Intended Use / Indications for Use
Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
The subject device, MAGNETOM Vida with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Open Workflow
Modified features and applications:
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified Features and Applications:
- GRE_PC
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified Features and Applications:
- GRE_PC
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Vida Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- Beat Sensor
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- GRE_PC
- Open Recon 2.0
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Deep Resolve for EPI
- Deep Resolve for HASTE
- Physiologging
- Complex Averaging
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- myExam Angio Advanced Assist (Test Bolus)
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
- Complex Averaging
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- myExam Angio Advanced Assist (Test Bolus)
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Altea with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Altea with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- myExam Implant Suite
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- myExam Implant Suite
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
The subject device, MAGNETOM Viato.Mobile with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato.Mobile with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
With the subject software version, Syngo MR XB10, we are also introducing the following new 1.5T devices, which are part of our MAGNETOM Flow. Platform:
MAGNETOM Flow.Elite
MAGNETOM Flow.Neo
MAGNETOM Flow.Rise
The subject device, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- Magnet
- MREF (Magnet Refrigerator)
- Gradient system
- Gradient Coil
- RF System
- System Cover
- Patient Table
- MaRS (Measurement and Reconstruction System)
- Select&GO Display (TPAN_3G) and Control Panel (CPAN_2G)
- Body Coil
- Head/Neck Coil
- BM Head/Neck Coil (with ComfortSound)
- BM Contour S Coil
- BM Contour M Coil
- BM Contour L Coil
- BM Contour XL Coil
- Foot/Ankle Coil
- BM Spine Coil
- iTx Extremity 18 Flare
- Multi-Index MR-RT Positioning (a part of "RT Pro Edition" marketing bundle) (not available for MAGNETOM Flow.Rise)
Modified Hardware:
- Gradient Power Amplifier (GPA)
- SAR Monitoring
- In-Vivo Shim
Software
New Features and Applications:
- CS Vibe
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- Quick Protocols
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements (not available for MAGNETOM Flow.Rise)
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
New (general) Software / Platform / Workflow:
- Select&GO extension (coil-based Iso Centering, Patient Registration at the touch display, Start Scan at the touch display)
- New Startup-Timer
- myExam RT Assist (not available for MAGNETOM Flow.Rise)
- myExam Brain RT-Autopilot (not available for MAGNETOM Flow.Rise)
- Eco Power Mode Pro
Modified (general) Software / Platform:
- Improved Gradient ECO Mode Settings
Furthermore, the following minor updates and changes were conducted for the subject devices MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Sola Fit and MAGNETOM Viato.Mobile:
- Off-Center Planning Support
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Marketing bundle "myExam Companion"
- Marketing Bundle "RT Pro Edition"(not available for MAGNETOM Flow.Rise)
This FDA 510(k) clearance letter pertains to several MAGNETOM MRI systems with software Syngo MR XB10. The document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing of new and modified hardware and software features, particularly those involving Artificial Intelligence (AI) such as "Deep Resolve" functionalities.
Here's an analysis of the acceptance criteria and the studies that prove the devices meet them, specifically for the AI features:
1. Table of Acceptance Criteria and Reported Device Performance for AI Features
The document does not explicitly state "acceptance criteria" for the AI features in a numerical format that would typically be seen for a device's performance metrics (e.g., minimum sensitivity, specificity). Instead, the acceptance criteria are implicitly defined by the evaluation methods and the "Test result summary" for each Deep Resolve feature, which aim to demonstrate equivalent or improved image quality compared to conventional methods.
