K Number
K962125

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-08-02

(60 days)

Product Code
Regulation Number
862.2160
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This section contains a brief statement of intended use for the method as a in vitro diagnostic procedure for quantitating analytes in human serum.

Device Description

The Technicon Immuno 100 system is an automated clinical immunoassav analyzer. All tests can be run in random access, batch, and STAT modes at a throughput rate of up to 120 tests per hour. Up to twenty-two (22) methods can be loaded on the system at one time. The system performs latex agglutination and magneticseparation enzyme immunoassays on serum. Magneticseparation enzyme immunoassays are divided into two (2) groups: sandwich assays and competitive assays.

AI/ML Overview

The provided text describes the operational aspects, calibration, and quality control of the Technicon Immuno 1 system, an automated clinical immunoassay analyzer. It focuses on the general procedures for handling reagents, calibrators, controls, and includes details about how performance characteristics like imprecision, correlation, and sensitivity are determined. However, the document does not contain a study or specific data proving that the device (the Technicon Immuno 1 system itself) meets an overall acceptance criterion.

Instead, the document details how individual assays/methods performed on the system have their performance characteristics established, and how laboratories should verify ongoing performance through quality control and calibration. The acceptance criteria mentioned are primarily for quality control and calibration, not for the overall device performance in a single study.

Therefore, many of the requested items cannot be directly extracted from this document, as it is a general manual for the system, not a specific performance validation study.

Here's an breakdown of what can be inferred or directly stated from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes how performance characteristics are defined and measured for individual assays, rather than providing an acceptance criterion for the overarching "device" or its overall performance in a single study. The key performance criteria for individual assays are:

Acceptance Criterion Type (for individual assays)Reported Device Performance (General description of how it's measured/evaluated)
Imprecision(Within-run, Total)Evaluated using human serum pools/control materials according to NCCLS EP5-T2, using two systems, 20 days, 1-2 runs/day, 2+ samples/run. Users should obtain data passing Chi Square test at 95% confidence. Values are "point estimates of average imprecision."
CorrelationDetermined using human serum samples from external evaluation sites, comparing Technicon Immuno 1 with reference/comparative systems. Regression statistics computed per NCCLS EP9-T.
Sensitivity (Minimum Detectable Concentration)Smallest concentration detected within a run, equivalent to two within-run standard deviations of the Level 1 (0.0) calibrator. Measured by running 0.0 calibrator 20 times on three systems.
SpecificityLists highest concentration of common interfering substances evaluated and found not to have a significant effect. Also lists % observed cross-reactivity.
Analytical RangeDescribes the concentration range for acceptable results.
Quality Control (QC) AcceptanceControl results must meet the laboratory's established criteria for acceptability. If not, patient test results from that run must be evaluated, and corrective actions (e.g., recalibration) taken. Controls should be within ±2 total standard deviations.
Calibration StabilityMinimum duration of calibration stability is established. Users may extend if QC indicates acceptable performance (control recovery within ±2 total standard deviations).
Reagent Stability (On-System)Minimum stable period stated in method sheet. Established by monitoring control recovery means within ±2 total standard deviations after activation.

2. Sample size used for the test set and the data provenance

For Imprecision:

  • Sample Size: For each method, imprecision was evaluated using "two (2) systems per method, twenty (20) days per system, one to two (1 - 2) runs per day, and two (2) or more samples per run." The samples are "human serum pools and control materials."
  • Data Provenance: Not explicitly stated (e.g., country of origin). The study design appears to be prospective in nature, following a protocol similar to NCCLS document EP5-T2.

For Correlation:

  • Sample Size: "human serum samples." The number is not specified.
  • Data Provenance: "from external evaluation sites." Not explicitly stated (e.g., country of origin, retrospective or prospective). Similar to NCCLS document EP9-T.

For Sensitivity:

  • Sample Size: The "0.0 calibrator is run twenty (20) times on three (3) systems." This refers to runs, not unique patient samples.
  • Data Provenance: Not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not mention the use of experts to establish ground truth for a test set in the context of device performance validation. The performance characteristics are determined using reference methods, calibrators, and control materials, according to established NCCLS guidelines.

For "Expected Values," asymptomatic donors' sera (approx. 100) were used, and the central 95% range was calculated. This is a statistical determination, not an expert panel establishing ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

No adjudication method is mentioned. The performance studies rely on quantitative measurements against reference methods or defined calibrator values.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was mentioned. The device is an automated immunoassay analyzer, not an AI-assisted diagnostic imaging tool with human interpretation. Therefore, this question is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The Technicon Immuno 1 system is an automated analyzer, meaning it operates in a standalone manner for performing immunoassays. Its performance characteristics (imprecision, correlation, sensitivity) described are standalone performance metrics of the instrument. Human intervention is primarily for sample loading, maintenance, calibration, quality control, and interpretation of results by laboratory personnel.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance characteristic studies appears to be based on:

  • Reference methods: For correlation studies, the Technicon Immuno 1 system is compared against "a reference method or a comparative system."
  • Defined calibrator values: For sensitivity, the "Level 1 (0.0) calibrator" is used. For imprecision and stability, control materials and human serum pools with known or expected values underpin the evaluation.
  • Statistical methods: For "Expected Values," statistical calculation (nonparametric or parametric) from asymptomatic donors' sera is used.

8. The sample size for the training set

The document refers to performance validation studies, not "training sets" in the context of machine learning. The system's operation is based on established biochemical principles and calibration, not machine learning models that require a distinct training set. Therefore, this question is not applicable in the typical sense.

9. How the ground truth for the training set was established

Not applicable, as no machine learning training set is described. The analytical methods are based on chemical reactions and optical density measurements, which are calibrated and controlled using established laboratory standards.

{0}------------------------------------------------

INTRODUCTION

AUG. 2 1906

INTRODUCTION The Technicon Immuno 100 system is an automated clinical immunoassav analyzer. All tests can be run in random access, batch, and STAT modes at a throughput rate of up to 120 tests per hour. Up to twenty-two (22) methods can be loaded on the system at one time.

The system performs latex agglutination and magneticseparation enzyme immunoassays on serum. Magneticseparation enzyme immunoassays are divided into two (2) groups: sandwich assays and competitive assays.

