K Number
K955380
Device Name
IBC QUICK CELL COMPONENT (1/2, 3/8, 1/4)
Date Cleared
1996-05-14

(172 days)

Product Code
Regulation Number
870.4330
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The IBC Quick Cell component was developed for use with the CDI Model 400 Blood Gas Monitoring System manufactured by 3M. The purpose of this report is to demonstrate the equivalence of the IBC Quick Cell component, as a substitute for the CDI Kwik Cell component.
Device Description
The IBC Quick Cell component was developed for use with the CDI Model 400 Blood Gas Monitoring System manufactured by 3M. The final geometry of the IBC Quick Cell component is identical to the final geometry of the 3M Kwik Cell component, and both are fabricated from the same plastic materials. Performance of the IBC Quick Cell component within the CDI Model 400 Blood Gas Monitoring System is identical to the CDI Kwik Cell component. The materials were evaluated for toxicity and sterilization compatability requirements as well as for function. The CDI Model 400 Blood Gas Monitoring System er ploys three photochemical sensors to measure pO3, pCO2 and pH. Additionally, there is a thermo-electronic sensor for the direct measurement of temperature. The Electronics also contain calculation programs which use the measured parameters to determine O2 Saturation (Venous side only) and Base Excess/[HCO3] (Arterial side only). To complete the necessary calculations for these approximations, the Hemoglobin content is also required. This value is internally set at a Hematocrit of 25%. This value is corrected by the user during the initial and any subsequent on line recalibrations. The functional properties of the Quick Cells are determined by the final assembly geometry and the materials employed in construction, especially the membrane material. The final geometry of the IBC and CDI components are identical. Using chemical analysis, electron microscopy and information in the public domain, the membrane was sourced from the same supplier used by 3M.
More Information

Not Found

Not Found

No
The document describes a component (Quick Cell) for a blood gas monitoring system. The system uses photochemical and thermo-electronic sensors and calculation programs based on measured parameters. There is no mention of AI or ML in the device description, intended use, or performance studies. The calculations are described as standard programs using measured values.

No
The device is a component of a blood gas monitoring system, used for measurement and monitoring, not for treating a condition or disease.

Yes

The device, specifically the CDI Model 400 Blood Gas Monitoring System (which the IBC Quick Cell component is designed for), measures various physiological parameters like pO2, pCO2, pH, and calculates O2 Saturation and Base Excess/[HCO3]. These measurements are used to provide information about a patient's blood gas status, which is a key aspect of diagnosis.

No

The device description explicitly states it is a "component" and describes its physical geometry, materials, and fabrication process, indicating it is a hardware part of a larger system.

Based on the provided text, the IBC Quick Cell component is not an IVD (In Vitro Diagnostic) device itself.

Here's why:

  • Intended Use: The intended use is as a component for the CDI Model 400 Blood Gas Monitoring System. It's a substitute for another component (the CDI Kwik Cell).
  • Device Description: It describes the physical characteristics and materials of the component, and how it functions within the larger blood gas monitoring system.
  • Function: The component's function is to facilitate the measurement of blood gas parameters (pO2, pCO2, pH) by the sensors within the CDI Model 400 system. It's not performing the diagnostic test itself.
  • Comparison: The performance studies compare the IBC Quick Cell component to the CDI Kwik Cell component within the context of the CDI Model 400 system. The focus is on the component's performance as part of the larger system, not as a standalone diagnostic test.

The CDI Model 400 Blood Gas Monitoring System, which uses this component, is likely an IVD device because it measures parameters in blood to provide diagnostic information. However, the IBC Quick Cell component itself is a part of that system, not the diagnostic device.

N/A

Intended Use / Indications for Use

The IBC Quick Cell component was developed for use with the CDI Model 400 Blood Gas Monitoring System manufactured by 3M.

Product codes (comma separated list FDA assigned to the subject device)

74DRY

Device Description

The IBC Quick Cell component was developed for use with the CDI Model 400 Blood Gas Monitoring System manufactured by 3M. The final geometry of the IBC Quick Cell component is identical to the final geometry of the 3M Kwik Cell component, and both are fabricated from the same plastic materials. Performance of the IBC Quick Cell component within the CDI Model 400 Blood Gas Monitoring System is identical to the CDI Kwik Cell component. The materials were evaluated for toxicity and sterilization compatability requirements as well as for function.