| AI Feature | Acceptance Criteria (Implied) | Reported Device Performance | Comments |
|---|---|---|---|
| Deep Resolve Swift Brain | - Quantitative quality metrics (PSNR, SSIM, NMSE) to demonstrate network impact.- Visual inspection to ensure no undetected artifacts.- Evaluation in clinical settings with collaboration partners. | - "Impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and normalized mean squared error (NMSE)."- "Images were inspected visually to ensure that potential artefacts are detected that are not well captured by the metrics."- "Work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners." | The results indicate successful performance in meeting these criteria, suggesting the AI feature performs as intended without negative impact on image quality and with acceptable quantitative metrics. |
| Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE | - Quantitative evaluations (SSIM, PSNR, MSE) showing convergence of training and improvements over conventional parallel imaging.- Visual inspection to confirm no negative impact on image quality.- The function should allow for faster acquisition or improved image quality. | - "Quantitative evaluations of structural similarity index (SSIM), peak signal-to-noise ratio (PSNR) and mean squared error (MSE) metrics showed a convergence of the training and improvements compared to conventional parallel imaging."- "An inspection of the test images did not reveal any negative impact to the image quality."- "The function has been used either to acquire images faster or to improve image quality." | The results indicate successful performance, demonstrating quantitative improvements and confirming user benefit (faster acquisition or improved image quality) without negative visual impact. |
| Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE | - Quantitative quality metrics (PSNR, SSIM, perceptual loss).- Rating and evaluation of image sharpness by intensity profile comparisons.- Demonstration of increased edge sharpness and reduced Gibb's artifacts. | - "The impact of the Deep Resolve Sharp network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss."- "The tests include rating and an evaluation of image sharpness by intensity profile comparisons of reconstruction with and without Deep Resolve Sharp. Both tests show increased edge sharpness and reduced Gibb's artifacts." | The results directly confirm improved image sharpness and reduced artifacts, meeting the implied performance criteria. |
| Deep Resolve Boost for TSE | - Similar metrics (PSNR, SSIM, LPIPS) to predicate (cleared) network, both outperforming conventional GRAPPA.- Statistically significant reduction of banding artifacts.- No significant changes in sharpness and detail visibility.- Radiologist evaluation confirming no difference in suitability for clinical diagnostics. | - "The evaluation on the test dataset confirmed very similar metrics in terms of peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and learned perceptual image patch similarity metrics (LPIPS) for the predicate and the modified network with both outperforming conventional GRAPPA as the reference."- "Visual evaluations confirmed statistically significant reduction of banding artifacts with no significant changes in sharpness and detail visibility."- "In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics between updated and cleared predicate network." | This AI feature directly demonstrates equivalent or improved performance compared to the predicate, with specific mention of "radiologist evaluation" ensuring clinical suitability. |
2. Sample Size Used for the Test Set and Data Provenance
Since the document distinguishes between training, validation, and testing datasets, the "test set" here refers to the data used for final evaluation of the AI model's performance.
-
Deep Resolve Swift Brain:
- Test Set Sample Size: The document lists "Validation: 3,616 slices (1.5T validation); 6,048 slices (3T validation)" as part of the split. It also mentions "work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners," implying additional testing, but a specific numerical sample size for this external validation is not provided in detail. However, the initial splits serve as the primary "test set" for performance metrics mentioned.
- Data Provenance: "in-house measurement," implying retrospective data collected at Siemens' facilities. The document notes that "attributes like gender, age and ethnicity are not relevant to the training data" due to network architecture, but no specific country of origin is stated beyond "in-house."
-
Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE:
- Test Set Sample Size: The document states 19% of 1265 measurements for validation. It also explicitly mentions "collaboration partners (testing)" indicating an external test set, but a specific numerical breakdown for this is not provided.
- Data Provenance: "in-house measurements (training and validation) and collaboration partners (testing)." This suggests a mix of retrospective data potentially from various countries where Siemens has collaboration, though specific locations are not listed.
-
Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE:
- Test Set Sample Size: 30% of the 500 measurements are listed for validation, which serves as a test set. This equates to 150 measurements.
- Data Provenance: "in-house measurements," implying retrospective data from Siemens' research facilities. Specific country not mentioned.
-
Deep Resolve Boost for TSE:
- Test Set Sample Size: "Additional test dataset for banding artifact reduction: more than 2000 slices."
- Data Provenance: "in-house measurements and collaboration partners" for training/validation. The "additional test dataset for banding artifact reduction" likely follows the same provenance. Retrospective data.
3. Number of Experts Used and Qualifications for Ground Truth
The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience") for any of the Deep Resolve features.
However, for Deep Resolve Boost for TSE, it mentions:
- "Visual evaluations confirmed statistically significant reduction of banding artifacts... "
- "In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics..."
