(79 days)
The HCL - 200 is intended for use as an in vitro diagnostic multichannel chemistry analyzer.
The HCL-200 is an automated discrete random access in vitro diagnostic multichannel chemistry analyzer. It is composed of two modules:
- The liquid handling system includes compartments of reagents and samples as well as the pipetting unit.
- The reaction module includes the cuvette turntable to which samples and reagents are transferred. A photometer to monitor reaction response is also contained in this unit.
Then analyzer is linked to a PC compatible computer which loads software into each of the two modules.
This document describes the performance of the HCL-200 Clinical Chemistry Analyzer, not an AI device. Therefore, many of the requested fields are not applicable.
Here's an analysis based on the provided text:
Acceptance Criteria and Reported Device Performance
The HCL-200 Clinical Chemistry Analyzer's performance was evaluated against the predicate device, the Roche COBAS MIRA, using NCCLS (National Committee for Clinical Laboratory Standards) evaluation protocols. The acceptance criteria were not explicitly stated as numerical targets in the provided text. Instead, the study aimed to demonstrate "substantially equivalent performance" to the predicate device. This equivalence was assessed through linearity, precision (simple and complex), and method comparison studies.
The tables below summarize the reported device performance for each evaluated reagent, which served as evidence for meeting the implicit acceptance criteria of substantial equivalence.
Table 1: HCL-200 Reported Linearity & Precision
| REAGENT NAME | LINEARITY (Linear Range) | SIMPLE PRECISION (Normal CV / Abnormal CV) | COMPLEX PRECISION (Normal CV / Abnormal CV) |
|---|---|---|---|
| Albumin | 2 to 8.5 g/dL | 1.8% / 1.7% | 8% / 6.4% |
| ALT (GPT) | 0 to 650 U/L | 3% / 1.5% | 4.6% / 2.1% |
| Amylase | 0 to 2500 U/L | 4.6% / 1.7% | 5.0% / 3.5% |
| AST (GOT) | 0 to 650 U/L | 3.5% / 2.6% | 6.1% / 4.6% |
| Bilirubin, Total | 0 to 20 mg/dL | 2.8% / 1.5% | 7.3% / 9.0% |
| BUN, Rate | 0 to 125 mg/dL | 4.1% / 1.9% | 5.8% / 5.2% |
| CK NAD, Imidazole | 0 to 2100 U/L | 2.3% / 2.3% | 4.6% / 2.1% |
| Creatinine, (Jaffe) | 0 to 12.5 mg/dL | 2.45% / 1.9% | 7.2% / 3.1% |
| CRP | not applicable | 6.3% / 1.9% | 6.7% / 2.6% |
| GGT | 5 to 129 mg/dL | 2.9% / 1.3% | 5.9% / 2.2% |
| Glucose, Hexokinase UV | 0 to 600 mg/dL | 2.6% / 2.3% | 1.0% / 2.9% |
| LDH-L | 0 to 1000 U/L | 2.1% / 0.5% | 2.8% / 1.9% |
| Phosphorus | 0 to 15.0 mg/dL | 2.9% / 1.2% | 5.5% / 3.1% |
| Total Protein | 0 to 12 g/dL | 1.6% / 0.9% | 2.4% / 1.9% |
| Triglycerides GPO | 0 to 1000 mg/dL | 2.8% / 1.6% | 4.2% / 3.3% |
| Uric Acid | 0 to 20 mg/dL | 1.7% / 1.8% | 4.2% / 4.8% |
Table 2: HCL-200 Comparison to Predicate Device (COBAS MIRA)
| REAGENT NAME | COMPARISON TO PREDICATE (COBAS MIRA) |
|---|---|
| Albumin | N: 100, Slope [m]: 1.000, Intercept [b]: -0.03, Correlation Coeff. [r]: 0.987 |
| ALT (GPT) | N: 104, Slope [m]: 1.08, Intercept [b]: 0.6, Correlation Coeff. [r]: 0.998 |
| Amylase | N: 101, Slope [m]: 1.09, Intercept [b]: -2.8, Correlation Coeff. [r]: 0.999 |
| AST (GOT) | N: 100, Slope [m]: 1.05, Intercept [b]: -0.6, Correlation Coeff. [r]: 0.997 |
| Bilirubin, Total | N: 100, Slope [m]: 0.902, Intercept [b]: 0.11, Correlation Coeff. [r]: 0.993 |
