K Number
K963127
Device Name
ELECSYS FT3
Date Cleared
1996-12-16

(126 days)

Product Code
Regulation Number
862.1710
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elecsys FT3 is an in vitro quantitative determination of free triiodothyronine (FT3) in human serum and plasma. Triiodothyronine is one of the thyroid hormones present in serum which regulate metabolism.

Device Description

The Elecsys FT3 employs a competitive test principle with polyclonal antibodies directed against T3 and with streptavidin microparticles and electrochemiluminescence detection.
1st Incubation: Sample (30 µl) and a specific anti-T3 antibody labeled with a ruthenium complex.
2nd Incubation: After addition of biotinylated T3 and streptavidin-coated microparticles, the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin.
The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode.

AI/ML Overview

The medical device in question is the Elecsys® FT3, an in vitro quantitative determination of free triiodothyronine (FT3) in human serum and plasma.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Elecsys FT3 device are not explicitly stated as distinct pass/fail thresholds in the provided document. Instead, the performance characteristics are presented in comparison to a predicate device (Enzymun-Test® FT3) and, for method comparison, to a reference method (Elkins Equilibrium Dialysis). The substantial equivalence is based on demonstrating comparable performance.

Here's a table summarizing the reported device performance, with implicit acceptance being comparable to the predicate or demonstrating robust performance metrics:

FeatureAcceptance Criteria (Implicit from Predicate/Reference)Elecsys FT3 Performance
Precision (Within-run %CV)Comparable to Enzymun-Test FT3 (3.8 - 2.3%)PreciControl Universal 1: 2.0% PreciControl Universal 2: 2.5% HS1: 2.3% HS2: 2.1% HS3: 2.3% (Generally better or comparable to predicate)
Precision (Total run %CV)Comparable to Enzymun-Test FT3 (4.5 - 2.4%)PreciControl Universal 1: 3.6% PreciControl Universal 2: 3.3% HS1: 4.7% HS2: 3.1% HS3: 2.7% (Generally comparable to or better than predicate)
Lower Detection Limit (LDL)Comparable to Enzymun-Test FT3 (0.47 pmol/l, 0.5 pg/ml)0.40 pmol/l (0.26 pg/ml) (Lower, indicating better sensitivity)
Assay RangeComparable to Enzymun-Test FT3 (0.47-46 pmol/l)0.40 - 50 pmol/l (0.26 - 32 pg/ml) (Slightly wider upper range and lower LDL)
Method Comparison (vs. Enzymun-Test FT3)Strong correlation (e.g., r > 0.90)Least Squares: y = -0.22 + 0.90x, r = 0.94 Passing/Bablok: y = -1.45 + 1.06x, r = 0.94 (Strong correlation, close to ideal y=x line)
Method Comparison (vs. Elkins Equilibrium Dialysis)Strong correlation (e.g., r > 0.95, given it's a reference method)Least Squares: y = 0.99x + 0.203, r = 0.998 (Excellent correlation with the reference method)
Interfering SubstancesNo significant interference at specified concentrationsHemoglobin: No interference up to 1 g/dl Lipemia: No interference up to 1500 mg/dl Bilirubin: No interference up to 25 mg/dl Biotin: No interference up to 30 ng/ml
Specificity (% cross-reaction)Low cross-reactivity for related substancesD-T3: 98.86% (high, as expected due to similarity to L-T3) L-T4: 0.115% D-T4: 0.115% L-rT3: 0.007% 3,3',5-tri-iodothyroacetic acid: 106.4% 3,3',5,5'-tetra-iodothyroacetic acid: 0.007% (Generally low for non-T3 thyroid hormones, with expected high for similar compounds)

