(50 days)
The Marker System is intended for establishing fixed reference points in patients requiring stereotactic surgery. Examples of situations necessitating stereotaxy could include patients in need of biopsy, resection, lesion generation, implantation of radioisotopes, and small field external beam radiotherapy. The Marker System can be used as an accessory with all currently marketed Howmedica Leibinger® stereotactic systems and can also be used with frameless navigation systems from other manufacturers.
The Marker System for Stereotaxic Navigation creates reference points fixed to the patient's skull. The system uses bone screws, adapted from craniofacial fracture fixation, to secure three to five markers into the calvarium. Each assembly consists of a screw, base component, and marker ball or optical marker. Screws are made from commercially pure titanium, have threaded lengths of 6 mm, and come in two overall lengths--9 mm and 18 mm--to accommodate different soft tissue thicknesses. A polysulfone base component screws onto the head of each screw and holds a transparent marker ball of the same material or an aluminum alloy optical marker. The marker ball, containing the medium visible during imaging, is 5 mm in diameter and has an internal cavity 2 mm in diameter. For angiography or CT imaging, these markers contain a gold ball for contrast. Markers intended for MRI and PET use are hollow. allowing injection of appropriate contrast media such as gadolinium, copper sulfate, or "F-labeled fluorodeoxyglucose solutions . This cavitv is accessed by a hole 0.6 mm in diameter descending from the top of the marker. Flat optical markers sit in the base component with the sa.ne profile as the marker balls. Fabricated from an aluminum alloy, they are anodized black and have a machined spot in the center where the true silver color of the alloy is visible; the position of this spot is identical to the center of the imaging marker ball.
Here's an analysis of the acceptance criteria and the studies performed for the Howmedica Leibinger, Inc. Marker System for Stereotaxic Navigation, based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device seem to be implicitly tied to the performance of predicate devices and general "clinically accepted accuracy levels for stereotaxic procedures." The primary performance measure is "image registration accuracy."
| Acceptance Criteria / Performance Metric | Predicate Device Performance (ZD Stereotactic Frame) | Reported Device Performance (Marker System) |
|---|---|---|
| Stereotaxic Accuracy | 1 mm | 0.6 mm - 2.2 mm (Phantom Study) |
| 0.6 mm - 1.5 mm (Clinical Study) | ||
| Image Clarity Artifacts | Not explicitly stated, implied minimal | Not impaired |
| Bone Screw Placement / Tissue Condition | Not explicitly stated, implied problem-free | No problems encountered |
| Infection | Not explicitly stated, implied low | No infection |
2. Sample Size Used for the Test Set and Data Provenance
-
Non-Clinical Study (Phantom Study):
- Sample Size: 35 markers placed on three platforms of a phantom assembly. 16 combinations of 4 markers and 16 combinations of 3 markers were tested for accuracy.
- Data Provenance: Retrospective (implicitly, as it's a controlled phantom study). The study was conducted at the University of Innsbruck, Innsbruck, Austria.
-
Clinical Study:
- Sample Size: 15 patients.
- Data Provenance: Prospective (implicitly, as it describes patients undergoing surgery). The study was conducted at the Neurosurgery Department, Hopitaux des Armees, Paris, France.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Non-Clinical Study: Not applicable. The ground truth for the phantom study was the known, precise positions of the holes on the phantom assembly.
- Clinical Study: Not explicitly stated how many neurosurgeons were involved in assessing the screw placement and clinical outcomes. However, the study was conducted by the "Neurosurgery Department, Hopitaux des Armees," implying involvement of qualified neurosurgical staff. The placement sites were "determined by the neurosurgeon."
4. Adjudication Method for the Test Set
- Non-Clinical Study: Not applicable for adjudication in the traditional sense. The accuracy was determined by manually placing a cursor in the center of each visible marker dot on the monitor and comparing it to the known positions. There's no mention of multiple readers or a consensus process.
