K Number
K250728

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-05

(55 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery.

The Arthrex Synergy Vision Endoscopic Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the system enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the Arthrex Synergy Vision Endoscopic Imaging System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Upon interstitial administration and use of ICG consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Upon administration and use of pafolacianine consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.

The Arthrex NanoNeedle Scope when used with the Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery. For pediatric patients, the Arthrex NanoNeedle Scope is indicated for laparoscopy and orthopedic procedures.

Device Description

The Arthrex Synergy Vision Endoscopic Imaging System includes a camera control unit (CCU) console, camera heads, endoscopes, and a laser light source. The system provides real-time visible light and near-infrared (NIR) illumination and imaging.

The Arthrex Synergy Vision Endoscopic Imaging System uses an integrated LED light to provide visible light illumination and imaging of a surgical site. For NIR imaging, the system interacts with the laser light source to visualize the presence of a fluorescence contrast agent, indocyanine green (ICG) and pafolacianine. The contrast agent fluoresces when illuminated through the laparoscope with NIR excitation light from the laser light source and the fluorescent response is then imaged with the camera, processed, and displayed on a monitor.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Arthrex Synergy Vision Endoscopic Imaging System (K250728) describes the device and its indications for use, but does not contain the detailed information necessary to fully answer all the questions regarding acceptance criteria and a study that proves the device meets those criteria.

Specifically, the document states: "The Arthrex Synergy Endoscopic Imaging System did not require animal testing or human clinical studies to support the determination of substantial equivalence." This implies that performance data demonstrating device capabilities against specific acceptance criteria (beyond general functional testing and compliance with standards) were not generated through clinical studies or animal studies for this submission.

However, based on the information provided, here's what can be inferred and what is missing:


Acceptance Criteria and Device Performance (Inferred/Missing)

Since no human or animal clinical studies were performed, there appears to be no specific clinical performance acceptance criteria listed in this document. The "Performance Data" section primarily focuses on engineering and regulatory compliance testing.

Acceptance CriterionReported Device Performance
Functional Performance (General)"The test results confirm the subject device met Arthrex product requirements and design specifications for the device."
Biocompatibility"Leveraged from the biocompatibility testing... of the additional predicate devices, as there were no modifications made to the subject device... that would affect the biocompatibility..."
Sterility"Leveraged from the... sterilization validation of the additional predicate devices, as there were no modifications made to the subject device... that would affect the... sterility of the device."
Electrical Safety (EMT)"The test results confirm the subject device conforms with EMT safety... standards."
Electromagnetic Compatibility (EMC)"The test results confirm the subject device conforms with... EMC standards."
Software Performance"Software testing was conducted and documentation was provided in this submission. The test results confirm the Arthrex software updates met product requirements and design specifications established for the device."
Image Resolution(Implicitly met by technical specifications matching predicate) 3840 x 2160, 400 x 400 (Nano), 720 x 720 (Nano)
Frame Rate(Implicitly met by technical specifications matching predicate) 60 fps, 30 fps (Nano)
NIR Wavelengths/Detection(Implicitly met by technical specifications matching predicate) Excitation Wavelength: 785 nm, Detection Bandwidth: 810 – 940 nm
Clinical Performance (e.g., Sensitivity, Specificity for specific disease/task)NOT APPLICABLE / NOT PROVIDED. The document explicitly states no human or animal studies were required for substantial equivalence. Therefore, there are no reported clinical performance metrics like sensitivity or specificity.

Study Details (Based on Provided Document and Inferences)

  1. Sample Size used for the test set and the data provenance:

    • Test set sample size: Not applicable for clinical performance as the document explicitly states: "The Arthrex Synergy Endoscopic Imaging System did not require animal testing or human clinical studies to support the determination of substantial equivalence."
    • Data provenance: Not applicable for clinical performance. The data provenance for engineering tests (e.g., EMT, EMC, software) would be internal Arthrex testing labs.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no human clinical studies establishing ground truth for clinical performance were conducted for this submission. Ground truth for engineering tests is established through technical specifications and industry standards.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for clinical performance.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. The device description does not indicate AI assistance features. The system is described as an endoscopic video camera and imaging system, not an AI-powered diagnostic aide.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is hardware (imaging system) with associated software, not a standalone algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the engineering and regulatory compliance tests: Ground truth was established by adherence to design specifications, industry standards (e.g., for biocompatibility, sterility, EMT, EMC), and internal product requirements.
    • For clinical performance (e.g., diagnostic accuracy for a specific disease): No such ground truth was established or presented as no clinical studies were performed.
  7. The sample size for the training set:

    • Not applicable, as this device does not appear to utilize machine learning for clinical interpretation or outcome prediction in a way that requires a "training set" in the context of AI/ML models. "Software testing" refers to verification against technical requirements, not AI model training.
  8. How the ground truth for the training set was established:

    • Not applicable for the same reasons as #7.

Summary of Key Takeaways from the Document:

The 510(k) clearance for the Arthrex Synergy Vision Endoscopic Imaging System (K250728) is based on substantial equivalence to existing predicate devices (K241361 and K243008). The justification for this clearance relies heavily on:

  • Similar intended use and indications for use.
  • Similar technological characteristics (e.g., components, imaging modes, specifications).
  • Successful completion of engineering and regulatory compliance testing (e.g., functional testing, biocompatibility, sterility, EMT, EMC, software).

