(50 days)
The Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit-Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only.
This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to a Fentanyl test result, particularly when a preliminary positive result is obtained.
The Healgen® Immunofluorescence analyzer OG-H180 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to detect test results from in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or point-of-care setting.
The AccuFluor Fentanyl FIA Test Kit-Qualitative is a rapid fluorescence immunoassay based on the principle of competitive binding, which uses fluorescent microspheres-labeled antibody as the indicator marker to qualitatively detect fentanyl in human urine. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody.
During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1.0 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody coated fluorescence particles will then be captured by immobilized Fentanyl conjugate, and the signal will be detected in the test line (T) region to show a negative result. The signal will not be detected in the test line (T) region if the Fentanyl level exceeds 1.0 ng/mL because all the binding sites for the anti-Fentanyl antibodies will be saturated and the result will show as positive. To serve as a procedural control, a signal will be detected at the control line (C) region indicating the proper volume of specimen has been added and membrane wicking has occurred. The test is interpreted by the Healgen® Immunofluorescence analyzer OG-H180 and the result will be interpreted by the analyzer.
The provided FDA 510(k) clearance letter pertains to the Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative and the Healgen® Immunofluorescence Analyzer (OG-H180). This document outlines the general regulatory approval and provides some performance characteristics, but it is not a comprehensive study report detailing all aspects of the acceptance criteria and the full study that proves the device meets those criteria.
Specifically, the document does not explicitly state "acceptance criteria" as a defined set of metrics and thresholds prior to presenting performance data. Instead, it presents results from various analytical performance studies which are implicitly used to demonstrate equivalence to a predicate device. Similarly, it does not describe "human expert ground truth establishment," "adjudication methods," or "MRMC comparative effectiveness studies" because these are typically relevant for AI/ML-based diagnostic devices utilizing image interpretation or complex decision support, which is not the primary function described for this immunoassay and analyzer.
This device is an in vitro diagnostic (IVD) test for qualitative detection of fentanyl in urine, which relies on a chemical reaction read by an analyzer. Therefore, the "study" described is a series of analytical performance tests, rather than a clinical study with human readers and ground truth established by medical experts in the way that would be done for an AI radiology device, for example.
Despite these limitations in the provided text for certain categories, I will extract and infer information where possible based on the provided document and common IVD device clearance practices.
Acceptance Criteria and Device Performance for Healgen® AccuFluor Fentanyl FIA Test Kit
1. Table of Acceptance Criteria and Reported Device Performance
As noted, the document does not explicitly list pre-defined "acceptance criteria" with specific numerical thresholds for all metrics. However, based on the provided performance data, here's an interpretation of the implied criteria and the reported performance. The "acceptance criteria" inferred here are based on what constitutes successful demonstration of performance for an IVD device of this type, often aiming for high accuracy, precision, and lack of interference, especially around the cutoff concentration.
| Performance Characteristic | Implied Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Analytical Precision (Around Cutoff) | High agreement (low false negatives/positives) at concentrations near the 1.0 ng/mL cutoff. | At -25% Cutoff (0.75 ng/mL): 56-57 out of 60 negative readings across 3 lots (93.3% - 95% negative agreement). At Cutoff (1.0 ng/mL): 22-24 negative and 36-38 positive readings out of 60 total across 3 lots. This demonstrates the expected transition around the cutoff. At +25% Cutoff (1.25 ng/mL): 60 out of 60 positive readings across 3 lots (100% positive agreement). At -100%, -75%, -50% Cutoff: 60 out of 60 negative readings. At +50%, +75%, +100% Cutoff: 60 out of 60 positive readings. |
| Analytical Specificity (Cross-Reactivity) | Minimal to no cross-reactivity with common related compounds (e.g., other opioids, metabolites) or other substances found in urine, beyond expected concentrations where some cross-reactivity is acceptable and quantified. | Norfentanyl: 0.003% cross-reactivity at 30,000 ng/mL. Carfentanil: 0.013% cross-reactivity at 8,000 ng/mL. Cyclopropyl fentanyl, Para-fluoro fentanyl, Acetyl fentanyl: 100% cross-reactivity at 1 ng/mL. Many other fentanyl analogs: various quantified cross-reactivity percentages. Numerous unrelated compounds/metabolites: No cross-reactivity at 100 µg/mL. |
