K Number
K964609

Validate with FDA (Live)

Date Cleared
1997-03-06

(108 days)

Product Code
Regulation Number
864.7140
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Actalyke Activated Clotting Time System and Test Tubes are intended for use in the measurement of the activated clotting time test (ACT). Current markets using the ACT include: bypass surgery, vascular surgery, transplant surgery, renal dialysis, ECMO, cardiac catheterization and Percutaneous Transluminal Coronary Angioplasty (PTCA), Critical Care, and other medical therapies requiring heparin anticoagulation.

Device Description

The Actalyke System is a portable, battery operated device designed to perform a range of whole blood coagulation tests at the patient's bedside using Activated Clotting Time (ACT) measurement techniques. Each test well of the Actalyke Instrument incorporates a highly sensitive clot detection mechanism. Operationally, the detection mechanism comprises a magnet located inside a test tube and two solid state magnetic detectors positioned within the test well - one situated at 0° and another situated at 90° with respect to the test tube. When a test tube is inserted into the test well, the detector at 0° senses the presence of the magnet as the tube slowly rotates. As clotting forms, the fibrin strands cause the magnet to rotate within the tube such that its presence is sensed by the detector at 90° (indicating the occurrence of clotting and triggering end-point detection). This two-point detection sensing system is an improvement in ACT detection methods and minimizes the possibility of a missed end-point. Each Actalyke test tube has a barcode label affixed to it, which is read by the Actalyke instrument to determine the test tube type (i.e., celite, kaolin, or glass bead), If the result is printed (optional), the normal range for the identified test tube type will be printed. The lot number and expiration date of each tube are also identified on the barcode label.

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Actalyke System appear to be based on demonstrating "normal ranges" and "high correlation" with the legally marketed predicate device, the HEMOCHRON Whole Blood Coagulation System, across different test tube types (Celite, Kaolin, Glass Bead). Since specific quantitative acceptance thresholds are not explicitly stated, the reported performance is compared directly to the predicate device's performance in the provided tables.

Test Tube TypeAcceptance Criteria (Implied)Reported Actalyke Performance (Ranges from studies)Predicate Device (HEMOCHRON) Performance (Ranges from studies)
CeliteSimilar normal range to predicate108-133 seconds101-129 seconds
High correlation with predicate for varying heparin concentrationsSee "Correlation with Predicate Device" tables and graphs.See "Correlation with Predicate Device" tables and graphs.
KaolinSimilar normal range to predicate118-133 seconds108-133 seconds
High correlation with predicate for varying heparin concentrationsSee "Correlation with Predicate Device" tables and graphs.See "Correlation with Predicate Device" tables and graphs.
Glass BeadSimilar normal range to predicate167-199 seconds145-191 seconds
High correlation with predicate for varying heparin concentrationsSee "Correlation with Predicate Device" tables and graphs.See "Correlation with Predicate Device" tables and graphs.

Summary of Reported Device Performance:

  • Normal Range Studies: The Actalyke System, using its own tubes, demonstrated normal ranges largely overlapping with or comparable to the Hemochron system using its own tubes, when tested on healthy patient samples (as shown in the detailed tables for Celite, Kaolin, and Glass Bead).
  • Correlation Studies (with varying heparin): The Actalyke Instrument and Actalyke ACT tubes showed "high correlation" with the Hemochron Instrument and Hemochron ACT tubes across different heparin concentrations for all three tube types (Celite, Kaolin, Glass Bead), indicating similar performance in measuring coagulation time under anticoagulation. This was visually represented and stated generally as "high correlation" in the DATA SUMMARY.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Size (Normal Range Studies):
    • Celite Normal Range Study: 15 patient samples (S011-S024 excluding some, plus S002).
    • Kaolin Normal Range Study: 15 patient samples (S033-S040, S004, S005, S009, S030, S031, S037, plus 11L001).
    • Glass Bead Normal Range Study: 15 patient samples (S051, S027, S025, S052, S028, S019, S035, S045, 11L001, S049, S048, S006, S016, S030, S009).
  • Sample Size (Correlation Studies):
    • The sample size for the heparin correlation studies is not explicitly stated as a number of individual patients. However, the data represents "mean data" for varying heparin concentrations (0, 1, 3, 5 u/ml for Celite & Kaolin; 0, 0.3, 0.6, 1, 2 u/ml for Glass Bead), suggesting multiple measurements per heparin concentration. Given the normal range studies used 15 patients, it's plausible similar numbers were used for correlation, but it's not confirmed.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The patient IDs (e.g., S011, S002) suggest individual patient data was collected, but we cannot deduce if it was prospective or retrospective from the provided text.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. This device measures a physical property (clotting time) directly, rather than relying on expert interpretation of images or other qualitative data. The "ground truth" is the measured Activated Clotting Time (ACT) itself, or comparison to an established, legally marketed device (HEMOCHRON system).

