K Number
K955602
Date Cleared
1996-04-19

(133 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This test system is for in vitro diagnostic use for the detection of antibodies to nuclear antigen Scl-70 in human serum.

Device Description

This is an enzyme immunoassay for the detection of antibodies to nuclear antigen Scl-70 in human serum.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Immuno Concepts RELISA® Scl-70 Antibody Test System:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state pre-defined acceptance criteria. Instead, it presents the performance of the new device relative to a predicate device. Therefore, the "acceptance criteria" are implied to be a level of performance that demonstrates substantial equivalence to the predicate. Based on the conclusion, the implied acceptance was met.

MetricAcceptance Criteria (Implied)Reported Device Performance
Relative SensitivityHigh agreement with predicate97.9%
Relative SpecificityHigh agreement with predicate100%
Overall AgreementHigh agreement with predicate99.4%

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The total sample size can be calculated from the
    contingency table: 37 (Pos) + 3 (Bor) + 0 (Neg) + 0 (Pos) + 7 (Bor) + 0
    (Neg) + 0 (Pos) + 1 (Bor) + 129 (Neg) = 177 samples.
  • Data Provenance: Not explicitly stated. There is no information
    regarding the country of origin of the data or whether it was retrospective
    or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable in this context. The ground truth for
    the new device's performance was established using a predicate device
    (Immuno Concepts RELISA® Screening Assay, K935129) as the reference method,
    not human experts.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The comparison was directly against
    a reference method (predicate device), not against multiple human
    interpretations requiring adjudication. The text mentions an assumption: "If
    we assume that "borderline" results are actually positive," which is a data
    interpretation choice, not an adjudication method by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No, an MRMC study was not performed. This study compares
    a new diagnostic assay to a predicate diagnostic assay, not the performance
    of human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Yes, this is essentially a standalone performance
    study for the diagnostic device. The RELISA® Scl-70 Antibody Test System
    is a laboratory assay, and its performance (sensitivity, specificity) is
    measured directly against a reference method (the predicate device) without
    human intervention in the final result interpretation for the purpose of
    this comparison.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for this study was established
    using a predicate device (Immuno Concepts RELISA® Screening Assay,
    K935129). This is a form of "reference standard" or "comparator method"
    ground truth.

8. The Sample Size for the Training Set

  • Training Set Sample Size: The document does not mention a training
    set or how the device was developed/trained. This is a traditional immunoassay
    kit, not an AI/ML algorithm that typically requires a training set. The data
    provided is solely for the validation/comparison of the new device against its
    predicate.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no mention of
    a training set for this device.

{0}------------------------------------------------

K955602

Image /page/0/Picture/1 description: The image shows the logo for Immuno Concepts. The logo consists of two horizontal lines on the left, followed by the letters "IC" with a stylized design, and the words "immuno concepts" in a smaller font size. The logo is black and white.

APR 1 9 1996

Date Prepared: December 4, 1995 Contact Person: Eric S. Hoy, Ph.D. Name of Device:

  • · Trade Name RELISA® Scl-70 Antibody Test System
  • Common Name Sc1-70 Antibody Test System
  • Classification Name Extractable Antinuclear Antibody (21 CFR 866.5100)

Legally marketed device with which this device has been shown to be equivalent: RELISA® ENA Antibody Screening Tests System, K935129

Description:

This is an enzyme immunoassay for the detection of antibodies to nuclear antigen Scl-70 in human serum.

Intended Use:

This test system is for in vitro diagnostic use for the detection of antibodies to nuclear antigen Scl-70 in human serum.

Summary of Technological Characteristics Compared to the Predicate Device:

This device is identical to the predicate device with the following exceptions :

a) The predicate device has six different autoantigens coated on individual microwells; the present device has only Scl-70 autoantigen coated on the microwells.

b) The predicate device includes a procedure control well on each strip of microwells, the present device includes a calibrator serum in the kit.

Description of Laboratory Data That Indicate Substantial Equivalence:

For direct determination of relative sensitivity and specificty, we used the Immuno Concepts RELISA® Screening Assay (K935129) as a reference method. The data obtained in this comparison are shown in the following Table.

{1}------------------------------------------------

K 955402

PositiveImmuno Concepts RELISA® Screening AssayBorderlineNegative
Immuno ConceptsRELISA® Scl-70Positive3730
Borderline070
Negative01129

Table 1. Detection of antibodies to the Scl-70 autoantigen.

If we assume that "borderline" results are actually positive, these data yield the following statistics: relative sensitivity, 97.9%; relative specificity, 100%; and overall agreement, 99.4%

In accordance with 21 CFR 807.92(b)(3), we conclude from these data that the present device is substantially equivalent to the predicate device.

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).