Latex agalutination assays utilize the competition between the analyte in the sample and the hapten-Ficoll" conjugate in Reagent 1 for binding sites on antibody-coated latex particles in Reagent 2. The binding of the hapten-Ficoll conjugate to the coated latex particles results in the formation of large aggregates which increases the turbidity of the reaction solution. The analyte in the sample inhibits the formation of these aggregates. The agglutination rate is measured by monitoring the optical density (turbidity) of the reaction liquid.

Magnetic separation competitive assays utilize the competition between the analyte in the sample and the analyte attached to the alkaline phosphatase enzyme in Reagent 2 for binding sites on the fluoresceinated antibody in Reagent 1.

Magnetic particles coated with an antibody to fluorescein are added to the mixture of sample, Reagent 1 and Reagent 2. The fluorescein molecule conjugated to the analyte antibody, bound to either the analyte in serum or to the analyte conjugated to the alkaline phosphatase enzyme, binds to the magnetic particles. After removal of excess reagent and the addition of substrate, the absorbance of the reaction liquid is read to determine the analyte concentration.

In magnetic separation sandwich assays, binding occurs at two different sites on a single analyte in the sample. Reagent 1 (antibody conjugated to fluorescein) and Reagent 2 (antibody conjugated to alkaline phosphatase enzyme) are added simultaneously to the sample. An excess of the two reagents drives the reaction. Binding of the resulting immune complex to a magnetic particle allows it to be separated from excess reagents. Upon addition of the substrate, the reaction absorbance is read to determine the analyte concentration. The reaction rate is directly proportional to the sample analyte level.

STATEMENT OF CONTENT

This Manual contains specific, assay-related information and data for Bayer methods performed on a Technicon Immuno 1 system.

The procedures necessary for the preparation, use, and storage of Technicon Immuno 1 system reagents and calibrators are provided in the applicable method sheet. Also refer to the tab titled "Reagent Information"; pages INTRO-8 through INTRO-11.

SAMPLE COLLECTION AND PREPARATION1-4

WARNING - POTENTIAL BIOHAZARDOUS MATERIAL Any samples of human blood, plasma, or serum should be handled cautiously as a biohazardous material according to good laboratory practices.

The blood sample should be collected in a commercially available collection tube using routine venipuncture techniques. During sample collection and preparation, hemolysis of erythrocytes should be avoided. The effect of hemolysis on a method is described in the "Interferences" section of the specific method sheet when it is significant. Samples should be processed in a manner that will prevent the introduction of clots into the system.

    1. After drawing the specimen from the patient, perform identification procedures consistent with your laboratory protocol. Place the collection tube in an upright vertical position and allow the specimen to stand undisturbed for a period of thirty (30) minutes prior to centrifugation.
    1. Before placing the specimen in the centrifuge, carefully rim the collection tube to free the clot from the glass. Perform this procedure carefully so that the cells are not hemolyzed.
    1. For optimum separation, the specimen should be centrifuged for twenty (20) minutes at a speed that produces a relative centrifugal force (RCF) of 2200 x Consult centrifuge manufacturers gravity ర్తో instructions for further information.
    1. The primary collection tube can be placed directly into the sample rack (refer to the Technicon Immuno 1 System Operator's Manual - UNIT 2 for allowable tube sizes) or carefully decant the serum from the primary collection tube into a clean Technicon Immuno 1 system sample cup or secondary tube. Make certain that no clots are decanted into the clean cup.

The serum should be separated from cells within two (2) hours after collection. If samples cannot be assayed immediately after separation, they should be stored in stoppered containers at 4 °C to 8 °C.

When determining the sample requirements, include the applicable dead volume: collection tubes 500 uL (maximum) and insert cups 1 mL (75 uL) or 2 mL (90 uL).

CAUTION!: To avoid having insufficient sample, carefully dispense specimen so that no air bubbles are entrapped in bottom of cup.

Special analyte-specific storage conditions such as freezing, protection from direct exposure to light, or the addition of a preservative will be noted in the method sheets where applicable.

Additional sample collection and preparation information is available in texts by Kaplan and Pesce and Tietz. In addition, the National Committee for Clinical Laboratory Standards (NCCLS) has published two documents on sample collection.

Ficoll is a trademark of Pharmacia Fine Chemical Inc., Piscataway, NJ Throughout this publication Bayer Corporation, Business Group Diagnostics is referred to as "Bayer."

{1}------------------------------------------------

ALIBRATOR AND CONTROL PRODUCTS

WARNING - POTENTIAL BIOHAZARDOUS MATERIAL

Any product prepared from human blood, plasma, or serum should be handled cautiously as a biohazardous material according to good laboratory practices.

Each donor unit used in the preparation of these products was tested by an FDA approved method for the presence of the antibody to Human Immunodeficiency Virus (HIV) as well as for hepatitis B surface antigen and found to be negative (was not repeatedly reactive).

Because no test method can offer complete assurance that HIV, hepatitis B virus, or other infectious agents are absent, these products should be handled at the Biosafety Level II as recommended for any potentially infectious human blood specimens in Protection of Laboratory Workers From Infectious Disease Transmitted by Blood, Body Fluids, and Tissue - Second Edition; Tentative Guideline (1991), Document M29-T2, oromulaated by the National Committee for Clinical aboratory Standards (NCCLS).

Table 1 lists the calibrator and control products available for use with the system. Refer to the current Price List, applicable method sheet, or package insert for further information. Also afer to the tab titled "Reagent Information"; pages INTRO-8 rough INTRO-11.