4. Adjudication Method (for the test set)

Not applicable. As noted above, the device measures a physical property directly. There is no qualitative assessment or interpretation by experts that would require an adjudication method. The study design involves direct comparison of measurements between the Actalyke system and the predicate device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is not applicable to the Actalyke System. An MRMC study typically involves human readers interpreting cases (e.g., medical images) with and without AI assistance to assess changes in diagnostic performance. The Actalyke System is a point-of-care coagulation analyzer, not an AI-powered diagnostic imaging tool that assists human readers.

6. Standalone Performance Study

Yes, a standalone performance was done in the "Normal Range Study" for each tube type. The Actalyke instrument was used with Actalyke tubes to establish normal ranges, which were then compared to the normal ranges obtained by the Hemochron instrument with Hemochron tubes. Additionally, the correlation studies implicitly demonstrate standalone performance when the Actalyke instrument is used with Actalyke tubes across varying heparin concentrations.

7. Type of Ground Truth Used

The ground truth used is predominantly:

  • Comparison to a Legally Marketed Predicate Device: The performance of the Actalyke System is evaluated by comparing its results directly to those obtained from the HEMOCHRON Whole Blood Coagulation System, which is the established standard.
  • Measured Activated Clotting Time (ACT): For the normal range studies, the ACT measurements from healthy individuals define the "normal range." For the correlation studies, the measured ACT values at different heparin concentrations serve as the primary data point for comparison.

8. Sample Size for the Training Set

Not applicable. The Actalyke System does not appear to be an AI/machine learning device that requires a "training set" in the conventional sense. It is a hardware device based on magnetic detection principles. The studies described are performance validation studies, not training of an algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

K964609

    1. 510(k) Summary of Information Respecting Safety and Effectiveness
  • A. Legally Marketed Device

Array Medical claims substantial equivalence to the HEMOCHRON Whole Blood Coagulation System and Activated Clotting Time Tests (pre-1976), currently in commercial distribution by International Technidyne Corporation.

  • B. Device Description

Actalyke System

The Actalyke System is a portable, battery operated device designed to perform a range of whole blood coagulation tests at the patient's bedside using Activated Clotting Time (ACT) measurement techniques.

ACT Measurement Principles

A step-by-step description of the Actalyke ACT measurement principle is provided below:

  • l ) Each test well of the Actalyke Instrument incorporates a highly sensitive clot detection mechanism.
    1. Operationally, the detection mechanism comprises a magnet located inside a test tube and two solid state magnetic detectors positioned within the test well - one situated at 0° and another situated at 90° with respect to the test tube. See below:

Image /page/0/Figure/12 description: This image shows a diagram of a test well with a tube inside. There is a detector at 90 degrees and a detector at 0 degrees. A magnet is also present in the diagram, positioned near the tube.

    1. When a test tube is inserted into the test well, the detector at 0° senses the presence of the magnet as the tube slowly rotates.
    1. As clotting forms, the fibrin strands cause the magnet to rotate within the tube such that its presence is sensed by the detector at 90° (indicating the occurrence of clotting and triggering end-point detection). This two-point detection sensing system is an improvement in ACT detection methods and minimizes the possibility of a missed end-point.

