(28 days)
This product shall be used in conjunction with the Endoscopic Video Image Processor and other peripheral equipment for observation, diagnosis, photography and treatment within the bladder, urethra, and kidney.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
This product shall not be used for other purposes.
The Single-Use Video Flexible Cysto-Nephroscope mainly consist of insertion portion, handle and connector section which include a detachable video cable which connect the endoscope to the Endoscopic Video Image Processor.
The provided FDA 510(k) clearance letter and summary describe a physical medical device (Single-Use Video Flexible Cysto-Nephroscope), not an AI/ML-driven software device. Therefore, the information requested about acceptance criteria and studies (especially those related to AI performance, ground truth, expert adjudication, MRMC studies, or standalone algorithm performance) is not applicable to this document.
The document discusses the substantial equivalence of the new device to a predicate device based on non-clinical performance testing of the physical properties and functionality of the endoscope itself. There is no mention of any AI component or software that processes images or provides diagnostic assistance.
Here's a breakdown of what is available in the document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document broadly states that "The proposed device has met the testing acceptance criteria in accordance with internal requirements and applicable standards to support substantial equivalence of the proposed device." However, it does not provide a specific table listing detailed acceptance criteria and the corresponding reported performance values for each physical test. It only lists the types of tests performed.
- Sterilization Validation:
- Acceptance Criteria: Determined by ISO 11135:2014 half-cycle method to establish routine control and monitoring parameters.
- Reported Performance: Method validated. (Specific parameters not provided).
- Shelf Life and Simulated Transportation Distribution followed by Sterile Packaging Integrity Test:
- Acceptance Criteria: Validated according to ASTM F1980-21, ISO11607-1:2019, ISO11607-2:2019, ASTM F1929-23, ASTM F88/F88M-23, ASTM F1886/F1886M-16, ASTM D4169-23.
- Reported Performance: Validated. (Specific results not provided).
- Performance Testing (using methods and acceptance criteria from K241500):
- Surface and Edges:
- Acceptance Criteria: Not specified but refers to previous submission K241500.
- Reported Performance: Met testing acceptance criteria.
- Deflection system and Fatigue test of Rocker and Bending Section: (Note: Downward deflection angle changed from 135° to 225° for all models).
- Acceptance Criteria: Not specified but refers to previous submission K241500.
- Reported Performance: Met testing acceptance criteria.
- Surface and Edges:
Information Not Applicable to this Device/Document (for AI/ML products):
The following points are typically relevant for AI/ML device submissions but are not found in this document because it pertains to a physical, non-AI medical instrument.
- Sample sized used for the test set and the data provenance: Not applicable. Evaluation was based on physical device testing, not a dataset.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth for an AI algorithm.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
Summary of Study Type:
The study described is a non-clinical performance testing study for a physical medical device (endoscope), designed to demonstrate substantial equivalence to a previously cleared predicate device. It involves component and system-level testing of physical characteristics, sterilization, packaging integrity, and functionality based on established industry standards and previous submission's methodologies. There is no AI or software performance evaluation in this clearance.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 8, 2025
Guangzhou Red Pine Medical Instrument Co., Ltd.
Ping Yang
Regulatory Affairs Director
12 F, No.87 Luoxuan Avenue
Guangzhou International Bioisland, Huangpu District
Guangzhou, Guangdong 510000
CHINA
Re: K252176
Trade/Device Name: Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FAJ
Dated: July 11, 2025
Received: July 11, 2025
Dear Ping Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K252176 - Ping Yang Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
Page 3
K252176 - Ping Yang Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK J. ANTONINO -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K252176
Device Name: Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS)
Indications for Use (Describe)
This product shall be used in conjunction with the Endoscopic Video Image Processor and other peripheral equipment for observation, diagnosis, photography and treatment within the bladder, urethra, and kidney.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
This product shall not be used for other purposes.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
Page 1 / 3
I. Contact Details
Submitter Name: Guangzhou Red Pine Medical Instrument Co., Ltd.
