K Number
K251406

Validate with FDA (Live)

Device Name
BriefCase-Triage
Date Cleared
2025-05-30

(24 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.

Device Description

Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

AI/ML Overview

Here's a detailed breakdown of the acceptance criteria and study findings for BriefCase-Triage, based on the provided FDA 510(k) clearance letter:


1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
Primary EndpointsA lower bound 95% Confidence Interval (CI) of 80% for Sensitivity and Specificity at the default operating point.Default Operating Point: - Sensitivity: 92.7% (95% CI: 88.2%, 95.8%). The lower bound (88.2%) is > 80%. - Specificity: 92.8% (95% CI: 89.2%, 95.4%). The lower bound (89.2%) is > 80%. Additional Operating Points (AOPs) meeting criteria: - AOP1: Sensitivity 95.6% (95% CI: 91.8%-98.0%), Specificity 88.2% (95% CI: 84.0%-91.6%) - AOP2: Sensitivity 94.1% (95% CI: 90.0%-96.9%), Specificity 89.8% (95% CI: 85.8%-93.0%) - AOP3: Sensitivity 89.3% (95% CI: 84.2%-93.2%), Specificity 94.7% (95% CI: 91.6%-97.0%) - AOP4: Sensitivity 86.3% (95% CI: 80.9%-90.7%), Specificity 97.7% (95% CI: 95.3%-99.1%)
Secondary Endpoints (Comparability with Predicate)Time-to-notification metric for the Briefcase-Triage software should demonstrate comparability with the predicate device.Briefcase-Triage (Subject Device): Mean time-to-notification = 10.7 seconds (95% CI: 10.5-10.9) Predicate AD: Mean time-to-notification = 38.0 seconds (95% CI: 35.5-40.4) The subject device's time-to-notification is faster than the predicate, demonstrating comparability and improvement in time savings.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 509 cases.
  • Data Provenance:
    • Country of origin: 5 US-based clinical sites.
    • Retrospective or Prospective: Retrospective.
    • Data Sequestration: Cases collected for the pivotal dataset were "all distinct in time or center from the cases used to train the algorithm," and "Test pivotal study data was sequestered from algorithm development activities."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Three (3) senior board-certified radiologists.
  • Qualifications: "Senior board-certified radiologists." (Specific years of experience are not provided.)

4. Adjudication Method for the Test Set

  • The text states "the ground truth, as determined by three senior board-certified radiologists." This implies a consensus-based adjudication, likely 3-0 or 2-1 (majority vote), but the exact method (e.g., 2+1, 3+1) is not explicitly detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done? No.
  • The study primarily focused on the standalone performance of the AI algorithm compared to ground truth and a secondary comparison of time-to-notification with a predicate device. It did not evaluate human reader performance with and without AI assistance.

6. Standalone Performance Study

  • Was it done? Yes.
  • The study evaluated the algorithm's performance (sensitivity, specificity, PPV, NPV, PLR, NLR) in identifying AD pathology without human intervention as a primary and secondary endpoint. The device's output is "flagging and communication of suspected positive findings" and "notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification," confirming a standalone function.

7. Type of Ground Truth Used

  • Ground Truth: Expert Consensus, specifically "as determined by three senior board-certified radiologists."

8. Sample Size for the Training Set

  • The document states, "The algorithm was trained during software development on images of the pathology." However, it does not specify the sample size for the training set. It only mentions that the pivotal test data was "distinct in time or center" from the training data.

9. How the Ground Truth for the Training Set Was Established

  • "As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding."
  • While it indicates images were "labeled ("tagged")" based on the "presence of the critical finding," it does not explicitly state who established this ground truth for the training set (e.g., experts, pathology, etc.). It's implied that medical professionals were involved in the labeling process, but no specific number or qualification is provided for the training set.

FDA 510(k) Clearance Letter - BriefCase-Triage

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 30, 2025

Aidoc Medical, Ltd.
℅ John Smith
Partner
Hogan Lovells US LLP
555 Thirteenth Street NW
Washington, District of Columbia 20004

Re: K251406
Trade/Device Name: BriefCase-Triage
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological computer aided triage and notification software
Regulatory Class: Class II
Product Code: QAS
Dated: May 6, 2025
Received: May 6, 2025

Dear John Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251406 - John Smith Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251406 - John Smith Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251406

Device Name
BriefCase-Triage

Indications for Use (Describe)

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Aidoc Medical, Ltd.'s Briefcase-Triage

Submitter:
Aidoc Medical, Ltd.
3 Aminadav St.
Tel-Aviv, Israel
Phone: +972-73-7946870
Contact Person: Amalia Schreier, LL.M
Date Prepared: May 28, 2025

Name of Device: Briefcase-Triage
Classification Name: Radiological computer-assisted triage and notification software device (21 CFR 892.2080)
Regulatory Class: Class II
Product Code: QAS

Primary Predicate Device: Briefcase-Triage for AD (K222329)
Reference Device: BriefCase-Triage for RibFx (K243548)

Device Description

Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.

