K Number
K252102

Validate with FDA (Live)

Device Name
Alinity m HCV
Date Cleared
2025-09-25

(84 days)

Product Code
Regulation Number
866.3170
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for both the detection and quantitation of hepatitis C virus (HCV) RNA, in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals. The assay is intended for use as an aid in the diagnosis of active HCV infection in individuals with antibody evidence of HCV infection, and to aid in the management of patients with known active HCV infection, including Sustained Virologic Response (SVR) determination.

The results from the Alinity m HCV assay must be interpreted within the context of all relevant clinical and laboratory findings.

The Alinity m HCV assay is not intended to be used in screening blood, plasma, serum, tissue or tissue donors for HCV.

Device Description

Alinity m HCV is an in vitro polymerase chain reaction (PCR) assay for both the detection and quantitation of HCV RNA in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals.

This device is similar to the predicate device originally approved (PMA P190025) with the exception that the subject device may use a new DNA Polymerase as an alternative to the original DNA Polymerase and a new Reverse Transcriptase as an alternative to the original Reverse Transcriptase in the reagent formulation of the assay. This formulation difference does not introduce any changes to sample processing, assay procedure, or data reduction.

Additional studies were initiated to support the formulation of the assay with alternative DNA Polymerase and Reverse Transcriptase. Supplemental data from these studies were used with data previously obtained from the analytical and clinical testing studies submitted in P190025.

The steps of the Alinity m HCV assay consist of sample preparation, real-time PCR assembly, amplification/detection, result calculation, and reporting. All steps of the Alinity m HCV assay procedure are executed automatically by the Alinity m System. Manual dilutions may be performed for low-volume specimens to meet the minimum volume requirement and for high-titer specimens above the upper limit of quantitation (ULoQ). The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m HCV assay in parallel with other Alinity m assays on the same instrument.

Alinity m HCV requires three separate assay specific kits as follows:

  • Alinity m HCV AMP Kit (List No. 08N50-095), consisting of 2 types of multi-well assay trays. The amplification trays (AMP Trays) contain lyophilized, unit-dose RT-PCR amplification/detection reagents and lyophilized, unit-dose IC in separate wells, and the activation trays (ACT Trays) contain liquid activation reagent. The intended storage condition for the Alinity m HCV AMP Kit is 2°C to 8°C.

  • Alinity m HCV CTRL Kit (List No. 08N50-085), consisting of negative controls, low positive controls, and high positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m HCV CTRL Kit is –25°C to –15°C.

  • Alinity m HCV CAL Kit (List No. 08N50-075), consisting of 2 calibrator levels, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m HCV CAL Kit is –25°C to –15°C.

HCV RNA from human plasma or serum is extracted automatically on-board in the Alinity m System using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash, and elution. The resulting purified nucleic acids are then combined with liquid unit-dose Alinity m HCV activation reagent and lyophilized unit-dose Alinity m HCV amplification/detection reagents and transferred into a reaction vessel. Alinity m Vapor Barrier Solution is then added to the reaction vessel which is then transferred to an amplification/detection unit for PCR amplification, and real-time fluorescence detection of HCV targets.

At the beginning of the Alinity m HCV sample preparation process, a lyophilized unit-dose IC on the AMP Tray is rehydrated by the Alinity m System and delivered into each sample preparation reaction. The IC is then processed through the entire sample preparation and real-time PCR procedure along with the specimens, calibrators, and controls to demonstrate proper sample processing and validity.

The Alinity m HCV amplification/detection reagents consist of enzymes, primers, probes, and activation reagents that enable reverse transcription, polymerization, and detection.

An HCV calibration curve is required for determination of HCV RNA concentration. Two levels of calibrators are processed through sample preparation and RT-PCR to generate the calibration curve. The concentration of HCV RNA in specimens and controls is then calculated from the stored calibration curve.

Assay controls are tested at or above an established minimum frequency to help ensure that instrument and reagent performance remains satisfactory. During each control event, a negative control, a low-positive control, and a high-positive control are processed through sample preparation and RT-PCR procedures that are identical to those used for specimens.

The Alinity m HCV assay also utilizes the following:

  • Alinity m HCV Application Specification File, (List No. 08N50-03B)
  • Alinity m System and System Software (List No. 08N53)
  • Alinity m Sample Prep Kit 2 (List No. 09N12-001)
  • Alinity m Specimen Dilution Kit I (List No. 09N50-001)
  • Alinity m System Solutions, (List No. 09N20):
    • Alinity m Lysis Solution (List No. 09N20-001)
    • Alinity m Diluent Solution (List No. 09N20-003)
    • Alinity m Vapor Barrier Solution, (List No. 09N20-004)
  • Alinity m Tubes and Caps (List No. 09N49):
    • Alinity m LRV Tube (List No. 09N49-001)
    • Alinity m Transport Tubes Pierceable Capped (List No. 09N49-010)
    • Alinity m Transport Tube (List No. 09N49-011)
    • Alinity m Pierceable Cap (List No. 09N49-012)
    • Alinity m Aliquot Tube (List No. 09N49-013)
AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Alinity m HCV

