(122 days)
da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
da Vinci SP Instruments:
Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
The da Vinci SP Surgical System, Model SP1098, is a software-controlled, electro-mechanical systems designed to enable complex surgery using a minimally invasive approach. The systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an endoscope, instruments, and accessories.
The provided FDA 510(k) Clearance Letter for the da Vinci SP Surgical System (K251324) does not contain the information requested regarding acceptance criteria and the study that proves the device meets the acceptance criteria in the context of an AI/ML device.
This document is for a robotic surgical system, not an AI/ML software device for diagnostic or analytical purposes. The clearance focuses on hardware, software for controlling the hardware, and human factors related to the physical use of the surgical system.
Therefore, I cannot extract the following information from the provided text:
- A table of acceptance criteria and the reported device performance: This document does not describe specific performance metrics for an AI/ML algorithm or corresponding acceptance criteria.
- Sample sizes used for the test set and the data provenance: No test set, data provenance, or sample sizes related to an AI/ML model are mentioned.
- Number of experts used to establish the ground truth... and qualifications: Not applicable as there's no AI/ML model for which ground truth would be established by experts for performance evaluation.
- Adjudication method for the test set: Not applicable.
- Multi-reader multi-case (MRMC) comparative effectiveness study: This type of study is relevant for evaluating the impact of AI assistance on human diagnostic performance. The document describes human factors evaluation related to the usability of the surgical system, not diagnostic performance with AI assistance.
- Standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as there's no standalone AI algorithm being cleared.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
The document mentions "Software verification and validation have been conducted at the unit, integration, and system level to confirm that the subject devices continue to meet design requirements and user needs" and "Human Factors Evaluation including usability validation has been conducted to demonstrate safety and effectiveness." These are general statements about software and usability testing for a medical device, but they do not detail the specific performance metrics, acceptance criteria, or study designs typically used for AI/ML devices as requested in the prompt.
FDA 510(k) Clearance Letter - da Vinci SP Surgical System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 29, 2025
Intuitive Surgical, Inc.
Amrit Rangan
Sr. Regulatory Affairs Specialist
1266 Kifer Road
Sunnyvale, California 94086
Re: K251324
Trade/Device Name: da Vinci SP Surgical System (SP1098)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: NAY
Dated: July 28, 2025
Received: July 28, 2025
Dear Amrit Rangan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K251324 - Amrit Rangan Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K251324 - Amrit Rangan Page 3
Sincerely,
Mark Trumbore -S
Digitally signed by Mark Trumbore -S
Date: 2025.08.29 11:58:30 -04'00'
Mark Trumbore Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K251324
Device Name: da Vinci SP Surgical System (SP1098)
Indications for Use (Describe)
da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
da Vinci SP Instruments:
Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) Summary (21 CFR § 807.92)
I. SUBMITTER INFORMATION
Submitter: Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086
Contact: Amrit Rangan
Senior Regulatory Affairs Specialist
Phone Number: 949-690-8799
Email: amrit.jaggi@intusurg.com
Date Summary Prepared: August 29, 2025
II. SUBMITTER INFORMATION
Trade Name: da Vinci SP Surgical System
Common Name: Endoscopes and accessories
Classification Name: System, Surgical, Computer Controlled Instrument (876.1500)
Regulatory Class: Class II
Product Code: NAY (System, Surgical, Computer Controlled Instrument)
Submission Type: Traditional 510(k)
III. PREDICATE DEVICE INFORMATION
Predicate Device: da Vinci SP Surgical System (K242318)
IV. Device Description
The da Vinci SP Surgical System, Model SP1098, is a software-controlled, electro-mechanical systems designed to enable complex surgery using a minimally invasive approach. The systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an endoscope, instruments, and accessories.
Indications for Use
Page 6
da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
da Vinci SP Instruments:
Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Comparison of Indications for Use Statement
The indications for use are identical between the predicate and subject device.
Comparison of Technological Characteristics
The subject da Vinci SP Surgical System, Model SP1098, is technologically similar to its predicate device. The technological differences that are the basis for this submission are limited to the updated system software for modifications to Adjust Mode while using an SP SureForm 45 Stapler and insertion jump detection algorithm for the SP SureForm 45 Stapler.
Non-Clinical Testing Summary and Conclusion
Software verification and validation have been conducted at the unit, integration, and system level to confirm that the subject devices continue to meet design requirements and user needs. Software documentation has been provided in accordance with FDA Guidance Content of Premarket Submissions for Device Software Functions, issued on June 14, 2023.
Human Factors Evaluation including usability validation has been conducted to demonstrate safety and effectiveness of the subject device. The results of this validation concluded that the subject device continues to be safe and effective for its intended use, by its intended users, and in its intended use environment. Human Factors documentation has been provided in accordance with FDA Draft Guidance Content of Human Factors Information in Medical Device Marketing Submissions issued on December 9, 2022 and FDA Guidance Document Applying Human Factors and Usability Engineering to Medical Devices issued on February 3, 2016.
Page 7
Based on the intended use, indications for use, technological characteristics, and non-clinical data, the subject da Vinci SP Surgical System, Model SP1098 is substantially equivalent to and is as safe, as effective, and performs as well as its predicate device.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.