{1}------------------------------------------------

Device Description (continued)

Actalyke Activated Clotting Time Tubes

Each Actalyke test tube has a barcode label affixed to it, which is read by the Actalyke instrument to determine the test tube type (i.e., celite, kaolin, or glass bead), If the result is printed (optional), the normal range for the identified test tube type will be printed. The lot number and expiration date of each tube are also identified on the barcode label.

Cat. No.Recomm. UseActivatorDescription
C-ACTBypass andVascular surgeryCelite (14 mg.)Glass tube with blackflip-top tubeand barcode label.2.0 ml blood sample
K-ACTBypass surgeryin the presenceaprotininKaolin (14 mg.)Glass tube with goldflip-top and barcodelabel. 2.0 ml bloodsample
G-ACTRenal DialysisGlass beads (50 mg.)Plastic tube with cleaflip-top and barcodelabel. 0.4 ml bloodsample

NOTE: Used Actalyke test tubes should be disposed of in accordance with each institution's policy regarding biohazardous materials.

C. Intended Use

The Actalyke Activated Clotting Time System and Test Tubes are intended for use in the measurement of the activated clotting time test (ACT). Current markets using the ACT include: bypass surgery, vascular surgery, transplant surgery, renal dialysis, ECMO, cardiac catheterization and Percutaneous Transluminal Coronary Angioplasty (PTCA), Critical Care, and other medical therapies requiring heparin anticoagulation.

{2}------------------------------------------------

D. Comparison with Predicate Device

(

Feature Comparison of Actalyke and HEMOCHRON Systems

ActalykeHEMOCHRON
Intended UseActivated Clotting Time TestActivated Clotting Time Test
Method of detectionMagnetic detectionMagnetic Detection
Reactive Ingredients (tubes)Celite, kaolin, glass-beadsCelite, kaolin, glass-beads
Storage of tubesRoom temperatureRoom temperature
Materials providedInstrumentTest TubesControlsInstrumentTest TubesControls
Instrument SpecificationsNo. Test Channels: 1 or 2 Timing Range: 0-1500 Operating Time on Battery (hours): 4-6 Test barcode identification Options: PrinterNo. Test Channels: 1 or 2 Timing Range: 0-1500 Operating Time on Battery (hours): 4-6
PerformanceC-ACT 105-130 sec.K-ACT 114-130 sec.G-ACT 165-212 sec.Normal Range studiesCA510 105-167 sec.K-ACT 91-151 sec.P214 145-195 sec.Normal Range studies

{3}------------------------------------------------

E. Performance Data

Specific Performance Characteristics

Normal Range Study

Celite Normal Range Study

InstrumentHemochronActalyke
TubesHemochronActalykeHemochronActalyke
NoPatient ID
1S011129133103117
2S012124126118125
3S013112118106113
4S014110110100109
5S015104112113125
6S016109114108117
7S002108117109112
8S017116119112123
9S018107114112117
10S019101108100110
11S020109115110116
12S021121129120130
13S022104106102112
14S023113124110122
15S024108112110115
mean111.67117.13108.87117.53
std7.837.835.956.20
%cv7.016.685.475.27
max129.00133.00120.00130.00
min101.00108.00100.00109.00
x-2std96.02101.4896.97105.14
x+2std127.32132.79120.77129.93
x-3std88.1993.6591.0198.94
x+3std135.14140.62126.72136.13

Kaolin Normal Range Study

(.