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 25, 2025

Abbott Molecular Inc.
Jennifer Peterson
Associate Director Regulatory Affairs
1350 East Touhy Avenue
Des Plaines, Illinois 60018

Re: K252102
Trade/Device Name: Alinity m HCV
Regulation Number: 21 CFR 866.3170
Regulation Name: Nucleic Acid-Based Hepatitis C Virus Ribonucleic Acid Tests
Regulatory Class: Class II
Product Code: MZP
Dated: July 2, 2025
Received: July 3, 2025

Dear Jennifer Peterson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252102 - Jennifer Peterson Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252102 - Jennifer Peterson Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bhawna Poonia -S

for

Uwe Scherf, M.Sc., Ph.D.
Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252102

Device Name: Alinity m HCV

Indications for Use (Describe)

The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for both the detection and quantitation of hepatitis C virus (HCV) RNA, in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals. The assay is intended for use as an aid in the diagnosis of active HCV infection in individuals with antibody evidence of HCV infection, and to aid in the management of patients with known active HCV infection, including Sustained Virologic Response (SVR) determination.

The results from the Alinity m HCV assay must be interpreted within the context of all relevant clinical and laboratory findings.

The Alinity m HCV assay is not intended to be used in screening blood, plasma, serum, tissue or tissue donors for HCV.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 1 of 25

510(k) Summary

Table of Contents

1.0 510(k) Summary ...................................................................................................... 2

  • 1.1 Submitter.......................................................................................................... 2
  • 1.2 Device Information.......................................................................................... 2
  • 1.3 Predicate Device .............................................................................................. 2
  • 1.4 Device Description .......................................................................................... 3
  • 1.5 Intended Use .................................................................................................... 6
  • 1.6 Similarities and Differences to Predicate Device ............................................ 6
  • 1.7 Performance Data ............................................................................................ 9
  • 1.8 Conclusions Drawn from the Studies............................................................ 25

Page 6

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 2 of 25

1.0 510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR Section 807.92(c).

1.1 Submitter

Applicant Name and Address: Abbott Molecular Inc.
1300 E. Touhy Avenue
Des Plaines, IL 60018

Contact Person: Jennifer Peterson
Associate Director Regulatory Affairs
Abbott Molecular, Inc.
1300 E. Touhy Avenue
Des Plaines, IL 60018
Phone: 224-361-7081
Mobile: 224-651-7305
Fax: 224-361-7269

Date Prepared: July 2025

1.2 Device Information

Trade NameRegulation NameProduct CodeRegulation No.Class
Alinity m HCVNucleic Acid-Based Hepatitis C Virus Ribonucleic Acid TestMZP21 CFR 866.3170II

1.3 Predicate Device

Device NamePredicate DevicePMAApproved
Alinity m HCVAlinity m HCVP19002503/23/2020

Page 7

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 3 of 25

1.4 Device Description

Alinity m HCV is an in vitro polymerase chain reaction (PCR) assay for both the detection and quantitation of HCV RNA in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals.

This device is similar to the predicate device originally approved (PMA P190025) with the exception that the subject device may use a new DNA Polymerase as an alternative to the original DNA Polymerase and a new Reverse Transcriptase as an alternative to the original Reverse Transcriptase in the reagent formulation of the assay. This formulation difference does not introduce any changes to sample processing, assay procedure, or data reduction.

Additional studies were initiated to support the formulation of the assay with alternative DNA Polymerase and Reverse Transcriptase. Supplemental data from these studies were used with data previously obtained from the analytical and clinical testing studies submitted in P190025.

The steps of the Alinity m HCV assay consist of sample preparation, real-time PCR assembly, amplification/detection, result calculation, and reporting. All steps of the Alinity m HCV assay procedure are executed automatically by the Alinity m System. Manual dilutions may be performed for low-volume specimens to meet the minimum volume requirement and for high-titer specimens above the upper limit of quantitation (ULoQ). The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m HCV assay in parallel with other Alinity m assays on the same instrument.

Page 8

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 4 of 25

Alinity m HCV requires three separate assay specific kits as follows:

  • Alinity m HCV AMP Kit (List No. 08N50-095), consisting of 2 types of multi-well assay trays. The amplification trays (AMP Trays) contain lyophilized, unit-dose RT-PCR amplification/detection reagents and lyophilized, unit-dose IC in separate wells, and the activation trays (ACT Trays) contain liquid activation reagent. The intended storage condition for the Alinity m HCV AMP Kit is 2°C to 8°C.

  • Alinity m HCV CTRL Kit (List No. 08N50-085), consisting of negative controls, low positive controls, and high positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m HCV CTRL Kit is –25°C to –15°C.

  • Alinity m HCV CAL Kit (List No. 08N50-075), consisting of 2 calibrator levels, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m HCV CAL Kit is –25°C to –15°C.

HCV RNA from human plasma or serum is extracted automatically on-board in the Alinity m System using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash, and elution. The resulting purified nucleic acids are then combined with liquid unit-dose Alinity m HCV activation reagent and lyophilized unit-dose Alinity m HCV amplification/detection reagents and transferred into a reaction vessel. Alinity m Vapor Barrier Solution is then added to the reaction vessel which is then transferred to an amplification/detection unit for PCR amplification, and real-time fluorescence detection of HCV targets.