The CDI Model 400 Blood Gas Monitoring System er ploys three photochemical sensors to measure pO3, pCO2 and pH. Additionally, there is a thermo-electronic sensor for the direct measurement of temperature. The Electronics also contain calculation programs which use the measured parameters to determine O2 Saturation (Venous side only) and Base Excess/[HCO3] (Arterial side only). To complete the necessary calculations for these approximations, the Hemoglobin content is also required. This value is internally set at a Hematocrit of 25%. This value is corrected by the user during the initial and any subsequent on line recalibrations. The purpose of this report is to demonstrate the equivalence of the IBC Quick Cell component, as a substitute for the CDI Kwik Cell component. The assumptions in the calculated values and any error which may be present as a result of the electronics, transducer or sensors is not to be evaluated. Consequently, only the measured parameters will be used in the functional analysis. The calculated values were recorded for general interest only

The functional properties of the Quick Cells are determined by the final assembly geometry and the materials employed in construction, especially the membrane material. The final geometry of the IBC and CDI components are identical. Using chemical analysis, electron microscopy and information in the public domain, the membrane was sourced from the same supplier used by 3M.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Cardiovascular surgical team

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A. Assembly of Test Samples.
1 . Twenty each frame /retainer assemblies were made using Standard Operating Procedures.
2. These assemblies were assembled to the 3/" flow through body .
3. The membrane guard was snapped into place and the assembly was leak tested.
4. Samples were subjected to the IBC standard sterilization cycle and placed in quarantine for 14 days.
5. Control samples of CDI manufactured products were purchased from commercial sources.

B. Test Equipment.
1 . Instrumentation Laboratories Model 1420 Blood Gas Analyzer
2. Instrumentation Laboratories Model 482 Co-Oximeter.
3. CDI Model 400 Blood Gas Monitor System (Monitor display, light head transducer, disposable photo-chemical sensors, and sensor calibrator).
4. IBC Model 1200 VenOsat Monitor.
Gambro Cardiovascular Heart-Lung Console Roller 5. Pump.
6. Cincinnati Sub Zero Model 400 Heater/Cooler.
7. Cobe CML Oxygenators.
8. Heart Lung Tubing and connectors.

C. Test Circuit and Set up.
1 Assemble the circuit as shown in Test Assembly 1. schematic drawing.
Using the 1/4" spike, prime the integral hard shell 2. reservoir on the Cobe CML with Human Blood which has been typed and cross matched and preserved with Acid Citrate Dextrose A U.S.P .. Fill to a level which is approximately 1" above the Heat Exchanger (Approximately four units).
Using the Roller Pump Head, prime the oxygenator 3. circuit with a clamp at point B. After the oxygenator is fully primed, move the clamp to Point A and continue to prime the oxygenator bypass line. Recirculate at 2 liters per minute.
4. Calibrate the CDI Kwik Cell Sensors per the operating instructions.
5. Draw a test sample for the Blood Gas Analyzer and C0-Oximeter.
6. Using the oxygenator and Gas Blender adjust the blood to physiological conditions. Using 5% Sodium Bicarbonate in Normal Saline, adjust the pF to physiological levels.

D. Test Procedure.

  1. Stop Pump Head to discontinue the recirculation.
  2. Remove the membrane guard from the two CDI Kwik Cells and snap the calibrated sensors in place and attach the light head transducers.
  3. Resume the operation of the Pump Head, setting a flow rate of 3.5 Liters per Minute.
  4. Observe the readings in the CDI Monitor Display and the IBC VenOsat Monitor Display. If the readings are suitable and stable, Take a sample from the sample port and run it through the IL Blood Gas Analyzer and the IL Co-Oximeter. Simultaneously document the readings from both on-line monitors and press the recalibration button on the CDI Model 400 Monitor.
  5. Recalibrate the monitor using the values obtained from the IL Co-Oximeter and Blood Gas Analyzer.
  6. Move the clamp from Point A to Point B and, using the Gas Blender, alter the blood gases. Move the clamp from Point B back to Point A and wart for the on line readings to stabilize(3 minutes).
  7. After the three minute period, draw another blood sample for the IL analyzers and document the readings on the CDI monitor. Repeat this procedure (steps 6. and 7.) until ten different readings are obtained.
  8. Stop the Pump Head .
  9. Calibrate two fresh Kwik Cell Sensors and snap them into the IBC flow through cells
  10. Resume operation of the Roller Pump at a flow rate of 3.5 Liters per Minute.
  11. Repeat steps 4., 5., 6. and 7. and then stop the pump and change the circuit using fresh disposables and blood for the next test samples.
  12. Repeat the test using IBC manufactured components first and CDI manufactured components second.
  13. Follow this alternating procedure for ten runs (until five pairs of cells have been tested from both manufacturers).