The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

K251406

Page 6

The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding.

Intended Use / Indications for Use

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.

Summary of Technological Characteristics

The subject Briefcase-Triage for AD and the algorithm analysis module for the primary predicate Briefcase-Triage for AD (K222329) are identical in most aspects and differ mostly with respect to changes in algorithm training process and their algorithm performance.

Both the primary predicate and subject devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated in software components for use with DICOM compliant CT scanners, PACS, and radiology workstations.

Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. Both devices are intended to provide the specialists with notifications and unannotated, compressed, low-quality, and grayscale preview images of suspect studies for the purpose of preemptive triage.

The subject and predicate Briefcase-Triage devices raise the same types of safety and effectiveness questions, namely, accurate triage of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor de-prioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases.

A table comparing the key features of the subject and the primary predicate devices is provided

Page 7

below.

Table 1. Key Feature Comparison

Subject DevicePredicate Device
Aidoc Briefcase-Triage for ADAidoc Briefcase-Triage for AD (K222329)
Intended Use / Indications for UseBriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of CT exams with contrast (CTA and CT with contrast) that include the chest in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.

Page 8

Subject DevicePredicate Device
Aidoc Briefcase-Triage for ADAidoc Briefcase-Triage for AD (K222329)
User populationHospital networks and appropriately trained medical specialistsHospital networks and appropriately trained medical specialists
Anatomical region of interestChest, abdomen and thoraco-abdominalChest, abdomen and thoraco-abdominal
Data acquisition protocolCT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast)CT exams with contrast (CTA and CT with contrast) that include the chest
Notification-only (/notification alerts), parallel workflow toolYesYes
Images formatDICOMDICOM
Interference with standard workflowNo. No cases are removed from desktop app or deprioritizedNo. No cases are removed from desktop app or deprioritized

Page 9

Subject DevicePredicate Device
Aidoc Briefcase-Triage for ADAidoc Briefcase-Triage for AD (K222329)
Inclusion/ Exclusion criteria for clinical performance testingInclusion Criteria• Scans performed on adults/transitional adolescents ≥ 18 years of age.• CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast)• Slice thickness 0.5 mm - 5.0 mmExclusion Criteria• All studies that have an inadequate field of viewInclusion Criteria• Scans performed on adults/transitional adolescents ≥ 18 years of age.• CT exams with contrast (CTA and CT with contrast) that include the chest• Slice thickness 0.5 mm - 5.0 mmExclusion Criteria• All studies that are technically inadequate, such as severe metal artifacts, or inadequate field of view.
Additional Operating Points4 Additional Operating Points2 Additional Operating Points
AlgorithmArtificial intelligence algorithm with database of images.Artificial intelligence algorithm with database of images.
Structure- Integrated with image routing module via image communication platform (ICP) (image acquisition).- Algorithm module (image processing)- Integrated with desktop application for workflow integration (feed and non-diagnostic Image Viewer).- AHS module ( image acquisition);- ACS module (image processing);- Aidoc Desktop Application for workflow integration (Feed/Worklist (alternate names) and non-diagnostic Image Viewer).

Performance Data

Pivotal Study Summary

Aidoc conducted a retrospective, blinded, multicenter study with the Briefcase-Triage software to evaluate the software's performance in identifying CT chest, abdomen, or chest/abdomen exams

Page 10

with contrast (CTA and CT with contrast) in 509 cases from 5 US-based clinical sites. The study compared the software's performance to the ground truth, as determined by three senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm, as was used for the most recent clearance (K222329). Test pivotal study data was sequestered from algorithm development activities, and use of the data is managed by appropriate Quality Management System procedures.

Primary endpoints were sensitivity and specificity with an 80% performance goal. Secondary endpoints were Briefcase-Triage time-to-notification compared to the predicate device. Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed.