.. '-

InstrumentHemochronActalyke
TubesHemochronActalykeHemochronActalyke
NoPatient ID
1S033118128107117
2S034118119108118
3S035116120105119
4S036122128115125
5S037114122110119
6S004116125118128
711L001119130115122
8S005122132119128
9S037108118108121
10S038112121114120
11S009122133122127
12S030119127121126
13S031115121116122
14S039110125106118
15S040109119108119
mean116.00124.53112.80121.93
std4.664.965.713.88
%cv4.023.985.063.18
max122.00133.00122.00128.00
min108.00118.00105.00117.00
x-2std106.68114.62101.38114.17
x+2std125.32134.44124.22129.70
x-3std102.02109.6795.67110.29
x+3std129.98139.40129.93133.58

{4}------------------------------------------------

Glass Bead Normal Range Study

InstrumentHemochronActalyke
TubesHemochronActalykeHemochronActalyke
NoPatient ID
1S051188180188203
2S027175168163185
3S025177167177199
4S052178199199198
5S028166169178178
6S019162167163167
7S035168197158199
8S045173172179172
911L001171190155189
10S049191186157203
11S048174169181197
12S006169180185185
13S016157176166177
14S030150178167183
15S009145179170188
mean169.60178.47172.40188.20
std12.5510.5612.7611.45
%cv7.405.917.406.08
max191.00199.00199.00203.00
min145.00167.00155.00167.00
X-2std144.50157.36146.88165.31
x+2std194.70199.58197.92211.09
x-3std131.95145.80134.12153.86
x+3std207.25210.13210.68222.54

{5}------------------------------------------------

Correlation with predicate device

Summary Celite tubes
--------------------------
Mean datau/ml heparin
Instrumenttube0135
HemochronHemochron120.6252.4448.8782.2
HemochronActalyke123264431.8709.2
ActalykeHemochron114214.2377.4548
ActalykeActalyke122235.2386.8598.6

Image /page/5/Figure/3 description: This figure is a line graph titled "Mean total data Celite tubes". The x-axis is labeled "u/ml heparin" and ranges from 0 to 5. The y-axis is labeled "ACT (seconds)" and ranges from 0 to 1000. There are 4 lines on the graph, representing Hemochron and Actalyke.

{6}------------------------------------------------

Summary Kaolin tubes

(

Mean datau/ml heparin
instrumenttube0135
HemochronHemochron129.6291533.4848.75
HemochronActalyke137297.6511780.2
ActalykeHemochron124264.4497735.6

Image /page/6/Figure/2 description: The image is a graph titled "Mean total data Kaolin tubes". The x-axis is labeled "u/ml heparin" and ranges from 0 to 5. The y-axis is labeled "ACT (seconds)" and ranges from 0 to 900. There are four lines on the graph, representing Hemochron and Actalyke, showing the relationship between heparin concentration and ACT.

{7}------------------------------------------------

Summary Glass Bead tubes

Mean datau/ml heparin
Instrumenttube00.30.612
HemochronHemochron179.0294.4345.6451.0636.2
HemochronActalyke181.0270.2336.0402.0584.4
ActalykeHemochron178.4272.6337.6378.4554.6

Image /page/7/Figure/2 description: The image is a graph titled "Mean total data Glass bead tubes". The x-axis is labeled "u/ml heparin" and ranges from 0 to 2. The y-axis is labeled "ACT (seconds)" and ranges from 0.0 to 800.0. There are multiple lines on the graph, representing Hemochron and Actalyke.

DATA SUMMARY

  • Actalyke Instrument/Actalyke ACT tubes yielded similar normal ranges when compared . to the Hemochron Instrument/Hemochron ACT tubes when utilizing the same normal patient samples
  • Actalyke Instrument/Actalyke ACT tubes yielded high correlation with Hemochron . Instrument/Hemochron ACT tubes when utilizing the same patient samples
  • Hemochron Instruments/Actalyke ACT tubes yielded high correlation with . Hemochron Instrument/Hemochron ACT tubes when utilizing the same patient samples
  • Actalyke Instruments/Hemochron ACT tubes yielded high correlation with . Hemochron Instrument/Hemochron ACT tubes when utilizing the same patient samples

. . .

ë;

§ 864.7140 Activated whole blood clotting time tests.

(a)
Identification. An activated whole blood clotting time tests is a device, used to monitor heparin therapy for the treatment of venous thrombosis or pulmonary embolism by measuring the coagulation time of whole blood.(b)
Classification. Class II (performance standards).