K Number
K251893

Validate with FDA (Live)

Device Name
SKAN C PULSAR
Date Cleared
2025-12-16

(179 days)

Product Code
Regulation Number
892.1650
Age Range
0 - 150
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SKAN C Pulsar, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic images of patients during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.

SKAN C Pulsar is not recommended for Cardiac Applications.

SKAN C Pulsar surgical C-Arm is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.

Device Description

The SKAN C Pulsar, a Mobile C-Arm X-Ray System, is intended to provide Fluoroscopic images of patients during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.

SKAN C Pulsar is a Mobile fluoroscopy C-Arm consisting of two main units:
a) C-arm main unit
b) A Workstation or Monitor Cart

The C-arm unit is composed of an X-ray tube, a flat panel detector, a collimator, a generator, a touch panel, foot switch, hand switch and a Console. C-arm has provision for mechanical movement of C-arm for Orbital and Yoke Rotation along with vertical and wig-wag movements.

Workstation or Monitor cart is composed of a monitor, keyboard and computing system.

The operating principle of the device is to expose X-ray, which are passed through the human body and falls on the sensor. The intensity of X-ray can be adjusted to required level. Detector follows two step conversion. It converts X-ray into light and Light is converted into electrical signal. Electrical signal is than digitized and stored. This stored information is processed and displayed on the monitor. The displayed images can be saved or transmitted to an external storage device.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Skan C Pulsar

Page 1

December 16, 2025

Skanray Technologies Limited
℅ Vasundhara R
Head - Regulatory
Plot# 15-17, Hebbal Industrial Area
MYSORE, KARNATAKA 570016
INDIA

Re: K251893
Trade/Device Name: Skan C Pulsar
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-intensified fluoroscopic x-ray system
Regulatory Class: Class II
Product Code: OWB OXO JAA
Dated: June 12, 2025
Received: November 18, 2025

Dear Vasundhara R:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251893 - Vasundhara R Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251893 - Vasundhara R Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251893

Device Name: SKAN C Pulsar

Indications for Use (Describe)

The SKAN C Pulsar, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic images of patients during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.

SKAN C Pulsar is not recommended for Cardiac Applications.

SKAN C Pulsar surgical C-Arm is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) summary: Skan C Pulsar

510(k) number: K251893

In accordance with 21 CFR 807.92, following information is provided:

FieldInformation
Date18 Nov 2025
510(k) ownerSkanray Technologies LimitedPlot #15-17, Hebbal Industrial Area,Mysore, Karnataka – 570016, INDIA
SubmitterVasundhara RHead – Regulatory AffairsSkanray Technologies LimitedContact number: +91 821 2415559Email: vasundhara.r@skanray.com
Trade nameSKAN C PULSAR
Device classII
Device common nameInterventional Fluoroscopic X-Ray System
Regulation nameImage-intensified Fluoroscopic x-ray System
Regulation number21 CFR 892.1650
Product codeOWB

Page 6

FieldInformation
Subsequent product codeOXO, JAA
Primary Predicate deviceExtron 5, Extron 7510(k) Clearance: K230871Device class: IIDevice common name: Interventional Fluoroscopic X-Ray SystemRegulation description: Image Intensified Fluoroscopic X-Ray SystemRegulation number: 21 CFR 892.1650Manufacturer: DRTECH Corporation
Secondary Predicate deviceSkan C Mobile C-arm X-Ray System – 230V Variant (3030-000187-1), Skan C Mobile X-Ray system – 110V Variant (3030-00187-2)510(k) Clearance: K170946Device class: IIDevice common name: Image Intensified Fluoroscopic X-Ray SystemRegulation description: Image Intensified Fluoroscopic X-Ray SystemRegulation number: 21 CFR 892.1650Manufacturer: Skanray Technologies Limited
Reference device 1Digiscan FDX510(k) Clearance: K200218Device class: 2Device common name: Interventional Fluoroscopic X-Ray SystemRegulation description: Image Intensified Fluoroscopic X-Ray SystemRegulation number: 21 CFR 892.1650Manufacturer: Allengers Medical Systems Limited

