(80 days)
The SYNCHRON Systems Methaqualone Reagent, in conjunction with the SYNCHRON System DAT Urine Calibrators, is intended for use in the qualitative determination of Methaqualone in human urine samples. This assay is designed for use with the family of SYNCHRON Systems, which include analyzers such as the SYNCHRON CX®4, CX6 DELTA, CX5, CX5CE, CX5 DELTA, CX7, and CX7 DELTA Systems.
The SYNCHRON Methaqualone Reagent is an enzyme immunoassay reagent test kit for the SYNCHRON Systems and provides a rapid screening procedure for determining the presence of Methaqualone and it's metabolites in urine. It is designed to complement the current panel of CX Drugs of Abuse Assays, previously cleared by FDA. The kit consists of reagent and calibrators.
This document describes the performance of the SYNCHRON Systems Methaqualone Reagent, an enzyme immunoassay for the qualitative determination of Methaqualone in human urine.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as distinct targets in the provided text. However, the study aims to demonstrate substantial equivalence to the predicate device (Emit II Methaqualone Assay) and highlights specific performance metrics. Below is a table summarizing the reported device performance, which implicitly serves as the 'met' acceptance criteria given the submission of the summary of safety and effectiveness.
| Performance Metric | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance |
|---|---|---|
| Overall Agreement | High agreement with predicate device (Emit II) | 94% |
| Relative Sensitivity | High sensitivity compared to predicate device (Emit II) | 88% |
| Relative Specificity | High specificity compared to predicate device (Emit II) | 100% |
| Shelf-life (unopened) | Comparable to predicate or established standard | 12 months |
| On-Instrument Stability | Comparable to predicate or established standard | 60 days |
| Calibration Frequency | Comparable to predicate or established standard | 7 days |
| Within-Run Imprecision (Negative Calibrator) | Low %CV | 0.6% |
| Within-Run Imprecision (Control 1: 200 ng/mL) | Low %CV | 0.5% |
| Within-Run Imprecision (Low Calibrator: 300 ng/mL) | Low %CV | 0.4% |
| Within-Run Imprecision (Control 2: 375 ng/mL) | Low %CV | 0.4% |
| Within-Run Imprecision (High Calibrator: 1000 ng/mL) | Low %CV | 0.3% |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: 103 samples.
- Data Provenance: Not explicitly stated, but the study compares the SYNCHRON Methaqualone Reagent to the Emit II Reagent and also mentions a comparison with GCMS. The nature of the samples (e.g., patient samples, spiked samples) and their geographical origin (e.g., country) are not specified. It is likely retrospective as it involves comparison to existing methods.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- The document does not mention the use of human experts or readers to establish the ground truth for the test set.
- The ground truth reference methods mentioned are the Emit II Reagent (a predicate device) and GCMS (Gas Chromatography-Mass Spectrometry).
4. Adjudication Method for the Test Set
- No human adjudication method is described, as the ground truth was established by laboratory methods (Emit II and GCMS).
- The study implicitly uses the Emit II Reagent as a primary comparitor, and GCMS as a confirmatory method for discordant results (as indicated by "Number of Disconcordant w/ GCMS: 0").
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
- This device is an automated laboratory assay (enzyme immunoassay) for drug detection, not an AI-assisted diagnostic imaging or classification tool that would involve human readers. Therefore, the concept of human reader improvement with/without AI assistance does not apply.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, the performance metrics reported (Overall Agreement, Relative Sensitivity, Relative Specificity, Imprecision, Stability) are indicative of standalone performance of the SYNCHRON Systems Methaqualone Reagent as an algorithm/assay.
- The device is an automated chemical analyzer component, and its reported performance reflects its intrinsic capability without direct human intervention in the result interpretation after the sample is processed by the instrument.
7. The Type of Ground Truth Used
- The primary ground truth for comparison in the "Relative Sensitivity and Specificity Study Results" was the Emit II Methaqualone Reagent (a predicate enzyme immunoassay).
- For discordant results, a more definitive analytical method, GCMS (Gas Chromatography-Mass Spectrometry), was used as a confirmatory ground truth. The fact that there were "0" discordant results with GCMS suggests GCMS was considered the gold standard for verifying the true status of the "disconcordant" samples when compared to Emit II.
8. The Sample Size for the Training Set
- The document does not mention a "training set" or "training data." This is because the device is a chemical reagent and assay system, not a machine learning model that requires a training phase.
- The described studies are performance validation studies for a diagnostic assay.
9. How the Ground Truth for the Training Set was Established
- As there is no mention of a training set, the establishment of ground truth for a training set is not applicable to this device.