This indicates that radiologists were involved in the evaluation of the Deep Resolve Boost for TSE feature, presumably as experts to establish the clinical suitability. The exact number and their detailed qualifications are not provided. For other features, the ground truth is primarily based on the acquired raw data or manipulated versions of it, without explicit mention of expert review in the ground truth establishment process.
4. Adjudication Method (for the test set)
The document does not specify an adjudication method like "2+1" or "3+1" for establishing ground truth or evaluating the test set for any of the AI features. The ground truth for training and validation is derived from the "acquired datasets" which are considered the ground truth due to data manipulation and augmentation from these high-quality source images. For Deep Resolve Boost for TSE, a "radiologist evaluation" is mentioned, implying expert review without detailing a specific adjudication protocol.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to measure the improvement of human readers with AI assistance versus without AI assistance. The evaluations focus on the standalone performance of the AI algorithms in improving image quality metrics and, in one instance (Deep Resolve Boost for TSE, radiologist evaluation), the suitability for clinical diagnostics, rather than the impact on human reader performance.
6. Standalone (Algorithm Only) Performance
Yes, standalone (algorithm only) performance was done. The descriptions for each Deep Resolve feature focus entirely on the algorithm's performance in terms of quantitative image quality metrics (PSNR, SSIM, NMSE, MSE, LPIPS), visual inspection for artifacts, and improvements over conventional techniques. There is no mention of a "human-in-the-loop" component in the described performance evaluations for these AI features, except for the "radiologist evaluation" for Deep Resolve Boost for TSE which assessed clinical suitability of the output images, not reader performance with the AI.
7. Type of Ground Truth Used
-
For Deep Resolve Swift Brain, Deep Resolve Boost for FL3D_VIBE & SPACE, and Deep Resolve Sharp for FL3D_VIBE & SPACE:
- The ground truth used was the acquired datasets (raw MRI data). The input data for the AI models was then "retrospectively created from the ground truth by data manipulation and augmentation" (e.g., undersampling k-space, adding noise, cropping, creating sub-volumes, cropping k-space to simulate low-resolution input from high-resolution output). This means the AI models were trained to learn the mapping from manipulated (e.g., noisy, low-resolution, undersampled) inputs to the original, high-quality acquired image data.
-
For Deep Resolve Boost for TSE:
- Similar to above, the "acquired training/validation datasets" were considered the ground truth. Input data was generated by "data manipulation and augmentation" (e.g., discarding k-space lines, lowering SNR, mirroring k-space data).
In essence, the AI models are trained to restore or enhance images to resemble the high-quality, fully acquired MRI data that serves as the reference ground truth.
8. Sample Size for the Training Set
- Deep Resolve Swift Brain: 20,076 slices
- Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE: 81% of 1265 measurements. (This equates to approximately 1024 measurements).
- Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE: 70% of 500 measurements. (This equates to 350 measurements).
- Deep Resolve Boost for TSE: More than 23,250 slices (93% of the total dataset).
9. How the Ground Truth for the Training Set Was Established
For all Deep Resolve features, the ground truth for the training set was established from acquired MRI datasets (either "in-house measurements" or from "collaboration partners"). These acquired datasets are implicitly considered the "true" or "high-quality" images. The AI models are designed to process inputs that mimic suboptimal acquisition conditions (e.g., undersampled k-space, lower SNR, lower resolution) and generate outputs that match these high-quality acquired images, which serve as the ground truth for learning. The process involved:
- Retrospective creation: Input data was created retrospectively from the acquired ground truth data.
- Data manipulation and augmentation: This involved techniques such as:
- Discarding k-space lines (undersampling).
- Lowering the SNR level by adding Gaussian noise to k-space data.
- Uniformly-random cropping of training data.
- Creating sub-volumes of acquired data.
- Cropping k-space to generate low-resolution inputs corresponding to high-resolution ground truth.
- Mirroring of k-space data.
This approach demonstrates an unsupervised or self-supervised learning paradigm where the ground truth is derived directly from the complete and high-fidelity raw data, and the AI is trained to reconstruct or enhance images from degraded inputs to match this ideal ground truth.