Table 1: CALIBRATORS AND CONTROLS

PARTNUMBERDESCRIPTIONPACKAGING(mL)
T03-3187-01Technicon SETpoint™ AFP Calibrator 1Technicon SETpoint™ AFP Calibrators 2-61 x 4.05 x 2.0
T03-3188-01Technicon SETpoint™ CEA Calibrator 1Technicon SETpoint™ CEA Calibrators 2-61 x 4.05 x 2.0
T03-3562-01Technicon SETpoint™ CA 125 II† Calibrator 1Technicon SETpoint™ CA 125 II Calibrators 2-61 x 4.05 x 2.0
T03-3554-01Technicon SETpoint™ CA 15-3† Calibrator 1Technicon SETpoint™ CA 15-3 Calibrators 2-61 x 4.05 x 2.0
T03-3563-01Technicon SETpoint™ CA 19-9† Calibrator 1Technicon SETpoint™ CA 19-9 Calibrators 2-61 x 4.05 x 2.0
T03-3562-01Technicon SETpoint™ CK-MB Calibrators 1-66 x 2.0
T03-3252-01Technicon SETpoint™ Cortisol Calibrator 1Technicon SETpoint™ Cortisol Calibrator 2-6Cortisol Calibrator Reconstitution Diluent1 x 4.0††5 x 2.0††1 x 15.0
T03-3251-01Technicon SETpoint™ Ferritin Calibrator 1Technicon SETpoint™ Ferritin Calibrators 2-61 x 4.05 x 2.0
T03-3596-01Technicon SETpoint™ Folate Calibrator 1Technicon SETpoint™ Folate Calibrators 2-6Folate Calibrator Reconstitution Diluent1 x 4.0††5 x 2.0††1 x 15.0
T03-3541-01Technicon SETpoint™ PSA Calibrator 1Technicon SETpoint™ PSA Calibrators 2-61 x 4.05 x 2.0
03-3148-01Technicon SETpoint™ Reproductive CalibratorsReproductive Calibrators Diluent6 x 2.0††1 x 15.0
PARTNUMBERDESCRIPTIONPACKAGING(mL)
T03-2864-01Technicon SETpoint™ TDM Calibrator 1Technicon SETpoint™ TDM Calibrators 2-6TDM Calibrator Reconstitution Diluent1 x 4.0††5 x 2.0††1 x 15.0
T03-3174-01Technicon SETpoint™ T4 Calibrator 1Technicon SETpoint™ T4 Calibrator 2-61 x 4.05 x 2.0
T03-3568-01Technicon SETpoint™ TSH Calibrator 1Technicon SETpoint™ TSH Calibrator 2-61 x 4.05 x 2.0
T03-2872-01Technicon SETpoint™ T3 Calibrator 1Technicon SETpoint™ T3 Calibrator 2-61 x 4.05 x 2.0
T03-3076-01Technicon SETpoint™ T Uptake Hypothyroid ControlTechnicon SETpoint™ T Uptake Euthyroid ReferenceTechnicon SETpoint™ T Uptake Hyperthyroid Control1 x 2.01 x 2.01 x 2.0
T03-3393-01Technicon Immuno 1 TESTpoint™ Ligand Control2 x 5.0(3 levels)
T03-3464-01Technicon SETpoint™ Vitamin B12 Calibrator 1Technicon SETpoint™ Vitamin B12 Calibrator 2 - 61 x 4.05 x 2.0
T03-3401-01Technicon SETpoint™ Free T4 Calibrators6 x 2.0

Table 1: CALIBRATORS AND CONTROLS (Cont)

†† When reconstituted as directed.

SYSTEM SOLUTIONS

Tr TO TO TO TO TO

To

Table 2 lists the products currently available for use with Technicon Immuno 1 systems. Those system solutions that are necessary to perform this method may also require preparation. Refer to section titled "Start of the Day" in the "OPERATION" section of the Technicon Immuno 1 System Operator's Manual for complete instructions on the preparation and use of the system solutions or refer to tab titled "Reagent Information"; pages INTRO-8 through INTRO-11.

PARTNUMBERDESCRIPTIONPACKAGING
T01-3130-01Substrate ReagentsReagent: 2 x 386.0 mgDiluent: 2 x 62.0 mL2 empty, labeled cassettes
T01-3543-01mIMP™ ReagentsmIMP Reagent1 x 23.5 mLIMP Buffer1 x 23.5 mL
T01-3601-01IMP® Buffer Reagent2 x 23.5 mL
T01-2879-52IMP® Wash Solution2 x 4.0 L
T01-3060-52Cuvette Wash Solution 12 x 4.0 L
T01-3570-52Cuvette Wash Solution 22 x 4.0 L
T01-3061-01Sample Probe Fluid6 x 30.0 mL
T01-3062-01Reagent Probe Fluid6 x 100.0 mL
T03-3574-01Sample Diluent-BReagent: 6 x 50 mL†

Table 2: SYSTEM SOLUTIONS

† - When reconstituted as directed.

A trademark of Centocor Inc., Malvem, PA 19355-1307

When reconstituted as directed.

{2}------------------------------------------------

Table 2: SYSTEM SOLUTIONS (Cont)

PARTNUMBERDESCRIPTIONPACKAGING
T01-3367-01System TroubleshootingReagentsNative ALP Reagent: 1 x 10 mLNative ALP Buffer: 1 x 15.0 mLBuffer Cassette: 2 x 15.0 mL(5X) ALP Reagent: 1 x 10.0 mLEnzyme Cassette: 1 x 15.0 mLEnzyme Cassette: 1 x 30.0 mL3 Empty, labeled cassettes

OPERATION

All information pertaining to the operation of the analyzer is included in the applicable Technicon Immuno 1 System Reference or Operator's Manual provided with your system.

NOTICE

Any modification of computer disks and/or the programs contained on such disks can adversely affect the control of instrument performance and thereby invalidate the results obtained and the claims that have been made regarding system performance.

Bayer cannot be responsible for errors that are introduced by, or result from, any modification or alteration of the computer disks and/or the programs contained on such disks by the user, or for any direct or indirect resulting consequences from such modification or alteration.

CALCULATION OF RESULTS

The system automatically performs all calculations that are necessary for obtaining final results. The system prints out the results for each sample in concentration units. The derivation of results from raw absorbance data is explained in the "Calculation of Results" section of the applicable Technicon Immuno 1 System Reference or Operator's Manual.

INTERPRETATION OF RESULTS

System operators and laboratory supervisors are responsible for operating and maintaining Bayer products in accordance with the procedures described in the applicable Product Labeling (Operator Manuals, Package Inserts, Reference Bulletins, etc.) and for determining that these Manuals, products conform to the applicable performance claims.

If, under these prescribed conditions of operation and maintenance, an aberrant or abnormal result, as defined by the laboratory protocol, occurs, laboratory personnel should first make certain that the system is performing and is being operated in accordance with the Product Labeling; then follow the laboratory protocol for advising the clinician of a result that appears to have deviated from the norms established by the laboratory.

Bayer products do not make diagnoses on patients. Bayer intends its diagnostic products (systems, reagents, software, hardware, etc) to be used to collect data reflecting the patient's chemical, hematological, or immunological status at a certain point in time. Such data must be used in conjunction with other diagnostic information and with the attending physician's evaluation of the patient's condition to arrive at a diagnosis and a clinical course of treatment.