| BUN, Rate | N: 100, Slope [m]: 0.948, Intercept [b]: 1.1, Correlation Coeff. [r]: 0.996 |
| CK NAD, Imidazole | N: 100, Slope [m]: 1.034, Intercept [b]: 5.6, Correlation Coeff. [r]: 0.999 |
| Creatinine, (Jaffe) | N: 103, Slope [m]: 0.994, Intercept [b]: 0.045, Correlation Coeff. [r]: 0.997 |
| CRP | N: 100, Slope [m]: Not specified, Intercept [b]: Not specified, Correlation Coeff: [r]: 0.992 |
| GGT | N: 100, Slope [m]: 1.07, Intercept [b]: -0.9, Correlation Coeff: [r]: 0.999 |
| Glucose, Hexokinase UV | N: 400, Slope [m]: 0.998, Intercept [b]: 7.5, Correlation Coeff: [r]: 0.992 |
| LDH-L | N: 100, Slope [m]: 1.06, Intercept [b]: 16.2, Correlation Coeff: [r]: 0.992 |
| Phosphorus | N: 111, Slope [m]: 1.06, Intercept [b]: 0.0, Correlation Coeff: [r]: 0.985 |
| Total Protein | N: 102, Slope [m]: 1.05, Intercept [b]: -0.09, Correlation Coeff: [r]: 0.983 |
| Triglycerides GPO | N: 102, Slope [m]: 1.025, Intercept [b]: -2.5, Correlation Coeff: [r]: 0.994 |
| Uric Acid | N: 100, Slope [m]: 1.02, Intercept [b]: -0.25, Correlation Coeff: [r]: 0.987 |
Study Details:
-
Sample size used for the test set and the data provenance:
-
Linearity (Number [N]):
- Albumin: 5
- ALT (GPT): 7
- Amylase: 9
- AST (GOT): 6
- Bilirubin, Total: 6
- BUN, Rate: 5
- CK NAD, Imidazole: 8
- Creatinine, (Jaffe): 5
- CRP: Not specified
- GGT: 8
- Glucose, Hexokinase UV: 5
- LDH-L: 9
- Phosphorus: 6
- Total Protein: 5
- Triglycerides GPO: 5
- Uric Acid: 5
-
Simple Precision (Number [N]): 20 for all reagents.
-
Complex Precision (Number [N]): 80 for all reagents.
-
Comparison to Predicate (Number [N]):
- Albumin: 100
- ALT (GPT): 104
- Amylase: 101
- AST (GOT): 100
- Bilirubin, Total: 100
- BUN, Rate: 100
- CK NAD, Imidazole: 100
- Creatinine, (Jaffe): 103
- CRP: 100
- GGT: 100
- Glucose, Hexokinase UV: 400
- LDH-L: 100
- Phosphorus: 111
- Total Protein: 102
- Triglycerides GPO: 102
- Uric Acid: 100
-
Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective/prospective). However, the tests were conducted according to NCCLS evaluation protocols commonly used for in vitro diagnostic devices. This implies lab-based testing.
-
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This study concerns a chemical analyzer, not an AI device relying on expert interpretation of images or clinical data. The "ground truth" for these tests would be the established reference methods of the predicate device (COBAS MIRA) and calibrated controls/standards used in analytical chemistry.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This is a measurement device, not an AI system requiring human adjudication of its output.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The HCL-200 is a "discrete random access in vitro diagnostic multichannel chemistry analyzer." The performance reported is that of the instrument itself (standalone performance) under laboratory control. There is no human-in-the-loop for the analytical process of measuring chemical concentrations.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this type of device is established clinical chemistry standards, controls, and comparisons to a legally marketed predicate device (Roche COBAS MIRA). The predicate device's measurements serve as the reference standard against which the HCL-200's results are compared.
