K Number
K252769
Date Cleared
2025-10-02

(30 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Upper Arm Electronic Blood Pressure Monitor, Models FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116 are intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

Device Description

The subject device, Upper Arm Electronic Blood Pressure Monitor, is an automatic non-invasive blood pressure monitor which can be driven by 4 AA batteries / 4×AAA batteries or type-C USB (optional). It uses an inflatable cuff which is wrapped around the patient's upper arm to measure the systolic and diastolic blood pressure as well as the pulse rate of adult at household, not for neonate or pregnancy.

The unit uses the oscillometric method of blood pressure measurement. It means the unit detects the movement of your blood through your brachial artery, and converts your blood pressure into a digital reading. The unit is simple to use because a stethoscope is not needed while using an oscillometric monitor.

The unit stores the blood pressure and pulse rate in the memory after completing a measurement each time. 2x120 sets of measurement values can be stored automatically. The earliest record will be deleted automatically to save the latest measurement value when more than the measurement values. The unit also calculates an average reading based on the values of the latest 3 times measurement.

This blood pressure monitor has the function of blood pressure classification, which is convenient for you to judge whether your blood pressure is normal or not.

This blood pressure monitor has voice broadcast function. During measurement and recall the memory, there will be voice operation tips.

During blood pressure measurement, if the body moves excessively, the body motion detection icon will appear on the display.

Sleeve with self-check: Detects whether the cuff is worn correctly. If the pressure can exceed 30mmHg within 15 seconds, this Sleeve with self-check icon will light up.

During the blood pressure measurement decrease process, if a pulse wave is detected, this Pulse Detected Icon will flash at the same frequency as the peak of the pulse wave appears.

The device detects an Irregular HeartBeat (IRB) (a Heartbeat that is more than 25% slower or 25% faster from the average Heartbeat) two or more times during the measurement, the irregular heartbeat Symbol will appear on the display with the measurement values.

The device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device(except model FC-BP107).

There is a maximum pressure safety setting at 300 mmHg, when the pressure is more than 300mmHg, the device will exhaust fast automatically.

No operation for 2 minute the device will shut down automatically.

The device includes model FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116. Model FC-BP107 and FC-BP116 can be powered by internal battery (4×AA batteries, 6.0V DC) or Type-C port 5VDC 1A, other models can be powered by internal battery (4×AAA batteries, 6.0V DC) or Type-C port 5VDC 1A.

All models have same MCU, Air pump, Solenoid valve and blood pressure core algorithm except for PCB layout, Display screen, function of the Button, dimension and power supply.

The subject devices (FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116) and the predicate devices (FC-BP100, FC-BP101, FC-BP102, FC-BP110, FC-BP111, FC-BP112) share the same schematic, circuitry, critical components, and measurement algorithm.

For the differences between the subject device models, the differences are PCB layout difference (physical board shape and/or component arrangement differences), Display screen, function of the Button, dimension, and power supply.

These changes are mainly cosmetic and do not affect the electrical design, safety, or essential performance of the devices.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Upper Arm Electronic Blood Pressure Monitor

Page 1

October 2, 2025

Shenzhen Finicare Co., Ltd.
℅ Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lujiazui Rd.,
Shanghai, 200120
China

Re: K252769
Trade/Device Name: Upper Arm Electronic Blood Pressure Monitor (FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive blood pressure measurement system
Regulatory Class: Class II
Product Code: DXN
Dated: September 2, 2025
Received: September 2, 2025

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K252769 - Boyle Wang Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K252769 - Boyle Wang Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JENNIFER W. SHIH -S

for LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252769

Device Name: Upper Arm Electronic Blood Pressure Monitor(Model FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116)

Indications for Use (Describe)

Upper Arm Electronic Blood Pressure Monitor, Models FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116 are intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

K252769

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name: Shenzhen Finicare Co., Ltd.
Address: 201 and 301 of Building A22, Building A22 and A23,
No.4 Industrial Park, Tantou Community, Songgang Street,
Bao'an District, Shenzhen 518105 China
Tel: 86-755-23013503
Contact: Chao Li

Designated Submission Correspondent

Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai,
200120 China
Tel: +86-21-50313932
Email: Info@truthful.com.cn

Date of Preparation: Oct.02,2025

2.0 Device Information

Trade name: Upper Arm Electronic Blood Pressure Monitor
Common name: Noninvasive Blood Pressure Measurement System
Classification name: Noninvasive Blood Pressure Measurement System
Model(s): FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116
Production code: DXN
Regulation number: 21 CFR 870.1130
Classification: Class II
Panel: Cardiovascular

3.0 Predicate Device Information

Manufacturer: Shenzhen Finicare Co., Ltd.

