K Number
K250749

Validate with FDA (Live)

Device Name
PhysCade System
Manufacturer
Date Cleared
2025-06-04

(84 days)

Product Code
Regulation Number
870.1425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician.

The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations.

Device Description

The PhysCade™ System (PhysCade) is an artificial intelligence (AI) enabled device intended to assist clinicians in their management of patients with heart rhythm disorders (arrhythmias). PhysCade is a medical decision support system which post-processes electrograms (EGMs) collected inside the heart during electrophysiology (EP) mapping procedures using compatible diagnostic EP catheters. The PhysCade software has advanced algorithms that analyze the collected EGMs to provide information on regions of interest in the heart that may be useful to the clinician to support clinical decisions together with other available patient-related information.

PhysCade provides specialized analyses of data from a compatible multipolar catheter. The primary output (coPILOT) indicates the predominant earliest site of activation relative to the catheter electrode array. Supporting outputs include determining activation times of successive beats in the EGM signal on each electrode (coMAP), voltage at each electrode, signal quality, and sequences of propagation over multiple beats of the arrhythmia on the catheter.

The PhysCade System consists of a computer workstation, display, and custom software and is not connected to other devices or medical equipment.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the PhysCade System gives some information about the device's development and testing, particularly regarding its AI algorithms. However, it does not explicitly detail the acceptance criteria or the specific results of a study proving the device meets those criteria, nor does it provide the requested levels of detail for the ground truth establishment, expert qualifications, or MRMC study results.

Based on the information provided, here's what can be extracted and inferred for the requested points. Where information is not present, it is explicitly stated.


Acceptance Criteria and Device Performance Study

The document states, "Design Validation confirmed that the AI/ML system is accurate for its intended use." This indicates that performance testing was conducted. However, the specific quantitative acceptance criteria (e.g., a specific F1 score, accuracy, sensitivity, or precision threshold) and the resulting performance metrics the device achieved are not explicitly stated in this document.

Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance Criteria (Threshold)Reported Device Performance
Specific quantitative metrics for AI/ML performanceNOT PROVIDEDNOT PROVIDED
Overall AI/ML system accuracy"Accurate for its intended use" (Qualitative)"Accurate for its intended use"

Study Details:

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size:
      • Number of electrograms: ~5 Million
      • Number of patients: 109
    • Data Provenance: The document does not explicitly state the country of origin.
    • Retrospective or Prospective: The document does not explicitly state whether the data was collected retrospectively or prospectively. It references "datasets with the following characteristics" suggesting pre-collected (likely retrospective) data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: NOT PROVIDED.
    • Qualifications of Experts: NOT PROVIDED. The document only mentions that the device is for "analysis by a physician" and is "intended to be operated by nurses, technicians, physicians, or other personnel who are trained and approved by each treating facility." It does not specify who established the ground truth or their qualifications.
  3. Adjudication method for the test set:

    • Adjudication Method: NOT PROVIDED. The document does not describe how disagreements, if any, among experts establishing ground truth were resolved.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The performance testing described focuses on the device's accuracy ("Design Validation confirmed that the AI/ML system is accurate for its intended use"), rather than human performance improvement.
    • Effect Size: NOT APPLICABLE as no MRMC study is mentioned.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: While not explicitly stated as "standalone performance," the phrase "Design Validation confirmed that the AI/ML system is accurate for its intended use" strongly implies a standalone performance evaluation of the algorithm's output against established ground truth. The device "provides information on regions of interest in the heart that may be useful to the clinician to support clinical decisions," suggesting the AI's output is evaluated directly.
  6. The type of ground truth used:

    • Type of Ground Truth: The document states the device "analyzes the collected EGMs to provide information on regions of interest in the heart that may be useful to the clinician." It further notes the "primary output (coPILOT) indicates the predominant earliest site of activation relative to the catheter electrode array." The ground truth would therefore pertain to the identification of these "regions of interest" or the "earliest site of activation" based on expert interpretation of electrophysiological data. The exact method of establishing this ground truth (e.g., expert consensus on EGM analysis, correlation with other diagnostic modalities, or clinical outcomes) is NOT EXPLICITLY STATED.
  7. The sample size for the training set:

    • Training Set Sample Size:
      • Number of electrograms: ~15 Million
      • Number of patients: 174
  8. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: The document does not explicitly state how the ground truth for the training set was established. It only describes the characteristics of the training, tuning, and test cohorts. It is commonly understood that ground truth for training data is established by similar means to test data (e.g., expert annotation), but this is not detailed here.