-
The sample size for the training set: Not applicable. This is not a machine learning or AI device that requires a training set in the conventional sense. The "training" in this context refers to optimizing application parameters for the reagents on the instrument after a "two-week device familiarization period." This is more akin to calibration and method development.
-
How the ground truth for the training set was established: Not applicable, as there's no traditional "training set" for AI. The process described is "Preliminary evaluations using NCCLS EP 10 - T2 were run on each reagent in order to optimize application parameters." This involves using laboratory standards and controls to ensure the instrument's methods are correctly set up and calibrated for each assay.
{0}------------------------------------------------
HCL Laboratory Systems, Inc. 510(k) Notification for HCL -200
JAN 22 1997
510 (k) Summary of Safety and Effectiveness
510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS OF THE HCL-200 CLINICAL CHEMISTRY ANALYZER:
Submitted by:
Gregory M. Hearne Executive Vice President HCL Laboratory Systems Inc. 362 Camino Bailen Escondido, CA 92029 Telephone: (619) 738 - 8371. FAX: (619) 738 - 8371.
Identification:
- Trade Name: HCL 200 Biochemistry Analyzer. .
- . Common or usual name: Clinical Chemistry Analyzer
- Classification name: Micro Chemistry Analyzer for Clinical Use. .
Description:
The HCL-200 is an automated discrete random access in vitro diagnostic multichannel chemistry analyzer. It is composed of two modules:
- The liquid handling system includes compartments of reagents and samples as well as the pipetting . unit.
- The reaction module includes the cuvette turntable to which samples and reagents are transferred. A . photometer to monitor reaction response is also contained in this unit.
Then analyzer is linked to a PC compatible computer which loads software into each of the two modules.
Intended Use:
The HCL - 200 is intended for use as an in vitro diagnostic multichannel chemistry analyzer.
Technological Characteristics Compared to the Predicate Device
The functions of the HCL-200 and the COBAS MIRA [510 (k) K851172] are fundamentally the same. The basic physical principles for measurement of radiant energy, absorbed or transmitted, under controlled conditions are the same. Both instruments are automated interference filter photometers capable of measuring monochromatic light with wavelengths between 340 and 700 mm.
{1}------------------------------------------------
Substantial Equivalence and Summary Performance Data:
NCCLS evaluation protocols were used to demonstrate substantially equivalent performance between the HCL-200 and the predicate device. NCCLS Documents used in the evaluation of HCL-200 and the comparison to the Roche COBAS MIRA are listed below:
| Table 3 NCCLS Protocols Used in Evaluation/ Comparison | |
|---|---|
| NCCLS Document Number | Document Title |
| EP 10 - T2Vol. 13 No, 10 | Preliminary Evaluation of Quantitative ClinicalLaboratory Methods, Second Edition |
| EP6-PVol. 6 No. 18 | Evaluation of the Linearity of QuantitativeAnalytical Methods |
| EP5 - T2Vol. 12, No.4 | Evaluation of Precision Performance of ClinicalChemistry Devices, Second Edition |
| EP9 - TVol. 13 No. 4 | Method Comparison and Bias Estimation UsingPatient Samples. |
Method protocols (applications) for OEM supplied reagents listed below were developed on the instrument after a two week device familiarization period as described in the NCCLS Documents. Preliminary evaluations using NCCLS EP 10 - T2 were run on each reagent in order to optimize application parameters.