At the beginning of the Alinity m HCV sample preparation process, a lyophilized unit-dose IC on the AMP Tray is rehydrated by the Alinity m System and delivered into each sample preparation reaction. The IC is then processed through the entire sample preparation and real-time PCR procedure along with the specimens, calibrators, and controls to demonstrate proper sample processing and validity.

The Alinity m HCV amplification/detection reagents consist of enzymes, primers, probes, and activation reagents that enable reverse transcription, polymerization, and detection.

Page 9

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 5 of 25

An HCV calibration curve is required for determination of HCV RNA concentration. Two levels of calibrators are processed through sample preparation and RT-PCR to generate the calibration curve. The concentration of HCV RNA in specimens and controls is then calculated from the stored calibration curve.

Assay controls are tested at or above an established minimum frequency to help ensure that instrument and reagent performance remains satisfactory. During each control event, a negative control, a low-positive control, and a high-positive control are processed through sample preparation and RT-PCR procedures that are identical to those used for specimens.

The Alinity m HCV assay also utilizes the following:

  • Alinity m HCV Application Specification File, (List No. 08N50-03B)
  • Alinity m System and System Software (List No. 08N53)
  • Alinity m Sample Prep Kit 2 (List No. 09N12-001)
  • Alinity m Specimen Dilution Kit I (List No. 09N50-001)
  • Alinity m System Solutions, (List No. 09N20):
    • Alinity m Lysis Solution (List No. 09N20-001)
    • Alinity m Diluent Solution (List No. 09N20-003)
    • Alinity m Vapor Barrier Solution, (List No. 09N20-004)
  • Alinity m Tubes and Caps (List No. 09N49):
    • Alinity m LRV Tube (List No. 09N49-001)
    • Alinity m Transport Tubes Pierceable Capped (List No. 09N49-010)
    • Alinity m Transport Tube (List No. 09N49-011)
    • Alinity m Pierceable Cap (List No. 09N49-012)
    • Alinity m Aliquot Tube (List No. 09N49-013)

Page 10

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 6 of 25

1.5 Intended Use

The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for both the detection and quantitation of hepatitis C virus (HCV) RNA, in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals. The assay is intended for use as an aid in the diagnosis of active HCV infection in individuals with antibody evidence of HCV infection, and to aid in the management of patients with known active HCV infection, including Sustained Virologic Response (SVR) determination.

The results from the Alinity m HCV assay must be interpreted within the context of all relevant clinical and laboratory findings.

The Alinity m HCV assay is not intended to be used in screening blood, plasma, serum, tissue or tissue donors for HCV.

1.6 Similarities and Differences to Predicate Device

The primary functional components of the Alinity m HCV assay are substantially equivalent to the current on-market Alinity m HCV assay (PMA 190025), which is a legally marketed Nucleic Acid-Based Hepatitis C Virus Ribonucleic Acid Test.

The Alinity m HCV assay has the same intended use as the predicate device. The subject device and predicate device differ only in the DNA polymerase and Reverse Transcriptase enzymes that may be used in manufacture of the assay; this formulation difference does not raise new types of safety or effectiveness questions.

These devices are similar in that they are designed to prepare nucleic acids for amplification, amplify specific HCV RNA sequences, detect the amplified products, and report quantitative results.

The primary similarities and differences between the Alinity m HCV assay and the predicate device are shown in Table 1 and Table 2.

Page 11

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 7 of 25

Table 1. Similarities Between Alinity m HCV and Predicate Device

DescriptionSubject DevicePredicate Device
Alinity m HCVAlinity m HCV (PMA 190025)
Intended UseSameThe Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for both the detection and quantitation of hepatitis C virus (HCV) RNA, in human plasma (EDTA, Acid Citrate Dextrose) or serum, from HCV antibody positive individuals. The assay is intended for use as an aid in the diagnosis of active HCV infection in individuals with antibody evidence of HCV infection, and to aid in the management of patients with known active HCV infection, including Sustained Virologic Response (SVR) determination.The results from the Alinity m HCV assay must be interpreted within the context of all relevant clinical and laboratory findings.The Alinity m HCV assay is not intended to be used in screening blood, plasma, serum, tissue or tissue donors for HCV.
Assay TypeSameQuantitative
Assay TargetsSamehighly conserved sequence within the 5' untranslated region (5'utr) of the HCV genome
Specimen TypesSameACD Plasma, EDTA Plasma, Serum
Sample Preparation ProcedureSameAutomated liquid handling and robotic manipulation platform
Amplification TechnologySameReal-time polymerase chain reaction

Page 12

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 8 of 25

Table 1. Similarities Between Alinity m HCV and Predicate Device (continued)

DescriptionSubject DevicePredicate Device
Alinity m HCVAlinity m HCV (PMA 190025)
Assay ControlsSame• Negative Control• Low Positive Control• High Positive Control• Internal Control (IC)

Table 2. Differences Between Alinity m HCV and Predicate Device

FeatureCurrent ApplicationPredicate Device
Alinity m HCVAlinity m HCV (P190025)
Amplification EnzymesOriginal DNA Polymerase or alternative DNA Polymerase enzyme used for DNA amplification.Original or alternative enzyme to convert target RNA to complimentary DNA (cDNA).Original DNA Polymerase enzyme used for DNA amplification.Original enzyme to convert target RNA to complimentary DNA (cDNA).