- Clinical Study: Not explicitly stated. Clinical outcomes (screw placement, tissue condition, infection, image clarity) were assessed, but the method of adjudication (e.g., whether multiple surgeons independently assessed and then converged) is not detailed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The provided text does not describe a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it quantify the effect size of human readers improving with AI assistance. This device is a passive marker system, not an AI-assisted diagnostic tool.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, in a sense, the non-clinical phantom study represents a standalone evaluation of the marker system's inherent accuracy in imaging. The "image registration accuracy" was measured by comparing the detected marker positions to known physical positions. The text notes that this testing "did not involve the use of an automatic detection algorithm for determining image center," meaning the measurement of the marker's center was a manual, human-in-the-loop task during the evaluation, but the device itself (the marker system) was being evaluated for its ability to provide accurate reference points, independently of human interpretation of those points for surgical navigation.
The clinical study also evaluated the device's accuracy in a standalone manner, with optical referencing accuracies being reported (0.6 mm to 1.5 mm).
7. Type of Ground Truth Used
- Non-Clinical Study (Phantom Study): The ground truth was based on the known, precise physical positions of the holes on the phantom assembly. This is a highly controlled, objective form of ground truth.
- Clinical Study: The ground truth for evaluating clinical performance aspects like screw placement, tissue condition, infection, and image clarity appears to be clinical observation and assessment by neurosurgeons. For optical referencing accuracy, the ground truth would be the actual physical target position during navigation.
8. Sample Size for the Training Set
No training set is mentioned or applicable for this device. This is a physical marker system for stereotaxic navigation, not a machine learning or AI algorithm that requires a training phase.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
Howmedica Leibinger, Inc. Summary of Safety and Effectiveness 510(k) Premarket Notification: Marker System for Stereotaxic Navigation
MAY 10 1996
SUMMARY OF SAFETY AND EFFECTIVENESS
1.0 Administrative Information
- 1.1 Submitter
Howmedica Leibinger, Inc. 14540 Beltwood Pkwy. East Dallas, TX 75244
1-800-962-6558 214-392-3636
- 1.2 Contacts
Andrew B. Rogers Director--Product Development, Special Projects
Kristyn R. Waski Product Engineer--Special Project;
- 1.3 Date Prepared
March 18, 1996
Device Identification 2.0
-
2.1 Trade/Proprietary Name
Ost-Reg Marker System for Sterectaxic Navigation -
2.2 Common Name
Marker System for Stereotaxic Navigation -
2.3 Classification
ﺳﮧ
Stereotaxic Instrument
| Classification Number | Class | Regulation Number |
|---|---|---|
| 84HAW | II | 882.4560 |
{1}------------------------------------------------
3.0 Predicate Device Identification
| STP Complete ModuleSet | Wurzburg TitaniumBone Screws | |
|---|---|---|
| 510(k) Number | K892425 | K854886 |
| Classification Name | Stereotaxic Instrument | Bone Screw |
| Classification Number | 84HAW | 87HWC |
| Class | II | II |
| Regulation Number | 882.4560 | 888.3040 |
4.0 Product Description
4.1 Background
Stereotaxy refers to a group of neurosurgical procedures involving the establishment of a three-dimensional coordinate system where any anatomical point can be determined as a set of coordinates (x,y,z). These procedures are used for resections, generating lesions, biopsy, implantation of radioisotopes, and small field external beam radiotherapy. Stereotactic systems all involve the same basic operating principle; by establishing known reference points and maintaining a fixed cosition between these points and the patient's anatomy throughout imaging, surgery, and/or radiotherapy, minimally invasive procedures can be conducted. Reference points are created using fiducial markers visible both during imaging (CT, MRI, angiography, etc.) and surgery or radiotherapy. Stereotaxic localization uses these points to assemble images into three-dimensional space with known coordinates at every position, facilitating surgical or treatment planning.
Fiducial markers have traditionally been incorporated into a stereotactic frame fixed to the patient's head with percutaneous pins before imaging and left in place for the duration of surgery or treatment. Several drawbacks are associated with this method. Patient comfort is compromised by the fixation pirs and the resulting immobilization. The frames are often cumbersome for the physician to work around, limiting visibility and approach angles. Depending upon the material used, interactions between the frame and magnetic fields for imagin: may occur, reducing system accuracy. Finally, access to the caudal twothirds of the brain is limited with some stereotactic frames. As a result, skull base tumors often necessitite highly-invasive surgery and general
{2}------------------------------------------------
Howmedica Leibinger, Inc. Summary of Safety and Effectiveness 510(k) Premarket Notification: Marker System for Stereotaxic Navigation
anesthetic. These factors have spurred interest in establishing stereotactic reference points without a frame.