Page 6

Trade name: Upper Arm Electronic Blood Pressure Monitor, Model: FC-BP100,FC-BP101,FC-BP102,FC-BP110,FC-BP111, FC-BP112
510(k) number: K220113

4.0 Indication for Use Statement

Upper Arm Electronic Blood Pressure Monitor, Models FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116 are intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

5.0 Device Description

The subject device, Upper Arm Electronic Blood Pressure Monitor, is an automatic non-invasive blood pressure monitor which can be driven by 4 AA batteries / 4×AAA batteries or type-C USB (optional). It uses an inflatable cuff which is wrapped around the patient's upper arm to measure the systolic and diastolic blood pressure as well as the pulse rate of adult at household, not for neonate or pregnancy.

The unit uses the oscillometric method of blood pressure measurement. It means the unit detects the movement of your blood through your brachial artery, and converts your blood pressure into a digital reading. The unit is simple to use because a stethoscope is not needed while using an oscillometric monitor.

The unit stores the blood pressure and pulse rate in the memory after completing a measurement each time. 2x120 sets of measurement values can be stored automatically. The earliest record will be deleted automatically to save the latest measurement value when more than the measurement values. The unit also calculates an average reading based on the values of the latest 3 times measurement.

This blood pressure monitor has the function of blood pressure classification, which is convenient for you to judge whether your blood pressure is normal or not.

This blood pressure monitor has voice broadcast function. During measurement and recall the memory, there will be voice operation tips.

During blood pressure measurement, if the body moves excessively, the body motion detection icon will appear on the display.

Sleeve with self-check: Detects whether the cuff is worn correctly. If the pressure can exceed 30mmHg within 15 seconds, this Sleeve with self-check icon will light up.

Page 7

During the blood pressure measurement decrease process, if a pulse wave is detected, this Pulse Detected Icon will flash at the same frequency as the peak of the pulse wave appears.

The device detects an Irregular HeartBeat (IRB) (a Heartbeat that is more than 25% slower or 25% faster from the average Heartbeat) two or more times during the measurement, the irregular heartbeat Symbol will appear on the display with the measurement values.

The device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device(except model FC-BP107).

There is a maximum pressure safety setting at 300 mmHg, when the pressure is more than 300mmHg, the device will exhaust fast automatically.

No operation for 2 minute the device will shut down automatically.

The device includes model FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116. Model FC-BP107 and FC-BP116 can be powered by internal battery (4×AA batteries, 6.0V DC) or Type-C port 5VDC 1A, other models can be powered by internal battery (4×AAA batteries, 6.0V DC) or Type-C port 5VDC 1A.

All models have same MCU, Air pump, Solenoid valve and blood pressure core algorithm except for PCB layout, Display screen, function of the Button, dimension and power supply.

The subject devices (FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116) and the predicate devices (FC-BP100, FC-BP101, FC-BP102, FC-BP110, FC-BP111, FC-BP112) share the same schematic, circuitry, critical components, and measurement algorithm.

For the differences between the subject device models, the differences are PCB layout difference (physical board shape and/or component arrangement differences), Display screen, function of the Button, dimension, and power supply.

These changes are mainly cosmetic and do not affect the electrical design, safety, or essential performance of the devices.

6.0 Technological Characteristics

Principle of Operation

There is no change in the principle of operation as part of this submission from the previous clearance under K220113. The module still utilizes the same principles of operation for blood pressure governed by the following principles:

Oscillographic method is used in the subject device to measure blood pressure, the

Page 8

concrete working principle is: the cuff is pressurized to block the brachial blood flow and then slowly decompress, with a small pulse of sound and pressure in the arm. In wave method is to rely on the instrument recognition from the arm to the cuff of small pulse, and the difference, through multiple processing, form a scan reflect the envelope of pulse peak, which blood pressure values are obtained.