Page 13

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 9 of 25

1.7 Performance Data

The following performance data were provided in support of substantial equivalence determination of the device.

1.7.1 Specific Performance Characteristics

The subject device is similar to the predicate device originally approved (P190025) with the exception that the subject device may use a new DNA Polymerase as an alternative to the original DNA Polymerase and a new Reverse Transcriptase as an alternative to the original Reverse Transcriptase in the reagent formulation of the assay. Apart from the DNA polymerase and Reverse Transcriptase enzymes used, the reagent formulation, including all oligonucleotide primers and probes (components and concentrations) is identical between the subject device and predicate device. There are no differences in the intended use, end-user workflow, instrument workflow, assay software, reagent manufacturing, or final product release specifications.

Additional analytical performance studies listed in Table 3 were conducted to confirm the ability of the Alinity m HCV assay formulated with alternative DNA Polymerase and alternative Reverse Transcriptase (subject device) to meet the performance claims established in the corresponding studies submitted in P190025 for the Alinity m HCV assay formulated with the original DNA Polymerase and the original Reverse Transcriptase (predicate device).

Table 3. Additional Supporting Studies (Alternate Enzymes Formulation)

Study Description / Performance Characteristic510(k) Summary Section
Limit of Detection1.7.1.1
Linear Range1.7.1.2
Precision1.7.1.3
Lower Limit of Quantitation1.7.1.4

Page 14

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 10 of 25

All other specific performance characteristics of the subject device are supported by the studies submitted in P190025. Refer to Table 4.

Table 4. Supporting Studies Submitted in P190025

Study Description / Performance CharacteristicReference
Traceability to the WHO StandardOriginal submission
Genotype Limit of DetectionOriginal submission
Genotype LinearityOriginal submission
Analytical Specificity – Potential Cross-ReactantsOriginal submission
Analytical Specificity – Potentially Interfering SubstancesOriginal submission
CarryoverOriginal submission
Alinity m HCV Testing Using Dilution ProcedureOriginal submission
Precision of Alinity m HCV Using Dilution ProcedureOriginal submission
Confirmation of LLoQ Using Dilution ProcedureOriginal submission

1.7.1.1 Limit of Detection

A limit of detection (LoD) study was conducted previously to support the approval of P190025. Please refer to the decision summary for P190025.

An additional study was conducted to confirm the LoD claim of Alinity m HCV (12 IU/mL) for the formulation of the assay using the alternative enzymes.

The LoD for HCV was evaluated by testing dilutions of a World Health Organization (WHO) HCV Standard (NIBSC code 18/184) prepared in HCV negative human plasma. Testing for each HCV DNA concentration was performed with 3 lots of amplification reagents across multiple days.

The data from the study demonstrated that the Alinity m HCV assay formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase had an overall detection rate of 98.6% at 12 IU/mL. Refer to Table 5.

Page 15

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 11 of 25

Table 5. Alinity m HCV (Alternate Enzymes Formulation) Limit of Detection (LoD) Results

Target Concentration (IU/mL)AMP Kit Reagent LotNumber of Total ReplicatesNumber of Replicates DetectedDetection Rate (%)95% Score Confidence Interval (%)
91484695.8(86.0, 98.8)
2484593.8(83.2, 97.9)
3484797.9(89.1, 99.6)
All Combined14413895.8(91.2, 98.1)
12ᵃ14848100.0(92.6, 100.0)
2484695.8(86.0, 98.8)
34747100.0(92.4, 100.0)
All Combined14314198.6(95.0, 99.6)
1614747100.0(92.4, 100.0)
24848100.0(92.6, 100.0)
34545100.0(92.1, 100.0)
All Combined140140100.0(97.3, 100.0)

ᵃ Alinity m HCV assay limit of detection (LoD) claim

1.7.1.2 Linear Range

A linearity study was conducted previously to support the approval of P190025. Please refer to the decision summary for P190025.

An additional study was conducted to confirm that the Alinity m HCV assay formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase was linear across the claimed quantitation range (12 IU/mL [1.08 Log IU/mL] to 100,000,000 IU/mL [8.00 Log IU/mL]).

Linearity was evaluated by testing 12 panel members that spanned the intended quantitation range of the assay(12 IU/mL to 100,000,000 IU/mL), including a panel member below the expected Lower Limit of Quantification (LLoQ) at 10 IU/mL and a panel member exceeding the expected Upper Limit of Quantification (ULoQ) at 200,000,000 IU/mL. Refer to Table 6.

Page 16

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 12 of 25

Table 6. Panel Members for Linearity

Panel MemberTargeted HCV ConcentrationSource
IU/mLLog IU/mL
01200,000,0008.30aRNA
0210,000,0007.00aRNA
031,000,0006.00aRNA
04100,0005.00aRNA
0510,0004.00Clinical Sample
065,0003.70aRNA
071,0003.00Clinical Sample
085002.70aRNA
091002.00Clinical Sample
10501.70aRNA
11ᵃ121.08Clinical Sample
12101.00Clinical Sample

ᵃ The assay's claimed LLoQ is 12 IU/mL per package insert.