FDA 510(k) Clearance Letter - MAGNETOM MRI Systems
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
December 19, 2025
Siemens Healthcare GmbH
℅ Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, Pennsylvania 19355
Re: K252838
Trade/Device Name: MAGNETOM Sola; MAGNETOM Altea; MAGNETOM Sola Fit; MAGNETOM Viato.Mobile; MAGNETOM Vida; MAGNETOM Lumina; MAGNETOM Vida Fit; MAGNETOM Flow.Elite; MAGNETOM Flow.Neo; MAGNETOM Flow.Rise
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, LNI, MOS
Dated: November 20, 2025
Received: November 20, 2025
Dear Alina Goodman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
Page 2
K252838 - Alina Goodman Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
Page 3
K252838 - Alina Goodman Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, PhD
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K252838
Device Name: MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise
Indications for Use (Describe)
Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
1. General Information
Establishment: Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Registration Number: 2240869
Date Prepared: September 29, 2025
Manufacturer: Siemens Healthineers AG
Magnetic Resonance (MR)
Allee am Röthelheimpark 2
91052 Erlangen
Germany
Registration Number: 3002808157
Siemens Shenzhen Magnetic Resonance LTD.
Siemens MRI Center
Hi-Tech Industrial Park (middle)
Gaoxin C. Ave., 2nd
Shenzhen 518057
P.R. CHINA
Registration Number: 3004754211
2. Contact Information
Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Phone: +1(317)371-8593
E-mail: alina.goodman@siemens-healthineers.com
3. Device Name and Classification
Device/ Trade name: MAGNETOM Vida
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Lumina
Page 6
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Vida Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Sola
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Altea
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Sola Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Viato.Mobile
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Elite
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Page 7
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Neo
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Rise
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
4. Legally Marketed Predicate and Reference Device
4.1. Predicate Device
Trade name: MAGNETOM Vida
510(k) Number: K231560
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Lumina
510(k) Number: K231560
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola
510(k) Number: K232535
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Page 8
Trade name: MAGNETOM Altea
510(k) Number: K232535
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Viato.Mobile
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola Fit
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
4.2. Reference Device
Trade name: MAGNETOM Vida Fit
510(k) Number: K220939
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Flow.Ace
510(k) Number: K250436
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Skyra Fit
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
Page 9
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Free.Max
510(k) Number: K231617
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: MOS
Trade name: syngo.via VB40A
510(k) Number: K191040
Product Code: Primary: LLZ
5. Intended Use / Indications for Use
Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Page 10
6. Device Description
The subject device, MAGNETOM Vida with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Open Workflow
Modified features and applications:
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
Page 11
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified Features and Applications:
- GRE_PC
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
Page 12
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified Features and Applications:
- GRE_PC
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
Page 13
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Vida Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- myExam 3D Camera
- Beat Sensor
- BM Contour XL Coil
Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- GRE_PC
- Open Recon 2.0
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Deep Resolve for EPI
- Deep Resolve for HASTE
- Physiologging
Page 14
- Complex Averaging
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- myExam Angio Advanced Assist (Test Bolus)
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Page 15
- Complex Averaging
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- myExam Angio Advanced Assist (Test Bolus)
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Page 16
Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Altea with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Altea with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
Software
New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Compressed Sensing Cardiac Cine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- GRE_PC
Page 17
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- myExam Implant Suite
- Deep Resolve Boost for TSE
Page 18
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam RT Assist workflow improvements
- myExam Implant Suite
- Deep Resolve Boost for TSE
Page 19
- "MTC Mode" for SPACE
The subject device, MAGNETOM Viato.Mobile with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato.Mobile with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- BM Contour XL Coil
Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
Software
New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
With the subject software version, Syngo MR XB10, we are also introducing the following new 1.5T devices, which are part of our MAGNETOM Flow. Platform:
MAGNETOM Flow.Elite
MAGNETOM Flow.Neo
MAGNETOM Flow.Rise
Page 20
The subject device, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
New Hardware:
- Magnet
- MREF (Magnet Refrigerator)
- Gradient system
- Gradient Coil
- RF System
- System Cover
- Patient Table
- MaRS (Measurement and Reconstruction System)
- Select&GO Display (TPAN_3G) and Control Panel (CPAN_2G)
- Body Coil
- Head/Neck Coil
- BM Head/Neck Coil (with ComfortSound)
- BM Contour S Coil
- BM Contour M Coil
- BM Contour L Coil
- BM Contour XL Coil
- Foot/Ankle Coil
- BM Spine Coil
- iTx Extremity 18 Flare
- Multi-Index MR-RT Positioning (a part of "RT Pro Edition" marketing bundle) (not available for MAGNETOM Flow.Rise)
Modified Hardware:
- Gradient Power Amplifier (GPA)
- SAR Monitoring
- In-Vivo Shim
Software
New Features and Applications:
- CS Vibe
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- DANTE blood suppression
Page 21
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- Quick Protocols
- myExam Autopilot Spine
- Open Workflow
Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements (not available for MAGNETOM Flow.Rise)
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
New (general) Software / Platform / Workflow:
- Select&GO extension (coil-based Iso Centering, Patient Registration at the touch display, Start Scan at the touch display)
- New Startup-Timer
- myExam RT Assist (not available for MAGNETOM Flow.Rise)
- myExam Brain RT-Autopilot (not available for MAGNETOM Flow.Rise)
- Eco Power Mode Pro
Modified (general) Software / Platform:
- Improved Gradient ECO Mode Settings
Furthermore, the following minor updates and changes were conducted for the subject devices MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Sola Fit and MAGNETOM Viato.Mobile:
- Off-Center Planning Support
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
Page 22
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Marketing bundle "myExam Companion"
- Marketing Bundle "RT Pro Edition"(not available for MAGNETOM Flow.Rise)
7. Substantial Equivalence
MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10 are substantially equivalent to the following devices:
| Predicate Devices | FDA Clearance Number and Date | Product Code | Manufacturer |
|---|---|---|---|
| MAGNETOM Vida with syngo MR XA60A | K231560, cleared on October 23, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Lumina with syngo MR XA60A | K231560, cleared on October 23, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Sola with syngo MR XA61A | K232535, cleared on December 22, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Altea with syngo MR XA61A | K232535, cleared on December 22, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Viato.Mobile with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Sola Fit with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| Reference Devices | FDA Clearance Number and Date | Product Code | Manufacturer |
|---|---|---|---|
| MAGNETOM Flow.Ace with syngo MR XA70A | K250436, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Shenzhen Magnetic Resonance Ltd. |
| MAGNETOM Free.Max with syngo MR XA60A | K231617, cleared on September 11, 2023 | LNH, MOS | Siemens Shenzhen Magnetic Resonance |
Page 23
Ltd. |
| MAGNETOM Skyra Fit with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Vida Fit with syngo MR XA50A | K220939, cleared on April 29, 2022 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| syngo.via VB40A | K191040 | LLZ | Siemens Healthcare GmbH |
8. Technological Characteristics
The subject devices, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10, are substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance.
The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.
While there are some differences in technological characteristics between the subject devices and predicate device, including new and modified hardware and software, these differences have been tested and the conclusions from the non-clinical data suggest that the features bear an equivalent safety and performance profile to that of the predicate device.
9. Nonclinical Tests
The following performance testing was conducted on the subject devices:
| Performance Test | Tested Hardware or Software | Source/Rationale for test |
|---|---|---|
| Software verification and validation | New or modified software features | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices |
| Sample clinical images | Coils, new or modified software features | Guidance for submission of Premarket Notifications for Magnetic Resonance |
| Image quality assessment by sample clinical images | - new / modified pulse sequence types.- comparison images between the new / modified features and the predicate device features | Diagnostic Devices |
| Performance bench test | New and modified hardware | |
| Biocompatibility | surface of applied parts | ISO 10993-1 |
| Electrical safety and electromagnetic compatibility (EMC) | Complete systems MAGNETOM Flow. Platform | IEC 60601-1-2 |
The results from each set of tests demonstrate that the subject devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.
Below table shows an executive summary of training and validation dataset of new AI features in the subject devices.
