K Number
K252670

Validate with FDA (Live)

Device Name
Alzevita
Date Cleared
2025-12-19

(116 days)

Product Code
Regulation Number
892.2050
Age Range
55 - 90
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Alzevita is intended for use by neurologists and radiologists experienced in the interpretation and analysis of brain MRI scans. It enables automated labelling, visualization, and volumetric measurement of the hippocampus from high-resolution T1-weighted MRI images. The software facilitates comparison of hippocampal volume against a normative dataset derived from MRI scans of healthy control subjects aged 55 to 90 years, acquired using standardized imaging protocols on 1.5T/3T MRI scanners.

Device Description

Alzevita is a cloud-based, AI-powered medical image processing software as a medical device intended to assist neurologists and radiologists with expertise in the analysis of 3D brain MRI scans. The software performs fully automated segmentation and volumetric quantification of the hippocampus, a brain structure involved in memory and commonly affected by neurodegenerative conditions.

Alzevita is designed to replace manual hippocampal segmentation workflows with a fast, reproducible, and standardized process. It provides quantitative measurements of hippocampal volume, enabling consistent outputs that can assist healthcare professionals in evaluating structural brain changes.

The software operates through a secure web interface and is compatible with commonly used operating systems and browsers. It accepts 3D MRI scans in DICOM or NIfTI format and displays the MRI image in the MRI viewer allowing trained healthcare professionals to view, zoom, and analyze the MRI scan alongside providing a visual and tabular volumetric analysis report.

The underlying algorithm used in Alzevita is locked, meaning it does not modify its behavior at runtime or adapt to new inputs. This ensures consistent performance and reproducibility of results across users and imaging conditions. Any future modifications to the algorithm including performance updates or model re-training will be submitted to the FDA for review and clearance prior to deployment, in compliance with FDA regulatory requirements and applicable guidance for AI/ML-based SaMD.

AI/ML Overview

Here's a detailed description of the acceptance criteria and the study proving the Alzevita device meets those criteria, based on the provided FDA 510(k) clearance letter:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance CriteriaReported Device Performance (Alzevita 95% Confidence Intervals)Criteria (Pass/Fail)
Overall Dice Score≥ 75%(0.85, 0.86)Pass
Overall Hausdorff Distance≤ 6.1 mm(1.43, 1.59)Pass
Overall Correlation Coefficient≥ 0.82Not explicitly given as CI, but stated as metPass
Overall Relative Volume Difference≤ 24.6%Not explicitly given as CI, but stated as metPass
Overall Bland-Altman Mean Difference (Total Hippocampus Volume)≤ 1010 mm³Not explicitly given as CI, but stated as metPass
Subgroup Dice Score (Clinical Subgroups)≥ 83% (implied from results)Control: (0.87, 0.88)MCI: (0.84, 0.85)AD: (0.82, 0.84)Pass
Subgroup Hausdorff Distance (Clinical Subgroups)≤ 3 mm (implied from results)Control: (1.32, 1.41)MCI: (1.44, 1.62)AD: (1.48, 2.10)Pass
Subgroup Dice Score (Gender)≥ 83% (implied)Female: (0.85, 0.87)Male: (0.84, 0.86)Pass
Subgroup Hausdorff Distance (Gender)≤ 3 mm (implied)Female: (1.40, 1.57)Male: (1.41, 1.66)Pass
Subgroup Dice Score (Magnetic Field Strength)≥ 83% (implied)3T: (0.86, 0.87)1.5T: (0.83, 0.85)Pass
Subgroup Hausdorff Distance (Magnetic Field Strength)≤ 3 mm (implied)3T: (1.38, 1.47)1.5T: (1.45, 1.79)Pass
Subgroup Dice Score (Slice Thickness)≥ 83% (implied)1 mm: (0.87, 0.88)1.2 mm: (0.84, 0.85)Pass
Subgroup Hausdorff Distance (Slice Thickness)≤ 3 mm (implied)1 mm: (1.35, 1.43)1.2 mm: (1.47, 1.72)Pass
Subgroup Dice Score (US Geographical Region)≥ 83% (implied)East US: (0.84, 0.86)West US: (0.85, 0.87)Central US: (0.85, 0.87)Canada: (0.82, 0.88)Pass
Subgroup Hausdorff Distance (US Geographical Region)≤ 3 mm (implied)East US: (1.44, 1.71)West US: (1.35, 1.55)Central US: (1.35, 1.47)Canada: (1.07, 2.34)Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 298 subjects.
  • Data Provenance: The test set data was collected from the publicly available ADNI (Alzheimer's Disease Neuroimaging Initiative) dataset. It is retrospective and sampled using stratified random sampling, with subjects recruited from ADNI 1 & ADNI 3 datasets.
  • Geographical Distribution: Approximately equal geographical distribution within the USA (East coast, Central US regions, West coast) and Canada.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three certified radiologists.
  • Qualifications of Experts: They are described as "certified radiologists in India, adhering to widely recognized and standardized segmentation protocols." Specific experience level (e.g., years of experience) is not provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: A consensus ground truth was established by integrating individual delineations from the three certified radiologists into a single consensus mask for each case. This integration was performed using the STAPLE (Simultaneous Truth and Performance Level Estimation) algorithm.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was a MRMC study done? No, the document describes a standalone performance evaluation of the Alzevita algorithm against a consensus ground truth. There is no mention of a human-in-the-loop study comparing human readers with and without AI assistance.
  • Effect size of human readers improvement: Not applicable, as no MRMC study was conducted.