E. Data Collection and Treatment.

  1. Record each CDI reading along with the appropriate IL readings.
  2. Calculate the mean variance and mean per cent variance for each test sample and tabulate at the bottom of the log sheets.
  3. Tabulate the mean error and % error from each of the ten runs on the Gas Stat Test Data Analysis Log.
  4. Calculate the average error for all 100 data points for all measured and calculated parameters. Record these calculations on the Analysis log
  5. Construct a graph for comparison of all 100 pairs of readings for the measured parameters, pO2, pCO2 and pH, obtained from testing components of both manufacturers.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The IBC Quick Cell Component was tested in comparison with the CDI Kwik Cell Component within the CDI Model 400 Blood Gas Monitoring System. The two products were identical relative to performance and function.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

SAMPLEpO2pCO2pH[HCO3]BESAT
IBC #1 ERROR500.0111.7
IBC #1 % ERR7%0%0%2%-189%3%
IBC #2 ERROR-8-10.030.31.0
IBC #2 % ERR-6%-3%0%1%-42%-2%
IBC #3 ERROR2-10.000-0.7
IBC #3 % ERR1%-2%0%2%13%2%
IBC #4 ERROR11-0.020-0.2
IBC #4 % ERR0%2%0%-2%7%-3%
IBC #5 ERROR-100.00-10.2
IBC #5 % ERR-1%-1%0%-3%-5%-2%
AVERAGE IBC ERROR000000
IBC AVG % ERROR0%0%0%0%-22%0%
CDI #1 ERROR520.0011.53
CDI #1 % ERR8%4%0%3%-170%4%
CDI #2 ERROR2-1.75-0.01-10.0-3
CDI #2 % ERR1%-4%0%-3%0%-3%
CDI #3 ERROR32-0.021-0.6-4
CDI #3 % ERR3%4%0%3%11%-5%
CDI #4 ERROR-61-0.010.10.40
CDI #4 % ERR-5%3%0%0%-14%0%
CDI #5 ERROR-700.0011.21-6
CDI #5 % ERR-6%1%0%6%-35%-6%
AVERAGE CDI ERROR00000-1
CDI AVG % ERROR0%1%0%1%-21%-1%

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Not Found

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.4330 Cardiopulmonary bypass on-line blood gas monitor.

(a)
Identification. A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.(b)
Classification. Class II (performance standards).

0

1 1996

Image /page/0/Picture/1 description: The image shows a handwritten number, "K955380". The number is written in black ink on a white background. The numbers are written in a bold, slightly irregular style. The number appears to be a serial number or some other type of identification number.

Image /page/0/Picture/2 description: The image shows a logo with the letters "IBC" in a bold, stylized font. Each letter is overlaid with a partial globe design, suggesting a global or international theme. The letters and globe outlines are in black, contrasting with a white background, which gives the logo a strong visual presence.

International Biophysics Corporation 4020 So. Industrial Drive, Suite 160 Austin, Texas 78744 (512) 326-3244, (512) 326-3299 fax Attn: H. David Shockley, jr., President November 20, 1995

Trade Name: IBC Quick Cell Component (Catalog Numbers 4000, 4010 & 4020) Common Name: Blood Gas Flow Through Connectors Classification Name: Sensor, Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass Product Code: (74DRY) C.F.R. Section: 870.4410 Equivalent Device: CDI Kwik Cell Component (Catalog Numbers 6720, 6730 & 6740)

Introduction

510 (K) SUMMARY

The IBC Quick Cell component was developed for use with the CDI Model 400 Blood Gas Monitoring System manufactured by 3M. The final geometry of the IBC Quick Cell component is identical to the final geometry of the 3M Kwik Cell component, and both are fabricated from the same plastic materials. Performance of the IBC Quick Cell component within the CDI Model 400 Blood Gas Monitoring System is identical to the CDI Kwik Cell component. The materials were evaluated for toxicity and sterilization compatability requirements as well as for function.

