K Number
K251579
Manufacturer
Date Cleared
2025-08-21

(90 days)

Product Code
Regulation Number
866.1645
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:

Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)

The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram- negative bacilli to antimicrobial agents when used as instructed.

Device Description

The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach (2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3).

Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

VITEK® 2 AST-GN Cefazolin has the following concentrations in the card: 1, 2, and 8 (equivalent standard method concentration by efficacy in µg/mL).

AI/ML Overview

The provided text describes the acceptance criteria and a study proving the device meets these criteria for the VITEK 2 AST-Gram Negative Cefazolin antimicrobial susceptibility testing system.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Generally Implied for AST Systems based on FDA Guidance)Reported Device Performance (VITEK® 2 AST-GN Cefazolin)
Essential Agreement (EA)Typically ≥ 90% (agreement between the MIC from the test device and the reference method, within a +/- 1 doubling dilution)97.5% (840/862)
Category Agreement (CA)Typically ≥ 90% (agreement between the interpretive category generated by the test device and the reference method)86.8% (748/862)
Very Major Errors (VME)As low as possible; specific limits usually apply (e.g., <1.5% and NME ≤30%)0.3% (1/299)
Major Errors (ME)As low as possible; specific limits usually apply (e.g., <3.0% and NME ≤30%)0.7% (3/414)
Minor Errors (mE)Not explicitly stated as an acceptance criterion percentage, but high mE often impacts CA.12.8% (110/862)
ReproducibilityTypically ≥ 95%100%

Note on Acceptance Criteria: The document references the "FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)" for defining acceptable performance. While specific numerical acceptance criteria (e.g., for EA, CA, VME, ME) are not explicitly listed in the provided text, they are implied by the reported performance metrics and the statement that the device "demonstrated substantially equivalent performance." For AST systems, the FDA typically looks for high Essential and Category Agreement, with very low rates of Major and Very Major Errors. The low Category Agreement (86.8%) is acknowledged and attributed mainly to minor errors.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Total isolates tested: 862 (derived from the denominator for EA and CA calculations).
    • The study included "fresh and stock clinical isolates" as well as "a set of challenge strains."
  • Data Provenance:
    • The study was an "external evaluation." This typically implies data collected from multiple sites outside of the manufacturer's primary lab.
    • The document does not specify the country of origin for the data.
    • The study included both "fresh" and "stock" clinical isolates, suggesting a mix of prospective (fresh isolates) and retrospective (stock isolates) data collection.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number or qualifications of experts used to establish the ground truth.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set. For AST devices, the "ground truth" is typically established by a reference method (broth microdilution in this case), not by expert consensus or adjudication in the way it might be for an imaging AI device.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Improvement

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. This type of study (human readers assisting vs. not assisting) is typically relevant for interpretative diagnostic devices, especially in imaging, where human interpretation is a key part of the workflow. The VITEK 2 AST system is an automated device for determining antimicrobial susceptibility, not directly for human interpretation or image reading.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the primary performance study presented is a standalone (algorithm only) performance study. The device's performance (VITEK 2 AST-GN Cefazolin) was compared directly against the CLSI broth microdilution reference method. There is no mention of human-in-the-loop performance evaluation, as the VITEK 2 system is automated.

7. The Type of Ground Truth Used

The type of ground truth used was: Reference Method Comparison.
Specifically, the performance of the VITEK 2 AST-GN Cefazolin was compared to the CLSI broth microdilution reference method, incubated for 16-20 hours. This is the gold standard for antimicrobial susceptibility testing.

8. The Sample Size for the Training Set

The document does not specify the sample size for the training set. This submission is for a 510(k) clearance, which typically focuses on the performance of the final device rather than the specifics of its development (like training data for AI/ML models). The VITEK 2 system relies on growth pattern analysis algorithms rather than deep learning models that require large, labeled training datasets in the conventional sense. Its "training" would likely involve optimizing these algorithms based on extensive internal validation.