Any malfunction of a Bayer diagnostic product (i.e., failure to meet a performance specification or to perform as intended) should be appropriately addressed by laboratory personnel. Various sections of the Product Labeling previously noted address malfunctions and their possible effect on results.

Automatic Reagent Cassette Switching

The system provides the capability of allowing the user to routinely place multiple reagent cassettes of the same reagent with the same or different lot number in the reagent tray. When the reagent inventory for the lead cassette counts down to zero, the system software will automatically switch to the subsequent cassette(s) to complete the remaining tests in the worklist for that method.

The lead cassette identification and the sequencing of subsequent cassettes follow these resident software rules:

    1. The cassette having the earliest installed date is the designated lead.
    1. If two (2) cassettes are installed on the same day, the cassette having the fewest number of remaining tests is the designated lead.
    1. If two (2) or more cassettes are installed on the same day and the number of tests remaining is equal, then the cassettes are sequenced in accordance with their ascending numerical reagent tray position.

In the event multiple cassettes of differing lot numbers are installed, the system will roll over to a cassette with a different lot number. If the different lot has a stored and accepted calibration, then test results will be printed using the lot specific calibration data. If a within lot calibration has not been accepted on the different lot then rate data will be obtained, but held internally. Results for the different lot number will be calculated and released when a within-lot calibration is performed and accepted. Until then, the results for that test remains suppressed and "WA" flags are printed in their place.

Automatic Substrate Cassette Switching

Up to five (5) substrate cassettes can be installed on the system. When the substrate cassette in use reaches "0 Tests Left," the system will automatically roll over and begin to use the next available substrate cassette.

The system has resident software rules to select which cassettes to roll over to or to use first:

    1. The substrate cassette with the earliest installation date and lowest number of tests remaining.
    1. If the installation date and number of tests remaining are equal, the substrate cassette in the lowest numerical position will be used first.

Recommendations for Use

To determine if multiple reagent cassettes are installed on the system, check the "REAGENT INVENTORY SCREEN." If the number of tests ordered is larger than the number of tests left in the primary cassette, then cassette roll over will occur.

NOTES

    1. Quality Control: As described in the analyte specific Method Sheets, control samples should be assayed at the beginning of each shift or at some other interval chosen by the laboratory; whenever a new reagent is used (same or different lot) prior to reporting patient results; following maintenance or cleaning of any detection system component. Refer to the "Quality Control" section of the analyte specific Method Sheet.

{3}------------------------------------------------

ﻣ ﺘﻘﺔ

Calibration: As described in the analyte specific Method Sheets, calibration should be performed when a method is implemented on the system. Recalibration is required after replacement of major components, a change in the lot number of reagents, mIMP Reagents, Substrate Reagents, or as indicated by quality control results. Refer to the "Calibration" section of the analyte specific Method Sheet for details.

  • If desired, cassette roll over may be disabled by not placing 3. secondary cassettes in the reagent tray.

WASTE DISPOSAL REQUIREMENTS

; and regulations enacted to protect the environment and to encourage resource conservation, require the disposal of hazardous and biohazardous wastes in a specified manner.

Some of the wastes from Technicon Immuno 1 system can be classified as hazardous or biohazardous wastes. It is essential that the laboratory take appropriate steps to determine the laws and regulations applicable to their disposal and to effect compliance.

It is necessary for you to sample instrument effluent in order to evaluate compliance with applicable regulations. The laboratory should contact a local licensed biohazardous waste disposal firm for assistance.

The principal wastes associated with the use of the Technicon Immuno 1 system are reagent cassettes, reaction travs, reagent containers, sample containers, and system waste.

CAUTION: Do not incinerate the reaction trays.

Sample containers with human specimens, control materials, and all reagents, also should be handled and disposed of in accordance with the prevailing regulations and quidelines of ~ncies holding jurisdiction over the laboratory.

  • er to the product label and to Material Safety Data Sheets letails concerning any special precautions related to the handling of reagents.

SYSTEM METHOD SHEETS

Each Method Sheet provides information relevant to an individual analyte. The Method Sheets are alphabetized within the manual and contain information and data described in NCCLS Publication GP2-A and by the FDA in 21 CFR. A laboratory procedure manual is mandatory for good laboratory practice and regulatory compliance. To aid in the preparation of a laboratory procedure manual, refer to Table 3 (pages INTRO-12 and INTRO-13) for a cross-reference of the information contained in our Method Sheets with the guidelines for Clinical Laboratory Procedure Manuals published by the National Committee on Clinical Laboratory Standards (NCCLS Approved Guideline GP2-A).

Bayer Method Sheets are presented in a consistent format with the headings listed below. The following is a brief description of the information included under each section heading:

TECH-CHECK" TABLE

This table contains information which can be used to reference assay specific information, i.e. method principle, CLIA complexity, range of reportable results, specimen type, sample test volume, minimum sensitivity, standardization, and units used in the assay.

INTENDED USE

This section contains a brief statement of intended use for the method as a in vitro diagnostic procedure for quantitating analytes in human serum.

SUMMARY AND EXPLANATION

This section contains a short history of the method with pertinent references to previous related methods.

PRINCIPLES OF THE PROCEDURE

This section provides a physical, chemical, and biological explanation of the method and a description of the chemical reactions, where applicable.

REAGENTS

This section describes the reagents required to perform the method including reagent packaging information, product descriptions, materials required but not provided, and reagent preparation instructions. Analyte specific calibrator information also may be included.

Throughout this Section three words are used to signal the presence of hazards associated with the use and handling of reagents. These words are - in descending order of severity - DANGER!, WARNING!, and CAUTION!

The following briefly explains the significance of each of these headings:

DANGER!

Signals the presence of a hazard that is life threatening. This hazard can result from the presence of, the handling of, or the use of substances that are highly toxic, extremely corrosive, or extremely flammable.

WARNING!

Signals the presence of a hazard that is injurious but not life threatening under ordinary circumstances. This hazard can result from the presence of, the handling of, or the use of substances such as toxic liquids or gases, irritants, or flammable materials.

CAUTION!

Signals the presence of a hazard that could cause illness, burns, skin reactions, etc. This hazard could result from the presence of, the handling of, or the use of substances such as toxic liquids or gases, irritants, or flammable materials.