Page 7

FieldInformation
Reference device 2Cios Alpha510(k) Clearance: K181560Device class: IIDevice common name: Interventional Fluoroscopic X-Ray SystemRegulation description: Image Intensified Fluoroscopic X-Ray SystemRegulation number: 21 CFR 892.1650Manufacturer: Siemens Medical Systems USA, Inc.
Device descriptionThe SKAN C Pulsar, a Mobile C-Arm X-Ray System, is intended to provide Fluoroscopic images of patients during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.SKAN C Pulsar is a Mobile fluoroscopy C-Arm consisting of two main units:a) C-arm main unitb) A Workstation or Monitor CartThe C-arm unit is composed of an X-ray tube, a flat panel detector, a collimator, a generator, a touch panel, foot switch, hand switch and a Console. C-arm has provision for mechanical movement of C-arm for Orbital and Yoke Rotation along with vertical and wig-wag movements.

Page 8

Workstation or Monitor cart is composed of a monitor, keyboard and computing system.

The operating principle of the device is to expose X-ray, which are passed through the human body and falls on the sensor. The intensity of X-ray can be adjusted to required level. Detector follows two step conversion. It converts X-ray into light and Light is converted into electrical signal. Electrical signal is than digitized and stored. This stored information is processed and displayed on the monitor. The displayed images can be saved or transmitted to an external storage device.

FieldInformation
Intended use / Indication for useThe SKAN C Pulsar, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic images of patients during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.SKAN C Pulsar is not recommended for Cardiac Applications.SKAN C Pulsar surgical C-Arm is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.

Primary and Secondary Comparable Properties Table

Predicate Device Manufacturer and Model510(k) NumberComparable Properties
Primary PredicateDRTECH Corporation: Extron 5, Extron 7K230871Indication for useSystem Image Acquisition and X-ray generationImage Acquisition and TransmissionDetector Technology
Secondary PredicateSkanray: Skan CK170946Indication for useConstructional features

Page 9

Predicate Device Manufacturer and Model510(k) NumberComparable Properties
Reference DeviceAllengers Medical Systems Limited: Digiscan FDXSimens Medical Systems USA: Cios AlphaK200218K181560Detector TechnologyDetector Technology

Selection Criteria for the Predicate Device

Valid Predicate DeviceWell Established methodMeets or exceeds expected predicate performanceUnmitigated use-related or design related safety issuesAssociated design related recall
K-183040Used relevant methods that were published in the public domainMeets the performanceNo known unmitigated use related of design related safety issuesFew Class II recalls made and were effectively addressed
K-200218Used relevant methods that were published in the public domainMeets the performanceNo known unmitigated use related of design related safety issuesNo design related recall identified
K-230871Used relevant methods that were published in the public domainMeets the PerformanceNo known unmitigated use related of design related safety issuesNo design related recall identified
K181560Used relevant methods that were published in the public domainMeets the PerformanceNo known unmitigated use related of design related safety issuesNo design related recall identified
K170946Used relevant methods that were published in the public domain and with Skanray internal dataMeets the PerformanceNo known unmitigated use related of design related safety issuesNo design related recall identified

Page 10

Discussion:

Primary Predicate Device 510k cleared (K230871), is considered as it is legally marketed device which is found to be consistent with IFU statements and device labeling. It has the same intended use when compared to subject device. Technological characteristics are similar.

Secondary Predicate Device K170946 from Skanray Technologies Ltd, considered as Secondary Substantial device, which has similar construction features when compared to subject device with the same intended use.