9. How the Ground Truth for the Training Set Was Established

The document does not describe how the ground truth for the training set was established, as details on the development of the device's analysis algorithms (likely proprietary) are not typically included in a 510(k) summary. Given the nature of AST systems, any "ground truth" used during development would also likely stem from comparisons to established reference methods like broth microdilution, similar to the test set's ground truth.

FDA 510(k) Clearance Letter - VITEK 2 AST-Gram Negative Cefazolin

Page 1

August 21, 2025

bioMerieux, Inc.
Jared Bronson
Regulatory Specialist
595 Anglum Rd.
Hazelwood, Missouri 63042

Re: K251579
Trade/Device Name: VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL)
Regulation Number: 21 CFR 866.1645
Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
Regulatory Class: Class II
Product Code: LON
Dated: May 22, 2025
Received: May 23, 2025

Dear Jared Bronson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251579 - Jared Bronson Page 2

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these

Page 3

K251579 - Jared Bronson Page 3

requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM)
Chief
General Bacteriology and Antimicrobial Susceptibility Branch
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251579

Device Name: VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL)

Indications for Use (Describe)

VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:

Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)

The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram- negative bacilli to antimicrobial agents when used as instructed.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) SUMMARY

VITEK® 2 AST-Gram Negative Cefazolin

A. 510(k) Submission Information:

Submitter's Name: bioMérieux, Inc.
Address: 595 Anglum Road
Hazelwood, MO 63042
Contact Person: Jared Bronson
Regulatory Affairs Specialist
Phone Number: 314 -731-8500
Fax Number: 314-731-8689
Date of Preparation: August 21, 2025

B. Device Name:

Formal/Trade Name: VITEK® 2 AST-Gram Negative Cefazolin (≤ 1– ≥32 µg/mL)
Classification Name: 21 CFR 866.1645
Fully Automated Short-Term Incubation Cycle
Antimicrobial Susceptibility System
Product Code: LON
Common Name: VITEK® 2 AST-GN Cefazolin

C. Predicate Device:

VITEK® 2 AST-Gram Negative Cefazolin (K222073)

D. Device Description:

The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach (2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3).

Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the

Page 6

incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

VITEK® 2 AST-GN Cefazolin has the following concentrations in the card: 1, 2, and 8 (equivalent standard method concentration by efficacy in µg/mL).

E. Substantial Equivalence Information

The similarities and differences of the VITEK® 2 AST-GN Cefazolin when compared to the predicate device, VITEK® 2 AST-GN Cefazolin (K222073), are described in the following table. The only difference between both devices are the Indications for Use. The below table provides the similarities and differences:

ItemDevice: VITEK® 2 AST-GN CefazolinPredicate: VITEK® 2 AST-GN Cefazolin (K222073)
Similarities
Intended UseVITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).The VITEK® 2 Gram Negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram-negative bacilli to antimicrobial agents when used as instructed.VITEK® 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of gram-negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.The VITEK® 2 Gram Negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram-negative bacilli to antimicrobial agents when used as instructed.

Page 7

Test MethodologyAutomated antimicrobial susceptibility test for use with the VITEK® 2 and VITEK® 2 Compact Systems to determine the in vitro susceptibility of microorganismsSame
InoculumSaline suspension of organismSame
Test CardGram Negative (AST-GN) Susceptibility Test CardSame
Analysis AlgorithmsGrowth Pattern AnalysisSame
Antimicrobial AgentCefazolinSame
Antimicrobial Concentrations1, 2, 8Same
Breakpoints for EnterobacteralesEnterobacterales: ≤2 (S), 4 (I), ≥8 (R)Same
Differences
Indications for UseVITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)Cefazolin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobialActive in vitro and in clinical infections:Escherichia coliProteus mirabilis
Instruments*VITEK® 2, VITEK® 2 Compact, and VITEK® 2 Compact Pro SystemsVITEK® 2 and VITEK® 2 Compact Systems

*At the time of the original clearance (K222073), the VITEK 2 AST-GN Cefazolin card (cz05n) was not intended to be used on the VITEK2 COMPACT PRO Instrument because the instrument was not yet available. The current clearance (K251579) of the VITEK 2 AST-GN Cefazolin card (cz05n) allows testing with the VITEK2 COMPACT PRO Instrument which was cleared in K234012.