The CDI Model 400 Blood Gas Monitoring System er ploys three photochemical sensors to measure pO3, pCO2 and pH. Additionally, there is a thermo-electronic sensor for the direct measurement of temperature. The Electronics also contain calculation programs which use the measured parameters to determine O2 Saturation (Venous side only) and Base Excess/[HCO3] (Arterial side only). To complete the necessary calculations for these approximations, the Hemoglobin content is also required. This value is internally set at a Hematocrit of 25%. This value is corrected by the user during the initial and any subsequent on line recalibrations. The purpose of this report is to demonstrate the equivalence of the IBC Quick Cell component, as a substitute for the CDI Kwik Cell component. The assumptions in the calculated values and any error which may be present as a result of the electronics, transducer or sensors is not to be evaluated. Consequently, only the measured parameters will be used in the functional analysis. The calculated values were recorded for general interest only

The functional properties of the Quick Cells are determined by the final assembly geometry and the materials employed in construction, especially the membrane material. The final geometry of the IBC and CDI components are identical. Using chemical analysis, electron microscopy and information in the public domain, the membrane was sourced from the same supplier used by 3M.

1

I. Functional Evaluation.

A. Assembly of Test Samples.

l . Twenty each frame /retainer assemblies were made using Standard Operating Procedures.

  1. These assemblies were assembled to the 3/ ," flow through body .

  2. The membrane guard was snapped into place and the assembly was leak tested.

  3. Samples were subjected to the IBC standard sterilization cycle and placed in quarantine for 14 days.

  4. Control samples of CDI manufactured products were purchased from commercial sources.

B. Test Equipment.

l . Instrumentation Laboratories Model 1420 Blood Gas Analyzer

  1. Instrumentation Laboratories Model 482 Co-Oximeter.

  2. CDI Model 400 Blood Gas Monitor System (Monitor display, light head transducer, disposable photo-chemical sensors, and sensor calibrator).

  3. IBC Model 1200 VenOsat Monitor.

Gambro Cardiovascular Heart-Lung Console Roller 5. Pump.

2

  1. Cincinnati Sub Zero Model 400 Heater/Cooler.

  2. Cobe CML Oxygenators.

  3. Heart Lung Tubing and connectors.

C. Test Circuit and Set up.

1

్లో ప్రత్యాల్యం

Assemble the circuit as shown in Test Assembly 1. schematic drawing.

Using the 1/4" spike, prime the integral hard shell 2. reservoir on the Cobe CML with Human Blood which has been typed and cross matched and preserved with Acid Citrate Dextrose A U.S.P .. Fill to a level which is approximately 1" above the Heat Exchanger (Approximately four units).

Using the Roller Pump Head, prime the oxygenator 3. circuit with a clamp at point B. After the oxygenator is fully primed, move the clamp to Point A and continue to prime the oxygenator bypass line. Recirculate at 2 liters per minute.

  1. Calibrate the CDI Kwik Cell Sensors per the operating instructions.

  2. Draw a test sample for the Blood Gas Analyzer and C0-Oximeter.

  3. Using the oxygenator and Gas Blender adjust the blood to physiological conditions. Using 5% Sodium Bicarbonate in Normal Saline, adjust the pF to physiological levels.

D. Test Procedure.

  1. Stop Pump Head to discontinue the recirculation.

3

Image /page/3/Figure/0 description: This image is a diagram of a test assembly with various components labeled. The assembly includes a COBE CML, a Gambro gas blender with inputs for O2, N2, and CO2, and a Gambro pump. The diagram also shows venous and arterial sensors, a Haemosat, and a sampling syringe, with flow indicated by an arrow.

4

  1. Remove the membrane guard from the two CDI Kwik Cells and snap the calibrated sensors in place and attach the light head transducers.

  2. Resume the operation of the Pump Head, setting a flow rate of 3.5 Liters per Minute.

  3. Observe the readings in the CDI Monitor Display and the IBC VenOsat Monitor Display. If the readings are suitable and stable, Take a sample from the sample port and run it through the IL Blood Gas Analyzer and the IL Co-Oximeter. Simultaneously document the readings from both on-line monitors and press the recalibration button on the CDI Model 400 Monitor.

  4. Recalibrate the monitor using the values obtained from the IL Co-Oximeter and Blood Gas Analyzer.

  5. Move the clamp from Point A to Point B and, using the Gas Blender, alter the blood gases. Move the clamp from Point B back to Point A and wart for the on line readings to stabilize(3 minutes).