4.2 Design
Rather than using fiducial markers external to the patient--consequently requiring the patient to be immobilized in a frame -- the Marker System for Stereotaxic Navigation creates reference points fixed to the patient's skull. The system uses bone screws, adapted from craniofacial fracture fixation, to secure three to five markers into the calvarium. Each assembly consists of a screw, base component, and marker ball or optical marker. Screws are made from commercially pure titanjum. have threaded lengths of 6 mm, and come in two overall lengths--9 mm and 18 mm--to accommodate different soft tissue thicknesses. A polysulfone base component screws onto the head of each screw and holds a transparent marker ball of the same material or an aluminum alloy optical marker. The marker ball, containing the medium visible during imaging, is 5 mm in diameter and has an internal cavity 2 mm in diameter. For angiography or CT imaging, these markers contain a gold ball for contrast. Markers intended for MRI and PET use are hollow. allowing injection of appropriate contrast media such as gadolinium, copper sulfate, or "F-labeled fluorodeoxyglucose solutions . This cavitv is accessed by a hole 0.6 mm in diameter descending from the top of the marker. Flat optical markers sit in the base component with the sa.ne profile as the marker balls. Fabricated from an aluminum alloy, they are anodized black and have a machined spot in the center where the true silver color of the alloy is visible; the position of this spot is identical to the center of the imaging marker ball. See Exhibit F for engineering drawings.
4.3 Biocompatibility
Marker System bone screws are made from commercially pure titanium, noted for its high biocompatibility, corrosion resistance, lack of toxicity. and minimal artifact in CT and MKI scans. Several of thes features are due to the oxide layer spontaneously formed on the titanium surface from exposure to oxygen and strongly bonded to the underiving metal. Titanium also allows for excellent osseointegration, creating very stable screw fixation due to the lack of connective tirsue between the screw and
1 Mark Levivier, M.D., et. al., "Diagnostic yield of stereotactic brain biopsy guided by positron emission tomography with [18F][luorodeoxyglucose," Journal of Neurosurgery, 82(3): 445-452, March 1995.
{3}------------------------------------------------
Howmedica Leibinger, Inc. Summary of Safety and Effectiveness 510(k) Premarket Notification: Marker System for Stereotaxic Navigation
bone . The base components and imaging markers are made from polysulfone complying with FDA regulation 177.1655. Other components supplied with the system--gold balls within CT/angiography imaging markers and aluminum alloy optical markers--are not implanted or skin-contacting. Additional material information can be found in Exhibit E.
4.4 Application
Each marker is placed by first making a small stab incision and then drilling a pilot hole for the bone screw with a twist drill. Skull thickness must be determined before drilling to prevent the pilot hole from extending into the dura. Proper screw length--9 mm or 18 mm--is selected based upon soft tissue thickness. The screw is advanced with a screwdriver, and the base component is fixed on the screw head. Depending upon imaging modality, the appropriate imaging marker is attached to the base component. In order to eliminate shift in marker position, care should be taken to completely seat the marker on the base component. At least 3 imaging markers are placed and will remain in position during scanning.
Angiography, CT, MRI, or PET image data can be transferred to navigation systems (e.g. Zeiss MKM System, Elektra Wand, Radionics OAS, etc.) for the surgical procedure. As with other neurosurgical procedures, the patient is first immobilized in a head rest such as a Mayfield to maintain constant patient position relative to the operating environment. For visual and want registration systems, the imaging markers are removed and replaced with the optical (black-anodized alloy) markers to provide visual or tactile reference points identical to those provided by the imaging markers during scanning. Additionally, the optical markers have the ontion of being used with a stereotactic frame, requiring use of special adanters also provided with the system. The Marker System is typically im )lanted for 1-2 days, spanning imaging and treatment procedures, but can he left in place for up to 30 days for fractionated radiotherapy Note that fractionated or intermittent procedures require re-registration of the Marker System using the same procedures and products previously indicated. The base components, marker balls, and optical markers are removable for fractionated radiotherapy, reducing the projection of the implant when the marker is not in use.