The customer must maintain a constant awareness of the hazards and warnings associated with all reagents. Where necessary, appropriate safety equipment should be used (e.g., eye protection, gloves, lab coats). Users should follow the handling and safety precautions specified in the Material Safety Data Sheets provided by Bayer.

Reagent and Calibrator Preparation

This section describes the preparation and use instructions for the reagents and calibrators that are unique to each method. The system solutions that are necessary to perform each method also require preparation. Refer to the section titled "Start of the Day" in the "OPERATION" section of the Technicon Immuno 1 System Operation Manual for complete instructions on the preparation and use of the system solutions or the tab titled "Reagent Information"; pages INTRO-8 through INTRO-11.

** * vailable from National Committee for Clinical Laboratory Standards, East Lancaster Avenue, Villanova, PA 19085

{4}------------------------------------------------

STORAGE AND STABILITY

Reagent and calibrator storage and expiration dates are printed on the product label. When stored unopened these products are stable through the last day of the month stated in the expiration date.

The on-system stability stated in the method sheet represents the minimum stable period for the working reagent. On-system reagent performance was established by monitoring control recovery means within ±2 total standard deviations after activation. The total standard deviation is obtained from the imprecision claims of the method. The user may elect to use reagents for a longer period if results from the recommended Quality Control Protocol indicate acceptable performance. Bayer assumes no responsibility for the performance of any reagents which are used beyond their expiration dates.

Method sheet minimum stability claims apply to all situations where the relative humidity is equal to or above 25%. At lower relative humidities between 15% and 25%, results from the recommended Quality Control Protocol may indicate that more frequent calibration may be required.

The system solutions when stored as indicated (unopened at appropriate temperature) are stable through the last day of the month stated in the expiration date.

The system solutions listed below are stable on-system after reconstitution or opening for the period indicated in Table 4.

Table 4: SYSTEM SOLUTIONS STABILTY
DESCRIPTIONSTABILITY (DAYS)
Substrate Reagent30
mIMP Reagent60
IMP Wash Solution30
Cuvette Wash Solution 130
Cuvette Wash Solution 230
Sample Probe Fluid30
Reagent Probe Fluid30
Sample Diluent-B30

Tebla 4: System sollutions stabli tv

If substrate cassettes are to be removed from the system, and temporarily stored in a 2 °C to 8 °C refrigerator, protect the contents from light. Evaporation covers provide adequate dust and evaporation protection for refrigerator storage.

SAMPLE HANDLING

For a detailed explanation of sample handling, refer to page INTRO-1 of this manual under the heading "Sample Collection and Preparation."

Refer to the specific Technicon Immuno 1 system, method sheets for details about special analyte specific sample handling requirements or dilutions.

MATERIALS REQUIRED BUT NOT PROVIDED

This section lists the additional materials other than the reagents and calibrators that are required to perform the methods.

PROCEDURE

This section refers the user to the "OPERATION" section of the Reference or Operation Manual for general operating instructions.

QUALITY CONTROL

This section specifies recommended quality control material and quality control frequency for the individual method. In general, Bayer recommends that Technicon Immuno 1 systems be monitored daily using the Technicon TESTpoint Ligand Controls (Prod. No. T03-3393-01) or, at least two (2) levels of commercially prepared control materials. Controls should be run at the beginning of each shift, or at some interval chosen by the laboratory whenever a new reagent is used, and following the performance of any detection system maintenance or cleaning. The laboratory must evaluate all control results prior to reporting patient results. If control results fail to meet the laboratory's established criteria for acceptability, all patient test results obtained in the unacceptable test run must be evaluated to determine if patient test results were adversely affected. The laboratory should take and document appropriate corrective actions, which may include recalibration, before reporting patient results.

If commercially prepared controls are used. the laboratory must verify that they respond properly on the system.

CALIBRATION

This section specifies the calibration type, frequency and procedure, for each method and identifies the calibration material to be used, where appropriate.

Calibration Procedure

This section refers the user to the "OPERATION" section of the Technicon Immuno 1 System Reference or Operation Manual for specific calibration instructions.

Calibration Schedule

This section lists the minimum duration of calibration stability with the use of multiple reagent cassettes within a single lot of reagent. By monitoring control recovery of newly activated cassettes within a single reagent lot, Bayer has determined that the results remain within ±2 total standard deviations. The total standard deviation is obtained from the imprecision claims of the method. The user may elect to use a calibration for a longer period of time if results from the recommended Quality Control Protocol continue to indicate acceptable performance.

RESULTS

This section describes the numeric printout, expected flags or messages, and specific conditions or cautions that apply to the method.

{5}------------------------------------------------

ﻟﻠﺒﻴﻌﺔ

ITATIONS OF THE PROCEDURE

The section describes substances, conditions, and patient states that have been reported to cause actual or apparent physiological changes in the analyte concentration.

Bayer makes every effort to identify in the Troubleshooting, Maintenance, and Bibliography sections of the Product Labeling, the potential effects caused by instruments, reagents, or endogenous substances, including known interferences on

nt results. However, it is impossible to provide information would be applicable to every clinical situation. This information is provided in lieu of attempting to define the need for additional testing which involves medical judgement in individual cases. The decision whether or not to report a diagnostic result rests with the laboratory.

EXPECTED VALUES

Expected values were determined from sera of asymptomatic donors Approximately one-hundred (100) adult donor specimens were assayed and the central 95% range was calculated by one of two procedures:

  • · Nonparametric (2.5 and 97.5 percentiles)
  • · Parametric (mean ± 2 standard deviations)

Any deviations from these procedures are noted in the method sheets

Ranges of expected values can vary depending on age, diet. location, etc; therefore, users should evaluate their usual population to determine their own laboratory specific ranges.

erapeutic Drug Assays

therapeutic ranges were taken from the literature cited in ... ch method sheet.

PERFORMANCE CHARACTERISTICS

This section provides technical information related to method performance characteristics, including imprecision, correlation, linearity, and sensitivity.

A description of each statistical parameter and the protocol followed to compute that parameter are contained in the paragraphs that follow.

Imprecision

Imprecision is defined as the degree of variability among repeated independent measurements of the same sample using external evaluation data.

Within-run imprecision refers to the reproducibility of the method when samples are assayed within a single run.

Total imprecision extends the within-run estimates to include between run and between day components of variation.