FDA 510(k) Clearance Letter - PhysCade System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 4, 2025

PhysCade, Inc.
Grace Li
Head of QA/RA
2100 Geng Road, Suite 210
Palo Alto, California 94303

Re: K250749
Trade/Device Name: PhysCade System
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK
Dated: March 12, 2025
Received: March 12, 2025

Dear Grace Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250749 - Grace Li Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250749 - Grace Li Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250749

Device Name
PhysCade System

Indications for Use (Describe)

The PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician.

The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250749 PhysCade System 510(k) Summary

Page 1

Submitter's information

PhysCade, Inc.
2100 Geng Road, Suite 210
Palo Alto, CA 94303
United States

Contact: Raphael Michel
President & CEO
PhysCade, Inc.
2100 Geng Road, Suite 210
Palo Alto, CA 94303

Classification

The classification for the new device is listed below.

21 CFR ReferenceProduct CodeClassGeneric Device NameClassification Description
§870.1425DQKIIComputer, Diagnostic, ProgrammableProgrammable diagnostic computer

New device

The new device's indications for use statement is listed in the table below.

Device NameIndications for Use
PhysCade SystemThe PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician.The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations.

Predicate device

The predicate device for the PhysCade System is shown in the table below.

K NumberProduct CodePredicate Device NameIndications for Use
K201298DQKVolta Medical VX1The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.The clinical significance of utilizing the VX1 software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.

Page 6

Device description

The PhysCade™ System (PhysCade) is an artificial intelligence (AI) enabled device intended to assist clinicians in their management of patients with heart rhythm disorders (arrhythmias). PhysCade is a medical decision support system which post-processes electrograms (EGMs) collected inside the heart during electrophysiology (EP) mapping procedures using compatible diagnostic EP catheters. The PhysCade software has advanced algorithms that analyze the collected EGMs to provide information on regions of interest in the heart that may be useful to the clinician to support clinical decisions together with other available patient-related information.

PhysCade provides specialized analyses of data from a compatible multipolar catheter. The primary output (coPILOT) indicates the predominant earliest site of activation relative to the catheter electrode array. Supporting outputs include determining activation times of successive beats in the EGM signal on each electrode (coMAP), voltage at each electrode, signal quality, and sequences of propagation over multiple beats of the arrhythmia on the catheter.

The PhysCade System consists of a computer workstation, display, and custom software and is not connected to other devices or medical equipment.

PhysCade is intended to be used in the cardiac electrophysiology lab, outside the sterile field. PhysCade is intended to be operated by nurses, technicians, physicians, or other personnel who are trained and approved by each treating facility to work in the electrophysiology lab.

PhysCade uses AI algorithms that were developed and tested using datasets with the following characteristics:

Page 7

Characteristics - comparison with predicate device

The table below lists the characteristics for both the new and predicate devices.

