K Number
K251854

Validate with FDA (Live)

Date Cleared
2025-08-06

(50 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SteadiSet infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The infusion set is indicated for single use.

Device Description

The subject device, SteadiSet infusion set, is a sterile, non-pyrogenic, intravascular administration set device used to administer insulin from a reservoir cartridge to a patient subcutaneously through a cannula. The infusion set administers insulin by means of a compatible external pump.

The infusion set consists of an inserter, tube set, and disconnect cover. The inserter consists of a housing, insertion buttons, an infusion set hub (with cannula) and adhesive patch with protective liner. The inserter facilitates insertion of the cannula subcutaneously. The cannula is a soft medical-grade polymer extruded over a stainless-steel coil.

The tube set provides the insulin pathway between the hub's indwelling cannula and an external insulin pump cartridge. The tube set consists of infusion set tubing with a reservoir connector (proximal end) and hub connector (distal end).

The disconnect cover can be connected to the hub to provide cover when the infusion set tubing is disconnected from the hub.

The device is sterilized by Ethylene Oxide (ETO) and is a single-patient, single-use device.

AI/ML Overview

This FDA 510(k) clearance letter pertains to a medical device, the SteadiSet infusion set, not an AI or software as a medical device (SaMD). Therefore, many of the requested elements for AI/SaMD studies (such as sample sizes for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable or not found within this type of documentation for a physical infusion set.

However, I can extract the relevant information regarding the clinical trial conducted for the device.


Acceptance Criteria and Study for SteadiSet Infusion Set

Based on the provided FDA 510(k) clearance letter for the SteadiSet infusion set, the "acceptance criteria" are not explicitly defined as specific numerical thresholds for performance metrics (such as sensitivity or specificity) in the way they would be for an AI model. Instead, the "acceptance criteria" are implied by the overall demonstration of safety and effectiveness as required for substantial equivalence to a predicate device.

The study that proves the device meets (these implied) acceptance criteria is a clinical study.

1. Table of Acceptance Criteria and Reported Device Performance

Performance Characteristic/Acceptance Criteria (Implied)Reported Device Performance (from Clinical Study)
Safe for a maximum of 7 days of useNo device-related serious adverse events reported.
Effective for a maximum of 7 days of useDemonstrated effectiveness for insulin infusion for a maximum of 7 days of use.
Substantially Equivalent to Predicate Device (K242692)Concluded to be substantially equivalent (SE) to the predicate device, despite some technological differences, as these differences "do not raise new questions of safety and effectiveness."

2. Sample Size and Data Provenance for Test Set (Clinical Study)

  • Sample Size: A total of 260 subjects were enrolled.
  • Data Provenance: The clinical study was conducted in the US at 15 investigational centers. This indicates prospective, real-world data collection in a clinical setting.

3. Number of Experts and Qualifications for Ground Truth

  • Not Applicable. For a physical medical device like an infusion set, "ground truth" is typically established through direct clinical observation of patient outcomes (safety and effectiveness) rather than expert consensus on diagnostic images or data. The assessment of safety and effectiveness would have been based on clinical endpoints and observed adverse events, which are measured directly.

4. Adjudication Method for Test Set

  • Not Applicable. Clinical trial outcomes (adverse events, device function) are typically evaluated by the investigational site staff and potentially reviewed by a clinical events committee, but the concept of "adjudication method" as applied to expert consensus for AI ground truth does not directly apply here.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This type of study is relevant for diagnostic imaging AI, comparing human reader performance with and without AI assistance. It is not applicable to a physical infusion set.

6. Standalone Performance Study

  • No, not in the context of an algorithm. The clinical study did evaluate the standalone performance of the device (the infusion set itself) in 260 subjects. However, this definition of standalone performance does not refer to an algorithm without human-in-the-loop, as the device is a physical product directly interacting with the patient.

7. Type of Ground Truth Used

  • Clinical Outcomes Data: The ground truth was established by direct observation of clinical outcomes, specifically the absence of device-related serious adverse events and the demonstration of effective insulin infusion over the study period (maximum of 7 days).

8. Sample Size for Training Set

  • Not Applicable. This device is a physical product, not an AI model requiring a training set.

9. How Ground Truth for Training Set Was Established

  • Not Applicable. As there is no AI model or training set, this question is not relevant.