Skan C Pulsar incorporates a Flat Panel Technology, for which two more 510k cleared reference devices K200218 and K181560, are used to evaluate the safety and performance of the Detectors with similar technologies

Technological characteristics of the subject device in comparison to those of the predicate device:

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
ManufacturerSkanray Technologies LimitedDRTECH CorporationSkanray Technologies Limited--
510(k) CodeK251893K230871K170946--
ModelSkan C PulsarExtron 5 and Extron 7Skan C--
Regulatory DescriptionImage-intensified fluoroscopic x-ray systemImage-intensified fluoroscopic x-ray systemImage-intensified fluoroscopic x-ray systemSame
Device NameInterventional fluoroscopic x-ray systemInterventional fluoroscopic x-ray systemImage-intensified fluoroscopic x-ray system, mobileSubject and primary predicate device are same.
Regulatory ClassClass IIClass IIClass IISame

Page 11

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
Product CodeOWB, OXO, JAAOWB, OXO, JAAOXO, OWBSubject and primary predicate devices are same.
Intended useThe Skan C Pulsar, a Mobile Surgical C-arm X-ray system, is intended to provide Fluoroscopic images of patient during Diagnostic, Surgical and Interventional procedures. SKAN C Pulsar is to be used by adequately trained, qualified and authorized healthcare professionals. Clinical Applications may include Orthopedic, GI Procedure, Neurology, Urology Procedures, Vascular in Critical Care and Emergency Room Procedures.SKAN C Pulsar is not recommended for Cardiac Applications.SKAN C Pulsar is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.EXTRON5 and EXTRON 7 are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high-resolution image processing system, which provides X-ray images of the patient's anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific image as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis. Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine (※ This device is not intended for mammography applications.)The Skan-C, a Mobile Surgical C-Arm X-Ray System, is intended to provide Fluoroscopic and Radiographic images of the patient during Diagnostic, Surgical and Interventional procedures. Examples of Clinical Applications may include Orthopaedic, GI Procedure like Endoscopy and Cholenography, Neurology, Urology Procedures, Vascular, Critical Care and Emergency Room Procedures. Skan-C is not recommended for Cardiac Applications. Skan-C Surgical C-Arm is indicated for visualization in real time and/or recording of surgical region of interest and anatomy, using X-ray imaging technique.Similar

Page 12

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
Target PopulationPatients who need X-Ray fluoroscopy images including Adults and pediatricsPatients who need X-Ray fluoroscopy images including Adults and pediatricsPatients who need X-Ray fluoroscopy images including Adults and pediatricsSame
System and Sub-system DesignDevice Major Subsystems are:X-Ray Generator with mono-block designFlat Panel DetectorC-Arm Mechanical systemC-Arm Orbital and Rotation, Horizontal travel, Wig-Wag movement and Vertical C-movement.User Console at C side for operator to useUser Console at Workstation to display the captured x-ray image on the screenBoth C-Arm and Workstation are mobileDevice Major Subsystems are:X-Ray Generator with mono-block designFlat Panel DetectorC-Arm Mechanical systemC-Arm Orbital and Rotation, Horizontal travel, Wig-Wag movement and Vertical C-movement.User Console at C side for operator to useUser Console at Workstation to display the captured x-ray image on the screenBoth C-Arm and Workstation are mobileDevice Major Subsystems are:X-Ray Generator with mono-block designFlat Panel DetectorC-Arm Mechanical systemC-Arm Orbital and Rotation, Horizontal travel, Wig-Wag movement and Vertical C-movement.User Console at C side for operator to useUser Console at Workstation to display the captured x-ray image on the screenBoth C-Arm and Workstation are mobileSame
Power OutputMax 6kWExtron 5: Max 4.8kWExtron 7: Max 15kWMax 2.2kW / 3.5kWSimilar:The specifications have been enhanced to enable operation at power levels up to 6kW, which is comparable to EXTRON 5 (primary predicate device) and Skan C (secondary predicate device). This does not give rise to any novel concerns regarding safety and effectiveness.