  6. After the three minute period, draw another blood sample for the IL analyzers and document the readings on the CDI monitor. Repeat this procedure (steps 6. and 7.) until ten different readings are obtained.

  7. Stop the Pump Head .

.

  1. Calibrate two fresh Kwik Cell Sensors and snap them into the IBC flow through cells

  2. Resume operation of the Roller Pump at a flow rate of 3.5 Liters per Minute.

5

  1. Repeat steps 4., 5., 6. and 7. and then stop the pump and change the circuit using fresh disposables and blood for the next test samples.

  2. Repeat the test using IBC manufactured components first and CDI manufactured components second.

  3. Follow this alternating procedure for ten runs (until five pairs of cells have been tested from both manufacturers).

E. Data Collection and Treatment.

ﻬﺎ ﻓﻬﻮ

  1. Record each CDI reading along with the appropriate IL readings.

  2. Calculate the mean variance and mean per cent variance for each test sample and tabulate at the bottom of the log sheets.

  3. Tabulate the mean error and % error from each of the ten runs on the Gas Stat Test Data Analysis Log.

  4. Calculate the average error for all 100 data points for all measured and calculated parameters. Record these calculations on the Analysis log

  5. Construct a graph for comparison of all 100 pairs of readings for the measured parameters, pO2, pCO2 and pH, obtained from testing components of both manufacturers.

  6. The following pages contain summary of the findings.

6

Image /page/6/Picture/0 description: The image shows a curved black line on a white background. The line is thickest in the middle and tapers off towards the ends. The curve is shallow, resembling a crescent shape.

IBC QUICK CELL AND CDI KWINGCELL TEST DATA ANALYSIS

ﻤﺴﺴﺴﺴ

:

SAMPLEpO2pCO2pH[HCO3]BESAT
IBC #1 ERROR500.0111.7
IBC #1 % ERR7%0%0%2%-189%3%
IBC #2 ERROR-8-10.030.31.0
IBC #2 % ERR-6%-3%0%1%-42%-2%
IBC #3 ERROR2-10.000-0.7
IBC #3 % ERR1%-2%0%2%13%2%
IBC #4 ERROR11-0.020-0.2
IBC #4 % ERR0%2%0%-2%7%-3%
IBC #5 ERROR-100.00-10.2
IBC #5 % ERR-1%-1%0%-3%-5%-2%
AVERAGE IBC ERROR000000
IBC AVG % ERROR0%0%0%0%-22%0%
CDI #1 ERROR520.0011.53
CDI #1 % ERR8%4%0%3%-170%4%
CDI #2 ERROR2-1.75-0.01-10.0-3
CDI #2 % ERR1%-4%0%-3%0%-3%
CDI #3 ERROR32-0.021-0.6-4
CDI #3 % ERR3%4%0%3%11%-5%
CDI #4 ERROR-61-0.010.10.40
CDI #4 % ERR-5%3%0%0%-14%0%
CDI #5 ERROR-700.0011.21-6
CDI #5 % ERR-6%1%0%6%-35%-6%
AVERAGE CDI ERROR00000-1
CDI AVG % ERROR0%1%0%1%-21%-1%

7

Image /page/7/Figure/0 description: This image is a scatter plot titled "GRAPH 1" that compares "IL DATA" to "CDI DATA WIT- IBC CELLS". The x-axis ranges from 0 to 300, representing the values for "CDI DATA WIT- IBC CELLS", while the y-axis ranges from 0 to 300, representing the values for "IL DATA". A dashed line runs diagonally across the plot, and multiple data points are clustered around this line, indicating a positive correlation between the two datasets.

PO2 COMPARISON BETWEEN INSTRUMENTATION LABORATORIES AND CDI USING IBC CELLS

Carana

8

Image /page/8/Figure/0 description: This image is a scatter plot comparing IL data to CDI data with CDI cells. The x-axis represents CDI data with CDI cells, ranging from 0 to 300, while the y-axis represents IL data, also ranging from 0 to 300. The plot includes a dashed line, and the data points are clustered around this line, indicating a positive correlation between the two datasets. The graph is labeled as "GRAPH 2" and titled "PO2 COMPARISON BETWEEN INSTRUMENTATION LABORATORIES AND CDI USING CDI CELLS".