The data from the study demonstrated that the Alinity m HCV assay formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase was linear across the quantitative range from 12 IU/mL to 100,000,000 IU/mL (1.08 Log IU/mL to 8.00 Log IU/mL). Results for Alinity m HCV linearity performance are shown in Figure 1.

Page 17

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 13 of 25

Figure 1. Alinity m HCV (Alternate Enzymes Formulation) Linearity

[THIS IS FIGURE: A scatter plot showing linearity data with Expected Concentration (Log IU/mL) on x-axis (0-9) and Alinity HCV (Log IU/mL) on y-axis (0-9). The plot shows a strong linear relationship with r = 1.000 and data points following a diagonal line from approximately (1,1) to (8,8).]

Page 18

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 14 of 25

1.7.1.3 Precision

A precision study was conducted previously to support the approval of P190025. Please refer to the decision summary for P190025.

An additional study was conducted to confirm the precision claim of Alinity m HCV (as follows) for the alternative DNA Polymerase and the alternative Reverse Transcriptase formulation of the assay:

  • Within-laboratory standard deviation (SD) of ≤ 0.25 Log IU/mL for samples between 100 IU/mL to 200,000,000 IU/mL (2.00 Log IU/mL to 8.30 Log IU/mL) HCV RNA
  • Within-laboratory SD of ≤ 0.35 Log IU/mL for samples with target concentrations from the claimed LLoQ to 3x LLoQ (12 IU/mL to 36 IU/mL or 1.08 Log IU/mL to 1.56 Log IU/mL) HCV RNA

Precision was evaluated by testing 8 panel members with HCV concentrations ranging from 12 IU/mL to 200,000,000 IU/mL (1.08 Log IU/mL to 8.30 Log IU/mL). Panel members were prepared by diluting HCV positive specimen or armored RNA (aRNA) in human EDTA plasma. aRNA was only used for panel members with high concentrations (> 5.00 Log IU/mL).

Three lots of Alinity m HCV amplification reagents manufactured with the alternative DNA Polymerase and the alternative Reverse Transcriptase were run on 1 Alinity m System. Three replicates of each panel member were run twice each day for 15 days for each lot with a minimum separation of 2 hours between successive runs of a panel member.

The overall precision analysis summary for the Alinity m HCV assay formulated with the alternative enzymes is presented in Table 7.

Page 19

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 15 of 25

The results demonstrated that the Alinity m HCV assay has a within-laboratory SD of 0.25 Log IU/mL or less for samples targeted between 100 IU/mL to 200,000,000 IU/mL (2.00 Log IU/mL to 8.30 Log IU/mL) HCV RNA, and a within-laboratory SD of 0.35 Log IU/mL or less for samples with target concentrations from the claimed LLoQ to 3x LLoQ (12 IU/mL to 36 IU/mL or 1.08 Log IU/mL to 1.56 Log IU/mL) HCV RNA

Table 7. Alinity m HCV (Alternate Enzymes Formulation) Precision Analysis – All Reagent Lots

PanelNMean Concentration (Log IU/mL)Within-Run ComponentBetween-Run ComponentBetween-Day ComponentWithin-Laboratory ᵃBetween-Lot ComponentTotal ᵇ
SD%CVSD%CVSD%CV
012690.960.2424.60.000.00.1010.2
022701.610.148.80.063.50.000.0
032701.800.116.20.063.10.000.0
042702.880.093.00.051.60.000.0
052703.780.071.90.041.10.000.0
062705.850.061.10.020.30.030.5
072706.890.040.50.020.20.000.0
082708.260.040.50.010.10.020.2

ᵃ Within-Laboratory includes Within-Run, Between-Run, and Between-Day Components.
ᵇ Total includes Within-Run, Between-Run, Between-Day, and Between-Lot Components.

Page 20

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 16 of 25

The Total SD and Total %CV for each panel member in this study were compared with the Total SD and Total %CV from the original precision study submitted in P190025. Refer to Table 8.

Table 8. Equivalence Evaluation of Total SD and Total %CV (New ᵃ vs. Original ᵇ)

Panel MemberMean ConcentrationTotal SDTotal %CVRatios and 95% CI
New ᵃ Mean (Log IU/mL)Original ᵇ Mean (Log IU/mL)New ᵃ SDOriginal ᵇ SDNew ᵃ % CVOriginal ᵇ % CVSD%CV
010.961.070.270.2128.219.71.28 (1.01, 1.63)1.43 (1.12, 1.81)
021.611.420.170.1810.612.60.96 (0.68, 1.34)0.85 (0.60, 1.19)
031.801.910.150.178.18.70.88 (0.53, 1.46)0.94 (0.57, 1.55)
042.882.880.120.174.15.80.71 (0.40, 1.26)0.71 (0.40, 1.26)
053.783.940.100.142.73.60.73 (0.47, 1.12)0.76 (0.49, 1.16)
06ᶜ5.856.140.090.111.51.80.77 (0.33, 1.80)0.81 (0.35, 1.89)
07ᵈ6.897.110.070.161.12.30.45 (0.22, 0.92)0.47 (0.23, 0.95)
08ᵉ8.268.550.110.161.31.90.69 (0.17, 2.87)0.72 (0.17, 2.97)