Page 24
Deep Resolve Swift Brain:
| Test result summary | The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and normalized mean squared error (NMSE). Additionally, images were inspected visually to ensure that potential artefacts are detected that are not well captured by the metrics.After successful passing of the quality metrics tests, work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners. |
|---|---|
| Test setup | Equipment: 3T and 1.5T MRI scanners.Protocols: representative protocols (T1, T2*, T2, T2 FLAIR).Body region: brain.Sample size: 29,740 2D slices.Sample source: in-house measurement.Note: Due to the network architecture, attributes like gender, age and ethnicity are not relevant to the training data.Dataset split:Training: 20,076 slicesValidation:1.5T Validation: 3,616 slices (1.5T validation);3T Validation: 6,048 slices (3T validation).Note: Data split maintained similar data distribution (e.g. contrast, orientation, field strength, …) in both training and validation datasets. |
| Patient Characteristics | No clinical subgroups have been defined for the datasets. |
| Reference standard | The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition of Gaussian noise to k-space data and uniformly-random cropping of the training data along the readout direction. |
| Data independency | Datasets for training and testing were split prior to training (see Test Setup field above). |
Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE:
| Test result summary | Quantitative evaluations of structural similarity index (SSIM), peak signal-to-noise ratio (PSNR) and mean squared error (MSE) metrics showed a convergence of the training and improvements compared to conventional parallel imaging. An inspection of the test images did not reveal any negative impact to the image quality. The function has been used either to acquire images faster or to improve image quality. |
|---|---|
| Test setup | Equipment: 0.55T, 1.5T and 3T scanners |
Page 25
| Protocols: representative measurement protocols (T1, T2 and PD with and without fat saturation) which have been altered for training (e.g. to increase SNR, increase resolution or reduce acceleration).Body regions: broad range of body regionsUsed coils: broad range of coils to cover the dedicated body regionsSample size: 27,679 3D patches from 1265 measurementsDataset split: Training: 81% of the 1265 measurementsValidation: 19% of the 1265 measurementsNote: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, …) in both training and validation datasets.Sample source: in-house measurements (training and validation) and collaboration partners (testing) | |
|---|---|
| Patient Characteristics | Gender distribution:- Male: 53%- Female 47%Age: for training and validation.- 19 - 45: 14%- 46 - 65: 43%- 66 - 89: 43%Clinical subgroups: No clinical subgroups have been defined for the datasets. |
| Reference standard | The acquired datasets (as described above) represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines as well as creating sub-volumes of the acquired data. |
| Data independency | Datasets determined for training and validation were split prior to training along individual acquisitions to ensure that there is no mixture of sub-volumes stemming from the same acquisition. |
Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE
| Test result summary | The impact of the Deep Resolve Sharp network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. The tests include rating and an evaluation of image sharpness by intensity profile comparisons of reconstruction with and without Deep Resolve Sharp. Both tests show increased edge sharpness and reduced Gibb's artifacts. |
|---|---|
| Test setup | Equipment: 0.55T, 1.5T and 3T MRI scannersProtocols: representative measurement protocols (T1, T2 and PD with and without fat saturation) which have been altered (e.g. to increase SNR, increase resolution or reduce acceleration)Body regions: broad range of different body regions.Used coils: broad range of coils to cover the dedicated body regionsSample size: approx. 13,000 high resolution 3D patches from 500 measurements.Dataset split: Training: 70% of the 500 measurements. |
Page 26
| Validation: 30% of the 500 measurementsNote: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, …) in both training and validation datasets.Sample source: in-house measurements | |
|---|---|
| Patient Characteristics | Gender distribution:- Male: 66.6%- Female 33.4%Age: for training and validation.- 19 - 45: 8.4%- 46 - 65: 40.2%- 66 - 89: 51.4%Clinical subgroups: No clinical subgroups have been defined for the datasets. |
| Reference standard | The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation. k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation. |
| Data independency | The high-resolution datasets were split to 70% training and 30% validation datasets before training to ensure independence of them. The input and output variables of the network have been derived from the same dataset so that no confounders exist for the training methodology. |
Deep Resolve Boost for TSE