Submitter Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China
Submitter Contact Telephone: +86 13902971205
Submitter Contact: Mr. Weihua Yang
Submitter Contact Email: regulation@gzredpine.com
Correspondent Name: Guangzhou Red Pine Medical Instrument Co., Ltd.
Correspondent Address: 12 F, No.87 Luoxuan Avenue, Guangzhou International Bioisland, Huangpu District Guangzhou Guangdong 510000 China
Correspondent Contact Telephone: +86 18565651640
Correspondent Contact: Ms. Ping Yang
Correspondent Contact Email: regulation@gzredpine.com
Date prepared: August 7, 2025
II. Subject Device
Device Trade Name: Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS)
Common Name: Endoscope and accessories
Classification Number: 21 CFR 876.1500
Classification Name: Cystoscope and accessories, flexible/rigid
Product Code: FAJ
III. Legally Marketed Predicate Device
Predicate 510(k) Number: K241500
Device name: Endoscopic Video Image Processor (RP-IPD-V2000EF); Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS)
Product Code: FAJ
IV. Device Description Summary
The Single-Use Video Flexible Cysto-Nephroscope mainly consist of insertion portion, handle and connector section which include a detachable video cable which connect the endoscope to the Endoscopic Video Image Processor.
Page 6
Page 2 / 3
V. Indications for use
This product shall be used in conjunction with the Endoscopic Video Image Processor and other peripheral equipment for observation, diagnosis, photography and treatment within the bladder, urethra, and kidney.
The device is suitable for professional healthcare facility environments such as hospitals and clinics.
This product shall not be used for other purposes.
VI. Indications for Use Comparison
The proposed device, Single-Use Video Flexible Cysto-Nephroscope, has the same indications for use in comparison to the predicate device, Single-Use Video Flexible Cysto-Nephroscope.
VII. Technological Comparison
The proposed Single-Use Video Flexible Cysto-Nephroscope is the same as the predicate Single-Use Video Flexible Cysto-Nephroscope (cleared under K241500) with respect to its indications for use, design, operating principle, and fundamental technology.
The proposed device contains the following differences to the predicate device:
-
All features of the new device (proposed device) are identical to the predicate, and only a detachable video cable is added to its immediate package.
-
The downward deflection angle for all models is changed from 135° to 225°.
-
The material of immediate packaging (Tyvek pouch) is changed from DuPont Tyvek 2FS paper to DuPont Tyvek 1073B paper. Along with this immediate packaging material change, the dimensions of the middle packaging and transportation packaging have been adjusted.
As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the above changes. Verification testing was conducted to evaluate the modifications. The proposed device has met the testing acceptance criteria in accordance with internal requirements and applicable standards to support substantial equivalence of the proposed device.
VIII. Non-Clinical Tests Summary
As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the design changes. Verification testing including performance testing using previous submitted method and acceptance criteria cleared under K241500, additional shelf-life testing and simulated transportation testing, performance testing including deflection system etc. were conducted to evaluate the modifications. The proposed device has met the testing acceptance criteria in accordance with internal requirements and applicable standards to support substantial equivalence of the proposed device.
Sterilization Validation
The sterilization method has been validated to ISO 11135:2014 half-cycle method, which has thereby determined the routine control and monitoring parameters.
Shelf Life and simulated transportation distribution followed by sterile packaging integrity test.
Shelf Life and simulated transportation distribution followed by packaging integrity testing have been validated according to the following applicable standards:
- ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
Page 7
Page 3 / 3
-
ISO11607-1:2019 Packaging for terminally sterilized Medical Device Part 1: Requirement for materials, sterile barrier systems and packaging systems
-
ISO11607-2:2019 Packaging for terminally sterilized Medical Device Part 2: Validation Requirement for forming, sealing and assembly process.
-
ASTM F 1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
-
ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
-
ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
-
ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
Performance Testing
The following performance testing was conducted using previous submitted method and acceptance criteria cleared under K241500 in support of the substantial equivalence, as shown below.
- Surface and Edges
- Deflection system and Fatigue test of Fatigue test of Rocker and Bending Section
IX. Clinical Evidence
N/A.
X. Conclusion
The results of the nonclinical performance testing demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.