{0}------------------------------------------------
BECKMAN
- 4 1996
Appendix C Summary of Safety & Effectiveness SYNCHRON® Systems Methaqualone Reagent
1.0 Submitted By
K 955676
Sheri Hall Manager, Product Submissions Beckman Instruments. Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92621 Telephone: (714) 993-8916 FAX: (714) 961-4457
2.0 Date Submitted
15 December 1995
3.0 Device Name(s)
3.1 Proprietary Names
SYNCHRON Systems Methaqualone (METQ) Reagent SYNCHRON Systems DAT Negative Urine Calibrator SYNCHRON Systems DAT Low Urine Calibrator II SYNCHRON Systems DAT High Urine Calibrator II SYNCHRON Systems DAT Low Urine Control II SYNCHRON Systems DAT High Urine Control II
3.2 Classification Names
Methaqualone test system (21 CFR 862.3630) Toxicology calibrator (21 CFR 862.3200) Toxicology control (21 CFR 862,3280)
4.0 Predicate Device(s)
Emit II Methaqualone Assay, Syva Company, K921013 Methaqualone Reagent, Diagnostic Reagents, Inc. K940123 Urine Calibrator and Controls, Diagnostic Reagents, Inc., K935101
5.0 Description
The SYNCHRON Methaqualone Reagent is an enzyme immunoassay reagent test kit for the SYNCHRON Systems and provides a rapid screening procedure for determining the presence of Methaqualone and it's metabolites in urine. It is designed to complement the current panel of CX Drugs of Abuse Assays, previously cleared by FDA. The kit consists of reagent and calibrators.
6.0 Intended Use
The SYNCHRON Systems Methaqualone Reagent, in conjunction with the SYNCHRON System DAT Urine Calibrators, is intended for use in the qualitative determination of
Beckman Instruments, Inc.
1wx. 910-592 1260 · The . 06-784 · 3
{1}------------------------------------------------
Methaqualone in human urine samples. This assay is designed for use with the family of SYNCHRON Systems, which include analyzers such as the SYNCHRON CX®4, CX6 DELTA, CX5, CX5CE, CX5 DELTA, CX7, and CX7 DELTA Systems.
7.0 Comparison to Predicate(s)
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| SYNCHRON Methaqualone Reagent | Emit II Methaqualone Reagent |
|---|---|
| Similarities | |
| Homogeneous enzyme immunoassayintended for the qualitative analysis ofMethaqualone in human urine | Same |
| Method based on competition betweendrug in the sample and drug labeled withthe enzyme G6PDH; enzyme activitymeasured as NAD is converted to NADH | Same |
| Reagent designed for chemistry analyzerswhich maintain constant reactiontemperature, use automated pipetting,measure enzyme rates, mix, andaccurately time reactions | Same |
| Reagent provides preliminary analyticaltest for screening purposes | Same |
| Differences | |
| SYNCHRON reagent is liquid stabilizedand requires no preparation | Emit II reagent is lyophilized and requiresreconstitution |
| SYNCHRON reagent is intended forqualitative determinations only | Emit II reagent has a specializedapplication for semi-quantitativedetermination of drug concentration |
| SYNCHRON reagent is stable for 60 daysonce opened, when stored properly | Emit II reagent is stable for 12 weeks onceopened, if handled as directed |
8.0 Summary of Performance Data
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method companson, stability, and imprecision experiments that relate results obtained from the SYNCHRON Systems Methaqualone Reagent to the Emit II Reagent.
| Overall Agreement | 94% |
|---|---|
| Relative Sensitivity | 88% |
| Relative Specificity | 100% |
| Number of Disconcordant w/ Emit II | 6 |
| Number of Disconcordant w/ GCMS | 0 |
| Total Number of Samples | 103 |
Relative Sensitivity and Specificity Study Results CX Methaqualone Reagent vs Emit II
{2}------------------------------------------------
| Product Claim | 510(k) Summary |
|---|---|
| Shelf-life | 12 months unopened |
| On-Instrument Stability | 60 days |
| Calibration Frequency | 7 days |
SYNCHRON Methaqualone Stability Study Results
| Material | Mean(mA/min) | SD(mA/min) | %CV | Number ofResults |
|---|---|---|---|---|
| Negative Calibrator | 117 | 0.8 | 0.6 | 20 |
| Control 1 (200 ng/mL) | 216 | 1.1 | 0.5 | 20 |
| Low Calibrator (300 ng/mL) | 243 | 0.9 | 0.4 | 20 |
| Control 2 (375 ng/mL) | 265 | 1.0 | 0.4 | 20 |
| High Calibrator (1000 ng/mL) | 290 | 0.9 | 0.3 | 20 |
Estimated Within-Run Imprecision SYNCHRON Methaqualone Reagent
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
§ 862.3630 Methaqualone test system.
(a)
Identification. A methaqualone test system is a device intended to measure methaqualone, a hypnotic and sedative drug, in urine. Measurements obtained by this device are used in the diagnosis and treatment of methaqualone use or overdose.(b)
Classification. Class II (special controls). A methaqualone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).