(178 days)
Intended use for the Flowmate LTE is for Pulmonary Function Testing (measuring a person's ability to move air into and out of their lungs). The volume and flow rates measured for the entire effort and then reported as numerical results with a printout of the effort if selected.
Flow sensing pneumotach and base that measures differential pressure as air passes through it (the same principle as the Flowmate, PC Flow +, Renaissance, and Flash devices)
Compact design molded from high impact UL rated ABS plastic. rechargeable Ni-Cad batteries for portable operation,
Automatic calculations of test results, and best test selection based on ATS criteria
Testing capabilities include; FVC, FEV1, FEV3, FEV1/FVC%, FEF.2-1.2, FEF25-75%, FEF75-85, PEF, FEF50, FIVC, FEF50/FIF50, PIF, MVV, Methacholine Challenge, and Lung Age
Testing Capabilities: Capacity 10 Liters
Volume Accuracy +/- 3% Max. Flow Rate +/- 16 L/s
Back Pressure Less than 1.5 cm H2O
Population Predicted Normals included: Composite Knudson Crapo (ITS) ECCS HSU
Customization allows user to change report format, input temperature and barometric readings, and select printer driver for printer being used
Reports printed on standard 8 1/2 " x 11" paper
Field calibration with a 1 to 9 Liter calibration syringe
Cross-contamination control via external filter system
Here's an analysis of the provided text regarding the acceptance criteria and study for the Flowmate LTE device:
Acceptance Criteria and Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Volume Accuracy | +/- 3% |
Max. Flow Rate | +/- 16 L/s |
Back Pressure | Less than 1.5 cm H2O |
FVC Accuracy | Exceeded ATS recommendations at 23 out of 24 waveforms. At waveform #17 (which exceeds the device's 15-second recording limit), accuracy was -1.8% at the 15-second mark, which is within ATS recommendations. |
FEV1 Accuracy | Exceeded ATS recommendations at 23 out of 24 waveforms. |
FEF 25-75 Accuracy | Exceeded ATS recommendations at 23 out of 24 waveforms. |
PEF Accuracy | Exceeded ATS recommendations at 23 out of 24 waveforms. |
MVV Accuracy | Acceptable results compared with ATS criteria for accuracy and precision to delivered values. |
Study Information
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size:
- Waveform Performance Testing: 24 waveforms were used.
- MVV Testing: A 2-liter volume at flow rates from 40 to 245 L/M was used. The number of specific tests or readings performed is not specified, but it refers to a range of flow rates.
- Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. The testing appears to be laboratory-based performance testing rather than human clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth for the test set (waveforms and MVV) was established based on "ATS recommendations" and "delivered values" from a 2-liter calibration syringe. No human experts are mentioned as directly establishing the ground truth for these performance tests.
4. Adjudication method for the test set:
- Not applicable, as the ground truth was based on pre-defined standards (ATS recommendations) and objective measurements from a calibration syringe, not on human interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case comparative effectiveness study was not done. This device is a diagnostic spirometer, which directly measures physiological parameters, not an AI-powered image analysis or diagnostic support tool that would typically involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance testing described for the Flowmate LTE is a standalone performance study. The device's measurements (volume, flow rates) are directly compared against established standards (ATS recommendations) and known delivered values (from a calibration syringe). There is no "human-in-the-loop" aspect to the performance evaluation itself, though the device is intended for human use.
7. The type of ground truth used:
- The ground truth used was based on expert consensus standards (ATS recommendations for FVC, FEV1, FEF 25-75, PEF, MVV accuracy and precision) and objective measurements from a calibrated device (a 1 to 9 Liter calibration syringe for field calibration, and a 2-liter volume for MVV testing).
8. The sample size for the training set:
- The document does not describe a "training set" in the context of machine learning or AI. This device appears to be a hardware-based medical device with integrated software for calculations, not a learning-based algorithm that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable, as there is no mention of a training set for a machine learning algorithm.
§ 868.1840 Diagnostic spirometer.
(a)
Identification. A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.(b)
Classification. Class II (performance standards).