Mechanism of Action for Achieving the Intended Effect

There is no change to the Mechanism of Action of the Upper Arm Electronic Blood Pressure Monitor from the previous clearance K220113- Upper Arm Electronic Blood Pressure Monitor (Model: FC-BP100, FC-BP101,FC-BP102,FC-BP110,FC-BP111, FC-BP112).

7.0 Summary of Technological Characteristics of the Subject Device Compared to the Predicate Device

The subject device, Upper Arm Electronic Blood Pressure Monitor (Model: FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116) and the predicate device, Upper Arm Electronic Blood Pressure Monitor (Model: FC-BP100,FC-BP101, FC-BP102,FC-BP110,FC-BP111,FC-BP112),have the following key similarities:

• Both devices have the same intended use
• Both devices are indicated for the same patient population
• Both devices have the same principle of operation and mechanism of action

The subject device and the predicate device have the following differences:
-- Memory capacity increased from 2×90 to 2×120 sets, validated through software verification and validation.
-- Bluetooth functionality was removed from all subject device models.
-- Power supply change: FC-BP123, FC-BP125, FC-BP126, FC-BP127 use 4×AAA batteries (predicate used 4×AA); this change also resulted in PCB layout differences, which triggered EMC testing (IEC 60601-1-2) and electrical safety testing (IEC 60601-1 and IEC 60601-1-11).
-- LED added in display screen: FC-BP107 and FC-BP127 incorporate an LED display instead of LCD display used in predicate.
-- Dimensions and weight difference

These changes have been evaluated and do not affect safety or essential performance as compared to the predicate.

8.0 Non-clinical Testing

In this current submission, just add 6 model FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116 to the legally marketed predicate device K220113. As there were no hardware, software or performance changes made to the subject

Page 9

device when compared to the predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence.

Performance Bench Testing

As there are no hardware or software changes as part of this submission from the previous clearance, no performance bench testing was included as part of this submission.

Biocompatibility Testing

The proposed Upper Arm Electronic Blood Pressure Monitor does not introduce any new direct or indirect patient contacting materials including new color additive, as compared to the previous clearance predicate Upper Arm Electronic Blood Pressure Monitor. So the biocompatibility testing reports provided and reviewed in K220113 remain valid and can support the subject Upper Arm Electronic Blood Pressure Monitor.

Electrical Safety, Environmental, Mechanical and Cleaning

The subject device has undergone changes limited to its display (LED non-backlight vs.LCD backlight), housing dimensions and PCB layout and power supply(AA vs.AAA batteries). In accordance with IEC 60601-1:2020 and IEC 60601-1-11:2020, representative subject device models were selected for testing to adequately cover the range of design differences:

-- Model FC-BP107 was selected to undergo the full suite of electrical safety testing, serving as a primary representative model.
-- Model FC-BP125 was tested for specific sub-clauses, including 4.11 Power input, 9.4.2 Stability and transportability, 11.1 Temperature test, 11.6.5 Harmful ingress of liquids, and 15.3 Enclosure mechanical strength. This model was selected due to its smallest dimensions, unique power supply configuration, LED display, button layout, and housing shape.
-- Models FC-BP123, FC-BP126, and FC-BP116 were tested for sub-clauses 9.4.2 Stability and transportability, 11.6.5 Harmful ingress of liquids, and 15.3 Enclosure mechanical strength. These units were selected to represent additional differences in housing shape among the subject devices.

Testing results confirmed that all subject device models met the applicable requirements of IEC 60601-1:2020 and IEC 60601-1-11:2020 for electrical safety.

Electromagnetic Compatibility

PCB layout (physical board shape or component arrangement) are slightly different among the subject models, full electromagnetic compatibility (EMC) testing per IEC 60601-1-2:2020 was conducted on all six new models.

Results confirmed that all subject device models comply with IEC 60601-1-2:2020 requirements for electromagnetic emissions and immunity.

Page 10

Software Verification and Validation Testing

Updated memory size to store 2x120 sets of blood pressure data. No changes made to the measurement algorithm, pressure sensor, cuff design, or calibration method. Verification and validation activities confirmed that the change is limited to storage capacity without introducing new risks or performance deviations.

