K Number
K963080
Manufacturer
Date Cleared
1996-09-27

(50 days)

Product Code
Regulation Number
862.1680
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This in vitro method is intended to quantitatively measure testosterone in human servin on a Technicon Immuno 1 system. Measurements of Testosterone are used in the diagnosis and treatment of various hormonal sexual disorders.

Device Description

This document describes the Immuno 1 Testosterone method, an in vitro method for quantitatively measuring testosterone in human serum on a Technicon Immuno 1 system. It compares the performance of this method to a predicate device, the DPC Coat-a-Count Total Testosterone RJA kit, in terms of minimum detection concentration, precision, and correlation. It also provides a specificity comparison by listing the crossreactivity percentages of various substances with both methods.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Bayer Immuno 1 Testosterone Method:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (from predicate device/comparison)Immuno 1 Testosterone Method PerformanceAcceptance Met?
Minimum Detection Concentration0.04 ng/mL (DPC predicate)0.04 ng/mLYes
Precision (Between-Run)0.93 ng/mL: 12%2.64 ng/mL: 6.4%6.72 ng/mL: 6.0% (DPC predicate)0.86 ng/mL: 6.3%3.13 ng/mL: 3.5%8.58 ng/mL: 2.2%Yes (Immuno 1 shows equal or better precision across all levels)
Correlation with Predicate Device (DPC)Implicitly, a strong linear correlation with the DPC Coat-a-Count Total Testosterone RJA kity = 0.98x + 0.01n = 335r = 0.990S = 0.41 ng/mLYes (Very strong correlation, indicating agreement)
Specificity (Cross-reactivity - examples)Aldosterone: 0%Androstenedione: 0.5%5α-Dihydrotestosterone: 3.3%19-Nortestosterone: 20%11β-Hydroxytestosterone: 0.8% (DPC predicate)Aldosterone: ND (None Detected)Androstenedione: 0.4%5α-Dihydrotestosterone: 4.7%19-Nortestosterone: 7.6%11β-Hydroxytestosterone: 14.2%Generally consistent or improved for many, some differences noted (e.g., 5α-Dihydrotestosterone higher, 19-Nortestosterone lower for Immuno 1). The overall intent is to show comparable specificity, which appears to be met given the wide range of cross-reactants examined.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 335 samples were used for the correlation study comparing the Immuno 1 method to the DPC predicate device.
  • Data Provenance: Not explicitly stated. The document does not mention the country of origin of the samples or whether they were retrospective or prospective. It only mentions "human serum."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes a comparison study between two in vitro diagnostic devices, not a study involving human interpretation of images or data. Therefore, the concept of "experts establishing ground truth" in the typical sense (e.g., radiologists reviewing scans) is not applicable here. The ground truth for the Immuno 1 device is implicitly established by its correlation with a predicate device (DPC Coat-a-Count Total Testosterone RJA kit), which itself is an established and cleared diagnostic method.

4. Adjudication Method for the Test Set

Not applicable. This is a direct comparison of quantitative measurements from two analytical instruments, not a subjective assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

Not applicable. This study concerns the performance of an in vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this is effectively a standalone performance study of the Immuno 1 Testosterone method. It evaluates the device's analytical performance (minimum detection, precision, correlation, specificity) independent of human interpretation or interaction beyond standard laboratory procedures for running the assay.

7. The Type of Ground Truth Used

The "ground truth" for the Immuno 1 Testosterone method in this submission is the results obtained from the DPC Coat-a-Count Total Testosterone RJA kit (predicate device). The study's goal is to demonstrate substantial equivalence by showing that the Immuno 1 method produces comparable results to a legally marketed and accepted device.

8. The Sample Size for the Training Set

The concept of a "training set" is not applicable in this context. This is a clinical validation study for an in vitro diagnostic device, not a machine learning algorithm that requires training data. The device's internal calibration and optimization would have occurred during its development, but the document does not provide details on that "training" process.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" with established ground truth is not relevant to this type of device and study as presented. The device is validated against an established predicate, and its internal mechanisms are based on biochemical principles rather than trained on data.

{0}------------------------------------------------

SEP 2 7 10

SUMMARY OF SAFETY AND EFFECTIVENESS

Testosterone Method for the Bayer Immuno 1 ° System

Listed below is a comparison of the performance between the Immuno 1 Testosterone method and a similar device that was granted clearance of substantial equivalence (DPC Coat-a-Count Total Testosterone RJA kit). The information used in the Summary of Safety and Effectiveness was extracted from the Immuno 1 Testosterone method sheet and the DPC Total Testosterone method sheet.

INTENDED USE

This in vitro method is intended to quantitatively measure testosterone in human servin on a Technicon Immuno 1 system. Measurements of Testosterone are used in the diagnosis and treatment of various hormonal sexual disorders.

METHODImmuno 1TestosteroneDPC(predicate Device)
Part No.Reagents T01-3694-51Calibrators T03-3695-01TKTT1 (100 tubes)
Minimum Det. Conc.0.04 ng/mL0.04 ng/mL
Precision (Between-Run)0.86 ng/mL 6.3%3.13 ng/mL 3.5%8.58 ng/mL 2.2%0.93 ng/mL 12%2.64 ng/mL 6.4%6.72 ng/mL 6.0%
Correlationy=0.98x + 0.01wherey=Immuno 1 Total Testosterone methodx=DPC Total Testosterone methodn=335r=0.990S =0.41 ng/mL

Gabe Muraca Bayer Corporation 511 Benedict Ave Tarrytown, NY 10591 Rev. Date: 09/16/96

{1}------------------------------------------------

Specificity

CrossreactantImmuno 1% CrossreactivityDPC Coat-a-Count% Crossreactivity
AldosteroneND0
5β-Androstan-3α,17β-diol0.30.4
Androstenedione0.40.5
5α-Androstan-3β,17β-diol1.20.04
5-Androsten-3β,17β-diol0.80.2
5α-Androstan-3,17-dioneND0.05
AndrosteroneND0
CorticosteroneND0
CortisolND0
CortisoneND0
DanazolND0.09
11-DeoxycortisolND0
DexamethasoneND0
DHEAND0
DHEA-SulfateND0.01
5α-Dihydrotestosterone4.73.3
Estradiol0.10.02
5(10)-Estren-17α-ethynyl-17β-ol-3-OneND0.2
4-Estren-7α-methyl-17β-ol-oneND1.1
19-Nortestosterone7.620
EstroneND0.01
Ethisterone0.50.7
FluoxymesteroneND0.01
19-HydroxyandrostenedioneND2
11-Ketotestosterone-1.216
11β-Hydroxytestosterone14.20.8
Methyltestosterone2.21.7
Norethindrone0.10.1
PrednisoneND0
ProgesteroneND0
SpironolactoneND0
Triamicinolone-0.10.2
5α-Androstan-3α-,17β-diol0.2NR
DesoxycortisoneNDNR
EstriolNDNR
11β-HydroxyandrostanedioneNDNR
11β-HydroxyandrostenedioneNDNR
17-HydroxyprogesteroneNDNR

NR= None reported
ND=None detected

ND=None detected

Gabe Muraca Bayer Corporation 511 Benedict Ave
Tarrytown, NY 10591 Rev. Date: 09/16/96

§ 862.1680 Testosterone test system.

(a)
Identification. A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.(b)
Classification. Class I.