FDA 510(k) Clearance Letter - SteadiSet Infusion Set

Page 1

August 6, 2025

Tandem Diabetes Care
Karen Mudd
Director, Regulatory Affairs
12400 High Bluff Drive
San Diego, California 92130

Re: K251854
Trade/Device Name: SteadiSet infusion set
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular Administration Set
Regulatory Class: Class II
Product Code: FPA
Dated: June 16, 2025
Received: June 17, 2025

Dear Karen Mudd:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251854 - Karen Mudd Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K251854 - Karen Mudd Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JOSHUA BALSAM -S

Joshua M. Balsam, Ph.D.
Branch Chief
Division of Chemistry and
Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251854

Device Name: SteadiSet infusion set

Indications for Use (Describe):
The SteadiSet infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The infusion set is indicated for single use.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Date: June 16, 2025

I. SUBMITTER

Address: Tandem Diabetes Care, Inc.
12400 High Bluff Drive
San Diego, California 92130 USA
+1 (858) 366-6900

Contact Person: Karen Mudd, PhD
Director, Regulatory Affairs
Phone: +1 (858) 366-6900

II. DEVICE

Proprietary / Trade Name: SteadiSet infusion set
Common Name: Infusion set
Classification Name: Intravascular administration set
Classification Regulation: 21 CFR 880.5440
Regulatory Class: Class II
Product Code: FPA

III. PREDICATE DEVICE

Predicate Device: K242692, SteadiSet infusion set

IV. DEVICE BRIEF DESCRIPTION

The subject device, SteadiSet infusion set, is a sterile, non-pyrogenic, intravascular administration set device used to administer insulin from a reservoir cartridge to a patient subcutaneously through a cannula. The infusion set administers insulin by means of a compatible external pump.

The infusion set consists of an inserter, tube set, and disconnect cover. The inserter consists of a housing, insertion buttons, an infusion set hub (with cannula) and adhesive patch with protective liner. The inserter facilitates insertion of the cannula subcutaneously. The cannula is a soft medical-grade polymer extruded over a stainless-steel coil.

The tube set provides the insulin pathway between the hub's indwelling cannula and an external insulin pump cartridge. The tube set consists of infusion set tubing with a reservoir connector (proximal end) and hub connector (distal end).

The disconnect cover can be connected to the hub to provide cover when the infusion set tubing is disconnected from the hub.

The device is sterilized by Ethylene Oxide (ETO) and is a single-patient, single-use device.

V. INDICATIONS FOR USE

The SteadiSet infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The infusion set is indicated for single use.

Page 6

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

ItemPredicate Device K242692Subject Device
Product CodeFPASame
Regulation Number21 CFR 880.5440Same
Indication for Use/Intended UseThe SteadiSet infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump. The infusion set is indicated for single useSame
Use TypeSingle UseSame
Time of Use3 daysMaximum of 7 days
Compatible DevicesTandem cartridges featuring the t:lock™ connectorSame
SterilizationEthylene Oxide (ETO)Same
Material Composition
Metallic ComponentsYesSame
PVC PlasticizersNoSame
AdditivesColorant PigmentsSame
Cannula MaterialPolyether amide TPE & Stainless-Steel coil CannulaSame
Physical Specifications
DimensionsID 0.38mmSame
OD 1.50mmSame
Tube Length5, 23, 32, and 43 inchesSame
Cannula Length13mmSame
Priming Volume≤0.2 mLSame
End ConfigurationDistal Hub ConnectorSame
Proximal Reservoir ConnectorSame
Connector TypeDistal Rotational hub connectorSame
Proximal t:lock reservoir connectorSame
Tubing ColorNo colorant Clear/transparentSame

VII. NON-CLINICAL PERFORMANCE DATA

Performance CharacteristicTesting Performed
Bench TestingInsertion Force and Depth
Strength of Materials, Joints, and Connectors
Tubing and Cannula Priming

Page 7

Performance CharacteristicTesting Performed
Tubing Elongation
Functional Performance
Tandem Pump Compatibility
Corrosion Testing
Insulin Compatibility
Usability/Human FactorsSimulated-Use Human Factors Validation
BiocompatibilityCytotoxicity
Sensitization
Irritation
Acute Systemic Toxicity
Material Mediated Pyrogenicity
Subacute Toxicity
Genotoxicity
Implantation
Sub-chronic Toxicity
Chemical Characterization (Extractables and Leachables)
Physicochemical, Morphological, And Topographical (PMT)
EO Residuals
Sterilization and ShippingSterility Assurance Level 10⁻⁶
Shipping, Shelf-Life, and Aging

VIII. CLINICAL PERFORMANCE DATA

A clinical study has been conducted in the US. A total of 260 subjects were enrolled at 15 investigational centers. The results of the clinical study demonstrated that the use of the SteadiSet infusion set was safe and effective for a maximum of 7 days of use. There were no device-related serious adverse events.

IX. SUBSTANTIAL EQUIVALENCE DISCUSSION SUMMARY

Guidance document, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] July 28, 2014, was consulted in determining substantial equivalence between the subject and predicate device. The predicate device is legally marketed. The predicate device labeling has been reviewed and is consistent with IFU statements for the subject device. The devices have the same intended use. As shown in comparison table above, the design, materials, energy source and other features of the devices have been reviewed. There are differences noted between the two devices, indicating that the devices do not have the same technological characteristics as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A). The different technological characteristics of the devices do not raise new questions of safety and effectiveness.

X. CONCLUSIONS

Per the comparison and analysis above, the subject device is determined to be substantially equivalent (SE) to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.