FeaturesDevelopment CohortTest Cohort
TrainingTuning
Number of electrograms~15M~5M~5M
Number of patients17462109
Age67.91 ± 8.8068.97 ± 8.2667.10 ± 7.66
AF Type
Non-Paroxysmal, n (%)124 (71.26)45 (72.58)82 (75.23)
Paroxysmal, n (%)50 (28.74)17 (27.42)27 (24.77)
Female, n (%)31 (17.82)6 (9.68)29 (26.61)
Racial category (non-white), n (%)*36 (20.93)16 (25.81)20 (20.20)
BMI, kg/m²29.45 ± 5.7229.45 ± 6.4629.67 ± 6.67
CHADS2VASC2.57 ± 1.572.65 ± 1.362.70 ± 1.66
LA Size Enlarged, n (%)*90 (55.21)32 (55.17)54 (58.70)
Coronary disease, n (%)35 (20.11)19 (30.65)27 (24.77)
Myocardial infarct, n (%)10 (5.75)6 (9.68)6 (5.50)
Hypertension, n (%)128 (73.56)47 (75.81)70 (64.22)
Diabetes Mellitus, n (%)32 (18.39)9 (14.52)21 (19.27)
Prior Stroke/TIA, n (%)16 (9.20)3 (4.84)8 (7.34)

*All data complete except for *minor missingness in the categorical variables of racial category (disclosed in N=172/174 in training, N=62/62 in tuning and 99/109 in test cohorts), and left atrial enlargement (variable recorded in N=163/174 in training, N=58/62 in tuning and 92/109 in test cohorts).

Page 8

ATTRIBUTESUBJECT DEVICE PhysCade SystemPREDICATE DEVICE Volta VX1 K201298COMPARISON
Indications for useThe PhysCade System is intended for the analysis, display, and storage of cardiac electrophysiological data and maps for analysis by a physician.The clinical significance of utilizing the PhysCade System to help identify areas with intra-cardiac atrial electrograms exhibiting local early activated potentials and other features of interest during atrial arrhythmias has not been established by clinical investigations.The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the VX1 software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.The PhysCade System has the same intended use and similar indications for use as the predicate device. Both devices assist in review and analysis of electrical maps of the human atria.The predicate device indications for use, while broader, identifies electrical activity in areas of interest for catheter ablation of atrial arrhythmias.
Regulation21 CFR 870.142521 CFR 870.1425Same
FDA Product codeDQKDQKSame
ClassificationProgrammable diagnostic computer, 21 CFR §870.1425Programmable diagnostic computer, 21 CFR §870.1425Same
PopulationAdultsAdultsSame
System TypeSignal processing based atrial mapping systemSignal processing based atrial mapping systemSame
Display(s)Color monitorColor monitorSame
ControlStandard keyboard / mouseStandard keyboard / mouseSame

Page 9

ATTRIBUTESUBJECT DEVICE PhysCade SystemPREDICATE DEVICE Volta VX1 K201298COMPARISON
Inputs RequiredIntra-cardiac atrial electrograms that are recorded with catheters, and surface ECG signals amplified and input to the PhysCade System on which they are displayed.Intra-cardiac atrial electrograms that are recorded with catheters, amplified and redirected to an acquisition system on which they are displayed.The subject and predicate devices analyze intra-cardiac atrial electrograms. The PhysCade System also inputs the ECG signal to assist in the analysis of intracardiac electrograms, which does not alter the intended use.
ConnectionsA USB flash drive is used to export data from compatible analog-to-digital converter systems and import data into the PhysCade System.A connection cable is used to connect the data acquisition system with an Advantech PCI-1713U analog-to-digital converter, which transmits the acquired information to a nearby computer that hosts the VX1 software.Both devices digitize intra-cardiac data for input to their analysis systems. While the subject and predicate devices use different brands of analog-to-digital converters, and the subject device uses a USB drive for export while the predicate device uses a connector cable, the transfer of data is equivalent.
Principal Mapping OutputDisplays adjudications as visual cues after analyzing intra-cardiac atrial electrograms using signal processing techniques.Displays adjudications as visual cues after analyzing intra-cardiac atrial electrograms in real-time using signal processing techniques.The predicate device and subject device provide adjudications during a procedure. While the predicate device inputs data from a connector cable, the subject device uploads data from a USB flash drive. Both systems provide output during an EP study to support clinical decision-making.