6. Standalone Performance Study

  • Was a standalone performance study done? Yes. The entire validation study described evaluates the Alzevita algorithm's performance in segmenting the hippocampus and calculating its volume against a ground truth, without human intervention in the segmentation process.

7. Type of Ground Truth Used

  • Type of Ground Truth: Expert consensus. Specifically, it was established through manual segmentation by three certified radiologists, with their individual segmentations integrated via the STAPLE algorithm. This STAPLE-derived consensus mask served as the ground truth.

8. Sample Size for the Training Set

  • Sample Size for Training Set: 200 cases.

9. How the Ground Truth for the Training Set Was Established

  • Training Set Ground Truth Establishment: "Expert radiologists manually segmented the hippocampus to create the ground truth, which is then used as input for training the Alzevita segmentation model." The number and specific qualifications of the expert radiologists for the training set's ground truth are not detailed beyond "expert radiologists." There is no mention of an adjudication method like STAPLE for the training set ground truth, suggesting individual expert segmentation or an unspecified consensus process.

FDA 510(k) Clearance Letter - Alzevita

Page 1

December 19, 2025

Topia MedTech Limited
Akshay Sojitra, Director
14 Havelock Place (MS)
Harrow, London HA11LJ
United Kingdom

Re: K252670
Trade/Device Name: Alzevita
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: November 3, 2025
Received: November 3, 2025

Dear Akshay Sojitra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252670 - Akshay Sojitra Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252670 - Akshay Sojitra Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, PhD
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252670
Device Name: Alzevita

Indications for Use (Describe)

Alzevita is intended for use by neurologists and radiologists experienced in the interpretation and analysis of brain MRI scans. It enables automated labelling, visualization, and volumetric measurement of the hippocampus from high-resolution T1-weighted MRI images. The software facilitates comparison of hippocampal volume against a normative dataset derived from MRI scans of healthy control subjects aged 55 to 90 years, acquired using standardized imaging protocols on 1.5T/3T MRI scanners.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Alzevita 510(k) Premarket Submission

Revision: 03
Date: DEC 19,2025

510(k) Summary

I. Submitter

FieldValue
NameTOPIA MEDTECH LIMITED
Address14 Havelock Place (MS), Harrow, United Kingdom, HA11LJ
Contact PersonAkshay Sojitra
Telephone Number+44 2081530878
Emailregulatory@topiamedtech.com

II. Device

FieldValue
Device Trade NameAlzevita
Common NameMedical Image Processing Software
Classification NameMedical Image Management and Processing System
Regulation Number21 CFR 892.2050
Regulation DescriptionPicture Archiving and Communications System
Product CodeQIH
Classification PanelRadiology

III. Predicate Device

FieldValue
DeviceNEUROShield
510(k) NumberK220034
ManufacturerIn Med Prognostics L3C
Product CodeLLZ

Page 6

IV. Device Description

Alzevita is a cloud-based, AI-powered medical image processing software as a medical device intended to assist neurologists and radiologists with expertise in the analysis of 3D brain MRI scans. The software performs fully automated segmentation and volumetric quantification of the hippocampus, a brain structure involved in memory and commonly affected by neurodegenerative conditions.

