(160 days)
The intended use of the Quinton Vital Statistics System (Q-Cath Accessory) is the acquisition, calculation, and transmission of physiological parameters as diastolic and systolic blood pressure, blood oxygen saturation levels, and heart rate to the Quinton Q-Cath during cardiac cath procedures.
The Vital Statistics System (VSS) is an accessory to the Quinton Q-Cath catheterization analysis system. The purpose of the VSS is to acquire, calculate, and transmit physiological data as blood pressure (acquired noninvasively), blood oxygen saturation data, and heart rate to the Q-Cath during cardiac cath procedures. Additionally, the VSS has 'pass-through' input and output connections for the purposes of acquiring and transmitting low voltage signal data to the Q-Cath from external devices and from the Q-Cath Signal Conditioning System component. Set-up of the VSS modules occurs at the Q-Cath via software controls.
The VSS is made up of a 3" x 20" x 12" metal enclosure that has compartments for insertion of the non-invasive blood pressure module, SP02 module, and two pass-through modules. A power supply within the device is capable of converting line voltage to regulated low voltage needed by the modules. The VSS weighs 35 Ibs when all modules are in place within the VSS.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mean percentage difference between measurements taken by the VSS NIBP module (adult mode) and the Colin BP-508 is less than 5% (at 0.05 confidence level). | VSS NIBP module (adult mode) is equivalent to Colin BP-508 at 95% confidence. |
| Mean percentage difference between measurements taken by the VSS NIBP module (neonate mode) and the Colin BP-508 is less than 5% (at 0.05 confidence level). | VSS NIBP module (neonate mode) is equivalent to Colin BP-508 at 95% confidence. |
| Mean percentage difference between measurements taken by the VSS SP02 module and the Nellcor N-180 pulse oximeter is less than 5% (at 0.05 confidence level). | VSS SP02 module is equivalent to Nellcor N-180 pulse oximeter at 95% confidence. |
Study Details
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document states "The number of randomized blood pressure, heart rate samples chosen were such that a meaningful statistical analysis and conclusion could be reached." and "The number of randomized blood pressure, heart rate samples chosen were such that a meaningful statistical analysis and conclusion could be reached." This indicates statistical power was considered, but an exact numerical sample size is not specified in the provided text for either NIBP or SP02 testing.
- Data Provenance: The data was generated using "Randomized simulated oscillometric NIBP and HR outputs" from an "NIBP simulator" and "Randomized simulated SP02 and HR outputs" from an "SP02 simulator." This implies simulated, laboratory-based data. The document does not specify a country of origin, but given the submitter's address is in Washington, USA, it is likely US-based. The study appears to be prospective in its design, as measurements were taken specifically for this comparison.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable. The ground truth in this study was not established by human experts but by the outputs of simulators, which are considered objective, controlled inputs for device testing.
-
Adjudication Method for the Test Set:
- Not applicable. Since the ground truth was based on simulator outputs, there was no need for human adjudication of results. The comparison was directly between device measurements and simulator outputs, or between the candidate device and the predicate device's measurements of the simulator outputs.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No. An MRMC study was not conducted as this device is a physiological monitor, not an interpretive diagnostic tool requiring human readers. The study focused on the accuracy and equivalence of measurements between devices.
-
Standalone (Algorithm Only Without Human-in-the-Loop) Performance:
- Yes. The performance testing described is inherently standalone as it compares the measurements produced by the VSS modules directly against those from the predicate devices, both operating on simulator outputs. There is no mention of a human operator influencing the measurement or calculation process in the performance test.
-
Type of Ground Truth Used:
- The ground truth was established by simulator outputs. Specifically, "Randomized simulated oscillometric NIBP and HR outputs" were obtained from an "NIBP simulator," and "Randomized simulated SP02 and HR outputs" were obtained from an "SP02 simulator."
-
Sample Size for the Training Set:
- Not applicable. This device is a physiological monitoring system, not an AI or machine learning algorithm that typically requires a training set. The performance testing involves direct comparison of measurements from the device against a known reference (simulator) or a predicate device.
-
How the Ground Truth for the Training Set Was Established:
- Not applicable. As a physiological monitoring device without AI/ML components, there's no concept of a "training set" or ground truth establishment for such a set in the context of this 510(k) submission.
{0}------------------------------------------------
510(k) SUMMARY
MAY 2 1 1997
| Submitter's Name: | Quinton Instrument Company |
|---|---|
| Submitter's Address: | 3303 Monte Villa ParkwayBothell, Washington 98021-8906 |
| Submitter's Phone Number: | (US) 206-402-2000 |
| Submitter's Fax Number: | (US) 206-402-2017 |
| Contact Person: | Matt Hedlund |
| Date Summary Prepared: | April 11, 1997 |
A. Device Name and Classification
-
- Device Trade Name: Vital Statistics System (Q-Cath Accessory)
-
Device Common Name: Physiological Monitoring and Analysis System
-
Device Classification Names 870.2700 Oximeter 870.1130 Noninvasive Blood Pressure Measurement System
B. Predicate Device
The legally marketed device to which we claim equivalence is the Nellcor Puritan Bennett N-180 Pulse Oximeter and the Colin Medical Instrument BP-508 Patient Monitor (PILOT model No. 9200) . The 510(k) numbers for these devices are K913695 and K922668, respectively.
{1}------------------------------------------------
C. Device Description
The Vital Statistics System (VSS) is an accessory to the Quinton Q-Cath catheterization analysis system. The purpose of the VSS is to acquire, calculate, and transmit physiological data as blood pressure (acquired noninvasively), blood oxygen saturation data, and heart rate to the Q-Cath during cardiac cath procedures. Additionally, the VSS has 'pass-through' input and output connections for the purposes of acquiring and transmitting low voltage signal data to the Q-Cath from external devices and from the Q-Cath Signal Conditioning System component. Set-up of the VSS modules occurs at the Q-Cath via software controls.