Imprecision estimates were computed at specified analyte levels using human serum pools and control materials in a protocol similar to that recommended in the NCCLS document EP5-T2, Precision Performance of Clinical Chemistry Devices -Second Edition; Tentative Guideline (1992).

All imprecision data were obtained over a single calibration period using a single lot of individual reagents and calibrator. Unless otherwise noted, imprecision was evaluated using two (2) systems per method, twenty (20) days per system, one to two (1 - 2) runs per day, and two (2) or more samples per run.

The imprecision values for each method, described on each of their respective Method Sheets, are point estimates of average imprecision that can be expected on the system. Since this is a point estimate, any single system tested with a similar protocol could vield a different estimate of imprecision and yet not be significantly different from the published average imprecision.

The NCCLS document, EP5-T2, describes a statistical test (Chi Square) for comparing an instrument's performance and the expected performance. A user should be able to obtain imprecision data which pass the Chi Square test at the 95% confidence level.

Correlation Data

Correlation data were determined using human serum samples from external evaluation sites by comparing the performance of Technicon Immuno 1 system with the performance of a reference method or a comparative system, where applicable.

Regression statistics were computed using a protocol similar to that recommended in the NCCLS document EP9-T, Method Comparison and Bias Estimation Using Patient Samples; Tentative Guideline (1993).

Regression data in each method indicate the linear least squares fit between the Technicon Immuno 1 system (y) and the comparative system/method (x), unless noted otherwise.

Sensitivity

This section lists the smallest concentration that can be detected within a run. It is equivalent to two (2) within-run standard deviations of the Level 1 (0.0) calibrator. The 0.0 calibrator is run twenty (20) times on three (3) systems.

Specificity

This section lists the highest concentration of common interfering substances evaluated and found not to have a significant affect on results. Substances that could affect results are shown by a listing of the percent (%) observed crossreactivity.

Analytical Range

This section describes the concentration range across which acceptable results can be obtained from the method.

Some specimens may exceed the range of the method and produce error flags or messages. The details of these flags and messages are available in the Reference Manual for the Technicon Immuno 1 system. Specimens that produce these errors should be diluted using a Class A volumetric pipette or equivalent and reassayed according to directions available in the specific method sheets.

{6}------------------------------------------------

Revision Bar

A vertical black bar appearing in the margin of the page indicating change(s) in the text and/or in the contents of a table in the document. The revision bar allows the reader to identify quickly and easily the revised material within the document.

BIBLIOGRAPHY

  • 1 Kaplan LA and Pesce AG: Clinical Chemistry -- Theory, Analysis, and Correlation, St. Louis, CV Mosby (1984) pp 43-50
  • 2 Tietz NW: Textbook of Clinical Chemistry, Philadelphia, PA, WB Saunders Company, pp 478-518 (1986)
  • 3 Standard Procedure for the Collection of Diagnostic Blood Specimens by Venipuncture. NCCLS Standard ASH-3. Villanova, PA, National Committee for Clinical Laboratory Standard (1977)
  • 4 Approved Standard Procedures for the Collection of Diagnostic Blood Specimens by Skin Puncture, NCCLS Standard H4-A. Villanova, PA, National Committee for Clinical Laboratory Standard (1982)

Image /page/6/Picture/9 description: The image shows the Bayer logo. The word "Bayer" is written in a bold, sans-serif font on the left side of the image. To the right of the word is a circular logo with the word "Bayer" written vertically and horizontally inside the circle, forming a cross shape.

{7}------------------------------------------------

্টি ই

REAGENT INFORMATION: A USER'S GUIDE TO REPLACING AND PREPARING TECHNICON IMMUNO 1 SYSTEM REAGENTS, CALIBRATOR, AND CONTROLS

The mIMP™ (monoclonal Immuno Magnetic Particle) Reagents and IMP Buffer that are used on the Technicon Immuno 1 system are listed in Table A.

PRODUCT NAMECOMPONENTPRODUCT NUMBERTOTAL NUMBER OF UNITS*QUANTITYREAGENT PREPARATION REQUIRED
mIMP ReagentsmIMP ReagentT01-3543-0140001Ready-to-use
IMP Buffer20001Ready-to-use
IMP Buffer Reagent*IMP BufferT01-3601-0140002Ready-to-use

Table A: mIMP REAGENT AND IMP BUFFER

  • For mIMP Reagent 1 unit = 5 uL; for IMP Buffer 1 unit = 10 µL.

WARNING! Contains 0.1% sodium azide. See Technical Bulletin TT6-0319-11. Harmful if swallowed. After contact with skin wash immediately with plenty of water.

Instructions to Replace mIMP Reagent / IMP Buffer

  • Remove IMP compartment cover. 1.
  • Mix new cup by inversion and remove gray stopper. 2.
    1. Replace desired cups. It is important to note the position of the cups.
  • Replace IMP compartment cover (directional arrow must face forward). 4.
  • E Reset number of tests on REAGENT INVENTORY screen.

The Substrate Reagents that are used on the Technicon Immuno 1 system are listed in Table B.

Table B: SUBSTRATE REAGENTS

PRODUCT NAMEPRODUCT NUMBERQUANTITYREAGENT PREPARATION REQUIRED
2 SubstrateReconstitution:

CAUTION! Avoid contact with eyes, skin, or clothing. Wash thoroughly after handling Avoid ingestion.

Instructions to Replace Substrate Reagent

  • Pour entire contents of one (1) Diluent into one (1) Substrate reagent bottle 1.
  • Gently swirl and allow to stand for five (5) minutes. 2.
    1. Gently mix reagent. Do not allow the reagent to foam.
  • Remove gray stopper from the fill port at the top of the cassette tower. 4.
    1. Slowly pour the contents of the Substrate Bottle into the cassette.
  • Reinsert the gray stopper into the fill port. દ.
    1. Remove and discard the gray stopper from the aspiration port.
    1. Open the substrate compartment cover; system must not be processing samples.
    1. Remove empty cassettes and install the full cassette(s). It is important to note the cassette(s).
    1. Up to five (5) substrate can be positioned on the system. For additional information,
  • refer to page INTRO-3 in the section titled "Automatic Substrate Cassette Switching."
  • 1 * Reset the number of tests on the REAGENT INVENTORY screen.

{8}------------------------------------------------

The Wash Solutions that are used on the Technicon Immuno 1 system are listed in Table C.