Page 10

ATTRIBUTESUBJECT DEVICE PhysCade SystemPREDICATE DEVICE Volta VX1 K201298COMPARISON
Map Types GeneratedThe operator is provided with a display of the multipolar electrode map indicating locations where wave direction is earliest, and markings of electrogram activation time, by the PhysCade System during recordings of atrial fibrillation.Real-time Dispersed Electrogram (DE) subtype of multipolar electrogram map- The operator is provided with display of the electrode locations where dispersed or non-dispersed electrograms have been recorded during atrial fibrillation or atrial tachycardia.Both the subject and predicate devices generate multipolar maps of their analysis of intra-cardiac data. The predicate device focuses on dispersed electrograms, while the subject device focuses on direction to the earliest site. Both devices produce a multipolar map of electrical locations of interest during atrial fibrillation.
Compatible CathetersAny compatible catheter meeting the requirements listed below:Compatibility requirements:Mapping catheter:- Electrode size: 0.5 mm to 3 mm- Inter-electrode spacing: 2 to 10mm (edge-to-edge)- Number of selected electrodes: between 8 and 64- Electrode configuration: a spatial grid of at least 2 electrodes along each side (e.g., 4X4, 8X8)Any compatible catheter meeting the requirements listed below:Compatibility Requirements:Mapping catheter:- Electrode size: 1 mm- Inter-electrode Spacing: 2 to 3 mm- Number of selected dipoles: 10Coronary sinus catheter:- Electrode size: 1 mm- Interelectrode Spacing: 2 to 3 mm- Number of selected dipoles: 5Both devices are compatible with catheters that are defined in their labeling which have been validated with their corresponding algorithms. The predicate device can also be used with Coronary Sinus catheters but this is not required for the subject device analysis.

Page 11

ATTRIBUTESUBJECT DEVICE PhysCade SystemPREDICATE DEVICE Volta VX1 K201298COMPARISON
Data Acquisition SystemThe PhysCade System works with three data acquisition systems:- LabSystem Pro Acquisition (Boston Scientific),- MacLab CardioLab Acquisition System (General Electric),- Ensite NavX/Precision System (Abbott)The VX1 works with two data acquisition systems:- LabSystem Pro Acquisition (Boston Scientific),- MacLab CardioLab Acquisition System (General Electric)The subject device is compatible with the same two data acquisition systems as the predicate device and adds compatibility with a third data acquisition system - Ensite. These 510(k) cleared data acquisition systems all produce digitized and amplified signals that are numerically similar.
Hardware Design and MaterialsOff-the-shelf computer and monitor, connection cable, acquisition system, proprietary software algorithm.Off-the-shelf analog / digital converter, computer and monitor, connection cable, acquisition system, proprietary software algorithm.Both devices use off-the-shelf hardware with a proprietary software algorithm.

Performance testing

The PhysCade System was subjected to extensive non-clinical testing including rigorous verification and validation testing. The testing consisted of:

Software/Firmware and cybersecurity verification and validation testing confirmed that the subject device meets the design input requirements and specifications
Design Verification testing confirmed that the subject device hardware meets performance specifications.
Design Validation confirmed that the AI/ML system is accurate for its intended use.
Usability testing demonstrated that the subject device meets the user needs specifications.

The PhysCade System was developed and tested in accordance with IEC 62304 Edition 1.1 2015-06.

Guidance documents referenced for testing

The following guidance documents were referenced for testing:
• Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions

Page 12

• Content of Premarket Submissions for Management of Cybersecurity in Medical Device
• Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
• Off-The-Shelf Software Use in Medical Devices
• Applying Human Factors and Usability Engineering to Medical Devices
• Good Machine Learning Practice for Medical Device Development: Guiding Principles

Conclusion

A comparison has been conducted of the intended use/indications for use, performance characteristics, design features, technological/functional characteristics, and labeling for the PhysCade System. Intended to provide the physician with additional information that supports standard clinical practice, no new or different questions of safety or effectiveness are raised.

The PhysCade System has the same intended use as the predicate device. Testing results demonstrate that minor differences in the technological characteristics do not raise questions of safety and effectiveness. Thus, the PhysCade System is substantially equivalent to the Volta VX1 predicate device (K201298).

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).