(35 days)
These Mirror Finish Bipolar Forceps are indicated for use in any surgical procedure that requires tissue coagulation and/or cauterization to achieve hemostasis. They are indicated for use in:
- general surgery,
- neurosurgery,
- urology, and nephrology.
The JJPI Bipolar Forceps Line is a modification of the bipolar forceps we currently have on the market. These forceps are made of either stainless steel or titanium. The only difference between our currently marketed line of bipolar forceps and the new line under consideration is the mirror finish proprietary coating that is applied to the tips.
The provided SSED (Summary of Safety & Effectiveness Data) for the Johnson & Johnson Professional, Inc. (JJPI) Mirror Finish Bipolar Forceps Line does not contain the level of detail typically found in reports for AI-based medical devices regarding acceptance criteria and detailed study metrics. This document refers to a traditional medical device (bipolar forceps) and therefore the questions provided for AI/ML device studies are not directly applicable.
However, I can extract the relevant information that is available about the "acceptance criteria" (which in this context refers to demonstrating substantial equivalence) and the "study" (which refers to functional testing and physical characteristic comparisons).
Here's an interpretation based on the provided text, addressing the questions where possible and indicating when the information is not present due to the nature of the device:
1. Table of Acceptance Criteria and Reported Device Performance
For traditional medical devices seeking substantial equivalence, "acceptance criteria" are not typically expressed as numerical performance metrics like sensitivity, specificity, or AUC, as they would be for AI/ML devices. Instead, they relate to demonstrating that the new device is as safe and effective as a predicate device. The "performance" is shown by functional equivalence and adherence to established design or material specifications.
Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance (Mirror Finish Bipolar Forceps) |
---|---|
Intended Use: Device must have similar indications for use as predicate devices. | Indicated for tissue coagulation and/or cauterization to achieve hemostasis in general surgery, neurosurgery, urology, and nephrology. This is identical to the predicate JJPI Bipolar Forceps Line and the Kirwan Non-Stick Bipolar Forceps. |
Fundamental Technology/Mechanism of Action: Device must function similarly to predicate devices. | Uses bipolar technology for tissue coagulation/cauterization. The only difference from the predicate JJPI Bipolar Forceps Line is a proprietary mirror finish coating on the tips. |
Design Characteristics: Key design features (tip size, shape, reusability) must be comparable to predicate devices. | Tip Size: 0.25 mm - 2 mm (matches JJPI Bipolar Forceps Line, comparable to Kirwan's 0.5 mm - 2 mm, which falls within this range). |
Forceps Shape: Bayonet and straight (matches both predicate devices). | |
Reusable: Yes (matches both predicate devices). | |
Material Composition: Materials should be safe and comparable, especially concerning patient contact. | Made of stainless steel or titanium. No nickel (matches JJPI Bipolar Forceps Line, Kirwan's materials not fully specified but no nickel noted). |
Non-Stick Qualities: Must possess or effectively implement non-stick properties where claimed, comparable to predicate devices. | Yes, explicitly stated to have non-stick qualities, and the report emphasizes the "mirror finish proprietary coating" for this purpose. This is a claimed similarity to both predicate devices. |
Functional Equivalence: Device must perform its intended function (e.g., coagulation, non-stick properties) as effectively as the predicate device(s). | The SSED states: "The functional testing and the basic mechanical properties exhibited by the JJPI non-stick forceps indicate that the JJPI proprietary coated tips are functionally equivalent to both the JJPI Bipolar Forceps Line and the KIRWAN Non-Stick Bipolar Forceps." |
2. Sample Size Used for the Test Set and the Data Provenance
This information is not present in the provided document. For traditional devices undergoing functional or mechanical testing, sample sizes would typically refer to the number of devices tested or the number of simulated uses. Data provenance (country, retrospective/prospective) is not relevant for this type of device submission focused on substantial equivalence through material and functional comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not present and is generally not applicable to the type of device and study described. There's no "ground truth" to establish through expert consensus in the sense of image interpretation for an AI/ML device. The "ground truth" here is rather the established performance of the predicate device (JJPI Bipolar Forceps Line and Kirwan Non-Stick Bipolar Forceps) through their own regulatory clearances and extensive market use, against which the new device is compared in terms of physical and functional characteristics.
4. Adjudication Method for the Test Set
This information is not present and not applicable. Adjudication methods are used in scenarios where multiple reviewers interpret data, which is not described here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This type of study is specific to AI-assisted diagnostic or interpretative devices and is not relevant to bipolar forceps.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
No, a standalone algorithm performance study was not done. This is not an AI/ML device.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" as typically applied to AI/ML devices (e.g., pathology for a cancer diagnosis algorithm) is not applicable here. The "ground truth" in this context is the established safety and effectiveness of the predicate devices. The study demonstrates that the "Mirror Finish Bipolar Forceps" are substantially equivalent to these predicates based on:
- Design and Material Specifications: Directly comparing physical characteristics and materials.
- Functional Testing: The statement "The functional testing and the basic mechanical properties exhibited by the JJPI non-stick forceps indicate that the JJPI proprietary coated tips are functionally equivalent..." implies that the device was subjected to tests demonstrating its ability to perform its intended function (e.g., coagulation, non-stick properties) at a level comparable to the predicate devices. The details of these tests are not provided but would typically involve benchtop testing of electrical properties, coagulation efficacy, tip adherence, durability, etc.
8. The Sample Size for the Training Set
This information is not present and not applicable, as this is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not present and not applicable, as this is not an AI/ML device.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.