(30 days)
The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging.
The da Vinci SP Firefly Imaging System consists of a near-infrared laser light source located in the Endoscope Controller and a connected endoscope. The Firefly imaging system uses near-infrared light in conjunction with the imaging agent Indocyanine Green (ICG), to create fluorescent images of tissue. ICG is administered to the patient according to its manufacturer's labeling, and the system is switched to Firefly imaging. Two modes are available in Firefly imaging: Standard mode and Sensitive mode. When Firefly is active in Standard mode, the system displays the resulting images as a fluorescent (green) overlay on a gray-scale background image. In Sensitive mode, the gray-scale background is no longer illuminated, resulting in increased sensitivity to fluorescent signal.
This FDA 510(k) clearance letter pertains to the da Vinci SP Firefly Imaging System, focusing on a new "Sensitive Firefly" imaging mode.
Based on the provided document, here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state specific quantitative acceptance criteria for the device's performance, nor does it provide a table of reported device performance metrics against such criteria. It generally states that "Verification and validation testing on the subject device confirmed that no issues of safety or effectiveness, analogous to the results of the predicate device verification and validation testing." This implies that the new mode performs comparably to the predicate device, but specific metrics are not detailed.
2. Sample Size Used for the Test Set and Data Provenance
The document states "clinical validation testing with an animal model" was performed.
- Sample Size: Not specified.
- Data Provenance: Animal model; implies prospective (testing performed specifically for this submission). The country of origin is not specified.
3. Number of Experts Used to Establish Ground Truth and Qualifications
Not specified in the document.
4. Adjudication Method for the Test Set
Not specified in the document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it quantify any improvement for human readers with AI assistance. The device description is for an imaging system, not an AI-assisted diagnostic tool in the typical sense that would involve human reader performance comparisons.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
The device is an imaging system designed for surgeons to visualize tissues in real-time. It's not an "algorithm only" device in a standalone diagnostic capacity. The testing described (bench and animal model) supports its functionality as an imaging tool to be used by a surgeon.
7. Type of Ground Truth Used
The document mentions "clinical validation testing with an animal model" to confirm "no issues of safety or effectiveness." This suggests the ground truth was likely based on observed physiological responses in the animal model, evaluating whether the system accurately visualized vessels, blood flow, and tissue perfusion using ICG, as intended. However, the specific method for establishing this ground truth (e.g., direct observation, histopathology) is not detailed.
8. Sample Size for the Training Set
The document does not explicitly mention a "training set" or its sample size. Given the description, the device appears to be an imaging hardware system with a new mode, rather than a machine learning model that would typically have a distinct training phase with a labeled dataset. The verification and validation testing focus on confirming the functionality of the imaging system itself.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned for an algorithm, the method for establishing its ground truth is not applicable or described in this document.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 28, 2025
Intuitive Surgical, Inc.
Connor Mccarty
Technical Lead, Regulatory Affairs
1266 Kifer Road
Sunnyvale, California 94086
Re: K252350
Trade/Device Name: da Vinci SP Firefly Imaging System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: NAY
Dated: July 28, 2025
Received: July 29, 2025
Dear Connor Mccarty:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Page 2
K252350 - Connor Mccarty
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K252350 - Connor Mccarty
Page 3
Sincerely,
Mark Trumbore -S
Digitally signed by Mark Trumbore -S
Date: 2025.08.28 08:32:17 -04'00'
Mark Trumbore Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K252350
Please provide the device trade name(s).
da Vinci SP Firefly Imaging System
Please provide your Indications for Use below.
The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
da Vinci SP Firefly Imaging System Page 8 of 26
Page 5
510(k) Summary
Prepared on: 2025-07-28
Contact Details
21 CFR 807.92(a)(1)
| Field | Value |
|---|---|
| Applicant Name | Intuitive Surgical, Inc. |
| Applicant Address | 1266 Kifer Road Sunnyvale CA 94086 United States |
| Applicant Contact Telephone | 805-798-4205 |
| Applicant Contact | Mr. Connor McCarty |
| Applicant Contact Email | connor.mccarty@intusurg.com |
Device Name
21 CFR 807.92(a)(2)
| Field | Value |
|---|---|
| Device Trade Name | da Vinci SP Firefly Imaging System |
| Common Name | Endoscope and accessories |
| Classification Name | System, Surgical, Computer Controlled Instrument |
| Regulation Number | 876.1500 |
| Product Code(s) | NAY, IZI |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|---|---|---|
| K250786 | da Vinci SP Firefly Imaging System | NAY |
Device Description Summary
21 CFR 807.92(a)(4)
The da Vinci SP Firefly Imaging System consists of a near-infrared laser light source located in the Endoscope Controller and a connected endoscope. The Firefly imaging system uses near-infrared light in conjunction with the imaging agent Indocyanine Green (ICG), to create fluorescent images of tissue. ICG is administered to the patient according to its manufacturer's labeling, and the system is switched to Firefly imaging. Two modes are available in Firefly imaging: Standard mode and Sensitive mode. When Firefly is active in Standard mode, the system displays the resulting images as a fluorescent (green) overlay on a gray-scale background image. In Sensitive mode, the gray-scale background is no longer illuminated, resulting in increased sensitivity to fluorescent signal.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use are identical between the predicate and subject devices.
Technological Comparison
21 CFR 807.92(a)(6)
The changes for the subject da Vinci SP Firefly Imaging System are limited to a new intraoperative fluorescence imaging mode called "Sensitive Firefly". When the surgeon uses the new, additional imaging mode "Sensitive Firefly", the endoscope emits only near-infrared light onto the surgical site. This allows the image sensors at the endoscope tip to detect ICG imaging agent in surgical site with greater
Page 6
sensitivity. However, without the additional blue light, the surgeon can only see the fluorescing ICG and not the tissue which is not infused with ICG. Because the strength of ICG perfusion into the surgical site can be dependent on the anatomy, the additional "Sensitive Firefly" imaging mode is designed to provide the surgeon with an option to visualize ICG in situations where concentrations are relatively low and not easily detected in Standard Firefly mode.
Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Both the subject and predicate devices underwent software verification testing, bench verification testing of the fluorescence-capable endoscope, and clinical validation testing with an animal model. Verification and validation testing on the subject device confirmed that no issues of safety or effectiveness, analogous to the results of the predicate device verification and validation testing. Therefore, the test results demonstrate that the subject device is substantially equivalent to its predicate device.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.