ᵃ "New" Mean, SD, and %CV data for Alinity m HCV assay formulated with alternative DNA Polymerase and alternative Reverse Transcriptase.
ᵇ "Original" Mean, SD, and %CV data for Alinity m HCV assay formulated with original DNA Polymerase and original Reverse Transcriptase (per P190025)
ᶜ Panel 06 in the New Precision protocol targeting 6.0 Log IU/mL is compared to panel 07 in the Original Precision protocol targeting 6.0 Log IU/mL
ᵈ Panel 07 in the New Precision protocol targeting 7.0 Log IU/mL is compared to panel 08 in the Original Precision protocol targeting 7.0 Log IU/mL.
ᵉ Panel 08 in the New Precision protocol targeting 8.3 Log IU/mL is compared to panel 09 in the Original Precision protocol targeting 8.3 Log IU/mL.
NOTE: Panel 06 in the Original Precision protocol targeting 5.0 Log IU/mL is not analyzed for comparison.

Page 21

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 17 of 25

1.7.1.4 Lower Limit of Quantitation (LLoQ)

A lower limit of quantitation (LLoQ) study was conducted previously to support the approval of P190025. Please refer to the decision summary for P190025.

An additional study was conducted to confirm the LLoQ claim of Alinity m HCV (12 IU/mL) for the alternative DNA Polymerase and alternative Reverse Transcriptase formulation of the assay.

The LLoQ is defined as the lowest concentration at which HCV RNA is reliably quantitated within an acceptable total error. Total error was estimated for detected samples from the LoD study (Section 1.7.1.1), the Linearity study (Section 1.7.1.2), and the Precision study (Section 1.7.1.3) by 2 methods:

  • Total Analytical Error (TAE) = |bias| + 2 × SD, (where bias = mean concentration – assigned target concentration)

and

  • Total Error (TE) = SQRT(2) × 2 × SD

Panel members were dilutions of a World Health Organization (WHO) HCV Standard (NIBSC code 18/184) prepared in HCV negative human plasma.

The results of the calculations are shown in Table 9.

The results of these analyses support a claimed LLoQ of 12 IU/mL (1.08 Log IU/mL), with an acceptable level of accuracy and precision (ie, TAE and TE less than or equal to 1.00 Log IU/mL).

Page 22

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 18 of 25

Table 9. Alinity m HCV (Alternate Enzyme Formulation) TAE and TE

StudyPanelTarget (IU/mL)NTarget (Log IU/mL)Mean (Log IU/mL)Bias (Log IU/mL)SDTAETE
Linearity1112241.080.90-0.180.270.720.76
1210241.000.72-0.280.320.930.92
Precision01122691.080.96-0.120.260.630.73
Sensitivity191380.950.77-0.180.260.700.74
2121411.080.85-0.230.270.770.76
3161401.201.06-0.150.210.570.59

Page 23

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 19 of 25

1.7.1 Performance Characteristics

1.7.1.1 Reproducibility

A reproducibility study was conducted previously to support the approval of P190025. Please refer to the decision summary for P190025.

An additional study was conducted to evaluate the reproducibility of the Alinity m HCV assay formulated with alternative DNA Polymerase and alternative Reverse Transcriptase.

Reproducibility performance of Alinity m HCV was evaluated by testing an 11-member reproducibility panel, which included 10 positive panel members and 1 negative panel member. Panel members were prepared in base-matrix (EDTA plasma) and spanned the linear range of Alinity m HCV assay. One panel member was negative for HCV. Ten positive panel members represented the two different genotypes 1a & 2. Panel members were prepared by diluting HCV positive clinical specimen or HCV armored RNA (aRNA) in HCV negative plasma. HCV aRNA was only used for panel members with high concentrations (>5.0 Log IU/mL).

One lot each of Alinity m HCV AMP Kit, Alinity m HCV CAL Kit, Alinity m HCV CTRL Kit, and Alinity m Sample Prep Kit 2 lots were used. Testing was performed at 3 clinical sites on 5 non-consecutive days with 2 runs per day. Four replicates of each panel member were tested on each of 5 days to ensure a minimum of 3 valid replicates for analysis.

The reproducibility results are summarized in Table 10 (for the positive panel members) and in Table 11 (for the negative panel member).

Page 24

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 20 of 25

Table 10. Alinity m HCV (Alternate Enzymes Formulation) Overall Variance Component Analysis

GenotypePanelNMean Concentration (Log IU/mL)Within-Run ComponentBetween-Run ComponentBetween-Day ComponentWithin-Laboratory ᵃBetween-Site ComponentTotal ᵇ
SD%CVSD%CVSD%CV
1a11187.710.202.60.081.10.101.4
1a21155.720.061.10.050.80.000.0
1a31153.530.061.60.041.10.030.9
1a41191.840.115.90.074.10.000.0
1a51160.890.2123.80.077.40.011.4
1a61160.800.2531.50.045.50.000.0
271206.970.060.90.030.50.050.6
281162.990.082.60.020.80.031.1
291181.650.148.50.000.00.052.8
2101191.050.2221.30.000.00.065.9

ᵃ Within-Laboratory includes Within-Run, Between-Run, and Between-Day Variance Components.
ᵇ Total (Overall) includes Within-Run, Between-Run, Between-Day, and Between-Site Variance Components.