PRODUCTNAMEPRODUCTNUMBERTOTALVOLUMEQUANTITYREAGENTPREPARATIONREQUIRED
Cuvette Wash 1T01-3060-524 Liters2Ready-to-use
Cuvette Wash 2T01-3570-524 Liters2Ready-to-use
IMP WashT01-2879-524 Liters2Ready-to-use

Table C: WASH SOLUTIONS

CAUTION! Avoid contact with eyes, skin, or clothing Wash thoroughly after handling. Avoid ingestion

Instructions to Replace Wash Solutions

  • Open center door of analytical console. 1.
  • Replace empty wash solutions. 2.
  • install new wash bottle onto shelf. 3.
  • Using a single edge razor blade, cut off the end of the straw allowing 0.25 inch to protrude from cap. 4.
    1. Using a sharp point object, puncture the vent hole.
    1. Attach appropriate colored tubing onto the end of straw.
    1. Close access door.
  • Reset the number of tests for each wash solution replenished on REAGENT INVENTORY screen. 8.
  • Perform FDT to prime lines. Refer to the Technicon Immuno 1 Maintenance / Troubleshooting Manual, ு. UNIT 3, for instructions.

The Reagent and Sample Probe Fluids that are used on the Technicon Immuno 1 system are listed in Table D.

Table D: REAGENT AND SAMPLE PROBE FLUIDS

PRODUCTNAMEPRODUCTNUMBERTOTALVOLUMEQUANTITYREAGENTPREPARATIONREQUIRED
Sample Probe FluidT01-3061-5630.0 mL / Bottle6Ready-to-use
Reagent Probe FluidT01-3062-56100.0 mL / Bottle6Ready-to-use

CAUTION! Avoid contact with eyes, skin, or clothing. Wash thoroughly after handling. Avoid ingestion.

Instructions to Replace Reagent and Sample Probe Fluids

  • Replace as required. 1.
  • Prime probes and lines by initializing the maintenance menu. Refer to the Technicon Immuno 1 Maintenance 2. / Troubleshooting Manual, UNIT 3, for instructions.

The Sample Diluent-B reagent that is used with the Technicon Immuno 1 system is listed in Table E.

PRODUCTNAMEPRODUCTNUMBERTOTALVOLUMEQUANTITYREAGENTPREPARATIONREQUIRED
Sample Diluent - BT03-3574-0150.0 mL / BottleLyophilized Bottle: 6 x 50 mLDiluent Bottle: 6 x 50 mLReconstitution;Lyophilized Material

Table E: SAMPLE DILUENT - B

CAUTION! Avoid contact with eyes, skin, or clothing. Wash thoroughly after handling. Avoid ingestion.

{9}------------------------------------------------

States

.ructions to Reconstitute Sample Diluent - B

  • Remove vial closure. 1.
  • Transfer entire contents of one (1) Reconstitution Diluent into one (1) Sample Diluent B bottle. 2.
  • Swirl gently and let stand for fifteen (15) minutes. 3.
    1. Mix by inversion at least five (5) times.
  • Refrigerate any unused material. Prior to reuse, mix contents thoroughly. 5.

Reagent Preparation Instructions

CAUTION! Avoid contact with eyes, skin, or clothing. Wash thoroughly after handling. Avoid ingestion.

Ready-to-use Liquid Reagent Cassettes

  • Remove the outer shrink wrap layer. 1.
    1. Swing evaporation covers to the side.
    1. Remove inner shrink wrap.
  • Remove and discard gray shipping stoppers. 4.
  • Position evaporation covers over aspiration ports and press down to "snap" into place. ક.

IMPORTANT Exercise covers gently to ensure proper attachment. Do not force the covers through the full length of travel.

  • Do not tip the cassette. When properly installed on the system, the tower will fill automatically. Store reagent(s) on Technicon Immuno 1 system or in refrigerator. Avoid overexposure to light.

Latex Liquid Reagents

  • Handle per the "Ready-to-use" liquid instructions as described above. 1.
  • Mix prior to use and remix weekly during use. 2.
  • Slowly lay the cassette on its side with the base pointing up and allow all the liquid to drain to the tower. 3.
  • Maintaining the cassette position (on its side), shake back and forth along the long axis about 25 times in 10 seconds. ব,
  • Return the cassette to the upright position. Do not over rotate. When properly installed on the base portion of the ર. tower will fill automatically.
    1. Store reagent(s) on Technicon Immuno 1 system or in refrigerator. Avoid overexposure to light.
  • If stored in refrigerator, mix prior to reuse. 7.

Lyophilized Reagents

  • Using a volumetric pipette, Class A or equivalent, pipette the specified volume of diluent into the bottle of lyophilized material. 1.
    1. Allow to stand for fifteen (15) minutes, then gently mix by inversion.
    1. Remove the gray stopper at the top of the cassette tower.
  • Transfer all of the reagent into the cassette. 4.
    1. Replace the cassette tower gray stopper.
    1. Prepare and handle the cassette as indicated above in the "Ready-to-use" instructions.

{10}------------------------------------------------

Calibrator and Control Preparation Instructions

For detailed application information refer to the Clinical Method Sheet and / or product numbers and specifications.

WARNING - POTENTIAL BIOHAZARDOUS MATERIAL

Any product prepared from human blood, plasma, or serum should be handled cautiously as a biohazardous material according to good laboratory practices. Each donor unit used in the preparation of these products was tested by an FDA approved method for the presence of the antibody to Human Immunodeficiency Virus (HIV) as well as for hepatitis B surface antigen and found to be negative (was not repeatedly reactive). Because no test method can offer complete assurance that HIV, hepatitis B virus, or other infectious agents are absent, these products should be handled at the Biosafety Level II as recommended for any potentially infectious human blood specimens in Protection of Laboratory Workers From Infectious Disease Transmitted by Blood, Body Fluids, and Tissue - Second Edition; Tentative Guideline (1991), Document M29-T2, promulgated by the National Committee for Clinical Laboratory Standards (NCCLS).