| Interference | No significant interference from common substances (physiological or exogenous) found in human urine at specified concentrations. | Many compounds (e.g., Acetaminophen, Ethanol, Glucose, Albumin, Hemoglobin) showed no interference at high concentrations (e.g., 100µg/mL or 1% for Ethanol) for both negative and ±50% Cut-Off fentanyl spiked samples. |
| Stability | Device maintains stated performance characteristics over its shelf life under specified storage conditions. | Stable at 2-30°C for 27 months based on real-time stability study. |
| Effect of Urine Specific Gravity & pH | Test performance (positive/negative call) should remain consistent across a physiological range of urine specific gravity and pH. | Results were consistent (all positive for samples at/above +50% Cut-Off, all negative for samples at/below -50% Cut-Off) across specific gravity 1.000-1.035 and pH 4-9. |
| Method Comparison (Clinical Samples) | High agreement (concordance) with a confirmed analytical method (LC-MS/MS) for clinical samples, especially for samples near the cutoff. | Overall Concordance: Across 3 sites, for 80 clinical samples (40 negative, 40 positive) compared to LC-MS/MS: True Negative Ranges: 7 negative, 19 low negative (less than -50% cutoff) correctly identified as negative. Near Cutoff Negative: 11-12 samples (between -50% and cutoff) correctly identified as negative, with 2-3 false positives. Discordant Results: 6 samples near cutoff showed discordance. For example, 3 samples (0.802, 0.841, 0.916 ng/mL) were LC-MS/MS negative but device positive. 3 samples (1.013, 1.092, 1.113 ng/mL) were LC-MS/MS positive but device negative. This indicates typical variability at the qualitative cutoff. Near Cutoff Positive: 20-21 samples (between cutoff and +50%) correctly identified as positive, with 2-3 false negatives. High Positive: 17 samples (greater than +50%) correctly identified as positive. |
2. Sample Size Used for the Test Set and Data Provenance
- Analytical Precision: 60 replicates per concentration (6 replicates/day for 10 days) per lot, across 9 concentration levels, for 3 device lots. Total: 60 * 9 * 3 = 1620 individual tests.
- Interference: Samples with various interfering substances were tested, each at both drug-free and ±50% Cut-Off spiked fentanyl concentrations, using three batches of device. (Exact number of tests not specified, but implies a comprehensive set).
- Specificity: Various drug metabolites and other compounds tested, each using three batches of device. (Exact number of tests not specified).
- Effect of Urine Specific Gravity and pH: Samples across the specified ranges were tested at -50% and +50% Cut-Off levels by three different operators using three device lots. (Exact number of tests not specified).
- Method Comparison (Clinical Samples): 80 unaltered clinical samples (40 negative, 40 positive). These samples were run at three different testing sites.
- Data Provenance: The document does not explicitly state the country of origin for the clinical samples. It does state they were "unaltered clinical samples," implying they were retrospective real-world samples collected from patients. It does not indicate if they were prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable in the traditional sense for this device. For this IVD device, the primary ground truth for its performance studies (precision, specificity, method comparison) is established by analytical gold standards, specifically:
- LC/MS-MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) for confirming fentanyl concentrations in precision studies and as the comparator method in the method comparison study.
- This is a highly accurate and precise laboratory method for quantifying drug concentrations, and its results are considered the "ground truth" for chemical concentration data.
- There were "three different operators" for the specific gravity/pH study, but these are not "experts" in the sense of medical professionals establishing a clinical diagnosis ground truth. They are laboratory personnel performing the test.
4. Adjudication Method for the Test Set
- Not Applicable in the traditional sense. Given that the ground truth is established by LC-MS/MS, there is no human "adjudication" process like consensus reading by multiple radiologists for image interpretation. The LC-MS/MS results serve as the definitive reference.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, an MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance on complex interpretation tasks) is not applicable to a qualitative immunoassay and analyzer like the Healgen AccuFluor Fentanyl FIA Test Kit, which determines the presence or absence of a substance based on a fluorescent signal. The device performance is assessed on its analytical accuracy against a gold standard method.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, the performance presented is primarily standalone. The Healgen® Immunofluorescence Analyzer (OG-H180) automatically interprets the fluorescent signal from the test kit. The performance data (precision, specificity, interference, method comparison) directly reflects the analytical capability of the device and test kit combination, without any human interpretation or intervention in the final "positive" or "negative" determination. A human loads the sample and the device performs the analysis and provides the result.
