(82 days)
The Innovasive Devices 8mm Ligament Fastener is intended for ligament to bone reattachment of Anterior and Posterior Cruciate Ligaments in patients with recurrent instability of the knee.
The Innovasive 8mm Ligament Fastener is an implantable tendon fixation device complete with the instrumentation needed to effect ligament to bone repair. The devices consist of a high density polyethylene sleeve with an acetal drive pin which is deployed into a pre-drilled hole in the bone. The Fastener is deployed through the use of a hex driver, for screwing in the sleeve, and an inserter which drives the pin into the sleeve.
In addition to the Fastener, a stainless steel drill is available to establish the proper hole in the bone for the Fastener. All of the instrumentation except the Fastener will be offered as reusable devices and be autoclaved in the sterilization tray provided for this purpose.
The Fastener will be available as a sterile, single use device for use in both an open and arthroscopic procedures.
Here's an analysis based on the provided text, focusing on the acceptance criteria and the study that demonstrates the device meets these criteria:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Ultimate holding strength in bone equivalent to predicate device | Innovasive Fastener holding strength found to be equivalent to the strength of the predicate device at a .01 significance level |
2. Sample Size Used for the Test Set and Data Provenance
The text mentions a "Cadaver study," implying human cadaveric tissue was used. However, the exact sample size (number of cadavers or individual tests) and the country of origin are not specified. It is a retrospective study in the sense that it uses pre-existing biological material (cadavers) rather than living patients.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
The document does not mention the use of experts to establish ground truth for the test set. The ground truth for holding strength appears to be derived from direct physical measurements in the cadaver study, rather than expert interpretation of data.
4. Adjudication Method for the Test Set
No adjudication method is mentioned, as expert consensus or interpretation was not the primary means of establishing ground truth for holding strength. The outcome was a direct measurement.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for diagnostic imaging or interpretation tasks where human readers' performance is being evaluated with and without AI assistance. This device is a physical implant, and its performance is measured mechanically.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
Yes, a standalone study was done in the sense that the device's mechanical performance (holding strength) was directly measured without continuous human intervention during the measurement process, beyond setting up the experiment. There is no "algorithm" in the context of this mechanical device.
7. Type of Ground Truth Used
The type of ground truth used was direct mechanical measurement of ultimate holding strength in cadaveric bone. This is a form of empirical or objective data rather than expert consensus, pathology, or outcomes data in the usual medical imaging or diagnostic sense.
8. Sample Size for the Training Set
The document does not mention a training set or any machine learning algorithm. The "study" described is a direct comparative mechanical test, not an AI-driven system requiring training data.
9. How the Ground Truth for the Training Set Was Established
Since there is no training set mentioned, the question of how its ground truth was established is not applicable.
{0}------------------------------------------------
K970316 510(K) SUMMARY
1. SUBMITTER:
Innovasive Devices, Inc. 734 Forest Street Marlboro, MA 01752 Telephone: 508-460-8229
APR 1 6 1997
Contact: Eric Bannon, V.P. of Quality Assurance and Regulatory Affairs
Date Prepared: April 14, 1997
2. DEVICE:
Innovasive 8 mm Ligament Fastener Classification Name: Single/multiple component bone fixation appliances and accessories. Trade Name: Innovasive Devices Ligament Fastener
3. PREDICATE DEVICE:
The predicate device used to determine substantial equivalence for the Innovasive Devices 8 mm Ligament Fastener was the Mitek 5 mm Ligament Anchor, marketed by Mitek Surgical, Norwood, MA.
4. DEVICE DESCRIPTIONS:
The Innovasive 8mm Ligament Fastener is an implantable tendon fixation device complete with the instrumentation needed to effect ligament to bone repair. The devices consist of a high density polyethylene sleeve with an acetal drive pin which is deployed into a pre-drilled hole in the bone. The Fastener is deployed through the use of a hex driver, for screwing in the sleeve, and an inserter which drives the pin into the sleeve.
In addition to the Fastener, a stainless steel drill is available to establish the proper hole in the bone for the Fastener. All of the instrumentation except the Fastener will be offered as reusable devices and be autoclaved in the sterilization tray provided for this purpose.
The Fastener will be available as a sterile, single use device for use in both an open and arthroscopic procedures.
{1}------------------------------------------------
The Innovasive Devices 8mm Ligament Fastener is intended for ligament to bone reattachment of Anterior and Posterior Cruciate Ligaments in patients with recurrent instability of the knee.
6. COMPARISON OF CHARACTERISTICS:
The Innovasive Devices 8mm Ligament Fastener is comprised of two polymer components; high density polyethylene and acetal plastic. This device is used to secure a ligament in a predrilled hole in bone. It remains fixed in the bone by the screw threads and also radial compression as the device is deployed.
The Mitek Ligament Anchor is comprised of titanium with nitinol arcs. The arcs are imbedded in the bone when placed in the predrilled hole. The intended use of the Mitek Ligament Anchor and the Innovasive Ligament Fastener are identical.
7. PERFORMANCE DATA:
The following performance data was provided in support of the substantial equivalence determination:
-
- Cadaver study: Comparison of the ultimate holding strength in the bone indicated, compared to the predicate device. The Innovasive Fastener holding strength was found to be equivalent to the strength of the predicate device at a .01 significance level.
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.