2. Sample Sizes and Data Provenance for Test Set

  • Precision Study (Elecsys FT3):
    • Sample Size (N): 60 for each of 5 different samples (PreciControl Universal 1, PreciControl Universal 2, HS1, HS2, HS3).
    • Data Provenance: Not explicitly stated, but the methodology refers to NCCLS (Modified) (EP5-T2), which are standard clinical laboratory guidelines. Implies clinical laboratory samples. Retrospective or prospective is not specified.
  • Precision Study (Enzymun-Test FT3, Predicate):
    • Sample Size (N): 120 for each of 3 different samples (1, 2, 3).
    • Data Provenance: Not explicitly stated. Methodology refers to NCCLS "Midi" (EP5-T).
  • Method Comparison vs. Enzymun-Test FT3:
    • Sample Size (N): 298.
    • Data Provenance: Not explicitly stated. Assumed to be human serum/plasma samples. Retrospective or prospective is not specified.
  • Method Comparison vs. Elkins Equilibrium Dialysis:
    • Sample Size (N): 22.
    • Data Provenance: Not explicitly stated. Assumed to be human serum/plasma samples. Retrospective or prospective is not specified. Given the small number, it might be a specific set of clinical samples where equilibrium dialysis was performed.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

This device is an in vitro diagnostic (IVD) assay for measuring a specific analyte (FT3). The "ground truth" for such devices is typically established through analytical methods and comparison to established reference methods, not by human expert opinion like in imaging studies.

  • Elkins Equilibrium Dialysis is referred to as a "reference method" for direct measurement of fT4 and fT3. This method itself serves as a highly accurate "ground truth" and does not involve human expert consensus or adjudication in interpreting the result. It's a laboratory gold standard for free thyroid hormone measurement.

Therefore, the concept of "number of experts" and their "qualifications" for establishing ground truth in the context of an IVD assay's performance evaluation (like for diagnostic accuracy in an imaging test) is not directly applicable here.

4. Adjudication Method for the Test Set

Not applicable. For an IVD device measuring an analyte, the "ground truth" is determined by the result of the reference method (e.g., Elkins Equilibrium Dialysis) or the predicate device, not by adjudication of human expert interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is an in vitro diagnostic (IVD) assay, not an imaging or interpretation-based diagnostic device. MRMC studies are not relevant for evaluating the performance of an analytical assay.

6. Standalone (Algorithm Only) Performance Study

Yes, the studies presented are all standalone performance evaluations of the Elecsys FT3 device (the "algorithm" in this context refers to the assay's chemical and detection methodology). The precision, sensitivity, assay range, method comparisons, and interference studies directly assess the device's inherent analytical performance without human intervention in the result determination process once the sample is loaded.

7. Type of Ground Truth Used

  • Reference Method: For the method comparison study, Elkins Equilibrium Dialysis served as a robust reference method for determining "true" FT3 concentrations.
  • Predicate Device: The Enzymun-Test® FT3 served as a comparator for demonstrating substantial equivalence. The results from the predicate were effectively a "ground truth" for comparison purposes in this context.
  • Known Concentrations: For precision and sensitivity studies, quality control materials (e.g., PreciControl Universal) with known or targeted concentrations and human serum samples (HS1, HS2, HS3) were used, providing a basis for evaluating repeatability and reproducibility.

8. Sample Size for the Training Set

The document describes performance characteristics, not the development of a machine learning algorithm that requires a "training set." This device is a biochemical assay. Therefore, the concept of a "training set" in the context of machine learning is not applicable. The development and optimization of the assay would involve various experiments and technical iterations to establish reagents, protocols, and instrument parameters, but this is distinct from "training data" for an AI model.

9. How Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the machine learning sense for this device. The biochemical principles and laboratory testing methods inherently define the assay's "truth."

{0}------------------------------------------------

K963127

Image /page/0/Picture/1 description: The image is a dark, low-resolution photograph. The image is mostly black, with a few small, blurry white spots scattered throughout. Due to the poor quality of the image, it is difficult to discern any specific objects or details.

510(k) Summary

DEC 1 6 1996

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1) Submitter name, address, contactBoehringer Mannheim Corporation9115 Hague Rd.Indianapolis, IN 46250(317) 845-2000
Contact Person: LeeAnn Chambers
Date Prepared: August 9, 1996
2) Device nameProprietary name: Elecsys® FT3
Common name: free triiodothyronine test system
3) Predicate deviceWe claim substantial equivalence to the Enzymun-Test® FT3.
4) Device DescriptionThe Elecsys FT3 employs a competitive test principle with polyclonal antibodies directed against T3 and with streptavidin microparticles and electrochemiluminescence detection.
1st Incubation: Sample (30 µl) and a specific anti-T3 antibody labeled with a ruthenium complex.2nd Incubation: After addition of biotinylated T3 and streptavidin-coated microparticles, the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin.

