone of the white membrane area, demonstrating that the reagents and the device are functioning correctly.
AI/ML Overview

This document describes the Excel OneStep Urine hCG Pregnancy DipStick Test, a qualitative in vitro diagnostic test for human chorionic gonadotropin (hCG) in urine. The study presented compares its performance to the Syntron Quikpac EIA, a currently marketed hCG assay.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for the Excel OneStep device in a quantitative manner (e.g., sensitivity, specificity thresholds). However, it establishes a comparison to an existing, marketed device, the Syntron Quikpac EIA. The implication is that the Excel OneStep device should perform comparably to the Syntron Quikpac EIA, particularly in its cutoff sensitivity.

Feature/MetricAcceptance Criteria (Implied by Comparison to Syntron Quikpac EIA)Reported Device Performance (Excel OneStep Urine hCG Pregnancy DipStick Test)
Cutoff25 mIU/mL (as established by the Syntron Quikpac EIA)25 mIU/mL (produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL)
SensitivityQualitatively detect hCG at or above 25 mIU/mL. The document indicates that the Excel OneStep "is based on comparative data with Syntron Quikpac EIA, an assay for human Chorionic Gonadotropin currently being marketed," implying the Excel OneStep should demonstrate equivalent qualitative detection at this cutoff.The device "produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL." This indicates it meets the established cutoff for detection. The document doesn't provide specific sensitivity or specificity percentages in a table, but the comparison to Syntron Quikpac EIA serves as the primary performance benchmark. The text states: "When the Excel OneStep Urine hCG Pregnancy DipStick is placed in a urine test cup containing the urine specimen, the urine will migrate by capillary action... This complex binds to the anti-hCG antibody immobilized in the positive reaction zone in the white membrane area and produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL."
Reaction Time5 minutes (as specified for the "QuikStrip (or E.Z. Strip) Pregnancy Home test" which is presented as similar in procedure to the Excel OneStep, and a general operational characteristic).5 minutes
Test Procedure"No difference in the testing procedure between the Excel OneStep Urine hCG Pregnancy DipStick and the QuikStrip (or E.Z. Strip) Pregnancy Home test." Implies simplicity and ease of use, maintaining a "one step" process. (Though a distinction is made for the Syntron Quikpac EIA which has 3 steps)."one step" (confirms the simple procedure, aligning with the QuikStrip/E.Z. Strip comparison).

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document does not specify the sample size used for the comparative study or test set.
  • Data Provenance: The document states the study is "based on comparative data with Syntron Quikpac EIA," but does not mention the country of origin of the data or whether it was retrospective or prospective. It appears to be an internal comparison or validation study, not a large-scale clinical trial with detailed demographics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • The document does not provide information about the number or qualifications of experts used to establish ground truth for the test set.

4. Adjudication Method for the Test Set

  • The document does not describe any adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is a qualitative diagnostic test (a dipstick for visual detection of a rose-pink band). It does not involve AI, human readers interpreting complex images, or an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This is a standalone diagnostic test in the sense that the device itself produces the result (a color change). There is no "algorithm only" performance separate from the physical dipstick reaction, as it's a chemical immunoassay, not an AI or software-based algorithm. The interpretation is visual by a human, but the test itself is complete upon color change.

7. The Type of Ground Truth Used

  • The ground truth is implied to be actual hCG concentrations in urine specimens, as demonstrated by a qualitative cutoff of "25 mIU/mL". This is a biochemical or analytical ground truth (presence/absence of hCG at a specific threshold) rather than expert consensus on images or pathology. The performance is gauged against an existing, marketed device, the Syntron Quikpac EIA, which presumably has its own established analytical performance standards.

8. The Sample Size for the Training Set

  • The document describes a comparative study, not a machine learning study that would typically involve distinct training and test sets. Therefore, there is no mention of a training set sample size.

9. How the Ground Truth for the Training Set was Established

  • As there's no mention of a training set, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "EXCEL" in white letters on a black rounded rectangle. Below the rectangle, the phrase "Commitment to Excellence" is written in a smaller font. The text and rectangle appear to be part of a logo or branding material.

Incubation

Reagent

Step

Cutoff

Urine Cup

Attachment #4

Summary of Safety and Effectiveness

MAR - 6 1997

Excel OneStep Urine hCG Pregnancy DipStick Test

Excel OneStep Urine hCG Pregnancy DipStick Test is Excel Scientific, Inc.'s name for professional use of solid Ekol of one of the more regioner in vitro diagnostic test kit for the qualitative detection of human phonic Gonadotropin (hCG) in urine. Excel OneStep Urine hCG Pregnancy DipStick Test is based on comparative data with Syntron Quikpac EIA, an assay for human Chorionic Gonadotropin currently being marketed.

The Principle of Excel OneStep Urine hCG Pregnancy DipStick Test is as follows:

  • Both the Excel OneStep Urine hCG Pregnancy DipStick and the QuikStrip (or E.Z. Strip) Pregnancy Home 1. tests are a sandwich solid phase dye-conjugated non-enzyme immunoassay. When the Excel OneStep Urine hCG Pregnancy DipStick is placed in a urine test cup containing the urine specimen, the urine will migrate by capillary action. The gold sol-antibody conjugate in the gold sol pad of the Excel OneStep Urine hCG Pregnancy DipStick binds to the hCG molecules in the specimen forming an antibody-antigen complex. This complex binds to the anti-hCG antibody immobilized in the positive reaction zone in the white membrane area and produces a rose-pink color band when the hCG concentration is at or greater than 25 mIU/mL. In the absence of detectable levels of hCG, there is no rose-pink band in the positive reaction zone. The reaction mixture continues to flow through the absorbent device, pass the positive reaction zone, producing a rose-pink color band in the control zone of the white membrane area, demonstrating that the reagents and the device are functioning correctly.
Test and the Syntron's QuikStrip (or E.Z. Strip) Pregnancy Home test as shown below :
QuikStrip (or E.Z. Strip)Pregnancy Home testExcel OneStep Urine hCGPregnancy DipStick Test
Specimen Volume6 drops6 drops of urine

5 minutes

one step

25 mIU/mL

. . . . . . . . . . . . . . . . . . . . . .

No

Yes

5 minutes

one step

25 mIU/mL

No

Yes

  • There is no difference in the testing procedure between the Excel OneStep Urine hCG Pregnancy DipStick 2.
3. The differences between the Excel OneStep Urine hCG Pregnancy DipStick Test and the Syntron Quikpa
EIA Pregnancy Test are as follows :
Syntron Quikpac EIA PregnancyTestExcel OneStep Urne hCGPregnancy DipStick Test
Specimen Volume3 drops of urine6 drops of urine
Incubation3 minutes with antibody and 3minutes with substrate5 minutes
Reagent2 bottlesNo
Step3 stepsone step
Cutoff25 mIU/mL25 mIU/mL

....

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "EXCEL" in a bold, sans-serif font inside a black rounded rectangle. Below the rectangle, the phrase "Commitment to Excellence" is written in a smaller, italicized font. The text is black against a white background.

  • For questions concerning the safety and effectiveness of the Excel OneStep Urine hCG Pregnancy DipStick, 4. please contact:
    Min-Lee Cheng, Ph.D. 23071 E. La Palma Avenue Yorba Linda, CA 92887 Tel: (714) 692-7151

Sales (1) 10.

,

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.