Alzevita is designed to replace manual hippocampal segmentation workflows with a fast, reproducible, and standardized process. It provides quantitative measurements of hippocampal volume, enabling consistent outputs that can assist healthcare professionals in evaluating structural brain changes.

The software operates through a secure web interface and is compatible with commonly used operating systems and browsers. It accepts 3D MRI scans in DICOM or NIfTI format and displays the MRI image in the MRI viewer allowing trained healthcare professionals to view, zoom, and analyze the MRI scan alongside providing a visual and tabular volumetric analysis report.

The underlying algorithm used in Alzevita is locked, meaning it does not modify its behavior at runtime or adapt to new inputs. This ensures consistent performance and reproducibility of results across users and imaging conditions. Any future modifications to the algorithm including performance updates or model re-training will be submitted to the FDA for review and clearance prior to deployment, in compliance with FDA regulatory requirements and applicable guidance for AI/ML-based SaMD.

V. Indications for Use

Alzevita is intended for use by neurologists and radiologists experienced in the interpretation and analysis of brain MRI scans. It enables automated labelling, visualization, and volumetric measurement of the hippocampus from high-resolution T1-weighted MRI images. The software facilitates comparison of hippocampal volume against a normative dataset derived from MRI scans of healthy control subjects aged 55 to 90 years, acquired using standardized imaging protocols on 1.5T/3T MRI scanners.

VI. Comparison of Technological Characteristics with the Predicate Device

DetailsSubject DevicePredicate DeviceRemark
Name of ManufacturerTOPIA MEDTECH LIMITEDIn Med Prognostics L3C-
Address of Submitter14 Havelock Place (MS), Harrow, United Kingdom, HA11LJ4918 September Street, San Diego, CA 92110, USA-
Brand NameAlzevitaNEUROShield-
510 (K) NumberK252670K220034-
Generic NameMedical Image Processing SoftwareMedical Image Processing softwareEquivalent
ClassificationClass IIClass IIEquivalent
Classification Product CodeQIHLLZSubstantially Equivalent
Intended UseAlzevita is intended for use by neurologists and radiologistsThe NEUROShield medical image processing software is intended forSubstantially Equivalent

Page 7

DetailsSubject DevicePredicate DeviceRemark
Intended Use (continued)experienced in the interpretation and analysis of brain MRI scans. It enables automated labelling, visualization, and volumetric measurement of the hippocampus from high-resolution T1-weighted MRI images. The software facilitates comparison of hippocampal volume against a normative dataset derived from MRI scans of healthy control subjects aged 55 to 90 years, acquired using standardized imaging protocols on 1.5T/3T MRI scanners.automatic labelling, visualization, and volumetric quantification of the Hippocampus brain structure from a set of MR images
Design and Incorporated Technology• A Software as a Medical Device (SaMD) designed for imaging and quantitative analysis of the hippocampal brain structure from MRI scans.• It is a fully automated, geometry-based brain analytics tool and cloud platform developed using advanced 3D U-Net++ methodologies• Software as a medical device to be used in the process of Imaging and quantification of the Hippocampus brain structure from a set of MR images• Fully automated brain geometry-based quantifying analytics tool/cloud platform developed using DeepNet / U-Net methodologiesSubstantially Equivalent
Physical Characteristics• Software Package (Accessible via Web Browser)• Operates on off the shelf hardware (multiple vendors)• Software Package (Accessible via Web Browser)• Operates on off the shelf hardware (multiple vendors)Equivalent
Supported Operating SystemsSupports Linux, Windows and Mac OS latestSupports Windows and Mac OS latestEquivalent
Data SourceAlzevita requires compressed DICOM files or NIFTI files as InputNEUROShield requires uncompressed DICOM files as inputEquivalent
OutputSoftware provides volumetric measurements of Hippocampus brain structuresProvides volumetric measurements of Hippocampus brain structuresEquivalent
Safety• Automated quality control functions - Scan protocol verification• Results must be reviewed by a trained clinicians/ Radiologist/ Neurologist• Automated quality control functions - Scan protocol verification• Results must be reviewed by a trained clinicians/ Radiologist/ NeurologistEquivalent

VII. Performance Testing

Page 8

Alzevita is a deep learning algorithm-based device. This algorithm is developed by training the Deep Learning based 3D U-Net++ model with the help of the training data.