{1}------------------------------------------------

Image /page/1/Picture/1 description: This image describes the device, intended use, and indications for use of a medical device. The device description explains that the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. The intended use is for the in vitro quantitative determination of free triiodothyronine (FT3) in human serum and plasma. The indications for use explain that triiodothyronine is one of the thyroid hormones present in serum which regulate metabolism.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image provides information on the indications for use and comparison to predicate devices. It mentions that a variety of methods are available for estimating free thyroid hormone levels, with direct measurement of fT4 and fT3 via equilibrium dialysis or ultrafiltration being a reference method. The image also lists references and similarities between the Boehringer Mannheim Elecsys FT3 and the Enzymun-Test® FT3, including intended use, test principle, sample type, antibody, and capture principle.

{3}------------------------------------------------

Image /page/3/Picture/10 description: The image is completely black. There are a few small white dots scattered across the image. The dots are not concentrated in any particular area. The image does not contain any discernible objects or shapes.

F. Substantial Differences: equivalence,

cont.

FeatureElecsys FT3Enzymun-Test FT3
Reaction testprinciplestreptavidin microparticles andelectrochemiluminescencestreptavidin-coated tubes andenzyme immunoassay
technologytechnology
Samplevolume30 µ1100 µl
InstrumentrequiredElecsys 2010ES 300
Calibrationa two point calibration renewalis recommended every 7 daysor 1 month if the same reagentlot is used and the reagent packis consumed within 7 daysa full calibration curve run isrecommended every 2 weeks

Performance Characteristics:

FeatureElecsys FT3Enzymun-Test FT3
PrecisionNCCLS (Modified) (EP5-T2)NCCLS "Midi" (EP5-T)
SamplePC U1*PC U2*HS1HS2HS3123
N6060606060120120120
Mean7.6420.85.1110.514.42.717.8628.97
wi/in run %CV2.02.52.32.12.33.82.32.7
total run %CV3.63.34.73.12.74.52.43.1
SensitivityLower Detection Limit:Lower Detection Limit:
0.40 pmol/l0.47 pmol/l
0.26 pg/ml0.5 pg/ml
Assay range(LDL to higheststandard)0.40 - 50 pmol/l0.47 - 46 pmol/l
0.26 - 32 pg/ml0.5 - 30 pg/ml
  • PC U1 and PC U2 = PreciControl Universal 1 and 2

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is completely black. There are no discernible shapes, objects, or figures within the frame. The image appears to be a solid, uniform color with no variations in tone or texture. It is impossible to provide any further details or analysis due to the lack of visual information.

F. Substantial

Performance Characteristics, continued:

equivalence,

cont.

Method Comparisonvs. Enzymun-Test FT3vs. Elkins Equilibrium Dialysis
Least SquaresLeast Squares
N = 298N = 22
y = - 0.22 + 0.90xy = 0.99x +0.203
r = 0.94r = 0.998
Passing/Bablok
N = 298
y = - 1.45 + 1.06x
r = 0.94
Interfering substance:No interference up to:No interference up to:
Hemoglobin1 g/dl1 g/dl
Lipemia1500 mg/dl1250 mg/dl
Bilirubin25 mg/dl17 mg/dl
Biotin30 ng/ml30 ng/ml
Specificity% cross reaction% cross reaction
D-T398.86140
L-T40.1150.16
D-T40.1150.07
L-rT30.0070.04
3,3',5-tri-iodothyroacetic acid106.47.5
3,3',5,5'-tetra-iodothyroacetic
acid0.0070.01

§ 862.1710 Total triiodothyronine test system.

(a)
Identification. A total triiodothyronine test system is a device intended to measure the hormone triiodothyronine in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases such as hyperthyroidism.(b)
Classification. Class II. This device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.