7. The Type of Ground Truth Used
- Analytical Gold Standard (LC-MS/MS): This is the primary method used to establish the true concentration of fentanyl in samples for precision studies and as the comparative reference for clinical samples.
- Spiked Samples: For analytical performance studies (precision, interference, specificity), known concentrations of fentanyl or interfering substances were added to negative urine samples, establishing a controlled ground truth.
8. The Sample Size for the Training Set
- Not explicitly stated in the document, and likely not applicable in the typical AI/ML sense. This device is an immunoassay, not an AI/ML diagnostic algorithm that undergoes a "training" phase with a large dataset. Immunoassays are based on biochemical principles and do not "learn" from data in the same way. Performance is optimized during development and validated analytically.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable / Not Described. As it's not an AI/ML device relying on a training set, the concept of establishing ground truth for training does not apply here. The analytical performance is characterized through rigorous testing under controlled conditions and comparison to established reference methods (like LC-MS/MS).
FDA 510(k) Clearance Letter - K251972
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 15, 2025
Healgen Scientific LLC
℅ Jenny Xia
Director
LSI International Inc
504E Diamond Ave., Suite H
Gaithersburg, Maryland 20877
Re: K251972
Trade/Device Name: Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit - Qualitative; Healgen® Immunofluorescence Analyzer (OG-H180)
Regulation Number: 21 CFR 862.3650
Regulation Name: Opiate test system
Regulatory Class: Class II
Product Code: DJG, KHO
Dated: June 26, 2025
Received: June 26, 2025
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K251972 - Jenny Xia Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K251972 - Jenny Xia Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.08.15 10:28:52 -04'00'
Joseph Kotarek
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K251972
Device Name: Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit - Qualitative; Healgen® Immunofluorescence Analyzer (OG-H180)
Indications for Use (Describe)
The Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit-Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only.
This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to a Fentanyl test result, particularly when a preliminary positive result is obtained.
The Healgen® Immunofluorescence analyzer OG-H180 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to detect test results from in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or point-of-care setting.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) SUMMARY
-
Date: August 12, 2025
-
Submitter: Healgen Scientific LLC.
3818 Fuqua Street
Houston, TX 77047 -
Contact person: Jenny Xia
LSI International Inc.
504E Diamond Ave., Suite H
Gaithersburg, MD 20877
Telephone: 301-525-6856
Email: jxia@lsi-consulting.org -
Device Names: Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative
Healgen Immunofluorescence Analyzer OG-H180Classification: Class 2
Product Code Classification Regulation Section Panel DJG II 21 CFR § 862.3650 Opiate Test System Toxicology (91) KHO I 21 CFR § 862.2560 Fluorometer for clinical use Clinical Chemistry -
Predicate Devices:
BioSieve™ Fentanyl FIA Test Kit, K240124
BioSieve™ ToxiSmart FIA Reader, K240124 -
Indications for Use
Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only.
This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to a Fentanyl test result, particularly when a preliminary positive result is obtained.
The Healgen Immunofluorescence analyzer OG-H180 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to detect test results from in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or point-of-care setting.
-
Device Description
The AccuFluor Fentanyl FIA Test Kit-Qualitative is a rapid fluorescence immunoassay based on the principle of competitive binding, which uses fluorescent microspheres-labeled antibody as the
Page 6
indicator marker to qualitatively detect fentanyl in human urine. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody.
During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1.0 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody coated fluorescence particles will then be captured by immobilized Fentanyl conjugate, and the signal will be detected in the test line (T) region to show a negative result. The signal will not be detected in the test line (T) region if the Fentanyl level exceeds 1.0 ng/mL because all the binding sites for the anti-Fentanyl antibodies will be saturated and the result will show as positive. To serve as a procedural control, a signal will be detected at the control line (C) region indicating the proper volume of specimen has been added and membrane wicking has occurred. The test is interpreted by the Healgen® Immunofluorescence analyzer OG-H180 and the result will be interpreted by the analyzer.