| Test result summary | The evaluation on the test dataset confirmed very similar metrics in terms of peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and learned perceptual image patch similarity metrics (LPIPS) for the predicate and the modified network with both outperforming conventional GRAPPA as the reference. Visual evaluations confirmed statistically significant reduction of banding artifacts with no significant changes in sharpness and detail visibility. In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics between updated and cleared predicate network.The function as on the predicate devices was modified to the subject devices but the training and testing from the predicate devices still fits. |
|---|---|
| Test setup | Equipment: 0.55T, 1.5T and 3T MRI scannersProtocols: representative protocols (T1, T2 and PD with and without fat saturation)Body regions: broad range of different body regionsUsed coils: broad range of coils to cover the dedicated body regionsTesting of the Deep Resolve Boost network has been described in the reference and predicate device submissions. Additional tests have been performed to evaluate the banding artifact reduction capabilities of the updated network.Dataset split: Training: more than 23250 slices (93%) |
Page 27
| Validation: more than 1750 slices (7%)Additional test dataset for banding artifact reduction: more than 2000 slicesFor training and validation of the network, the identical data was used as for the initial device submission (K213693).Note: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, …) in both training and validation datasets.Sample source: in-house measurements and collaboration partners | |
|---|---|
| Patient Characteristics | Due to reasons of data privacy, gender, age and ethnicity during data collection have not been recorded. Due to the network architecture, attributes like gender, age and ethnicity are not relevant to the training data.No clinical subgroups have been defined for the collected dataset |
| Reference standard | The acquired training/validation datasets (identical to the initial submission K213693) represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data. |
| Data independency | Training and validation datasets were kept independent from each other during training and validation. The acquired datasets (one dataset consists of a group of multiple slices) were split into 93% training and 7% validation data prior to the training. A similar distribution was maintained for training and validation data. The test dataset for banding artifact reduction was acquired after the release of the predicate network and is therefore independent of the training/validation data. |
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
10. Clinical Tests
No clinical tests were conducted to support substantial equivalence for the subject devices.
Sample clinical images were provided to support substantial equivalence for the subject devices.
11. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards.
Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
The subject devices with Syngo MR XB10 conform to the following FDA recognized and international IEC, ISO and NEMA standards:
Page 28
| Recognition Number | Product Area | Title of Standard | Reference Number and date | Standards Development Organization |
|---|---|---|---|---|
| 19-46 | General II (ES/ EMC) | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | ES60601-1:2005 /(R)2012 & A1:2012, C1:2009/(R)2012 &A2:2010/(R)2012 (Cons. Text) [Incl.AMD2:2021] | ANSI AAMI |
| 19-36 | General | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 60601-1-2 Edition 4.1 2020-09 | IEC |
| 12-347 | Radiology | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis | 60601-2-33 Edition 4.0 2022-08 | IEC |
| 5-125 | General I (QS/ RM) | Medical devices - Application of risk management to medical devices | 14971 Third edition 2019-12 | ISO |
| 5-129 | General I (QS/ RM) | Medical devices - Part 1: Application of usability engineering to medical devices | 62366-1: 2015 + AMD1:2020 | IEC |
| 13-79 | Software/ Informatics | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] | IEC 62304:2006 + AMD1:2015 | IEC |
| 12-232 | Radiology | Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices | MS 4-2010 | NEMA |
| 12-288 | Radiology | Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images | MS 9-2008 (R2020) | NEMA |
| 12-352 | Radiology | Digital Imaging and Communications in Medicine (DICOM) | PS 3.1 - 3.20 (2023e) | NEMA |
| 2-258 | Biocompatibility | Biological evaluation of medical devices - part 1: evaluation and testing within a risk management process. (Biocompatibility) | 10993-1: 2018 | ANSI AAMI ISO |
Page 29
12. Conclusion as to Substantial Equivalence
MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10 have the same intended use and same basic technological characteristics than the respective predicate devices, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate devices, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that the subject devices MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10 are substantially equivalent to the currently marketed devices MAGNETOM Vida and MAGNETOM Lumina with syngo MR XA60A (K231560, cleared on October 23, 2023), MAGNETOM Sola and MAGNETOM Altea with syngo MR XA61A (K232535, cleared on December 22, 2023) and MAGNETOM Sola Fit and MAGNETOM Viato.Mobile with syngo MR XA70A (K250443, cleared on June 16, 2024).
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.