K Number
K961647

Validate with FDA (Live)

Date Cleared
1996-10-08

(161 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DeRoyal Cautery Pencil is indicated for use whenever an electrosurgical procedure requires the removal of tissue or the control of bleeding by either cutting or coagulation.

Device Description

The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil functions in the same manner as predicate devices in that it is intended to be used in electrosurgical procedures to remove tissue and control bleeding. The device can be used for both cutting and coagulating. Device Design/Materials Used/Physical Properties: The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil is made of materials commonly used for their purpose. The concept of use is intended to remove tissue and control bleeding by use of high frequency electrical current.

AI/ML Overview

This document describes a 510(k) summary for the DeRoyal Industries, Inc. Electrosurgical Cautery Pencil, which appears to be a traditional medical device and not an AI/ML-driven device.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, which are typically relevant for AI/ML device evaluations, are not applicable to this submission.

The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (DeBusk Technology Corporation) based on design, materials, intended use, and technological characteristics. The "predicate or legally marketed devices" section and the "technological comparison with predicate or legally marketed device(s)" table confirm this approach.

In summary, as this is a traditional electrosurgical cautery pencil and not an AI/ML device, the specific types of acceptance criteria and performance study details asked for in the prompt (e.g., related to accuracy, sensitivity, specificity, expert consensus, training sets, etc.) are not present in the provided text and are not typically required for this class of device submission.

{0}------------------------------------------------

DEROY AL INDUSTRIES, INC. ELECTROSURGICAL CAUTERY PENCIL

K 961647

510(k) Summary

OCT - 8 1996 SUMMARY OF THE SAFETY AND EFFECTIVENESS INFORMATION UPON WHICH AN EQUIVALENCE DETERMINATION COULD BE BASED

SUBMITTER INFORMATION

ﻨﻬﺎ

NAME:DeRoyal Industries, Inc.TELEPHONE:(423) 938-7828
ADDRESS:200 DeBusk LaneCONTACT:Debra F. Manning
Powell, TN 37849DATE OF PREPARATION:April 22, 1996

DEVICE NAMES DeRoyal Industries, Inc. Electrosurgical Cautery Pencil NAME: COMMON/USUAL NAME: Cautery Pencil CLASSIFICATION NAME (if known): Electrosurgical Cutting & Coagulation Device and Accessories (79 GEI)

PREDICATE OR LEGALLY MARKETED DEVICES

DeBusk Technology Corporation

DEVICE DESCRIPTION

The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil functions in the same manner as predicate devices in that it is intended to be used in electrosurgical procedures to remove tissue and control bleeding. The device can be used for both cutting and coagulating. Device Design/Materials Used/Physical Properties: The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil is made of materials commonly used for their purpose. The concept of use is intended to remove tissue and control bleeding by use of high frequency electrical current.

DEVICE INTENDED USE

The DeRoyal Cautery Pencil is indicated for use whenever an electrosurgical procedure requires the removal of tissue or the control of bleeding by either cutting or coagulation.

TECHNOLOGICAL COMPARISON WITH PREDICATE OR LEGALLY MARKETED DEVICE(S)

CharacteristicDeRoyal DeviceDeBusk Technology DeviceRationale
Length of Pencil Body6.0"6.0"
Length of Cable10'0"10'0"
Material of Outer Cable InsulationPVCPVC
Method of Manufacturing of TipStamped and formed fromsheet of stainless steelStamped and formed from sheetof stainless steel
Tip CoatingTitanium NitrideNoneProvides a non-toxic, anti-stick, lubricated surface.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.