Page 25

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 21 of 25

Table 11. Alinity m HCV (Alternate Enzymes Formulation) Overall Negative Agreement Analysis

PanelSiteNNegative Agreement Rate95% CI
TotalPositiveNegative
11All1200120100.0%

The Total SD and Total %CV for each panel member in this study were compared with the Total SD and Total %CV from the original reproducibility study submitted in P190025. Refer to Table 12.

Page 26

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 22 of 25

Table 12. Equivalence Evaluation of Total SD and Total %CV (New ᵃ vs. Original ᵇ)

GenotypePanelMean ConcentrationTotal SDTotal %CVRatios and 95% CI
New Mean (Log IU/mL)Original Mean (Log IU/mL)New SDOriginal SDNew %CVOriginal %CVSD%CV
1a17.718.500.250.153.21.71.68 (0.97, 2.92)1.85 (1.06, 3.21)
1a25.726.150.100.161.82.60.64 (0.44, 0.93)0.69 (0.48, 1.00)
1a33.533.890.080.182.44.50.47 (0.36, 0.61)0.52 (0.40, 0.68)
1a41.841.950.140.227.811.00.66 (0.52, 0.84)0.70 (0.55, 0.90)
1a50.891.060.240.3527.533.00.70 (0.53, 0.94)0.83 (0.63, 1.11)
1a60.800.650.260.3732.956.80.71 (0.58, 0.87)0.58 (0.47, 0.71)
276.977.300.090.111.31.50.83 (0.59, 1.16)0.87 (0.62, 1.21)
282.993.060.090.213.06.70.44 (0.36, 0.54)0.45 (0.36, 0.56)
291.651.440.160.249.717.00.66 (0.51, 0.85)0.57 (0.44, 0.74)
2101.050.880.240.3023.134.20.80 (0.66, 0.98)0.67 (0.55, 0.83)

ᵃ "New" Mean Concentration, Total SD, and Total %CV data for Alinity m HCV assay formulated with alternative enzymes.
ᵇ "Original" Mean Concentration, Total SD, and Total %CV data for on-market Alinity m HCV assay (submitted in P190025).

Page 27

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 23 of 25

1.7.1.2 Method Comparison

A method comparison study was conducted with the Alinity m HCV assay formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase to demonstrate equivalence to the FDA-approved (on-market) Alinity m HCV assay formulated with the original DNA Polymerase and the original Reverse Transcriptase (P190025).

Regression and bias analysis included a total of 222 samples with results that were within the common quantitation range of both the Alinity m HCV assay formulated with alternative enzymes formulation (Alinity m HCV IUO) and the comparator assay (on-market Alinity m HCV). Figure 2 shows the results of the Deming regression analysis with a slope of 1.01, an intercept of 0.03, and a correlation coefficient of 0.996.

The mean bias between Alinity m HCV IUO and the comparator (Alinity m HCV IUO minus comparator) was 0.06 Log IU/mL with a 95% Confidence Interval of (0.04, 0.09).

Page 28

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 24 of 25

Figure 2. Deming Regression Analysis

[THIS IS FIGURE: A scatter plot showing the correlation between Comparator HCV Quantitation (Log IU/mL) on x-axis and Alinity m HCV IUO Quantitation (Log IU/mL) on y-axis. The plot shows a strong linear relationship with the equation y = 1.01x + 0.03, r = 0.996, n = 222. Data points range from approximately 1 to 8 on both axes, following a tight linear pattern.]

The Systematic difference between Alinity m HCV assay formulated with alternate enzymes and the on-market Alinity m HCV assay (comparator) at 4 selected viral load levels is shown in Table 13.

Page 29

Alinity m HCV 510(k) K252102
September 2025
Attachment 022_Alinity m HCV_510(k) Summary_mw006 Page 25 of 25

Table 13. Systematic Difference at Selected Viral Load Levels

Target Viral Load Level (per comparator)Systematic Difference95% CI (Log IU/mL)
1.08 Log IU/mL0.04 Log IU/mL(-0.00, 0.09)
1.40 Log IU/mL0.05 Log IU/mL(0.00, 0.09)
3.60 Log IU/mL0.06 Log IU/mL(0.04, 0.09)
5.80 Log IU/mL0.08 Log IU/mL(0.05, 0.11)

1.8 Conclusions Drawn from the Studies

The additional analytical and clinical study results demonstrate that the Alinity m HCV assay formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase performs comparably to the FDA approved Alinity m HCV assay formulated with the original DNA Polymerase and the original Reverse Transcriptase. The results support a substantial equivalence decision for Alinity m HCV formulated with the alternative DNA Polymerase and the alternative Reverse Transcriptase.

§ 866.3170 Nucleic acid-based hepatitis C virus ribonucleic acid tests.

