AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MX7/MX7T/Vaus7/Zeus/ME7/Anesus ME7/Anesus ME7T/MX7P/MX7W/MX8/MX8T/Vaus8/ME8/Nuda/Aquila/Malody Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in ophthalmic, fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic,adult cephalic,trans-rectal, trans-vaginal, musculo-skeletal(conventional), musculo-skeletal(superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative(abdominal, thoracic, and vascular) , Laparoscopic, trans-esoph(Cardiac) exams.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Power Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape, TDI, color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

Device Description

The MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include: B, M, PWD, CWD, Color Doppler, Power Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape, TDI, color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

The MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.

AI/ML Overview

The provided FDA 510(k) Clearance Letter states that clinical studies were Not Applicable for the Mindray Diagnostic Ultrasound System (K251192) to support substantial equivalence. Therefore, there is no information in this document about acceptance criteria or a study proving the device meets specific performance criteria through clinical data.

The letter explicitly states:
"8. Clinical Studies
Not applicable. The subject of this submission, MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence."

Instead, the submission relies on:

  • Comparison with Predicate Devices: The document extensively details how the subject device is comparable and substantially equivalent to existing cleared predicate devices (e.g., MX7 (K241432), TEX20 (K241201), Resona I8W (K240115), TE Air (K240906), Consona N6 (K221300), QLAB (K190913)) in terms of intended use, imaging modes, features, functions, and technological characteristics.
  • Non-clinical Tests: The device was evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and thermal, electrical, and mechanical safety, in compliance with various recognized standards (e.g., NEMA UD 2-2004, ANSI AAMI ES60601-1, IEC 60601-1-2, ISO 14971, ISO 10993-1).

Therefore, I cannot populate the table or answer the specific questions about acceptance criteria and a study proving device performance as they would apply to clinical efficacy/performance for a new AI/software component, because this information is not present in the provided 510(k) summary. The clearance is based on substantial equivalence to existing predicate ultrasound systems and non-clinical safety/performance testing.

FDA 510(k) Clearance Letter - K251192

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

Shenzhen Mindray Bio-medical Electronics Co., LTD.
Tang Jing
Technical Regulatory Affairs
Mindray Building, Keji 12th Road South
Hi-tech Industrial Park, Nanshan
Shenzhen, Guangdong 518057
CHINA

Re: K251192

Trade/Device Name: Diagnostic Ultrasound System (MX7); Diagnostic Ultrasound System (MX7T); Diagnostic Ultrasound System (Vaus7); Diagnostic Ultrasound System (Zeus); Diagnostic Ultrasound System (ME7); Diagnostic Ultrasound System (Anesus ME7); Diagnostic Ultrasound System (Anesus ME7T); Diagnostic Ultrasound System (MX7P); Diagnostic Ultrasound System (MX7W); Diagnostic Ultrasound System (MX8); Diagnostic Ultrasound System (MX8T); Diagnostic Ultrasound System (Vaus8); Diagnostic Ultrasound System (ME8); Diagnostic Ultrasound System (Nuda); Diagnostic Ultrasound System (Aquila); Diagnostic Ultrasound System (Malody)

Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: July 24, 2025
Received: July 24, 2025

Dear Tang Jing:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

August 22, 2025

Page 2

K251192 – Tang Jing
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K251192 – Tang Jing
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MARJAN NABILI -S for

Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K251192

Please provide the device trade name(s).

Diagnostic Ultrasound System (MX7);
Diagnostic Ultrasound System (MX7T);
Diagnostic Ultrasound System (Vaus7);
Diagnostic Ultrasound System (Zeus);
Diagnostic Ultrasound System (ME7);
Diagnostic Ultrasound System (Anesus ME7);
Diagnostic Ultrasound System (Anesus ME7T);
Diagnostic Ultrasound System (MX7P);
Diagnostic Ultrasound System (MX7W);
Diagnostic Ultrasound System (MX8);
Diagnostic Ultrasound System (MX8T);
Diagnostic Ultrasound System (Vaus8);
Diagnostic Ultrasound System (ME8);
Diagnostic Ultrasound System (Nuda);
Diagnostic Ultrasound System (Aquila);
Diagnostic Ultrasound System (Malody)

Please provide your Indications for Use below.

The MX7/MX7T/Vaus7/Zeus/ME7/Anesus ME7/Anesus ME7T/MX7P/MX7W/MX8/MX8T/Vaus8/ME8/Nuda/Aquila/Malody Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in ophthalmic, fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic,adult cephalic,trans-rectal, trans-vaginal, musculo-skeletal(conventional), musculo-skeletal(superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative(abdominal, thoracic, and vascular) , Laparoscopic, trans-esoph(Cardiac) exams.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Power Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape, TDI, color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Diagnostic Ultrasound System
Page 1/1

Page 5

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).

