(144 days)
The JMMAGE Immunochemistry System (TDM) CAR. PHE. PHY. and THE Reagents in conjunction with Beckman Drug Calibrator 1, are intended for use in the quantitative determination of carbamazepine. phenobarbital, phenytoin, and theophylline concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
The IMMAGE Immunochemistry System Carbamazepine (CAR) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of carbamazepine in human serum or plasma bv rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Phenobarbital (PHE) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of phenobarbital in human serum or plasma bv rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Phenytoin (PHY) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of phenytoin in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Theophylline (THE) reagent when, used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of theophylline in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE™ Immunochemistry Systems Drug Calibrator 1, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE Immunochemistry Systems for the calibration of Carbamazepine, Phenobarbital, Phenytoin, and Theophylline test systems.
The JMMAGE Immunochemistry System (TDM) CAR. PHE. PHY. and THE Reagents in conjunction with Beckman Drug Calibrator 1, are intended for use in the quantitative determination of carbamazepine. phenobarbital, phenytoin, and theophylline concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
Here's an analysis of the provided text, focusing on acceptance criteria and study details:
Device: IMMAGE™ Immunochemistry System Therapeutic Drug Monitoring Reagents [Carbamazepine (CAR), Phenobarbital (PHE), Phenytoin (PHY), and Theophylline (THE)]
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria provided in this document are primarily related to method comparison (correlation with a predicate device), stability, and imprecision. However, explicit numerical acceptance criteria are not stated; instead, the reported performance is presented, demonstrating equivalence to already marketed devices.
| Analyte | Acceptance Criteria (Implicit) | Reported Device Performance (Slope) | Reported Device Performance (Intercept) | Reported Device Performance (r-value) | Reported Device Performance (n, samples) |
|---|---|---|---|---|---|
| Carbamazepine | Substantial equivalence to predicate | 0.982 | 0.17 | 0.992 | 111 |
| Phenobarbital | Substantial equivalence to predicate | 0.998 | -1.18 | 0.996 | 103 |
| Phenytoin | Substantial equivalence to predicate | 1.051 | -0.76 | 0.996 | 107 |
| Theophylline | Substantial equivalence to predicate | 0.992 | 0.12 | 0.995 | 144 |
| Reagent | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| IMMAGE CAR, PHE, PHY, and THE | Substantial equivalence to predicate, demonstrating claimed stability | 24 month shelf-life |
| IMMAGE CAR, PHE, PHY, and THE | Substantial equivalence to predicate, demonstrating claimed stability | 14 day open container stability |
| IMMAGE CAR, PHE, PHY, and THE | Substantial equivalence to predicate, demonstrating claimed stability | 14 day calibration stability |
Note: The document only provides "Estimated Within-Run Imprecision" without specific numerical values, making it impossible to directly compare to acceptance criteria in a table format. It states generally: "These reagents exhibit acceptable imprecision for the intended use."
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Method Comparison - Test Set):
- Carbamazepine (CAR): 111 samples
- Phenobarbital (PHE): 103 samples
- Phenytoin (PHY): 107 samples
- Theophylline (THE): 144 samples
- Data Provenance: Not specified. It is likely internal data generated by Beckman Instruments, Inc. The document does not mention external sources or specific geographic locations for the samples.
- Retrospective or Prospective: Not specified. Standard practice for method comparison studies like these would typically involve prospective testing of patient samples against the predicate method, but this is not explicitly stated.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. This is a diagnostic device for quantitative determination of drug concentrations, not an AI or imaging device requiring human expert consensus for ground truth. The "ground truth" for the method comparison study is established by the results from the predicate device (Abbott TDx reagents).
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is not an AI or imaging device requiring human adjudication. The comparison is directly between the new device's measurements and the predicate device's measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI device or an imaging device involving human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The studies described are standalone performance studies. The IMMAGE™ Immunochemistry System, including these reagents, performs the quantitative determination of drug concentrations without human-in-the-loop interaction for result generation. The performance data presented (method comparison, stability, imprecision) reflects the device's intrinsic analytical capabilities.
