K Number
K963521
Manufacturer
Date Cleared
1997-08-06

(337 days)

Product Code
Regulation Number
872.3640
Panel
DE
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IMZ implants are indicated for single tooth replacement, as immediate abutments on long span to bridgework, as distal abutments on free-end edentulousareas to be restored with fixed bridgework, to support overdentures in totally or partially edentulous arches, and as abutments supporting a full arch fixed prosthesis in the totally edentulous mandible or maxila.

Device Description

Not Found

AI/ML Overview

I am sorry, but based on the provided document, there is no information available regarding the acceptance criteria for a device, the study that proves the device meets the acceptance criteria, or any of the specific details requested in your prompt (e.g., sample size, expert qualifications, adjudication method, MRMC study, standalone performance, ground truth type, training set details).

The document is a 510(k) clearance letter from the FDA for a device called "IMZ Twin Plus - Cylinder," which are dental implants. It states that the device is substantially equivalent to legally marketed predicate devices and outlines the indications for use. However, it does not contain the detailed performance study information you are seeking.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.