| REAGENT NAME | 510k Number | REAGENT NAME | 510k Number | REAGENT NAME | LINEARITY | SIMPLE PRECISION | COMPLEX PRECISION | COMPARISON TOPREDICATE (COBAS MIRA) |
|---|---|---|---|---|---|---|---|---|
| Albumin | K831585 | CRP | K902593 | Albumin | Number [N]: 5Linear Range:2 TO 8.5 g/dL | Number [N]: 20Normal: CV: 1.8%Abnormal CV: 1.7% | Number [N]: 80Normal: CV: 8%Abnormal CV: 6.4% | Number [N]: 100Slope [m]: 1.000Intercept [b]: -0.03Correlation Coeff. [r]: 0.987 |
| ALT (GPT) | K820525 | GGT | K905085 | ALT (GPT) | Number [N]: 7Linear Range:0 to 650 U/L | Number [N]: 20Normal: CV: 3%Abnormal CV: 1.5% | Number [N]: 80Normal: CV: 4.6%Abnormal CV: 2.1% | Number [N]: 104Slope [m]: 1.08Intercept [b]: 0.6Correlation Coeff. [r]: 0.998 |
| Amylase | K861718 | Glucose, HexokinaseUV | K801765 | Amylase | Number [N]: 9Linear Range:0 to 2500 U/L | Number [N]: 20Normal: CV: 4.6%Abnormal CV: 1.7% | Number [N]: 80Normal: CV: 5.0%Abnormal CV: 3.5% | Number [N]: 101Slope [m]: 1.09Intercept [b]: -2.8Correlation Coeff. [r]: 0.999 |
| AST (GOT) | K820270 | LDH-L | K822312 | AST (GOT) | Number [N]: 6Linear Range:0 to 650 U/L | Number [N]: 20Normal: CV: 3.5%Abnormal CV: 2.6% | Number [N]: 80Normal: CV: 6.1%Abnormal CV: 4.6% | Number [N]: 100Slope [m]: 1.05Intercept [b]: -0.6Correlation Coeff. [r]: 0.997 |
| Bilirubin, Total | K841892 | Phosphorus | K870776 | Bilirubin, Total | Number [N]: 6Linear Range:0 to 20 mg/dL | Number [N]: 20Normal: CV: 2.8%Abnormal CV: 1.5% | Number [N]: 80Normal: CV: 7.3%Abnormal CV: 9.0% | Number [N]: 100Slope [m]: 0.902Intercept [b]: 0.11Correlation Coeff. [r]: 0.993 |
| BUN, Rate | K812417 | Total Protein | K841466 | BUN, Rate | Number [N]: 5Linear Range:0 to 125mg/dL | Number [N]: 20Normal: CV: 4.1%Abnormal CV: 1.9% | Number [N]: 80Normal: CV: 5.8%Abnormal CV: 5.2% | Number [N]: 100Slope [m]: 0.948Intercept [b]: 1.1Correlation Coeff. [r]: 0.996 |
| CK NAD, Imidazole | K905084/A | Triglycerides GPO | K842029/A | CK NAD, Imidazole | Number [N]: 8Linear Range:0 to 2100 U/L | Number [N]: 20Normal: CV: 2.3%Abnormal CV: 2.3% | Number [N]: 80Normal: CV: 4.6%Abnormal CV: 2.1% | Number [N]: 100Slope [m]: 1.034Intercept [b]: 5.6Correlation Coeff. [r]: 0.999 |
| Creatinine, (Jaffe) | K840872 | Uric Acid | K812641 | Creatinine, (Jaffe) | Number [N]: 5Linear Range:0 to 12.5 mg/dL | Number [N]: 20Normal: CV: 2.45%Abnormal CV: 1.9% | Number [N]: 80Normal: CV: 7.2%Abnormal CV: 3.1% | Number [N]: 103Slope [m]: 0.994Intercept [b]: 0.045Correlation Coeff. [r]: 0.997 |
A summary of pertinent statistical results are presented in the table on the following two (2) pages:
{2}------------------------------------------------
HCL Laboratory Systems, Inc.
510(k) Notification for HCL -200
!