Page 13

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
X-Ray TubeRotating AnodeRotating AnodeStationary AnodeSame: Both Subject device and Primary Predicate device make use of Rotating Anode X-Ray tube technology.Similar: Secondary Predicate device is lower powered x-ray system making use of stationary anode x-ray tube.
Fluoroscopic Mode Parameters range1. kV Range: 40 – 120kV2. mA Range: Max 60 mA1. kV Range: 40 – 120kV2. mA Range: Max 60mA (EXTRON 7) and Max 40mA (Extron 5)1. kV Range: 40 – 110kV2. mA Range: Max 12 mASimilar:Primary Predicate device and subject device have the same range of kV and mA.Secondary Predicate device parameters are limited to 110kV and 12mA due to X-Ray tube type and low powered device.
Radiographic Mode Parameters rangeNot ApplicableEXTRON 5: Not ApplicableEXTRON 7: kV Range : 40 to 120kVmA Range : Up to 150mA1. kV Range: 40-110kV2. mAs Range: Max 200 mAsSimilar:Radiographic Mode is not provided in both Primary Predicate EXTRON 5 and Subject device with comparable power outputs.Secondary predicate device has lower power Radiographic output.Since the intended function is to provide Fluoroscopy x-ray system, Radiographic mode is deemed as a feature.This does not give rise to any novel safety concerns, as in both cases, single image is displayed and recorded.

Page 14

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
DimensionImmersion Depth: 74 cmImmersion Depth: 74 cmImmersion Depth: 65 cmSimilar:Immersion depth is the distance from the beam center to the C section surface used to reach the far side of the human body.This is same as Primary Predicate device.Better when compared to the secondary predicate device.Alteration in the dimension does not give rise to any novel concerns regarding safety and effectiveness.
DimensionFree Space: 82.5 cmFree Space: 80 cmFree Space 75 cmSimilar:Free Space is the space available between Source and the detector to accommodate Human body for the purpose of fluoroscopy exposure during the procedure.This is same when compared to Primary Predicate Device.Slightly better than when compared to Secondary Predicate Device.Minor alteration in the dimension does not give rise to any novel concerns regarding safety and effectiveness.
DimensionOrbital Movement: Total: 147ºOrbital Movement: Total: 165ºOrbital Movement: Total: 125ºSimilar:Orbital Rotation is the rotation around the patient in a plane perpendicular to c-arm moving in an arc like path. This facilitates capturing images from different arc angles for convenience of positioning.Minor changes in Orbital movement does not give rise to any novel concerns regarding safety and effectiveness.

Page 15

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
Exposure SwitchWired Foot switch and Hand SwitchWired and Wireless Foot switch and wired Hand SwitchWired Foot switch and Hand SwitchSimilar except for wireless option.Predicate devices and subject devices have same wired Foot Switch and Hand Switch.Wireless Hand Switch and Hand Switch are additional features for convenience doing the same intended function of wired one.This does not give rise to any novel concern over the Safety and Performance of the device.
Laser GuideYesYesYesSame
CollimatorAvailable, Rectangular CollimationAvailable, Rectangular CollimationAvailable, Rectangular with Circular CollimationSame
Deployment MethodUsed for visualizing internal organ of the body.Used inside the procedure room with controlled environmentUsed for visualizing internal organ of the body.Used inside the procedure room with controlled environmentUsed for visualizing internal organ of the body.Used inside the procedure room with controlled environmentSame