9

Image /page/9/Figure/0 description: This image is a scatter plot comparing IL data to CDI data with IBC cells. The x-axis represents CDI data with IBC cells, ranging from 0 to 75, while the y-axis represents IL data, also ranging from 0 to 75. The plot shows a positive correlation between the two datasets, with data points clustered around a dashed line that indicates a strong linear relationship. The title of the graph is "GRAPH 3".

pc02 COMPARISON BETWEEN INSTRUMENTATION LABORATORIES AND CDI USING IBC CELLS
. Analana

10

Image /page/10/Figure/0 description: This image is a scatter plot comparing 'IL DATA' and 'CDI DATA WITH CDI CELLS'. The x-axis represents 'CDI DATA WITH CDI CELLS', ranging from 0 to 75, while the y-axis represents 'IL DATA', also ranging from 0 to 75. A dashed line runs diagonally across the plot, and the data points are clustered closely around this line, indicating a positive correlation between the two variables.

GRAF 2 4

pC02 COMPARISON BETWEEN INSTRUMENTATION LABORATORIES AND CDI USING CDI CEN ﻤﺴﺘﺸﺎﻋﺎ

11

Image /page/11/Figure/0 description: This image is a scatter plot titled "GRAPH 5". The x-axis is labeled "CDI DATA WITH IBC CELLS" and ranges from 7.00 to 7.50. The y-axis is labeled "IL DATA" and ranges from 7.00 to 7.50. There is a dashed line going through the data points.

ph Comparison between instrumentation laboratories and cdi using ibc cells.

Ages on

12

Image /page/12/Figure/0 description: This image is a scatter plot titled "Graph 6". The x-axis is labeled "CDI DATA WITH CDI CELLS" and ranges from 7.00 to 7.50. The y-axis is labeled "IL DATA" and also ranges from 7.00 to 7.50. There are many data points plotted on the graph, and a dashed line runs diagonally across the plot.

ph Comparison between instrumentation laboratories and cdi using cod cells.

13

II Assembly Integrity.

ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

A. Assembly Leak Test.

  1. Assemble sixty IBC Quick Cell Components using Standard Operating Procedure, using the semi-opaque membrane as designed.

  2. Subject the assemblies to the IBC standard sterilization procedure and allow a fourteen day degassing period.

  3. Store twenty of the assemblies 24 hrs at 4℃, 25℃ and 60°C.

Place the assemblies to fit loosely in a drum type 4. Shake the container in a paint mixer for 5 container. minutes.

న్. Package the assemblies per Standard Operating Procedure and drop from a height of fifteen feet ten times.

  1. Pressurize the assemblies at 10 P.S.I. under water and check for leaks with the membrane support in place.

  2. Remove the membrane support clip from the assemblies, fully wet the semi-opaque membrane to establish the bubble point and check for leaks at 10 P.S.I. under water.

  3. No leaks were detected.

B. Clinical Simulation Leak Test.

l . Remove the sixty samples from the Assembly leak test section and allow them to dry for use in this section.

14

Image /page/14/Figure/0 description: This image is a diagram of a leak test assembly. The diagram includes a COBE CML, a Gambro gas blender, and a pump. The gas blender is connected to sources of O2, N2, and CO2. There are also 60 total cells and a sampling syringe.

15

  1. Assemble the components in series using a Cobe CML Membrane Oxygenator and extracorporeal tubing and connectors as shown in Leak Test Assembly schematic drawing.

  2. Prime the circuit with Bovine Blood which has been freshly collected and preserved in ACD A U.S.P. for less than twenty four hours.

  3. Recirculate at 6 Liters per Minute flow rate for one hour at 37℃, for four hours at 25℃ and for one hour at 37℃.

  4. Stop the pump, remove the Membrane Support Guards and observe for leaks. There should be a straw colored plasma present on top of the membrane but no cellular (Red) components.

  5. No leaks were detected.

C. Sensor Seal Integrity.

  1. Select ten samples from the Assembly leak test section.

  2. Remove the membranes completely using a scalpel.

  3. Insert the used sensors from the functional test section.

  4. Pressurize under water at 10 P.S.I. and observe for leaks.

  5. No leaks were detected.

III. Toxicity Testing.

A. U.S.P. Plastic Class 6 Testing.

  1. Prepare plastic samples of the clear plastic housing

16

material, the elastomer seal material and the semi-opaque white membrane as needed.

