K Number
K962626

Validate with FDA (Live)

Date Cleared
1996-10-02

(89 days)

Product Code
Regulation Number
880.6760
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Kenad Limb Holders are indicated for medical purposes where a patient's arm and leg movement must be limited to the extent necessary for treatment, examination or protection of the patient or others.

Device Description

The Kenad Limb Holders subject to this Premarket Notification are depicted in the figure at ATTACHMENT 6. As depicted in this figure. Kenad Limb Holders are assemblies of the following modules: Ankle/Wrist Cuff, Ankle/Wrist Strap, Connector Strap, Quick Release Buckle and Bed Strap.

The Ankle/Wrist Cuff is a rectangular, two sided material. On the side which contacts skin, the cuff is made from blue or white polyurethane foam. At one end of the rectangle, a patch of Velcro hooks are stitched to the foam with 12 pound nylon thread. Heat bonded to the foam back is a piece of blue or white Velcro pile.

Except for Black Limb Holder with Removable Cuff (Catalog # 50-1600), the Ankle/Wrist Strap is affixed to the Ankle/Wrist Cuff by sewing using 12 pound nylon thread. For Black Limb Holder, a Velcro hook strip is sewn to one end of the Ankle/Wrist Strap. The Velcro hook strip then affixes to the Velcro pile of the cuff. Affixed by sewing to one end of the Ankle/Wrist Strap is a polished metal buckle for locking the Ankle/Wrist strap in place around the ankle/wrist.

The Connector Strap is affixed by sewing with 12 pound nylon thread to the Ankle/Wrist Strap. Affixed by sewing to the other end of the Connector Strap is the female or male end of the Quick Release Buckle.

The Bed Strap comes as either a loop able or threadable end for anchoring to a fixed object. Affixed by sewing to one end of the Connector Strap is the complementary female or male end of the Quick Release Buckle.

The material used to make the straps is woven polypropylene. The polypropylene color is either black or white, and its width is either a nominal 1 or 1.5 inch. The variations are outlined in more detail in the Table also included at ATTACHMENT 6.

AI/ML Overview

This 510(k) summary describes a medical device called the "Patient Protective Limb Holder (Universal)" and its variations. This product is a limb restraint system. The submission relies on a predicate device (Danek Medical, Inc.'s "Patient Protective Limb Holder", K823815). The information provided does not describe a clinical study in the traditional sense, but rather relies on a historical performance review.

Here's an analysis of the provided text with respect to your questions:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Safety (No patient injury)"During this time [over 5 years], Kenad received no patient injury or product failure complaints." (Implied: The absence of complaints over a significant usage period indicates safety.)
Effectiveness (No product failure)"During this time [over 5 years], Kenad received no patient injury or product failure complaints." (Implied: The absence of complaints over a significant usage period indicates effectiveness for its intended use of limiting limb movement.)

Explanation: The document does not explicitly state quantitative acceptance criteria or a formal study design with measured outcomes against those criteria. Instead, it leverages a retrospective review of complaint data as evidence of safety and effectiveness.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "More than 150,000 pairs of Kenad Limb Holders were sold over the last 5 years. Assuming 100% usage, there have been more than 300,000 contacts between ankles and/or wrists over the last 5 years." This suggests the "test set" consists of all these instances of product use.
  • Data Provenance: The data is retrospective, derived from the company's internal complaint records over a 5-year period. The country of origin of the data is not explicitly stated, but given the company's address (Memphis, TN), it is likely US-based.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not applicable. This submission does not involve expert-established ground truth as it is not a diagnostic device or a device requiring human interpretation of results. The "ground truth" here is the absence of recorded complaints over extensive real-world use.

4. Adjudication Method for the Test Set

  • Not applicable. There was no formal adjudication process for a "test set" in the context of this submission. The performance relies solely on the company's complaint logging system.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study was not conducted as this is a medical restraint device, not a diagnostic imaging device typically requiring human reader performance analysis.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This is a physical medical device, not an algorithm or AI. The "performance" assessment is based on real-world usage without reported failures or injuries.

7. The Type of Ground Truth Used

  • Real-world performance/Complaint Data: The ground truth is established by the absence of patient injury or product failure complaints over a large volume of actual product usage in clinical settings. This serves as evidence that the device performs as intended for its stated indications without causing harm.

8. The Sample Size for the Training Set

  • Not applicable. This submission does not involve a "training set" in the context of an algorithm or machine learning. The device's design and manufacturing are based on established engineering principles and prior experience, not data-driven machine learning.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set mentioned, there is no ground truth established for it.

{0}------------------------------------------------

510(k) SUMMARY

PERSONNEL AND DATE (807.92(a)(1)) 1.

1.1. Submitter

Kenad SG Medical Inc. 2692 Huntley Memphis, TN 38132

1.2.. Contact Person

Kenneth Elmore President and CEO Kenad SG Medical Inc.

1.3.. Date Of summary Preparation

This summary was prepared in April and May 1996.

2. DEVICE NAME (807.92(a)(2))

  • As noted in the Cover Letter, five similar devices are grouped within .. .. this 510(k) notification. These devices differ only in minor design aspects. Their function, application and use is identical.

2.1. Trade Or Proprietary Name

Patient Protective Limb Holder (Universal) is the trade or proprietary name for the devices which are the subject of this Premarket Notification Submission. Within this name, there are five Kenad Limb Holders.