Specifications of the Training dataset

Data is collected from India, between May 2024 to July 2025. The training dataset, consisting of 200 cases, is meticulously curated by considering various factors such as image variance, and quality. All cases are acquired using a 1.5 Tesla MRI scanner. Expert radiologists manually segmented the hippocampus to create the ground truth, which is then used as input for training the Alzevita segmentation model.

Table 1 presents the distribution of subjects based on key imaging and demographic parameters.

SubgroupsCount
Magnetic field strength 1.5T200
Slice thickness 1200
Equipment GE200
Gender Male110
Gender Female90

Validation Study:

The performance of Alzevita is evaluated by a validation study summarized as follows:

A. Data Description

The performance validation dataset is collected from the publicly available ADNI (Alzheimer's Disease Neuroimaging Initiative) dataset. This dataset is independent of the training data and is not used to develop Alzevita's algorithm.

  • Data Size: 298 subjects
  • Study Type: Analytical & Cross-Sectional
  • Data collection type: Retrospective
  • Data Sampling: Stratified Random Sampling
  • Recruitment factors: ADNI 1 & ADNI 3 dataset (Alzheimer's Disease Neuroimaging Initiative)
  • MRI Equipment manufacturers: GE medical systems, Philips medical systems, Siemens Healthineers
  • Magnetic Field Strength: 1.5T and 3T
  • MRI Sequences/ protocol: 3D T1 MPRAGE
  • Slice thickness: 1, 1.2
  • Approximately equal geographical distribution in USA: East coast, Central US regions, West coast and Canada

The distribution for age bands is as follows:

Page 9

The mean age of subjects is found to be 76 ± 7 years.

Age groupCount
55-6420
65-6937
70-7474
75-7982
80-8458
85-9027

The table below provides the categorization of subjects according to selection criteria:

SubgroupsCountMean (mL)Standard deviation (mL)
Clinical Sub-groups
ADNI-control1327.01.0
ADNI-MCI1035.61.3
ADNI-AD634.90.9
Gender
Male1505.81.3
Female1486.41.4
Magnetic field strength
1.5T1285.51.3
3T1706.051.1
Slice thickness
11266.91.3
1.21725.61.3
Region
East USA1336.11.4
West USA566.21.5
Central USA906.21.4
Canada196.31.5

Page 10

B. Ground Truth

i. A consensus ground truth is established through manual segmentation of MRI brain scans from 298 subjects. This task is performed by three certified radiologists in India, adhering to widely recognized and standardized segmentation protocols. The individual delineations are subsequently integrated into a single consensus mask for each case utilizing the STAPLE (Simultaneous Truth and Performance Level Estimation) algorithm. The reliability and accuracy of the consensus ground truth are statistically validated by comparing it with the individual segmentations done by the radiologists.

ii. The Alzevita algorithm performed automated segmentation of the hippocampal brain structure across all subjects, generating labelled segmentations with precise anatomical markings and calculating hippocampal volumes using advanced algorithmic techniques.

C. Statistical Analysis

Ground Truth Validation: Using the Dice coefficient, and Hausdorff distance, the STAPLE-derived ground truth (via ITK-SNAP) is validated through comparison with segmentations from three radiologists. Statistical analysis indicated no significant discrepancies (p > 0.05) between these individual segmentations and the consensus ground truth, thereby substantiating the consistency of the radiologist annotations and the reliability of the ground truth itself.

i. Geometric comparison of Alzevita with Ground Truth: An evaluation of Alzevita's automated hippocampus segmentations is conducted by comparing them to the STAPLE-derived ground truth, utilizing Dice scores and Hausdorff distance as quantitative measures. The algorithm's performance met the established criteria for both metrics, signifying a high level of geometric correspondence with the annotations provided by expert radiologists.

ii. Quantitative Volume Comparison: A comparison is made between hippocampal volumes derived from the STAPLE ground truth and those generated by Alzevita. Utilizing correlation analysis, Bland-Altman (BA) plots, and relative volume difference, Alzevita successfully met the criteria across all three statistical evaluation methods, demonstrating strong agreement between the hippocampal volumes computed by Alzevita and those derived from the STAPLE consensus ground truth.