  1. Subject the samples to the IBC standard ethylene oxide sterilization cycle.

  2. Allow 14 days for aeration.

  3. Provide samples to an independent laboratory for U.S.P. Testing per current GLP's.

  4. The samples were found to meet U.S.P. Plastic Class 6.

B. Hemolysis.

్రాల్లో ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ప్రాథమిక ఆరోగ్య కేంద్రం, ప్రాథమిక ఆరోగ్య కేంద్రం, ప్రాథమిక ఆరోగ్య కేంద్రం, ప్రాథమిక

  1. Assemble and sterilize 10 IBC Quick Cell components and allow 14 days for aeration

Place all ten connectors in series using a Cobe CML 2. Membrane Oxygenator and extracorporeal tubing and connectors as shown in the Hemolysis schematic diaphragm.

Prime the circuit with fresh Bovine blood that is 3. preserved in ACD A U.S.P. and stored for less than 24 hours.

  1. Recirculate at 6 Liters per minute for 6 Hours at 37℃.

  2. Repeat the test with 10 CDI Kwik Cell components.

  3. Centrifuge a representative sample from each test run and from the uncirculated blood.

  4. Observe the plasma fraction for signs of free hemoglobin. It should be straw colored if free of hemolysis induced plasma hemoglobin. Plasma hemoglobin will give the plasma a distinct pink to red tint.

17

Image /page/17/Figure/0 description: This image shows a diagram of a Hemolysis Test Assembly. The assembly includes a COBE CML, a Gambro gas blender with inputs for O2, N2, and CO2, and a Gambro pump. There are 10 total cells labeled "IBC" connected in a series, with an arrow indicating the direction of flow. A sampling syringe is also included in the assembly.

18

    1. The samples were found to be Non-Hemolytic.

IV. Bioburden and Sterilization Evaluation.


ﻬﺎﺗﻬﺎ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

A. Bioburden. Per C.G. Laboratories standard methods and GLP's, the samples were found to contain an average bioburden of 20 colony forming units per assembly.

B. Ethylene Oxide Residues. Per C. G. Laboratories standard methods and GLP's, the following Ethylene Oxide Residue levels were noted.

| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Company Company County County Completent possible on any post and poss and | THE COLORIAL COLLECTION COLL COLL COLL COLL COLL COLL COLL COLL CLAIM A . AND A AND A AND A AND A AND A AND A AND A AND A AND A AND A AND A A

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ----------
And Call of Children and Children
.
.
CARD LEAST LE LEBELLER LE BERREAL CONSULT CARDING LEAST LEAST LEAST LAND AND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LAND LA
Standa Comments comes

and and the control control of the control of the control of the contribution of the contribution of the contribution of the contribution of the contribution of the contribut |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ़
t
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -
œ |

C. Pyrogenicity. Per the United States Pharmacopeia, the samples were found to be Non-Pyrogenic, U.S.P.

Discussion

  • Functionally, the CDI Model 400 Blood Gas Monitoring System is reasonably accurate when used in accordance with the Operator's Manual. The primary variations seen during the testing were attributed to variations from sensor to sensor. The calibration of the sensor using calibration gases and the buffer solution would in most cases yield reasonable results. The device became exceedingly accurate after one on line recalibration. The largest variations as seen on the graphs are generally attributable to the first readings taken after going on-line.
    The oxygen sensor is located on one semi-opaque membrane window. This membrane is porous and less susceptible to error than the 300 series. This was particularly noticeable at high pO2 levels. This is probably attributable to the lack of porosity with the 300 series window. This sensor, however does require the clarity of the membrane to function properly. The Model 400 employs a porous membrane which is clear when wetted. The apparent superiority of the IBC cells relative to the CDI cells is a result of a greater number of high readings with CDI. The products actually showed no significant difference relative to the pO, readings.

The remainder of the sensors are located on the second semi-opaque membrane. This

19

membrane is microporous (0.2 u) and allows a free flow of ions and gases through it while maintaining a sterile barrier. The free flow of Hydrogen ions to the sensor provides the stimulus that allows the transducer to measure pH. On the arterial channel, this pH measurement is factored into the equation for Bicarbonate and Base Excess calculations. The pH measurements are comparably accurate when one compares the IBC and CDI flow through cells.

