(302 days)
Stockert Instrumente (Sorin) Pump System
Not Found
No
The document describes a mechanical peristaltic pump system with electronic controls for monitoring and adjusting speed, flow, and pressure. There is no mention of AI or ML in the device description, intended use, or performance studies.
Yes.
The device is used in extracorporeal perfusion circuits to pump blood during cardiopulmonary bypass, which is a therapeutic procedure to support heart and lung function.
No
The device is a blood pump system intended for use in cardiopulmonary bypass procedures, which is a therapeutic function rather than a diagnostic one. While it monitors flow rate, pump speed, and pressure, these are operational parameters for the pump itself, not diagnostic indicators of a patient's condition.
No
The device description explicitly lists multiple hardware components including a console, rotor, holder, chamber, and connecting tube. It also describes the physical mechanism of the peristaltic pump.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "pump blood during routine cardiopulmonary bypass procedures." This is a procedure performed on a patient's body (extracorporeal), not a test performed on a sample taken from a patient's body.
- Device Description: The description details a mechanical pump system designed to move blood in a circuit outside the body. It does not describe any components or functions related to analyzing biological samples.
- Lack of IVD Characteristics: The text does not mention any of the typical characteristics of an IVD, such as:
- Analyzing biological samples (blood, urine, tissue, etc.)
- Detecting or measuring substances in samples
- Providing diagnostic information based on sample analysis
The device is clearly intended for a therapeutic/supportive function during surgery, not for diagnostic testing.
N/A
Intended Use / Indications for Use
The AVECOR Affinity Blood Pump System is intended for use in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Product codes
74 DTQ
Device Description
The AVECOR Cardiovascular Affinity Pump System consists of the Model 2000 Affinity Pump Console, Affinity Pump Rotor, Affinity Pump Holder, Affinity Pump Chamber, 5/8" Connecting Tube and an AVECOR venous reservoir with 5/8" outlet. The AVECOR Affinity Pump has been designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyurethane chamber incorporating polycarbonate inlet and outlet connectors, which fills passively due to the filling pressure exhibited by the fluid height in a reservoir. The peristaltic action of the rollers against the filled pump chamber results in forward flow. Flow rate is affected by the combination of reservoir volume, RPM's of the pump rotor and the total circuit resistance.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
The following tests were conducted by AVECOR to establish the safety and effectiveness of the Affinity Blood Pump System:
- In Vitro Operating Characteristics (Section III)
- In Vitro Cellular Damage (Section III)
- Material Biocompatibility (Section IV)
- Sterility (Section V)
- Pyrogenicity (Section V)
- Sterilant Residuals (Section V)
- Software Validation & Verification (Appendix E)
The results indicate that the Affinity Blood Pump System is substantially equivalent in performance to the Stockert Instrumente (Sorin) Pump System. Blood trauma testing showed equivalence in hemolysis, platelet and white cell response.
Key Metrics
Not Found
Predicate Device(s)
Stockert Instrumente (Sorin) Pump System
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.4220 Cardiopulmonary bypass heart-lung machine console.
(a)
Identification. A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.(b)
Classification. Class II (performance standards).
0
K 964017
SECTION II
AVECOR CARDIOVASCULAR INC.
AFFINITY BLOOD PUMP SYSTEM
AUG - 5 1997
SUPPORTING SUMMARY FOR 510 (k) NOTIFICATION
- September 30, 1996 1 . Date Prepared:
-
- AVECOR Cardiovascular Inc. Submitter: 13010 County Road 6 Plymouth, MN 55441
-
- Contact: Dennis E. Steger Director Regulatory Affairs/ Quality Assurance (800) 328-3320
-
- Device Name: Cardiopulmonary Bypass Roller-Type Pumps, Consoles and Bypass Pump Tubing have been classified by the Cardiovascular Device Classification Panel as Class II devices, as proposed in Section 870.4220/4370/4380/4390, Title 21 of the Code of Federal Regulations
- న్. Device Description & Comparison to Predicate Device:
To establish the substantial equivalency of the AVECOR Affinity Pump System to the Stockert Instrumente (Sorin) Pump design, specifications, materials, indications for use, and performance characteristics were compared.
DESIGN
The AVECOR Affinity Blood Pump System is intended for use in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The AVECOR Cardiovascular Affinity Pump System consists of the Model 2000 Affinity Pump Console, Affinity Pump Rotor, Affinity Pump Holder, Affinity Pump Chamber, 5/8" Connecting Tube and an AVECOR venous reservoir with 5/8" outlet.
The AVECOR Affinity Blood Pump Console and components' shall conform with all the applicable requirements of IEC 601, UL 544 and will display the European Community CE marking.
The Affinity Pump Console, Model 2000 provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure, timed events and battery status.