(a)
Identification. A nucleic acid-based hepatitis C virus (HCV) ribonucleic acid (RNA) test is identified as an in vitro diagnostic device intended for prescription use as an aid in the diagnosis of HCV infection in specified populations, and/or as an aid in the management of HCV-infected patients including guiding the selection of genotype-specific treatment in individuals with chronic HCV infection. The test is intended for use with human serum or plasma. The test is not intended for use as a donor screening test for the presence of HCV antibodies in blood, blood products, or tissue donors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) For all nucleic acid-based HCV RNA tests, the labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the test is not intended for use as a donor screening test for the presence of HCV RNA from human cells, tissues, and cellular and tissue-based products.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. These limitations must include, but are not limited to, statements that indicate:
(A) The specimen types for which the device has been cleared and that use of this test kit with specimen types other than those specifically cleared for this device may result in inaccurate test results.
(B) When applicable, that assay performance characteristics have not been established in populations of immunocompromised or immunosuppressed patients or, other populations where test performance may be affected.
(C) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with the individual's clinical presentation, history, and other laboratory results.
(2) For all nucleic acid-based HCV RNA tests, the design verification and validation must include:
(i) Detailed device description, including the device components, ancillary reagents required but not provided, and an explanation of the device methodology. Additional information appropriate to the technology must be included such as design of primers and probes, rationale for the selected gene targets, specifications for amplicon size, and degree of nucleic acid sequence conservation.
(ii) For devices with assay calibrators, the design and nature of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
e.g., a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
e.g., determination of the identity, supplier, purity, and stability) of all critical reagents (including nucleic acid sequences for primers and probes) and protocols for maintaining product integrity.(iv) Detailed documentation of analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including, but not limited to, limit of detection (LoD), upper and lower limits of quantitation (ULoQ and LLoQ, respectively), linearity, precision, endogenous and exogenous interferences, cross reactivity, carryover, matrix equivalency, and sample and reagent stability. Samples selected for use in analytical studies or used to prepare samples for use in analytical studies must be from subjects with clinically relevant circulating genotypes in the United States. Cross-reactivity studies must include samples from HCV RNA negative subjects with other causes of liver disease, including autoimmune hepatitis, alcoholic liver disease, chronic hepatitis B virus, primary biliary cirrhosis, and nonalcoholic steatohepatitis, when applicable. The effect of each claimed nucleic-acid isolation and purification procedure on detection must be evaluated.
(v) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on test performance.
(vi) Final release criteria to be used for manufactured test lots with appropriate evidence that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(vii) Multisite reproducibility study that includes the testing of three independent production lots.
(viii) All stability protocols, including acceptance criteria.
(ix) Final release test results for each lot used in clinical studies.
(x) Analytical sensitivity and specificity of the test must be the same or better than that of other cleared or approved tests.
(xi) Lot-to-lot precision studies, as appropriate.
(3) For devices intended for the qualitative detection of HCV RNA, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, the design verification and validation must include detailed documentation of performance from a multisite clinical study. Performance must be analyzed relative to an FDA cleared or approved qualitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with appropriate numbers of HCV positive and negative samples in applicable risk categories. Additional genotypes must be validated using appropriate numbers and types of samples. The samples may be a combination of fresh and repository samples, sourced from within and outside the United States, as appropriate. The study designs, including number of samples tested, must be sufficient to meet the following criteria:
(i) Clinical sensitivity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 95 percent.
(ii) Clinical specificity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 96 percent.
(4) For devices intended for the quantitative detection of HCV RNA, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Labeling required under § 809.10(b) of this chapter must include a prominent statement that the test is not intended as a diagnostic test to confirm the presence of active HCV infection, when applicable.
(ii) Design verification and validation must include the following:
(A) Detailed documentation of the following analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including but not limited to: LoD, ULoQ and LLoQ. LoD, LLoQ, and linearity studies must demonstrate acceptable device performance with all HCV genotypes detected by the device.
(B) Detailed documentation of clinical performance testing from either:
(
1 ) A multisite clinical study with an appropriate number of clinical samples from chronically HCV infected patients in which the results are compared to an FDA-cleared or approved quantitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must include a sufficient number of HCV positive samples containing an analyte concentration near the LLoQ to describe performance at this level. Clinical samples must cover the full range of the device output and must be consistent with the distribution of these genotypes in the U.S. population. Clinical samples may be supplemented with diluted clinical samples for those viral load concentrations that are not sufficiently covered by natural clinical specimens, or(
2 ) A clinical study with prospectively collected samples demonstrating clinical validity of the device.(C) Detailed documentation of a qualitative analysis near the lower end of the measuring range demonstrating acceptable performance when used as an aid in diagnosis.
(5) For devices intended for HCV RNA genotyping, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, design verification and validation must include the following:
(i) Detailed documentation of an analytical performance study demonstrating the LoD for all HCV genotypes detected by the device.
(ii) Detailed documentation, including results, of a multisite clinical study that assesses genotyping accuracy (
i.e., the proportion of interpretable results that match with the reference method result) and the genotyping rate (i.e., the proportion of results that were interpretable).(6) For any nucleic acid-based HCV RNA test intended for Point of Care (PoC) use, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Clinical studies must be conducted at PoC sites.
(ii) Additional labeling must include a brief summary of the instructions for use that are appropriate for use in a PoC environment.