KENAD Catalog #Description
50-1400White Limb Holder
50-1500Deluxe Blue Limb Holder
50-1500eEconomy White Limb Holder
50-1600Black Limb Holder w/Removable Cuff
50-1700Pediatric Limb Holder

{1}------------------------------------------------

2.2. Common Or Usual Name

Protective Restraint 21CFR 880.6760

3. LEGALLY MARKETED DEVICE (807.92(a)(3))

The Kenad Limb Holders that are the subject of this Premarket Notification Submission are comparable to Patient Protective Limb Holder from Danek Medical. Inc. Danek successfully filed a premarket notification for Patient Protective Limb Holder at Document Control number K823815, January 17, 1983.

4. DEVICE DESCRIPTION (807.92(a)(4)) (p 11)

4.1. General Description

The Kenad Limb Holders subject to this Premarket Notification are depicted in the figure at ATTACHMENT 6. As depicted in this figure. Kenad Limb Holders are assemblies of the following modules: Ankle/Wrist Cuff, Ankle/Wrist Strap, Connector Strap, Quick Release Buckle and Bed Strap.

The Ankle/Wrist Cuff is a rectangular, two sided material. On the side which contacts skin, the cuff is made from blue or white polyurethane foam. At one end of the rectangle, a patch of Velcro hooks are stitched to the foam with 12 pound nylon thread. Heat bonded to the foam back is a piece of blue or white Velcro pile.

Except for Black Limb Holder with Removable Cuff (Catalog # 50-1600), the Ankle/Wrist Strap is affixed to the Ankle/Wrist Cuff by sewing using 12 pound nylon thread. For Black Limb Holder, a Velcro hook strip is sewn to one end of the Ankle/Wrist Strap. The Velcro hook strip then affixes to the Velcro pile of the cuff. Affixed by sewing to one end of the Ankle/Wrist Strap is a polished metal buckle for locking the Ankle/Wrist strap in place around the ankle/wrist.

The Connector Strap is affixed by sewing with 12 pound nylon thread to the Ankle/Wrist Strap. Affixed by sewing to the other end of the Connector Strap is the female or male end of the Quick Release Buckle.

{2}------------------------------------------------

The Bed Strap comes as either a loop able or threadable end for anchoring to a fixed object. Affixed by sewing to one end of the Connector Strap is the complementary female or male end of the Quick Release Buckle.

The material used to make the straps is woven polypropylene. The polypropylene color is either black or white, and its width is either a nominal 1 or 1.5 inch. The variations are outlined in more detail in the Table also included at ATTACHMENT 6.

4.2. How Limb Holders Function

The Ankle/Wrist Cuff is placed around the ankle/wrist, and held in place by Velcro. The cuff is then surrounded by the Ankle/Wrist Strap, and held in place by a buckle. The Bed Strap is affixed to an immovable object either by looping or threading, and the Bed Strap is connected to the Connector Strap via the Quick Release Buckle. Slack is removed from the Connector Strap so that the limb is comfortably held close to the bed.

INTENDED USE (807.92(a)(5)) 5.

5.1. ... Indications

Kenad Limb Holders are indicated for medical purposes where a patient's arm and leg movement must be limited to the extent necessary for treatment, examination or protection of the patient or others.

5.2. Patient Population

Kenad Limb Holders are used with adults and children.

5.3. Indications Different From Substantially Equivalent Device

The indications and patient population for the Kenad devices are identical to the legally marketed device cited above at Section 10.

{3}------------------------------------------------

6. COMPARISON TO LEGALLY MARKETED DEVICE (807.92(a)(6))

The Kenad Limb Holders that are the subject of this Premarket Notification Submission are comparable to Patient Protective Limb Holder from Danek Medical, Inc. A comparison of these devices was provided at Section 6 in the Premarket Notification Submission.

7. PERFORMANCE DATA (807.92(b))

More than 150,000 pairs of Kenad Limb Holders were sold over the last 5 years. Assuming 100% usage, there have been more than 300,000 contacts between ankles and/or wrists over the last 5 years. During this time, Kenad received no patient injury or product failure complaints.

8. SAFETY, EFFICACY AND PERFORMANCE CONCLUSION (807.92(b)(3))

The absence of patient injury or product failure complaints following more than 300,000 product uses over 5 years indicates these products are safe and effective.

{4}------------------------------------------------

ATTACHMENT 6

:

/

:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a diagram of a restraint system. The system includes an ankle/wrist cuff and strap, a connector strap, and a quick-release buckle. The system also includes a bed strap that is loopable and a bed strap that is threadable.

ﯿﻨﭧ- ﭘﺎﮐﺴﺘﺎﻥ ﮐﮯ ﻟﯿﻨﮉ ﮐﮯ ﺍﻧﺘﻈﺎﻣﯿﺎﺕ ﮐﮯ ﺷﮩﺮ

i

{6}------------------------------------------------

CATALOG #50-1400WhiteLimb Holder50-1500eEconomyLimb Holder50-1500DeluxeLimb Holder50-1700PediatircLimb Holder50-1600DetachableCuffLimb Holder
CUFF
ColorWhiteWhiteBlueBlueBlue
Dimensions (Nominal)12 X 312 X 313 X 39 X 213 X 3
STRAP (Ankle/Wrist, Bed,Connector)
ColorWhiteWhiteWhiteWhiteWhite
Width1.5"1"1.5"1"1.5"
BED STRAP
AttachmentThreadableLoopableLoopableLoopableLoopable
QUICK RELEASE

and the comments of the comments of the comments of the comments of

the submit and the submit of the submit of the subject of the subject of

្រី

§ 880.6760 Protective restraint.

(a)
Identification. A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.