1
The AVECOR Affinity Pump has been designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyurethane chamber incorporating polycarbonate inlet and outlet connectors, which fills passively due to the filling pressure exhibited by the fluid height in a reservoir. The peristaltic action of the rollers against the filled pump chamber results in forward flow. Flow rate is affected by the combination of reservoir volume, RPM's of the pump rotor and the total circuit resistance.
The Stockert Instrumente (Sorin) Pump Console Model No. 10-10-00, provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure and timed events.
The Stockert Instrumente (Sorin) Pump is designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyvinyl chloride tubing segment with polycarbonate connectors at each end. Flow rate is affected by the RPM's, of the pump rotor, total circuit resistance and the size of the tubing segment used.
SPECIFICATIONS & MATERIALS
A detailed comparison of specifications and materials can be found in Table 1. Specifications for the Stockert (Sorin) console are taken from Stockert's labeling.
Table 1 Pump Console Comparison of Specifications
Affinity Console | Stockert Console | |
---|---|---|
Input Power | • 100 or 110-120 Vac, 50-60 Hz, | |
• 220-240 Vac, 50-60 Hz | • 100 or 110-120 Vac, 50-60 Hz | |
• 220-240 Vac, 50-60 Hz | ||
Internal Batteries | Yes | No (External Battery Pack Optional) |
Case Dimensions | 13" H x 7" W x 17" D | 12" H x 7" W x 19" D |
Weight | 50 lb. | 48 lb. |
2
Pump Chamber Comparison of Materials
Affinity | Stockert (Sorin) | |
---|---|---|
Chamber | Polyurethane | Polyvinyl Chloride |
Connectors | Polycarbonate | Polycarbonate |
Protective Caps | PVC | PVC |
INDICATIONS FOR USE
The Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are indicated for use in an extracorporeal perfusion circuit to pump blood during cardiopulmonary bypass procedures.
PERFORMANCE CHARACTERISTICS
The following tests were conducted by AVECOR to establish the safety and effectiveness of the Affinity Blood Pump System.
Test Description | Location |
---|---|
In Vitro Operating Characteristics | Section III |
In Vitro Cellular Damage | Section III |
Material Biocompatibility | Section IV |
Sterility | Section V |
Pyrogenicity | Section V |
Sterilant Residuals | Section V |
Software Validation & Verification | Appendix E |
Protocols describing the test methods utilized in this application may be found in the appropriate Appendix.
The results indicate that the Affinity Blood Pump System is substantially equivalent in performance to the Stockert Instrumente (Sorin) Pump System.
BLOOD TRAUMA SUMMARY STATEMENT
Blood trauma testing shows that in all aspects evaluated (hemolysis, platelet and white cell response) the Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are equivalent.
3
CONCLUSION
WE HAVE CONSIDERED DESIGN, MATERIALS, INDICATIONS FOR USE, PERFORMANCE CHARACTERISTICS, ALONG WITH A DIRECT COMPARISON OF TEST RESULTS BETWEEN THE AFFINITY AND STOCKERT PUMP SYSTEMS IN OUR APPLICATION. IT IS BASED UPON THE REVIEW OF ALL OF THESE PARAMETERS THAT AVECOR CARDIOVASCULAR INC. DETERMINES THAT THE AFFINITY BLOOD PUMP SYSTEM IS SUBSTANTIALLY EQUIVALENT TO THE CURRENTLY MARKETED STOCKERT INSTRUMENTE (SORIN) BLOOD PUMP SYSTEM.
4
Image /page/4/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's commitment to health, human services, and the future. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
Mr. Dennis E. Steger Director Requlatory Affairs/ Quality Assurance Avecor Cardiovascular, Inc. 7611 Northland Drive Minneapolis, Minnesota 55428 -
AUG - 5 1997 -
K964017 Re: AFFINITY™ Blood Pump System Regulatory Class: II (Two) Product Code: 74 DTQ Dated: May 9, 1997 Received: May 12, 1997
Dear Mr. Steger:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
5
Page 2 - Mr. Dennis E. Steger
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labelinq regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callehan
Thomas J. Callawan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
870.4370- DWB II - Roller-type CPB Blood Pump
Labels | Values |
---|---|
510(k) Number (if known): | K.964017 |
Device Name: | AFFINITY Blood Pump System |
Indications For Usc:
The Affinity™ Blood Pump System is intended to be used in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Dennis E. Hughes
Dennis E. Steger Director Regulatory Affairs/ Quality Assurance AVECOR Cardiovascular Inc.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bette Lempuele
Division Sign-Off) Division of Cardiovascular, Respiratory, nd Neurological Devices
510(k) Number: K964017
Prescription Use
(Per 21 CFR 801.109)
(){{
Over-The-Counter Use
(Optional Format 1-2-96)