2 Elof Eriksson, M.D., and Per-Ingvar Branematk, M.D., "Osseointegration from the Perspective of the Plastic Surgeon," Plastic and Reconstructive Surgery, March 1994, 93(3): 628-629, March 1994.
{4}------------------------------------------------
The Marker System is compatible with the following stereotactic products:
Howmedica Leibinger® ZD Stereotactic Frame Howmedica Leibinger® STP Stereotactic Planning Software
5.0 Intended Use
Draft labeling can be found in Exhibit A.
- 5.1 Indications
- . The Marker System is intended for establishing fixed reference points in patients requiring stereotactic surgery. Examples of situations necessitating stereotaxy could include patients in need of biopsy, resection, lesion generation, implantation of radioisotopes, and small field external beam radiotherapy.
- . The Marker System can be used as an accessory with all currently marketed Howmedica Leibinger® stereotactic systems and can also be used with frameless navigation systems from other manufacturers.
5.2 Contraindications
Cranial markers should not be used in the following situations:
- . Bone infection
- . Skin infection
5.3 Precautions
- . Imaging and optical markers must be completely seated into the base components or frame adapters in order to reproduce exact positioning after removal and replacement.
- When filling hollow imaging markers with contrast medium, air bubbles must be avoided to optimize imaging accuracy.
- . Markers should be placed in calvarial bone of thickness greater than the threaded length of the selected implant to prevent risk of implant protrusion into the dura.
{5}------------------------------------------------
- . Markers placed for MRI imaging should be located near the target site to facilitate re-registration.
- . Hollow imaging markers containing contrast medium should be removed from the base components after scanning to reduce the potential for contrast medium to be spilled.
- . Contact between the imaging marker balls and surgical cloth during scanning should be prevented to reduce the risk of fluid absorption.
Non-Clinical Study 6.0
Departments of Neurosurgery, Computerized Tomography, and Neurology University of Innsbruck, Innsbruck, Austria
A study was conducted at the University of Innsbruck (Innsbruck, Austria) to evaluate the image registration accuracy of the Marker System in CT and MRI use. 35 markers were placed into holes of known position on three platforms of a phantom assembly. The central cylinder of the phantom was filled with diluted contrast agent, and the test assembly was placed into a standard CT head coil along with a control assembly. Gold-filled imaging markers were used for CT scans, while the contrast agent used for MRI imaging was not reported. The marker images present on the monitor after scanning were registered by manually placing a cursor in the center of each visible marker dot. Note that this testing did not involve the use of an automatic detection algorithm for determining image center; this tool, commonly incorporated into imaging software, improves registration accuracy 16 combinations of 4 markers and 16 combinations of 3 markers were tested for accuracy. The following results were obtained:
| Mean Correlation Error | ||
|---|---|---|
| CT | MRI | |
| 3 markers | 1.6 mm | 2.2 mm |
| 4 markers | 1.2 mm | 1.1 mm |
Considerably higher accuracies-0.6 mm : 0 0.8 mm--were obtained with only three markers when these markers were all on the same platform of the phantom. This indicates that magnetic field distortion--independent of the marker system used-reduces accuracy as distance increases between the plane defined by the markers and the feature to be registered. The pooled mean, combining 3-marker and 4-marker results from both imaging methods. was
{6}------------------------------------------------
under 2 mm. This agrees with clinically accepted accuracy levels for stereotaxic procedures. See Exhibit C for additional information.