7.0 Clinical Test Conclusion

As the subject device utilizes the same monitoring technology as the predicate device, there is no change on the sensors, cuffs, algorithms, measurement accuracy of the candidate device, additional testing was not considered necessary to support the substantial equivalence.

8.0 Technological Characteristic Comparison Table

Page 11

Table1-General Comparison

ItemSubject Device K252769Predicate Device K220113Remark
ManufacturerShenzhen Finicare Co., Ltd.Shenzhen Finicare Co., Ltd./
Product NameUpper Arm Electronic Blood Pressure Monitor: FC-BP107, FC-BP123, FC-BP125, FC-BP126, FC-BP127, FC-BP116Upper Arm Electronic Blood Pressure Monitor: FC-BP100, FC-BP101, FC-BP102, FC-BP110, FC-BP111, FC-BP112/
Product CodeDXNDXNSame
Regulation No.21 CFR 870.113021 CFR 870.1130Same
ClassIIIISame
Intended Use/Indication for UseUpper Arm Electronic Blood Pressure Monitor, Models FC-BP107, FC-BP123, FC-BP125, FC-BP126,FC-BP127, FC-BP116 are intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.Upper Arm Electronic Blood Pressure Monitor, Models FC-BP100,FC-BP101,FC-BP102,FC-BP110,FC-BP111,FC-BP112 are intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected. The devices' feature includes Bluetooth function to transmit data to an external Bluetooth device with wireless communication.Same* (Bluetooth functionality was removed from all subject device models)
Application SiteUpper ArmUpper ArmSame
Cuff Circumference22-42cm22-42cmSame
Patients Contacting MaterialsPatient contact materials of the cuff: 210D Nylon TPU According to ISO-10993Patient contact materials of the cuff: 210D Nylon TPU According to ISO-10993Same
Patient PopulationAdultAdultSame
Measurements ItemSYS,DYS,PulseSYS,DYS,PulseSame
DisplayFC-BP107,FC-BP127:LED Digital Display Other Models: LCD Digital DisplayLCD Digital DisplayDifferent
Design MethodOscillometric MethodOscillometric MethodSame

Page 12

Table 2 Performance Comparison

ItemSubject Device K252769Predicate Device K220113Remark
Max Cuff pressure300mmHg300mmHgSame
BP Range0-299 mmHg0-299 mmHgSame
BP Accuracy±3 mmHg±3 mmHgSame
PR Range40-180 beats/min40-180 beats/minSame
Pulse Accuracy±5% of reading value±5% of reading valueSame
Irregular heartbeat detectionMore than ±25% to the mean interval of pulse intervals.More than ±25% to the mean interval of pulse intervals.Same
Inflation MethodAutomatic inflation by pumpAutomatic inflation by pumpSame
Deflation MethodAutomatic rapid deflationAutomatic rapid deflationSame
Memory Size2x120 set of data2x90 set of dataDifferent
Operation Condition1040℃ 1585%RH1040℃ 1585%RHSame
Storage Condition-2055 ℃ 095% RH-2055 ℃ 095% RHSame
Performance StandardComply with IEC 80601-2-30Comply with IEC 80601-2-30Same
Power SupplyFC-BP107 and FC-BP116: 4 AA alkaline batteries (not included) , DC6V or USB Type C(DC5V1A); FC-BP123, FC-BP125, FC-BP126 and FC-BP127: 4×AAA batteries (not included), 6.0V DC or USB Type C(DC5V1A)4 AA batteries or DC6V,or USB Type C(DC5V1A)Different

Table 3 Safety Comparison

ItemProposed Device K252769Predicate Device K220113Remark
Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1Same
Home UseComply with IEC 60601-1-11Comply with IEC 60601-1-11Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same
FCC conformityN/AFCC 47 part 15 subject BDifferent
ERM conformityN/AEN3014891:2017; EN301489-17:2017Different
RF conformityN/AEN300328:2016Different
HealthN/AEN62479:2010Different
BiocompatibilityComply with ISO 10993-1, FDA GuidanceComply with ISO 10993-1, FDA GuidanceSame

Page 13

9.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).