-
Substantial Equivalence Information
A summary comparison of features of the Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative and the predicate devices is provided in following table.
Table 1: Features Comparison of Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit - Qualitative and the Predicate Devices
Item Device Predicate - K240124 Indication(s) for Use For the qualitative determination of fentanyl in human urine. Same Calibrator and Cut-Off Values Fentanyl (FTY) 1 ng/ml Same Methodology Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. Same Type of Test Qualitative Same Specimen Type Human Urine Same Intended Use For prescription use Same Configurations Cassette Same Platform Required Immunofluorescence Analyzer Same Storage 2-30˚C Same Table 2: Instrument Similarities and Differences
Item Device Predicate - K240124 Intended Use/ Indication for Use Immunofluorescence analyzer designed to perform in vitro diagnostic tests on clinical specimens including drug urine test. Same
Page 7
| Principles of Assay Operation | Competitive immunofluorescence immunoassay | Same |
|---|---|---|
| Calibration Check | A self-test card is supplied with the analyzer and used to check instrument functions including QR code recognition and test cassette scanning. | A Quality control test device is supplied with the Reader and used to check the Reader optics and calculation systems. |
| Test Modes | Standard test Non-standard test (Quick test). | Same |
| User interface | 2.8 inch LCD Screen display | 1.54 inch LCD Screen display |
| Light Source | LED Light | Same |
| Power Supply | Rechargeable 3.7V - 2200mA Li-ion Battery | Same |
| Dimensions | 13.52 cm x 8.38 cm x 4.3 cm | 12.45 cm x 7.25 cm x 4 cm |
| Weight | ~0.54 lbs | ~0.36 lbs |
-
Test Principle
The AccuFluor Fentanyl FIA Test Kit-Qualitative uses the principle of competitive and fluorescence immunochromatography assay. The nitrocellulose membrane test area (T) of the test strip is correspondingly coated with fentanyl-bovine serum albumin conjugate, and the quality control area (C) is coated with goat anti-rabbit lgG polyclonal antibody. Both Fentanyl monoclonal antibody and rabbit lgG polyclonal antibody labeled with fluorescent microspheres are embedded on the conjugate pad. The labeled antibody will flow forward with the sample, when the urine sample is applied to the sample well of the test device. When the concentration of fentanyl in the sample is higher than or equal to the cut-off of the product, it will compete with the corresponding conjugate coated on the test area (T) to bind to the fluorescently labeled monoclonal antibody, the fluorescence signal rendering of the test line is inhibited and the result is positive; while when the sample does not contain fentanyl or its concentration is lower than the cut-off of the product, the corresponding conjugate on the test line reacts with sufficient fluorescently-labeled monoclonal antibodies, the test line will have fluorescence signal and the result is negative. The quality control area (C) will develop fluorescence signal, which is the criteria for judging whether the test process is normal or not. Signal intensity of fluorescence is detected by Healgen Immunofluorescence analyzer OG-H180.
-
Performance Characteristics
-
Analytical Performance
a. Precision
Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut
-
Page 8
off and +100% cut off. These samples were prepared by spiking fentanyl in negative samples. Each fentanyl concentration was confirmed by LC/MS-MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed six replicates per day for 10 days using three device lots in a randomized order. Each device was read on one Healgen Immunofluorescence analyzer. The results obtained are summarized in the following tables.
| Lot Number | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | 60-/0+ | 60-/0+ | 60-/0+ | 57-/3+ | 22-/38+ | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
| Lot 2 | 60-/0+ | 60-/0+ | 60-/0+ | 57-/3+ | 24-/36+ | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
| Lot 3 | 60-/0+ | 60-/0+ | 60-/0+ | 56-/4+ | 23-/37+ | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
c. Stability
The devices are stable at 2-30°C for 27 months based on the real time stability study.
d. Interference
Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drug fentanyl urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following tables.