F. Intended Use:

VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

Page 8

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).

VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:

Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)

The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram- negative bacilli to antimicrobial agents when used as instructed.

G. Performance Overview:

VITEK® 2 AST-GN Cefazolin demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009).

The Premarket Notification (Traditional 510[k]) presents data in support of VITEK® 2 AST-GN Cefazolin. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-GN Cefazolin by comparing its performance with the CLSI broth microdilution reference method incubated for 16-20 hours. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms.

VITEK® 2 AST-GN Cefazolin demonstrated acceptable performance of 97.6% overall Essential Agreement and 85.1% overall Category Agreement with the reference method.

AntimicrobialAntimicrobial CodeAntibiotic VersionBp¹CommentEssential AgreementCategory Agreement% Reproducibility
% EAVMEME
CefazolinCZcz05nCLSI (FDA)#, E Enterobacterales(840/862) 97.5N/AN/A

Page 9

The low Category Agreement as listed above is mainly the result of one doubling dilution minor errors with E. coli and P. mirabilis.

VITEK 2 AST-Gram Negative Cefazolin MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Klebsiella oxytoca by VITEK 2 auto-dilution method compared to the CLSI reference broth microdilution method.

VITEK 2 AST-Gram Negative Cefazolin MIC values tended to be in exact agreement or at least one doubling dilution lower when testing Klebsiella pneumoniae by VITEK 2 manual-dilution method compared to the CLSI reference broth microdilution method.

¹Abbreviations — Bp = breakpoint committee; EA = essential agreement; CA = category agreement; VME = Very Major Error (susceptible result with resistant reference result); ME = Major Error (resistant result with susceptible reference result); mE = minor Error (susceptible or resistant result with an intermediate reference result, or an intermediate result with a susceptible or resistant reference result).

# = US Food and Drug Administration 510(k) cleared
CLSI = Clinical and Laboratory Standards Institute
E = External performance data
N/A = Not applicable

Reproducibility and Quality Control demonstrated acceptable results.

H. Limitations:

A VITEK® 2 AST card cannot be used with a direct clinical specimen or sample or other sources containing mixed flora. Any change or modification in the procedure may affect the results.

A result for an antibiotic/organism combination, which may have a limitation, may be suppressed from reporting. This can be accomplished through the use of bioART rules in the VITEK® 2 Systems software. Refer to the software user manual for instructions.

Due to the occurrence of minor errors (resulting in a category agreement below 90% for the following species), perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination: Cefazolin (cz05n): K. oxytoca when the VITEK2 MIC is 8 μg/mL

The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing: Cefazolin (cz05n): Citrobacter koseri

I. Conclusion:

The performance data presented in this submission support a substantial equivalence decision. VITEK® 2 AST-GN Cefazolin (≤1–≥32 µg/mL) is substantially equivalent to VITEK® 2 AST-GN Cefazolin (K222073).

Page 10

References:

  1. MacLowry, J.D. and Marsh, H.H., Semi-automatic Microtechnique for Serial Dilution Antibiotic Sensitivity Testing in the Clinical laboratory, Journal of Laboratory Clinical Medicine, 72:685-687, 1968.

  2. Gerlach, E.H., Microdilution 1: A Comparative Study, p. 63-76. Current Techniques for Antibiotic Susceptibility Testing. A. Balows (ed.), Charles C. Thomas, Springfield, IL, 1974.

  3. Barry, A.L., The Antimicrobic Susceptibility Test, Principles and Practices, Lea and Febiger, Philadelphia, PA, 1976.

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”