ﺔ ﺍﻟﻤﺮﺍﺟﻊ
트
{3}------------------------------------------------
HCL Laboratory Systems, Inc.
510(k) Notification for HCL -200
| REAGENT NAME | LINEARITY | SIMPLE PRECISION | COMPLEX PRECISION | COMPARISON TO PREDICATE (COBAS MIRA) |
|---|---|---|---|---|
| CRP | Number [N]:Linear Range: not applicable.Limits of Standard Curve: | Number [N]: 20Normal: CV: 6.3%Abnormal CV:1.9% | Number [N]: 80Normal: CV: 6.7%Abnormal CV: 2.6% | Number [N]: 100Slope [m]:Intercept [b]:Correlation Coeff: [r]: 0.992 |
| GGT | Number [N]: 8Linear Range:5 to 129 mg/dL | Number [N]: 20Normal: CV: 2.9%Abnormal CV: 1.3% | Number [N]: 80Normal: CV: 5.9%Abnormal CV:2.2% | Number [N]: 100Slope [m]: 1.07Intercept [b]: -0.9Correlation Coeff: [r]: 0.999 |
| Glucose, Hexokinase UV | Number [N]: 5Linear Range:0 to 600 mg/dL | Number [N]: 20Normal: CV: 2.6%Abnormal CV: 2.3% | Number [N]: 80Normal: CV: 1.0%Abnormal CV: 2.9% | Number [N]: 400Slope [m]:0.998Intercept [b]: 7.5Correlation Coeff: [r]: 0.992 |
| LDH-L | Number [N]: 9Linear Range:0 to 1000 U/L | Number [N]: 20Normal: CV: 2.1%Abnormal CV: 0.5% | Number [N]: 80Normal: CV: 2.8%Abnormal CV: 1.9% | Number [N]: 100Slope [m]: 1.06Intercept [b]: 16.2Correlation Coeff: [r]: 0.992 |
| Phosphorus | Number [N]: 6Linear Range:0 to 15.0 mg/dL | Number [N]: 20Normal: CV: 2.9%Abnormal CV: 1.2% | Number [N]: 80Normal: CV: 5.5%Abnormal CV: 3.1% | Number [N]: 111Slope [m]: 1.06Intercept [b]: 0.0Correlation Coeff: [r]: 0.985 |
| Total Protein | Number [N]: 5Linear Range:0 to 12 g/dL | Number [N]: 20Normal: CV: 1.6%Abnormal CV: 0.9% | Number [N]: 80Normal: CV: 2.4%Abnormal CV: 1.9% | Number [N]: 102Slope [m]: 1.05Intercept [b]: -0.09Correlation Coeff: [r]: 0.983 |
| Triglycerides GPO | Number [N]: 5Linear Range:0 to 1000 mg/dL | Number [N]: 20Normal: CV: 2.8%Abnormal CV: 1.6% | Number [N]: 80Normal: CV: 4.2%Abnormal CV: 3.3% | Number [N]: 102Slope [m]: 1.025Intercept [b]: -2.5Correlation Coeff: [r]: 0.994 |
| Uric Acid | Number [N]: 5Linear Range:0 to 20 mg/dL | Number [N]: 20Normal: CV: 1.7%Abnormal CV: 1.8% | Number [N]: 80Normal: CV: 4.2%Abnormal CV: 4.8% | Number [N]: 100Slope [m]: 1.02Intercept [b]: -0.25Correlation Coeff: [r]: 0.987 |
{4}------------------------------------------------
Claim of Substantial Equivalence to Predicate Device
The safety and effectiveness of the HCL-200 Clinical Chemistry Analyzer is evidenced by:
- Its equivalent performance to the predicate device, the Roche COBAS MIRA, as summarized in table 1. 5 above and,
- the fact that the two devices are virtually identical in terms of their use of basic principles of radiant 2. energy measurement, automated liquid handling and data management capabilities.
§ 862.2170 Micro chemistry analyzer for clinical use.
(a)
Identification. A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.