7.0 Clinical Study
Neurosurgery Department, Hopitaux des Armees, Paris, France
The Marker System was assessed in 15 patients undergoing surgery with the Zeiss MKM navigation system after CT or MRI imaging. Number of markers placed and imaging modality is shown in the following table:
| 3.8.8.8.8.6.6.6.6.6.6.6.4.6.8.4.8.1.8.1.6.1.6.6.8.4.1.1.1.5.6.4.1.1000 . Cor 100 and 100000 . Concession Comments of Concessional Comparis Comments of100 / Auto BAntiled development a li . IC. L. L. L. L. M. M. M. M. M. M. M. M. M. M. M. M. M. M. M. M. M. B. B..A LAT.The Color Color College of the Section and Comments of Concession of ConcessionSURALLABORARIA And Antiques and CARA A walla A walla di also a managin'in tana a mini al program and any and any and any and on the mail of the manage.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ||
|---|---|---|
| A MARANAA APark a com a comments of consistence of the comments of the comments of1888 887 18 100 31 6 1 6 3 1 6 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1A Comments of Children Comments of Children Comments of ChildrenCAPAGE FOR DOCT A CARRET CONSULTION CONSULTION COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECTсобстав сертари и как така в въ как ч ч в свежно в числе видео. | ||
| ENTRE TEARLY AND START PRODUCT CARRECT CONSTITUTION ATT THEThe Court of Children Comments of Children Comments of Children.TE LE TEA Shows 1 . F . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8 . 8A Same Comments of ChildrenAN CHEAD PAR COLLECT STATE CT STORES CLASS COLLEGENo. of Cattle.The College of Children of the Callery of Children of Children College of | ||
| Same / A R . L . L . Land a the organization and other the country accounted1998 - 1998 - 1998 - 1998 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -Contract of the contribution and any and any and the comments of the comments of the |
All screws were placed one day pre-operatively under local anesthesia. Placement sites were determined by the neurosurgeon based upon the preplanned surgical access route. In each case, all screws were placed close to the access site with the exception of a control placed at a distance to improve referencing precision.
No problems were encountered with screw placement, tissue condition, or infection. CT or MRI image clarity was not impaired by artifacts. Optical referencing accuracies ranged from 0.6 mm to 1.5 mm--this compares very favorably with accuracies attainable with stereotactic frames.
{7}------------------------------------------------
8.0 Substantial Equivalence
Components of the Marker System are substantially equivalent to the ZD stereotactic frame incorporated in the STP Complete Module Set for stereotaxy [510(k) K892425] and to Wurzburg titan:um bone screws [510(k) K854886]. Both devices are listed as predicates because the Marker System merges the long history of bone screw use in the craniofacial area with stereotaxy to provide a frameless marker system with accuracy equivalent to stereotactic frames. Specific comparisons are made in the following table:
| Feature | Cranial Marker | ZD Stereotactic Frame(STP Complete ModuleSet) | Wurzburg TitaniumBone Screws |
|---|---|---|---|
| 510(k) Number | n/a | K892425 | K854886 |
| Indication | establish a three-dimensional coordinatesystem for stereotacticsurgery | establish a three-dimensional coordinatesystem for stereotacticsurgery | bone fixation incraniofacial area |
| StereotaxicAccuracy | 0.6-2.2 mm(see sections 3.0 and4.0) | 1 mm | n/a |
| Marker Diameter | 2.0 mm | 1.6 mm for MRI0.2 mm for CT | n/a |
| Material | titanium screwsUdel marker balls | pins: titanium point,carbon fiber and plastichousing | titanium |
| contrast agent dependentupon imaging modality--CT, angiography:gold ball--MRI, PET: atdiscretion of physician | contrast agent dependentupon imaging modality--CT, angiography:stainless steel wire--MRI, PET: atdiscretion of physician | ||
| Screw Diameter | 2.0 mm | n/a | 2.0 mm |
| Threaded Length | 4 or 5 mm | n/a | 3.5-13.5 mm |
| Thread Pitch | 1 mm | n/a | 1 mm |
{8}------------------------------------------------
9.0 Bibliography
- Eriksson, Elof, M.D., and Branemark, Per-Ingvar, M.D. "Osseointegration for the perspective of the plastic surgeon." Plastic and Reconstructive Surgery, 93(3): 626-637, March 1994.
Levivier, Marc, M.D., et. al. "Diagnostic yield of stereotactic brain biopsy guided by positron emission tomography with [18F]fluorodeoxyglucose." Journal of Neurosurgery, 82(3): 445-452, March 1995.
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).