The VSS is made up of a 3" x 20" x 12" metal enclosure that has compartments for insertion of the non-invasive blood pressure module, SP02 module, and two pass-through modules. A power supply within the device is capable of converting line voltage to regulated low voltage needed by the modules. The VSS weighs 35 Ibs when all modules are in place within the VSS.
D. Intended Use of Device
The intended use of the Quinton Vital Statistics System (Q-Cath Accessory) is the acquisition, calculation, and transmission of physiological parameters as diastolic and systolic blood pressure. blood oxygen saturation levels, and heart rate to the Quinton Q-Cath during cardiac cath procedures.
E. Summary of Technological Characteristics Compared with the Predicate Device
The NIBP and SP02 modules are manufactured by Nellcor Puritan Bennett and Colin Medical, respectively. These modules are modifications of the devices that are used as the predicate devices in this 510k notification, as referenced in Section B, supra. The SP02 module acquires, calculates, and transmits oximetry and heart rate data as the Nellcor N-180. The SP02 and heart rate measurement system in the module and in the N-180 are 'equivalent' within certain statistical parameters. [See 'Performance Test and Conclusions' in Section F, infra.] Differences between the Nellcor N-180 and the VSS SP02 module are that the N-180 is a self-contained unit having controls, alarms, and a display. For the VSS module these features are contained on the Q-Cath. The VSS NIBP module acquires, calculates, and transmits diastolic and systolic blood pressure and heart rate data, as the Colin BP-508. The blood pressure and heart rate measurement system in the module and in the BP-508 are 'equivalent' within certain statistical parameters. [See 'Performance Test and Conclusions' in Section F, infra.] Differences between the Colin BP-508 and
000023
{2}------------------------------------------------
E. Summary of Technological Characteristics Compared with the Predicate Device (Continued)
the VSS NIBP module are that the BP-506 is a self-contained unit having controls, alarms, and a display. For the VSS NIBP module these features are contained on the Q-Cath. Additionally, the BP-508 has measurement and recording capability covering ECG, invasive blood pressure, SP02, tonometry blood pressure, and C02. These features are not on the VSS NIBP module.
F. Performance Testing and Conclusions
-
- Performance Testing
Performance testing was an equivalence study comparing blood pressure measurements and heart rate obtained by the VSS NIBP module with blood pressure measurements and heart rate obtained by the predicate device, the Colin BP-508 NIBP monitor, over both adult and neonate modes. Randomized simulated oscillometric NIBP and HR outputs over the specified range of both the NIBP module and the predicate device were obtained from a NIBP simulator. The test was conducted by the VSS NIBP module taking blood pressure and heart rate measurements from the simulator for a certain number of randomized samples. This was then repeated for the predicate device at the same blood pressure levels and heart rate. The number of randomized blood pressure, heart rate samples chosen were such that a meaningful statistical analysis and conclusion could be reached.
- Performance Testing
The SP02 performance testing was conducted in a similar fashion. Randomized simulated SP02 and HR outputs over the specified range of both the SP02 module and the Nellcor N-180 Pulse Oximeter were obtained from a SP02 simulator. The test was conducted by the VSS SP02 module taking SP02 data and heart rate measurements from the simulator for a certain number of randomized samples. This was then repeated for the predicate device at the same SP02 levels and heart rate. The number of randomized blood pressure, heart rate samples chosen were such that a meaningful statistical analysis and conclusion could be reached.
It was formulated that both modules would be considered subtantially equivalent to their respective predicate devices if it could be shown at the 0.05 confidence level that the mean percentage difference between each pair of measurements taken by the devices being compared is less than 5%.
000024
{3}------------------------------------------------
F. Performance Testing and Conclusions (Continued)
2) Conclusions
Using a +/- 5% equivalence region, the study results showed the following:
- The VSS NIBP module operating in adult mode can be regarded as . equivalent to the Colin BP-508 at the 95% confidence level (e.g., 0.05 significance level);
- The VSS NIBP module operating in neonate mode can be regarded as . equivalent to the Colin BP-508 at the 95% confidence level; and
- The VSS SP02 module can be regarded as equivalent to the Nellcor N-180 . pulse oximeter at the 95% confidence level.
都能
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes. The eagle is positioned within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 1 1997
Mr. Matt Hedlund Quinton Instrument Company 3303 Monte Villa Parkway Bothell, Washington 98021-8906
Re : K964978 Vital Statistics System (Q-Cath Accessory) Requlatory Class: II (two) Product Code: 73 DQA Dated: April 15, 1997 Received: April 17, 1997
Dear Mr. Hedlund:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (CMP) regulation (21 CFR Part 820) and that, through perioding GMP 2 inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Matt Hedlund
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K96 4978 510 K Number (if known):
Device Name: Vital Statistics System (Q-Cath Accessory)
Indications For Use:
The intended use of the Quinton Vital Statistics System (Q-Cath Accessory) is the acquisition, calculation, and transmission of physiological parameters as diastoic and systolic blood pressure, blood oxygen saturation levels, and heart rate to the Quinton Q-Cath during cardiac cath procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-OH) Division of Cardiovascular, Respiratory, and Neurological Devices
K964978 510(k) Number
Prescription Use_ X Per 21 CFR 801.109) ·
OR
Over-The-Counter Use
000012
(Optional Format 1-2-96)
ー
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).