iii. Pass/Fail criteria: The Alzevita algorithm met all predefined performance thresholds, demonstrating its accuracy and reliability in hippocampal segmentation and volume estimation:

  • Dice Score: ≥ 75%
  • Hausdorff Distance: ≤ 6.1 mm
  • Correlation Coefficient: ≥ 0.82
  • Relative Volume Difference: ≤ 24.6%
  • Bland-Altman Mean Difference (Total Hippocampus Volume): ≤ 1010 mm³

These results validate the robustness of the algorithm's automated segmentation and volume computation capabilities.

Page 11

iv. Subgroup Error Analysis: Subgroup analyses are conducted across variations such as magnetic field strength, gender, slice thickness, clinical subgroups, and geographic regions within the U.S. In all cases, Dice scores exceeded 83%, and Hausdorff distances remained below 3 mm, indicating consistently high segmentation accuracy. Additionally, Alzevita met the criteria for both correlation and relative volume difference across all subgroups, demonstrating robustness and generalizability of the model.

D. Results

i. The average dice coefficient and Hausdorff distance is found to be 0.86 and 1.51 mm respectively. The following table shows the 95% confidence interval for both.

MeasureThresholdAlzevita 95% confidence intervalsCriteria (Pass/Fail)
Dice0.75(0.85, 0.86)Pass
Hausdorff distance6.1(1.43, 1.59)Pass

ii. The outcomes of subgroup error analysis are as follows:

a) Clinical subgroups

Dice ScoreHausdorff Distance (mm)
Clinical subgroupsControlMCIADControlMCIAD
Measured value0.880.850.831.361.531.79
Alzevita 95% confidence intervals(0.87, 0.88)(0.84, 0.85)(0.82, 0.84)(1.32, 1.41)(1.44, 1.62)(1.48, 2.10)
CriteriaPassPassPassPassPassPass

Page 12

b) Gender

Dice scoreHausdorff Distance (mm)
GenderFemaleMaleFemaleMale
Measured value0.860.851.481.54
Alzevita 95% confidence intervals(0.85, 0.87)(0.84, 0.86)(1.40, 1.57)(1.41, 1.66)
CriteriaPassPassPassPass

c) Magnetic field strength

Dice scoreHausdorff Distance (mm)
MRI strength3T1.5T3T1.5T
Measured value0.870.841.431.62
Alzevita 95% confidence intervals(0.86, 0.87)(0.83, 0.85)(1.38, 1.47)(1.45, 1.79)
CriteriaPassPassPassPass

d) Slice Thickness

Dice scoreHausdorff Distance (mm)
Slice thickness1 mm1.2mm1 mm1.2mm
Average value0.870.841.391.60
Alzevita 95% confidence intervals(0.87, 0.88)(0.84, 0.85)(1.35, 1.43)(1.47, 1.72)
CriteriaPassPassPassPass

Page 13

e) US Geographical Region

Dice scoreHausdorff Distance (mm)
RegionEast USWest USCentral USCanadaEast USWest USCentral USCanada
Average value0.850.860.860.851.571.451.411.71
Alzevita 95% confidence intervals(0.84,0.86)(0.85, 0.87)(0.85, 0.87)(0.82, 0.88)(1.44,1.71)(1.35,1.55)(1.35, 1.47)(1.07, 2.34)
CriteriaPassPassPassPassPassPassPassPass

Performance analysis of Alzevita for hippocampus segmentation revealed consistently strong results, with overall Dice scores surpassing 83% and Hausdorff distances remaining under 3 mm. Such outcomes affirm the algorithm's high accuracy and reliability, substantiating its precision in both structural correspondence and relative hippocampal volume assessment.

VIII. Conclusions

The evaluation provides objective evidence supporting the reliability, accuracy, and reproducibility of Alzevita for automated hippocampal segmentation. Validation and performance testing demonstrate that the device meets predefined acceptance criteria for its intended use in clinical neuroimaging workflows. The results support the device's suitability for use by qualified healthcare professionals in the assessment of neurological structures. Furthermore, Alzevita has been demonstrated to be substantially equivalent to its predicate device, NEUROShield (K220034), with respect to intended use, technological characteristics, and performance.

N/A