The pCO, sensor is likewise located on the second semi-opaque membrane. Dissolved CO2, carried in the bicarbonate buffer system of the plasma fraction, provides the stimulus to the pCO2 sensor which allows the transducer to measure pCO2. The pCO2 measurement on the arterial channel is another of the factors that is fed into the Bicarbonate and Base Excess calculations. The pCO2 measurements are comparably accurate when one compares the IBC and CDI flow through cells.

The Oxygen Saturation Percent is far less accurate when compared to the IL Co-Oximeter as opposed to the IL Blood Gas Analyzer. For the purpose of this evaluation, the Co-Oximeter values were used since those values are generally used by perfusionists. The IL Blood Gas Values are calculated just as the CDI values are and they were found to be in very close agreement. Perhaps the generally accepted formulae that these companies employed for their calculated values should be reviewed. In any case, these values were not used in comparing the IBC and CDI flow through cells. The two cells seemed to perform comparably relative to % Oxygen Saturation.

The only measured parameter that is not mentioned in the data section is temperature. The temperature was maintained throughout the functional testing phase at 37℃ by using the Cincinnati Sub Zero Heater/Cooler in association with the Cobe CML integral heat exchanger. The reading on the CDI Monitor remained at 37℃ throughout the experiments regardless of whether the monitor was reading the Arterial or Venous channel (Only one channel can be read at a time). The thermo electronic sensor is held in intimate contact with a hemispherical shaped shell of stainless steel molded into the body of the sensor. This shell is in turn held in intimate contact with the semi-opaque membrane. The mass of the materials is minimal and capable of rapidly transferring heat adequately to yield accurate temperature readings. There was no difference noted ' between the IBC flow through cells and the CDI flow through cells.

1988

The response time of the CDI Monitor seemed to be two to three minutes depending on the size of the incremental change in the measured parameters. A large swing, e.g.,

20

pO, 60 mmHg to 180 mmHg and pCO, 60 mmHg to 40 mmHg and pH 7.20 to 7.40, required a full three minutes to stabilize. Regardless of the amount of change in parameters, the full three minutes was used throughout the data gathering.

The assemblies of IBC product were found to be sound and free of leaks. The assembly methods yield bond strengths far in excess of the requirements that might be encountered in a clinical setting. The bond between the membrane and the support frame was evaluated very closely. The semi-opaque membrane is porous and easily bonded to the polycarbonate with the UV adhesive we selected, but the membrane is made of a material that is inherently fragile when bonded with most adhesives.

The materials, three in all, used to fabricate the IBC Quick Cell Component all meet U.S.P. Plastic Class 6 Testing. The materials are non toxic after Ethylene Oxide Sterilization and meet the FDA recognized standards for ethylene oxide residues after 14 day aeration. The fabrication methods result in low bioburden assuring adequate margins of safety in normally validated ethylene oxide sterilization cycles.

The hemolysis testing was designed to be five times more stringent than the exposure the blood would get in the most severe clinical instance. In spite of this extreme protocol, there was no detectable free hemoglobin in the plasma fraction. The use of the IBC product in place of the CDI product will result in no differences relative to trauma to the blood. While hemolysis is not the only form of damage that the blood might encounter in the extracorporeal setting, it is generally recognized as representative of the level of trauma the blood is experiencing.

Conclusion

్లో ప్రాథమిక

  1. The IBC Quick Cell Component was tested in comparison with the CDI Kwik Cell Component within the CDI Model 400 Blood Gas Monitoring System. The two products were identical relative to performance and function.

  2. The IBC Quick Cell component was tested for its integrity and found to be sound and free of leaks and assembly weaknesses. The product is safe for its intended use in extracorporeal circuits.

  3. The IBC Quick component was evaluated for Toxicity and found to be nontoxic. It is suitable for use in the intended extracorporeal applications.

21

21

  1. The IBC Quick component was evaluated for its compatibility with Ethylene Oxide Sterilization. The IBC component may be safely used in tubing packs which are gas sterilized and may be sold after gas sterilization for insertion into the appropriate tubing line by the Cardiovascular surgical team.

  2. The IBC Quick component is not hemolytic when used in the extracorporeal circuit in accordance with normal clinical practice.

  3. The membrane used in the IBC Quick Cell Component is identical to that used in the CDI product.

The IBC Quick Cell Component may be substituted for the CDI Kwik Cell 7. Component in the construction of Extracorporeal Custom Tubing Packs either by the manufacturer of such packs or by the end user to modify his or her pack as needed.