| Acetaminophen | Doxepin | Nortriptyline |
| Acetone (1000 mg/dL) | Ecgonine methyl ester | Noscapine |
| Acetophenetidin | Ephedrine | O-Hydroxyhippuric acid |
| Acetylsalicylic acid | Erythromycin | Octopamine |
| Albumin (100mg/dL) | Ethanol (1%) | Oxalic acid (100mg/dL) |
| Albuterol | Fenoprofen | Oxazepam |
| Aminopyrine | Fluphenazine | Oxolinic acid |
| Amitriptyline | Furosemide | Oxymetazoline |
| Amobarbital | Galactose (10mg/dL) | Papaverine |
| Amoxicillin | Gamma Globulin (500mg/dL) | Penicillin G |
| Ampicillin | Gentisic acid | Perphenazine |
| Apomorphine | Glucose (3000mg/dL) | Phencyclidine |
| Ascorbic acid | Hemoglobin | Phenelzine |
| Aspartame | Hydralazine | Phenobarbital |
| Atropine | Hydrochlorothiazide | Prednisone |
| Benzilic acid | Hydrocortisone | Propoxyphene |
| Benzoic acid | Hydroxytyramine | Propranolol |
| Benzoylecgonine | Ibuprofen | Pseudoephedrine |
| Bilirubin | Imipramine | Quinine |
| Boric Acid (1%) | Isoproterenol | Ranitidine |
| Bupropion | Isoxsuprine | Riboflavin (10mg/dL) |
| Caffeine | Ketamine | Salicylic acid |
| Carbamazepine | Ketoprofen | Secobarbital |
| Chloral hydrate | Labetalol | Serotonin (5-Hydroxytyramine) |
| Chloramphenicol | Lidocaine | Sulfamethazine |
| Chlorothiazide | Loperamide | Sulindac |
| Chlorpromazine | Maprotiline | Tetrahydrocortisone 3-D-glucuronide) |
Page 9
| Cholesterol | Meperidine | Tetrahydrocortisone 3-acetate |
|---|---|---|
| Clomipramine | Meprobamate | Tetrahydrozoline |
| Clonidine | Methapyrilene | Thiamine |
| Cortisone | Methaqualone | Thioridazine |
| Cotinine | Methoxyphenamine | Triamterene |
| Creatinine | Metronidazole (300μg/mL) | Trifluoperazine |
| Cyclobenzaprine | N-Acetylprocainamide | Trimethoprim |
| Deoxycorticosterone | NaCl (4000mg/dL) | Tyramine |
| Desipramine | Nalidixic acid | Urea (2000mg/dL) |
| Dextromethorphan | Naloxone | Uric acid |
| Diclofenac | Naltrexone | Valproic acid (250μg/mL) |
| Diflunisal | Naproxen | Venlafaxine |
| Digoxin | Niacinamide | Verapamil |
| Diphenhydramine | Nicotine | Zomepirac |
| DL-Tryptophan | Nifedipine | β-Estradiol |
| DL-Tyrosine | Norethindrone |
e. Specificity
To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device. The lowest concentration that caused a positive result for each compound are listed below.
| Components | Concentration (ng/mL) | Cross- Reactivity (%) |
|---|---|---|
| Norfentanyl | 30,000 | 0.003% |
| Carfentanil | 8,000 | 0.013% |
| Sufentanil | 50,000 | 0.002% |
| Cyclopropyl fentanyl | 1 | 100% |
| Furanyl Fentanyl | 10 | 10% |
| Para-fluorobutyryl fentanyl | 10 | 10% |
| 4-Fluoro-isobutyrylfentanyl | 5 | 20% |
| O-Fluorofentanyl | 10 | 10% |
| 2'-Fluoro ortho-Fluorofenyanyl | 10 | 10% |
| Valeryl Fentanyl | 5 | 20% |
| (±) β-hydroxythiofentanyl | 3 | 33.33% |
| Tetrahydrofuranyl fentanyl | 1.56 | 64.10% |
| 2-Thiofuranyl fentanyl | 5 | 20% |
| Methoxyacetyl fentanyl | 1.56 | 64.10% |
| 4-methoxybutyryl fentanyl (para) | 20 | 5% |
| N-methyl norfentanyl | 20,000 | 0.005% |
| 3',4'-dimethoxy Fentanyl | 125 | 0.8% |
| Acetyl-α-methyl fentanyl | 62.5 | 1.6% |
| 4'-methyl acetyl fentanyl | 125 | 0.8% |
| Benzyl fentanyl | 125 | 0.8% |
| Meta-methoxy Furanyl fentanyl | 100 | 1% |