Page 16

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
Operating PrincipleThe operating principle of the device is designed to expose the patient to X-ray beams. The range of X-ray irradiation are adjusted by the collimator.X-rays can penetrate into the human and hits the detector surface.X-ray photons are converted into light. The light is then converted into electrical signals through the sensor. The electrical charges are transmitted as the sensor output and converted into signals. These signals are digitized and captured by memory. The captured images are processed and displayed on the monitor.The operating principle of the device is designed to expose the patient to X-ray beams. The range of X-ray irradiation are adjusted by the collimator.X-rays can penetrate into the human body through a two-step conversion process.X-ray photons are converted into light. The light is then converted into electrical signals through the sensor. The electrical charges are transmitted as the sensor output and converted into signals. These signals are digitized and captured by memory. The captured images are processed and displayed on the monitor.The operating principle of the device is designed to expose the patient to X-ray beams. The range of X-ray irradiation are adjusted by the collimator.X-rays can penetrate into the human body and hits the detector surface.X-ray photons are converted into light. The light is then converted into electrical signals through the sensor. The electrical charges are transmitted as the sensor output and converted into signals. These signals are digitized and captured by memory. The captured images are processed and displayed on the monitor.Same
Detector TechnologyFlat Panel DetectorFlat Panel DetectorNot ApplicableSame

Page 17

ParameterSubject DevicePrimary PredicateSecondary PredicateRemark
Detector Details9" Panels:CMOS panel: Pixel matrix: 1416 x 1416ASi panel: Pixel matrix 1024 x 1024IGZO Panel: Pixel matrix 1536 x 153612" Panels:CMOS panel: Pixel matrix 1943 x 1943ASi panel: Pixel matrix 1536 x 1536IGZO Panel: Pixel matrix 2048 x 20489" PanelSize: 210 x 210 mmIGZO Panel: EXPD 2121PPixel Matrix 1500 x 1500DQE 70% @ 0.5lp/mmMTF 60% @ 1lp/mmPixel Pitch : 140umGray Scale: 16 bit12" PanelIGZO Panel: EXPD 3030PPixel matrix 2048 x 2048Not ApplicableIGZO Detector: Similar Technology. Slight difference in pixel size and size do not impact safety and performance.CMOS and Amorphous Silicon (AsI) technology and specification are compared with reference devices mentioned in subsequent table.

Page 18

Submission of 510(k) for Solid State X-Ray Imaging Devices

Submission of 510(k) for Solid State X-Ray Imaging Devices includes the evaluation for detectors specified as per the following performance specifications, identified in the SSXI guidance, showing comparable performance of Skan C Pulsar to its reference devise. Following table indicates comparative information.

Details of Comparable specifications with cleared detector's specifications.

Subject DeviceReference Device 1: K200218Reference Device 2: K181560Remarks
Device Classification NameImage-intensified fluoroscopic x-ray systemImage-intensified fluoroscopic x-ray systemImage-intensified fluoroscopic x-ray system
Primary Product CodeOWBOWBOWB
ClassificationIIIIII
Detector Sizes / Variants20 cm x 20 cm (8inch) aSi (amorphous silicon TFT technology)Size: 20.5 cm x 20.5 cmDetector matrix: 1,024 x 1,024 pixelsDQE> 75%MTF>50%@1lp/mmPixel size 200μmGray scale 16bit20 cm x 20 cm (8inch) aSi (amorphous silicon TFT technology)Size: 20.5 cm x 20.5 cmDetector matrix: 1,024 x 1,024 pixelsDQE: 78%MTF>50%@1lp/mmPixel size 200μmGray scale 16bit--
20 cm x 20 cm (8inch) (CMOS TFT technology)Size: 21.5 cm x 21.5 cmDetector matrix: 1,416 1,416 pixelsDQE: 70%MTF>50%@1lp/mmPixel size 151.8 umGray scale 14bit--20 cm x 20 cm (8inch) (CMOS TFT technology)Size: 21.5 cm x 21.5 cmDetector matrix: 1,416 1,416 pixelsDQE: 70%MTF>50%@1lp/mmPixel size 151.8 umGray scale 14bit