The assigned 510(k) number: K251192

1. Submitter

Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan,
Shenzhen, 518057, P. R. China
Tel: +86 755 8188 6129
Fax: +86 755 2658 2680

Contact Person

Tang Jing
Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, Shenzhen, 518057, P. R. China

Date Prepared: April 16, 2025

2. Device Name

  • Diagnostic Ultrasound System (MX7)
  • Diagnostic Ultrasound System (MX7T)
  • Diagnostic Ultrasound System (Vaus7)
  • Diagnostic Ultrasound System (Zeus)
  • Diagnostic Ultrasound System (ME7)
  • Diagnostic Ultrasound System (Anesus ME7)
  • Diagnostic Ultrasound System (Anesus ME7T)
  • Diagnostic Ultrasound System (MX7P)
  • Diagnostic Ultrasound System (MX7W)
  • Diagnostic Ultrasound System (MX8)
  • Diagnostic Ultrasound System (MX8T)
  • Diagnostic Ultrasound System (Vaus8)
  • Diagnostic Ultrasound System (ME8)
  • Diagnostic Ultrasound System (Nuda)
  • Diagnostic Ultrasound System (Aquila)

Page 6

Diagnostic Ultrasound System (Malody)

Classification

Regulatory Class: II
Review Category: Tier II
21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

3. Predicate devices

MX7 series Diagnostic Ultrasound System is comparable with and substantially equivalent to the predicate devices listed below. MX7(K241432) is the main predicate device.

DeviceManufacturerModel510(k) Number
1. Main predicate deviceMindrayMX7K241432
2. Reference deviceMindrayTEX20K241201
3. Reference deviceMindrayResona I8WK240115
4. Reference deviceMindrayTE AirK240906
5. Reference deviceMindrayConsona N6K221300
6. Reference devicePhilips HealthcareQLABK190913

The result shows the conformance of subject device to the predicate devices.

Regulation name and code

21 CFR 892.1550 Ultrasonic Pulsed Dopple Imaging System (IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

4. Device Description:

The MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include: B, M, PWD, CWD, Color Doppler, Power Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape, TDI, color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

The MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic

Page 7

Ultrasound System can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.

5. Intended Use:

The MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in ophthalmic, fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic,adult cephalic,trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative(abdominal, thoracic, and vascular) , Laparoscopic, trans-esoph(Cardiac) exams.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Power Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape, TDI, color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

6. Comparison with Predicate Devices:

Subject device MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices mentioned in Predicate Devices with regards to intended use, imaging modes, features and functions and technological characteristics.

All systems transmit ultrasonic energy into patients, perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, as well as calculations.

Subject device MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the same intended uses as the predicated device MX7 (K241432) except Laparoscopic application. At the same time, TEX20(K241201) has the Laparoscopic application instead.

The patient contact materials of the new added transducers (L15-3RCs, L24-6RCs, L24-6Hs, SC10-2RCs, ELC13-4s, L14-3Ws, DE11-3s, LAP13-4Cs, 7LT4s, i3P, i5M, i8L) and new added needle-guided brackets (NGB-058, NGB-051, NGB-

Page 8

054, NGB-027, NGB-010) of subject device MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System are the same to the predicate devices.

The acoustic power levels of MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System are below the levels of FDA, which are the same as the predicated device MX7 (K241432).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is designed in compliance with the FDA recognized electrical and physical safety standards, which are the same as the predicated device MX7 (K241432).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the equivalent features and functions as the predicated devices. Among these features, the VA Grid function in proposed MX7 series has been cleared in predicate device TEX20 (K241201).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the equivalent features and functions as the predicated devices. Among these features, the Smart Face, Face+and UR Grid (Biopsy Grid) supported in proposed MX7 series has been cleared in predicate device Resona I8W (K240115).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the equivalent features and functions as the predicated devices. Among these features, the TTQA-RV supported in proposed MX7 series has been cleared in predicate device QLAB (K190913).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the equivalent features and functions as the predicated devices. Among these features, the X-link, iVocal Plus, Air Station, Air Charge Module and iVocal Plus Microphone Array supported in proposed MX7 series has been cleared in predicate device TEX20 (K241201).

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has the equivalent features and functions as the predicated devices. Among these features, the Air Capsule supported in proposed MX7 series has been cleared in predicate device TE Air (K240906).

For the differences compared to the predicate devices:

The subject device introduces new options, which in the below table that either the improvements or enhancement based on the cleared functions to facilitate user, no

Page 9

intended uses are added and passed the related tests, no clinical risks introduced.

Subject DeviceFunction explanation
Auto VMThe Auto VM can automatically measure the vessel diameter and depth.
Smart FVThe Smart FV can automatically measure the intravascular hemodynamic parameters.
HD Access worksHD Access works is used to set the workflow template protocol for vessel application to guide the doctor's operation steps.
Probe Adapter PCM-CWS01Probe Adapter PCM-CWS01 is used to transfer the pencil probe interface to S interface, so that it can be connected to the main unit.

7. Non-clinical Tests:

MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and this device has been designed to conform with applicable medical safety standards.

This device has been tested and evaluated under the following standards:

  • NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic ultrasound equipment revision 3.
  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)].
  • IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • IEC TS 60601-4-2 Edition 1.0 2024-03 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
  • IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
  • IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
  • ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices.

Page 10

  • ISO 10993-1 Fifth edition 2018-08, biological evaluation of medical devices - part 1: evaluation and testing within a risk management process.
  • IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, medical device software - software life cycle processes.
  • IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices - Part 1: Application of usability engineering to medical devices.

These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.

8. Clinical Studies

Not applicable. The subject of this submission, MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.

9. Summary

Based on the performance data as documented in the study, MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System was found to have a safety and effectiveness profile that is similar to the predicate device.

10. Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the MX7, MX7T, Vaus7, Zeus, ME7, Anesus ME7, Anesus ME7T, MX7P, MX7W, MX8, MX8T, Vaus8, ME8, Nuda, Aquila, Malody Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to the predicate devices.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.