7. The Type of Ground Truth Used
The ground truth for the method comparison studies was established by the predicate device's measurements. For example, for Carbamazepine, the Abbott TDx Carbamazepine reagent was used as the reference method.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI-based system that requires a "training set." It is a chemical immunoassay system where reagents react to determine concentrations.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
BECKMAN
Summary of Safety & Effectiveness
IMMAGE™ Immunochemistry System Therapeutic Drug Monitoring Reagents [Carbamazepine (CAR), Phenobarbital (PHE), Phenytoin (PHY), and Theophylline (THE)]
FEB - 4 1997
1.0 Submitted By:
Annette Hellie Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 993-8767 FAX: (714) 961-4457
2.0 Date Submitted:
12 September 1996
3.0 Device Name(s):
3.1 Proprietary Names
IMMAGE™ Immunochemistry System Carbamazepine (CAR) Reagent IMMAGE™ Immunochemistry System Phenobarbital (PHE) Reagent IMMAGE™ Immunochemistry System Phenytoin (PHY) Reagent IMMAGE™ Immunochemistry System Theophylline (THE) Reagent IMMAGE™ Immunochemistry Systems Drug Calibrator 1
3.2 Classification Names
Carbamazepine Test System (Not Classified) Phenobarbital Test System (21 CFR §862.3660) Phenytoin Test System (Not Classified) Theophylline Test System (21 CFR §862.3880) Calibrator (21 CFR §862.1150)
Predicate Device(s): 4.0
| IMMAGE SystemReagent | Predicate | Manufacturer | DocketNumber |
|---|---|---|---|
| Carbamazepine (CAR)Reagent | TDx®** CarbamazepineReagent | Abbott* Laboratories,Inc. | K904226 |
| Phenobarbital (PHE)Reagent | TDx Phenobarbital IIReagent | Abbott Laboratories,Inc. | K904226 |
| Phenytoin (PHY)Reagent | TDx PhenytoinReagent | Abbott Laboratories,Inc. | K904226 |
| Theophylline (THE)Reagent | TDx Theophylline IIReagent | Abbott Laboratories,Inc. | K922991 |
*Abbott Laboratories, Abbott Park, IL 60064
** Trademark of Abbott Laboratories
Beckman Instruments, Inc.
{1}------------------------------------------------
5.0 Description:
The JMMAGE Immunochemistry System (TDM) CAR. PHE. PHY. and THE Reagents in conjunction with Beckman Drug Calibrator 1, are intended for use in the quantitative determination of carbamazepine. phenobarbital, phenytoin, and theophylline concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
6.0 Intended Use:
The IMMAGE Immunochemistry System Carbamazepine (CAR) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of carbamazepine in human serum or plasma bv rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Phenobarbital (PHE) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of phenobarbital in human serum or plasma bv rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Phenytoin (PHY) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of phenytoin in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Theophylline (THE) reagent when, used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 1, is intended for the quantitative determination of theophylline in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE™ Immunochemistry Systems Drug Calibrator 1, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE Immunochemistry Systems for the calibration of Carbamazepine, Phenobarbital, Phenytoin, and Theophylline test systems.
{2}------------------------------------------------
Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System TDM (CAR, PHE, PHY, & THE) Reagents Summary of Safety & Effectiveness
7.0 Comparison to Predicate(s):
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| Reagent | Aspect/Characteristic | Comments |
|---|---|---|
| SIMILARITIES | ||
| IMMAGE System(CAR, PHE, PHY,THE) Reagents | Intended use. | Same as Abbott TDxReagents |
| Liquid stable reagents. | ||
| Initial analytic ranges. | ||
| DIFFERENCES | ||
| IMMAGE System(CAR, PHE, PHY,THE) Reagents | IMMAGE utilizes nephelometricinhibition immunoassay | Abbott TDx reagents utilizefluorescence polarizationimmunoassay |
| IMMAGE System(CAR, PHE, PHY)Reagents | Antibody source for IMMAGE CAR,PHE, and PHY reagents is mouse. | Antisera source for TDxCarbamazepine, Phenytoin,and Phenobarbital is sheep. |
{3}------------------------------------------------
8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method companson, stability, and imprecision experiments that relate results obtained from the Abbott TDx Reagents to the IMMAGE System Reagents.