Page 19

Subject DeviceReference Device 1: K200218Reference Device 2: K181560Remarks
Detector Sizes / Variants30 cm x 30 cm (12inch) aSi (amorphous silicon TFT technology)Size: 30 cm x 30 cmDetector matrix: 1536 x 1536 pixelsDQE > 70%MTF > 50%@1lp/mmPixel size 200μmGray scale 16bit30 cm x 30 cm (12inch) aSi (amorphous silicon TFT technology)Size: 30 cm x 30 cmDetector matrix: 1536 x 1536 pixelsDQE > 70%MTF > 50%@1lp/mmPixel size 200μmGray scale 16bit--
30 cm x 30 cm (12inch) (CMOS TFT technology)Size: 29.5 cm x 29.5 cmDetector matrix: 1943 x 1943 pixelsDQE: 70%MTF > 50%@1lp/mmPixel size 151.8 μmGray scale 14bit--30 cm x 30 cm (12inch) (CMOS TFT technology)Size: 29.5 cm x 29.5 cmDetector matrix: 1943 x 1943 pixelsDQE: 70%MTF > 50%@1lp/mmPixel size 151.8 μmGray scale 14bit

Page 20

Subject DeviceReference Device 1: K200218Reference Device 2: K181560Remarks
Detector Sizes / Variants30 cm x 30 cm (12inch) aSi (amorphous silicon TFT technology)Size: 30 cm x 30 cmDetector matrix: 1536 x 1536 pixelsDQE > 70%MTF > 50%@1lp/mmPixel size 200μmGray scale 16bit30 cm x 30 cm (12inch) aSi (amorphous silicon TFT technology)Size: 30 cm x 30 cmDetector matrix: 1536 x 1536 pixelsDQE > 70%MTF > 50%@1lp/mmPixel size 200μmGray scale 16bit--
30 cm x 30 cm (12inch) (CMOS TFT technology)Size: 29.5 cm x 29.5 cmDetector matrix: 1943 x 1943 pixelsDQE: 70%MTF > 50%@1lp/mmPixel size 151.8 μmGray scale 14bit--30 cm x 30 cm (12inch) (CMOS TFT technology)Size: 29.5 cm x 29.5 cmDetector matrix: 1943 x 1943 pixelsDQE: 70%MTF > 50%@1lp/mmPixel size 151.8 μmGray scale 14bit

Page 21

Similar Detector/Technology Models used in 510(k) cleared devices.

Similar Detector Models used in 510(k) cleared device/s

Detector TFT TechnologySubject Detector Skan C PulsarK181560 Cios AlphaK200218 Digiscan FDXK230871 Extron 5/7
CMOSXineos-2222HSXineos-3030HSXineos-2222HSXineos-3030HS
aSiPixium 2121S-AUPixium 3030S-AUPixium 2121S-AUPixium 3030S-AU

Similar Detector Technology used in 510(k) cleared device/s

| IGZO | FXDD-2323GFXDD-3131G | | | EXPD 2121PEXPD 3030P |

Discussion of similarities and differences:

The Subject Device, Skan C Pulsar, and its Predicate Devices have same Intended Use of providing fluoroscopic images of patients during diagnostic, surgical and interventional procedure or surgery or treatment.

The Subject Device and Predicate devices are comparable in terms of mobility, patient positioning, C-arm housing X-ray source and detector, a workstation to view the images. Skan C Pulsar provides display of images, saving patient data and transmitting images over the PACS system are same.

Skan C Pulsar and its predicates share similar system and sub-system design and operating principles. Minor differences like power output with predicate device shall not impact any safety concern nor usage of the device for similar applications.

Page 22

Detector technology, which is comparable to already 510k cleared devices with minor differences in pixels and sizes, are addressed in the FDA Solid State Device Guidance Evaluation and found to be not impacting any safety or performance of Skan C Pulsar. Where applicable as in IGZO detectors, were evaluated under the clinical expert opinion method, to demonstrate adequate image quality for clinical purposes.

Overall, with the above analysis, it is deemed acceptable that the Skan C Pulsar is substantially equivalent to the predicate device in terms of its intended use, safety and effectiveness.