Primary Endpoint

Sensitivity was 92.7% (95% CI: 88.2%, 95.8%) and Specificity was 92.8% (95% CI: 89.2%, 95.4%). Because the lower bound of each 95% CI exceeded 80%, the study's primary endpoints were achieved.

Secondary Endpoint

In addition, the time-to-notification metric observed for the Briefcase-Triage software, when integrated with a compatible image communication platform, was compared to the equivalent metric of the predicate devices. The Briefcase-Triage time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.

The Briefcase-Triage time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the Briefcase-Triage device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate Briefcase-Triage AD.

The time-to-notification results obtained for the subject Briefcase-Triage device show comparability with the primary predicate with regard to time savings to the standard of care review. The Briefcase-Triage mean time-to-notification for the subject AD triage was 10.7 seconds (95% CI: 10.5-10.9) . The time-to-notification for the predicate AD was 38.0 seconds (95% CI: 35.5-40.4).

Table 2. Time-to- notification comparison for Briefcase-Triage devices (Seconds)

Page 11

Time-to-notificationMean Estimate (seconds)N95% Lower CL95% Upper CLMedian
Predicate K222329 Processing Time38.049935.540.431.1
Briefcase-Triage + Image Communication Platform Time-To-Notification10.721210.510.910.4

NPV was 99.8% (95% CI: 99.7%- 99.9%) and PPV was 24.7% (95% CI: 18.0%- 33.0%).
PLR was 12.8 (95% CI: 8.5- 19.2) and NLR was 0.08 (95% CI: 0.05- 0.13).

Thus, the reported similar time-to-notification data demonstrates that when using the subject Briefcase-Triage for AD the clinician may have the same benefit in time saving as with the predicate Briefcase-Triage for AD.

As can be seen in Table 3 the mean age of patients whose scans were reviewed for AD was 60.1 years, with a standard deviation of 17.3 years. Gender distribution was 50.9% male, and 49.1% female (Table 4). Scanner distribution can also be found in Table 5 below.

Table 3. Descriptive Statistics for Age

MeanStdMinMedianMaxN
Age (Years)60.117.3186290212

Page 12

Table 4. Frequency Distribution of Gender

Ground Truth ResultsGender
FemaleMaleAll
N%N%N%
Positive7314.3%13225.9%20540.3%
Negative17734.8%12725.0%30459.7%
All25049.1%25950.9%509100.0%

Table 5. Frequency Distribution of Manufacturer

ManufacturerN%
GE MEDICAL SYSTEMS14127.7%
Philips11222%
SIEMENS17434.2%
TOSHIBA8216.1%
Total509100%

Clinical Subgroups and Confounders:

Pathologies present in negative cases: Inflammatory; Oncology; Trauma; Heart & vascular; Chronic diseases; None of the above and Fully negative.

Additional Operating Points:

In addition to the default operating point that was selected to maximize both sensitivity and specificity, four additional operating points (AOP) were selected to maximize specificity or sensitivity

Page 13

while maintaining a lower bound 95% confidence interval of 80% for sensitivity and specificity respectively:

Table 6. Additional Operating Points

Operating PointsSensitivity % (95% CI)Specificity % (95% CI)
AOP195.6% (95% CI: 91.8%-98.0%)88.2% (95% CI: 84.0%-91.6%)
AOP294.1% (95% CI: 90.0%-96.9%)89.8% (95% CI: 85.8%-93.0%)
AOP389.3% (95% CI: 84.2%-93.2%)94.7% (95% CI: 91.6%-97.0%)
AOP486.3% (95% CI: 80.9%-90.7%)97.7% (95% CI: 95.3%-99.1%)

In summary, performance goals were achieved for the default and four additional operating points. Combined with the comparison results of time-to-notification metric with the predicate device, these data establish the achievement by the subject Briefcase-Triage of preemptive triage in the range of several minutes.

Conclusions

The subject Briefcase-Triage for AD and the predicate Briefcase-Triage for AD (K222329) are intended to aid in prioritization and triage of radiological images for the indications for suspected positive findings of incidental pulmonary embolism pathologies. Both devices are software components consisting of deep learning AI algorithms that process images and produce analysis results, which are displayed to the user by a prioritization alert and a compressed, low-quality, grayscale, unannotated preview image. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.

The subject Briefcase-Triage device for AD is thus substantially equivalent to the primary predicate Briefcase-Triage for AD (K222329).

N/A