Ready-to-use Calibrator or Control

  • Remove vial closure. 1.
    1. Mix by inversion at least five (5) times.
    1. Refrigerate any unused liquid. Gently mix prior to reuse.
MATER FARACTER FREE SECARES RECARRENAL CONTRACTORIAL CONSULARANT
CONTRACT FOR COLLECT COLLEGO COLLEGO COLLECT COLLEGO COLLECT COLLEGION.
FRITUL BRITAN - FARM FALL - FARM FALL SE - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F - F -.PART PARTY PARTY CONTRACT CONTRACT CONSULTER CONSULTERS AND THE CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CO
EST 1 8 1 8 1 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
A Comments of the control control control concerner comparent of the comments of the comments of the comments of the comments of the comments of the comments of the comments------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Corporation
.

Lyophilized Calibrator or Control

    1. Remove vial closure.
    1. Using a volumetric pipette, Class A or equivalent, pipette specified volume of calibrator reconstitution diluent.
******************************************************************************************************************************************************************************F PATE FEE = Exe = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =NE KALLER PART PART THE LED CONTRACT LE BEAR AND LE LE CARACT LE CARACT POST AND CONSULTION CONTENT LE COLLECT COLLEGION COLLEGO CONTRACT CONTENT CONTENT CONTENT CONTENT CONT.THE FECTED TO FER FLAQUES ANGELEA A A A A A A C COLOCA OF COLOCA COLL COLOCA COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT C.CLASS CLASS CONCENTRAL CALL TOPEL PRESENT PARTIES PARTY CONTRACTER.TATTATTAL.
Comments of the consisted on the consisted on the collection of the comment of the country of the county of the comment of the fight for theCLI. A. Amerikante . LA . Annual . L' REBRILLI . Management . BarasCONSTITUTION CONSULTION COLLEGIAL CONSULTION CONSULTION CONSULTION CONSULTION COLLEGIONAL CONSULTION COLLEGIONAL CONSULTION COLLEGIONAL CONSULTION CONSULTION CONSULTION CONSU.
.AP.L.K. FREE B S.A.A.A.A.A. TRE ES S. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B. B
    1. Swirl gently and let stand for fifteen (15) minutes.
    1. Mix by inversion at least five (5) times.
    1. Refrigerate any unused liquid. Gently mix prior to reuse.

{11}------------------------------------------------

INTRODUCTION

・アッ ; ;

NCCLS Approved Guideline GP2-ABayer Method Sheet Headings
5.1.2Title
5.1.2.1TitleTitle
5.1.2.2Type of SpecimenIntended Use
5.1.2.3Method or instrumentation in subtitle.Intended Use
5.1.3Principle
5.1.3.1Type of reaction(s) involved.Principles of the Procedure
5.1.3.2Clinical reasons for performing the test.Intended Use
5.1.4Specimen
5.1.4.1State the conditions for patient preparation.Sample Handling / Limitations
5.1.4.2Type of specimenIntended Use / Sample Handling
5.1.4.3Special handling conditionsSample Handling
5.1.5Reagents
5.1.5.1List reagents, supplies, and equipmentReagents
5.1.5.2Directions for preparationReagent Preparation
5.1.5.3Parameters used to determine acceptable reagent performanceStorage and Stability
5.1.5.4Storage requirementsStorage and Stability
5.1.6Calibration
5.1.6.1Standard preparationCalibration / Calibrator Package Insert
5.1.6.2Calibration procedureCalibration
5.1.7Quality Control
5.1.7.1Identify control materials to be usedQuality Control
5.1.7.2Instructions for preparing and handling control materialsControl Package Insert
5.1.7.3Control frequencyQuality Control
5.1.7.4Description of control tolerance limitsControl Package Insert
5.1.7.5Corrective actions to be taken when limits are exceededAccording to Individual Laboratory Practice
Reference or Operator's Manual /
5.1.7.6How QC data are recorded and storedAccording to Individual Laboratory Practice
5.1.7.7If no controls are available, list alternativesNot applicable
5.1.8Procedure - Stepwise
5.1.8.1Detailed stepwise instructionsProcedure
5.1.8.2Centrifugation requirementsSample Handling (Intro-1)
5.1.8.3Specify special glasswareNot applicable
NCCLS Approved Guideline GP2-ABayer Method Sheet Headings
5.1.8.4Specify the following for photometric measurements:Type of instrumentWavelengthCuvette sizeSolution used as a blankLinearityHow the raw data are usedAllowable time intervalsStability of final solutionReference or Operator's Manual / Description and SpecificationsChemistry Program Data SheetChemistry Program Data SheetReagents / ProcedureLinearityReference or Operator's Manual / Theory of OperationReference or Operator's Manual / Theory of OperationNot applicable
5.1.9Calculations
5.1.9.1Give stepwise instructionsResults
5.1.9.2Give the equationReference or Operator's Manual / Theory of Operation
5.1.9.3Give an exampleReference or Operator's Manual / Theory of Operation
5.1.10Reporting results
5.1.10.1State reference rangesExpected or Therapeutic Values
5.1.10.2Identify procedures to be used in reporting resultsto the physicianAccording to Individual Laboratory Practice
5.1.10.3Guidelines on acceptable reporting formatResults / According to Individual Laboratory Practice
5.1.11Procedure Notes
5.1.11.1Explain reasons for special precautionsSample Handling
5.1.11.2List possible sources of errorSample Handling / Limitations of the Procedure / Results
5.1.11.3Include helpful hintsProcedure
5.1.11.4Clinical situations that may influence test validityLimitations of the Procedure
5.1.11.5Acceptable alternative procedures, expecteddifferencesCorrelation
5.1.11.6Enlarge upon clinical applicationsAccording to Individual Laboratory Practice
5.1.12Limitation of Procedure
5.1.12.1State linearity and / or detection limitsAnalytical Range / Minimum Detectable Concentration
5.1.12.2State known interfering substancesInterfering Substances / Cross Reactivity
5.1.13Bibliography
5.1.13.1Literature referencesBibliography
5.1.13.2Manufacturer product literatureEach Method Sheet contains a method number and revisiondate on each page
5.1.13.3Textbooks
5.1.13.4Standard Publications
5.1.13.5Written personal communications
5.1.13.6Research including supporting data for validationof methods

Table 3: NCCLS GUIDELINE (GP2-A) AND BAYER METHOD SHEET HEADINGS CROSS-REFERENCE

{12}------------------------------------------------

Table 3: NCCLS GUIDELINE (GP2-A) AND BAYER METHOD SHEET HEADINGS CROSS-REFERENCE (Cort)

§ 862.2160 Discrete photometric chemistry analyzer for clinical use.

(a)
Identification. A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.