| α-methyl fentanyl | 62.5 | 1.6% |
| Para-fluoro fentanyl | 1 | 100% |
| Ocfentanil | 5 | 20% |
| Isobutyryl fentanyl | 2.5 | 40% |
| Butyryl fentanyl | 3 | 33.33% |
| Acetyl fentanyl | 1 | 100% |
Page 10
| Acrylfentanyl | 0.9 | 111.11% |
|---|---|---|
| Risperidone | 50,000 | 0.002% |
| 9-Hydroxyrisperidone | 10,000 | 0.01% |
| (±)-3-cis-methyl fentanyl | 50 | 2% |
| Despropionyl fentanyl (4-ANPP) | 7000 | 0.014% |
| ω-1-Hydroxyfentanyl | 50,000 | 0.002% |
| Acetyl norfentanyl | > 100 μg/mL | <0.001% |
| Norcarfentanil | > 100 μg/mL | <0.001% |
| Remifentanil | > 100 μg/mL | <0.001% |
| Alfentanil | > 100 μg/mL | <0.001% |
The following other Metabolites and opioids compounds were tested at a concentration of 100 µg/mL. Negative results were obtained for all these compounds. There is no cross-reactivity for these compounds using the Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit.
| 6-Acetyl morphine | Ketamine | Noroxycodone |
| Amphetamine | Levorphanol | Oxycodone |
| Buprenorphine | Meperidine | Oxymorphone |
| Buprenorphineglucuronide | Methadone | Pentazocine (Talwin) |
| Codeine | Morphine | Pipamperone |
| Dextromethorphan | Morphine-3-glucuronide | Trazodone |
| Dihydrocodeine | Naloxone | Buspirone |
| EDDP | Naltrexone | Tapentadol |
| EMDP | Norbuprenorphine | Thioridazine |
| Fluoxetine | Norcodeine | Tilidine |
| Heroin | Norketamine | Tramadol |
| Hydrocodone | Normeperidine | Tramadol-O- Desmethyl |
| Hydromorphone | Normorphine | Tramadol-N- Desmethyl |
f. Effect of Urine Specific Gravity and Urine pH
To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target fentanyl at 50% below and 50% above Cut-Off levels. These samples were tested by three different operators using three device lots. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.
- Comparison Studies
Method comparison studies for the Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit were performed at three different testing sites. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the tables below.
| Negative | Low Negative by LC-MS/MS (less than -50%) | Near Cutoff Negative by LC-MS/MS (Between -50% and cutoff) | Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%) | High Positive by LC-MS/MS (greater than +50%) | |
|---|---|---|---|---|---|
| Site 1 | |||||
| Positive | 0 | 0 | 3 | 21 | 17 |
| Negative | 7 | 19 | 11 | 2 | 0 |
Page 11
| Positive | 0 | 0 | 2 | 20 | 17 | |
|---|---|---|---|---|---|---|
| Site 2 | Negative | 7 | 19 | 12 | 3 | 0 |
| Site 3 | Positive | 0 | 0 | 2 | 20 | 17 |
| Negative | 7 | 19 | 12 | 3 | 0 |
Discordant Results
| Site | Sample Number | LC-MS/MS Result | Result |
|---|---|---|---|
| Site 1 | FENMCS079 | 0.802 | + |
| Site 2 | FENMCS070 | 0.841 | + |
| Site 3 | FENMCS036 | 0.916 | + |
| Site 1, 2 | FENMCS001 | 0.952 | + |
| Site 1, 3 | FENMCS066 | 0.976 | + |
| Site 2, 3 | FENMCS071 | 1.013 | - |
| Site 1, 2 | FENMCS051 | 1.092 | - |
| Site 2, 3 | FENMCS042 | 1.098 | - |
| Site 1, 3 | FENMCS028 | 1.113 | - |
- Clinical Studies
Not applicable.
- Conclusion
Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, and method comparison studies of the devices, it's concluded a substantial equivalence decision.
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).