Performance Data:

SKAN C PULSAR is designed and manufactured under the established quality management system in compliance with the following international and FDA recognized consensus standards and FDA guidance document. Risk analysis, design requirement review, design verification, design validation was conducted. Predefined acceptance criteria were met, and the subject device is safe and effective as the predicate device.

Summary of non-clinical tests:

Testing is carried out for SKAN C PULSAR and it is meeting the applicable requirements of the standards mentioned below:

  1. IEC 60601-1 Edition 3.2 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  2. IEC 60601-1-2 Edition 4.1 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard
  3. IEC 60601-1-3 Edition 2.2 - Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
  4. IEC 60601-1-6 Edition 3.2 - Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  5. IEC 60601-2-43 Edition 3.0 - Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures
  6. IEC 60601-2-54 Edition 2.0 - Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
  7. IEC 60601-2-28 Edition 3.0 2017-06 - Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis

Page 23

  1. IEC 62304 Edition 1.1 - Medical device software - Software life cycle processes
  2. IEC 62366-1 Edition 1.1 - Medical devices - Part 1: Application of usability engineering to medical devices
  3. IEC TS 60601-4-2 Edition 1.0 - Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
  4. ISO 14971 Edition 3 - Medical devices - Application of risk management to medical devices
  5. NEMA XR 27 Amendment 1-2013 - X-ray equipment for interventional procedures - User Quality Control Mode
  6. IEC 61910-1 Edition 1.0 2014-09 - Medical electrical equipment - Radiation dose documentation - Part 1: Radiation dose structured reports for radiography and radioscopy

Following FDA guidance documents were used in this premarket notification submission:

  1. The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff, 2014
  2. Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff, 2016
  3. Post market Management of Cybersecurity in Medical Devices: Guidance for Industry and Food and Drug Administration Staff, 2016
  4. Pediatric Information for X-ray Imaging Device Premarket Notifications: Guidance for Industry and Food and Drug Administration Staff, 2017
  5. Electromagnetic Compatibility (EMC) of Medical Devices: Guidance for Industry and Food and Drug Administration Staff, 2022
  6. Content of Premarket Submissions for Device Software Functions: Guidance for Industry and Food and Drug Administration Staff, 2023
  7. Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: Guidance for Industry and Food and Drug Administration Staff, 2023

The proposed Skan C Pulsar demonstrated substantial equivalence to the currently marketed and predicate device DR Tech Corporation, Model Extron 5 / Extron 7 was demonstrated with the following attributes:

  • Indications for use.
  • Technological characteristics.
  • Non-clinical performance testing; and

Page 24

  • Safety and effectiveness.

Furthermore, on the Solid-State Detector evaluation with reference to similar technology used in currently 510(k) cleared devices are mentioned below.

CMOS detectors used Xineos-2222HS and Xineos-3030HS utilizes the same design and technology used in currently marketed and cleared device "Cios Alpha" with 510(k) number K181560.

Amorphous Silicon (AsI) detectors Pixium 2121S-AU and Pixium 3030S-AU utilizes the same design and technology with minor changes in Pixel and panel size used in the currently marketed and cleared device is being used in "Digiscan FDX" with 510(k) approval K200218.

IGZO detectors FXDD-2323G and FXDD-3131G utilizes similar design and technology with minor changes in size and specifications used in the currently marketed and 510(k) cleared devices like DRTech Corporation, Extron5/Extron7 with 510(k) number K230871.

All technical detector characteristics that potentially have an influence on image quality are assessed and verified according to "FDA Guidance for Industry and Food and Drug Administration Staff: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices"; issued on September 1, 2016.

For the detectors using IGZO technology FXDD-2323G and FXDD-3131G are further evaluated through clinical considerations under the qualified expert opinion methodology to substantiate the image quality, safety and performance of Skan C Pulsar.

Conclusion:

Based on the above details, SKAN C PULSAR is safe and effective when the device is used as labelled and is substantially equivalent to the currently marketed predicate device Extron 5.

N/A