| Analyte | Slope | Intercept | r | n | Predicate Method |
|---|---|---|---|---|---|
| Carbamazepine | 0.982 | 0.17 | 0.992 | 111 | Abbott TDxCarbamazepine |
| Phenobarbital | 0.998 | -1.18 | 0.996 | 103 | Abbott TDxPhenobarbital II |
| Phenytoin | 1.051 | -0.76 | 0.996 | 107 | Abbott TDxPhenytoin |
| Theophylline | 0.992 | 0.12 | 0.995 | 144 | Abbott TDxTheophylline II |
Method Comparison* Study Results IMMAGE System CAR. PHE. PHY, and THE Reagents
- Deming Regression statistics used
Stability Study Results
| Reagent | Product Claim |
|---|---|
| IMMAGE CAR, PHE, PHY, and THE | 24 month shelf-life14 day open container stability14 day calibration stability |
4
{4}------------------------------------------------
Estimated Within-Run Imprecision
| CLELE BLOUIS BLOGS (1) 1931 11 1994 11 1994 11 11 464 11 11 11 11 11 11 11 11 5AN AN ALANDARI AND AND AND AND------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.BRANNAGERThe for the contract and the contribution of the comments of the comments of the....and the comments of the consisted on the contribution of the comments of1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ARE JUN JUR - 7 & 15 PMLance and...2000 706. V. 2002 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1STERT THE COMMENT CONTRACTOR.11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1:1- Free & No------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AND COLLEGION COLLEGION. | |||||
|---|---|---|---|---|---|
| BE FIEMILE . F.. No Ar speases someone as a ca.----------------------------------- | |||||
| --------------------------------------------------------------------------------------------------------------------- | -------------------------------------------------------------------------------------------------------------- | ||||
| . | |||||
| Concession | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Haller State Controller------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------No. on a man manager from a from the first |
| The first and as an a consider of I T 1 T 1 T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T T. LEBEL LINE BE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. 1 . 1 1a parties parties and controlA CarA111 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 | .------------------------------------..And Charles of College of Concession of Concession of Concession of Concession of Concession of Concession of Concession of Concession of Concession of Concessional Concessio------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A | AND AN AND ANLINER OR ONE AND AND..Caroline Company | A CALL ALL ANDContraction of the Career Carder Comments of the Comments of10 8 10011 | ...Canada Career Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come. |
|---|---|---|---|---|
| money computer and of the late to the least to the least to the least toCall & Callery & Children & Children CalleryAND AND A BOOK OF COLLECTION15 On AmountCALLER OF CHILDRING------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ||||
| ----------A COLLECT OF CARD AND | ||||
| . | ||||
| -------------------- | Comments of the program and the management of the management of the many of | Call Resident Line Laures Bron |
| CON REAL CONSULER OR COLLECT AND LE CONSUL..1 %C.V. 1 %C.V. 1 - 1 - 1 - 1 - 1 - 1 - 1 -And Address of 10ARLA STANKLER.CAR BARA" " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "Sample Barren. BE STATE:.21222 8 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------Comprehension of the control of the contribution of the larges and any and a many a.TOIN (PHY) R.P 1143 | |||||||
|---|---|---|---|---|---|---|---|
| -------------------------------------------------------------------------------------------------------------------------- | |||||||
| ...CON COLOR LANDER EXCHINE IS CONSUMICALLER & CCHELES MERCERS OF BOOKS OF CHICLE LEASE | .1.1 4 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1A . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Parties . Portugues . Portugue.A-MOND AND AND A. | 111 B R B R F R B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B B..------------------------------------------------------------------------------------------------------------------------------------------------------------------------------And and the property of | 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 -100 1000 1000 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 | A BAR BERRELL E E A R REAR A REAL S CARRENA MARK MA..Componential Confident Advance Col Concession Call Concession Leich Learning | ||
|---|---|---|---|---|---|---|
| ALMER IN ARE | ||||||
| -------------------. | ALL RESERUCTION | |||||
| CARDE LEAR & LE LEASE & AND ANDRESS OF ANTIQUE OF CANADIAN OF CANADIAN OF CHARACT AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | B.A. MANAGE A. LANS OR SENSION, OF AND | ||||
| The Comments of Children of Children |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
§ 862.3645 Neuroleptic drugs radioreceptor assay test system.
(a)
Identification. A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.(b)
Classification. Class II.