Crucially, the document explicitly states that no animal testing or human clinical studies were required or performed to demonstrate the device meets acceptance criteria related to clinical performance. This means the clearance is not based on a study proving diagnostic accuracy or clinical effectiveness in patients, but rather on the device's technological similarity and safety/performance in a non-clinical, engineering test environment compared to previously cleared devices.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 5, 2025

Arthrex, Inc.
Lai Saeteurn
Senior Regulatory Affairs Specialist
1370 Creekside Boulevard
Naples, Florida 34108

Re: K250728
Trade/Device Name: Arthrex Synergy Vision Endoscopic Imaging System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ, IZI
Dated: March 10, 2025
Received: March 11, 2025

Dear Lai Saeteurn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250728 - Lai Saeteurn
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250728 - Lai Saeteurn
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K250728

Device Name
Arthrex Synergy Vision Endoscopic Imaging System

Indications for Use (Describe)

The Arthrex Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery.

The Arthrex Synergy Vision Endoscopic Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the system enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the Arthrex Synergy Vision Endoscopic Imaging System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Upon interstitial administration and use of ICG consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Upon administration and use of pafolacianine consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.

The Arthrex NanoNeedle Scope when used with the Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery. For pediatric patients, the Arthrex NanoNeedle Scope is indicated for laparoscopy and orthopedic procedures.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

FORM FDA 3881 (8/23)
Page 1 of 2

Page 5

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 2 of 2

Page 6

K250728 510(k) Summary

Date Prepared: April 21, 2025

Submitter: Arthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945

Contact Person: Name: Lai Saeteurn
Phone: 239-643-5553
Email: Lai.Saeteurn@Arthrex.com

Trade Name: Arthrex Synergy Vision Endoscopic Imaging System

Classification Name: 21 CFR 876.1500: Endoscope and accessories,
21 CFR 892.1600: Angiographic x-ray systems

Product Code: GCJ, IZI

Common Name: Laparoscope, General & Plastic Surgery

Regulatory Class: Class II

Primary Predicate Device: K241361 Arthrex Synergy Vision Endoscopic Imaging System

Additional Predicate Device: K243008 Arthrex NanoScope System

Purpose of Submission: This Traditional 510(k) premarket notification is submitted to expand indications for the Arthrex Synergy Vision Endoscopic Imaging System. Arthrex is also reporting minor software updates and additional device models made since the previous clearance (K241361).

Device Description: The Arthrex Synergy Vision Endoscopic Imaging System includes a camera control unit (CCU) console, camera heads, endoscopes, and a laser light source. The system provides real-time visible light and near-infrared (NIR) illumination and imaging.

The Arthrex Synergy Vision Endoscopic Imaging System uses an integrated LED light to provide visible light illumination and imaging of a surgical site. For NIR imaging, the system interacts with the laser light source to visualize the presence of a fluorescence contrast agent, indocyanine green (ICG) and pafolacianine. The contrast agent fluoresces when illuminated through the laparoscope with NIR excitation light from the laser light source and the fluorescent response is then imaged with the camera, processed, and displayed on a monitor.

Indications for Use: The Arthrex Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the

Page 1 of 4

Page 7

K250728

thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery.

The Arthrex Synergy Vision Endoscopic Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the system enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the Arthrex Synergy Vision Endoscopic Imaging System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Upon interstitial administration and use of ICG consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Upon administration and use of pafolacianine consistent with its approved label, the Arthrex Synergy Vision Endoscopic Imaging System is used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.

The Arthrex NanoNeedle Scope when used with the Synergy Vision Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including laparoscopy, orthopedic, plastic surgery, sinuscopy, spine, urology, and procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery. For pediatric patients, the Arthrex NanoNeedle Scope is indicated for laparoscopy and orthopedic procedures.

Performance Data: Design verification testing included inspection, engineering analysis and functional testing of the subject device. The test results confirm the subject device met Arthrex product requirements and design specifications for the device.

The biocompatibility and sterility data for the subject device was leveraged from the biocompatibility testing and sterilization validation of the additional predicate devices, as there were no

Page 2 of 4

Page 8

K250728

modifications made to the subject device since prior clearances that would affect the biocompatibility or sterility of the device.

EMT and EMC testing was conducted on the subject device. The test results confirm the subject device conforms with EMT safety and EMC standards.

Software testing was conducted and documentation was provided in this submission. The test results confirm the Arthrex software updates met product requirements and design specifications established for the device.

Technological Comparison

Device & Predicate DevicesArthrex Synergy Vision Endoscopic Imaging System K250728Arthrex Synergy Vision Endoscopic Imaging System K241361Arthrex NanoScope System K243008
System ComponentsCamera Control Unit, Camera Head, Light Source, EndoscopeCamera Control Unit, Camera Head, Light Source, EndoscopeCamera Control Unit, Endoscope
Imaging ModesWhite Light, NIR FluorescenceWhite Light, NIR FluorescenceWhite Light
Power Rating100-240V ~50/60 Hz100-240V ~50/60 Hz100-240V ~50/60 Hz
Light SourceIntegrated (Visible), External (NIR)Integrated (Visible), External (NIR)Integrated (Visible)
Image Resolutions3840 x 2160, 400 x 400 (Nano), 720 x 720 (Nano)3840 x 2160400 x 400
Frame Rate60 fps, 30 fps (Nano)60 fps30 fps
Scope TypeRigid Rod Lens, Chip on Tip (Nano)Rigid Rod LensChip on Tip
Image SensorCMOSCMOSCMOS
Angle of View0°, 30°, 45° 0° (Nano)0°, 30°, 45°0
Field of View75°, 120° (Nano)75°120°
Diameter5.5 mm, 10 mm 1.9 mm (Nano)5.5 mm, 10 mm1.9 mm
Length302 – 333 mm 125, 180, 250 mm (Nano)302 – 333 mm125, 180, 250 mm
Fluorescence Imaging Excitation SourceNIR LaserNIR Laser--
Excitation Wavelength785 nm785 nm--
Detection Bandwidth810 – 940 nm810 – 940 nm--
Excitation Light Source Intensity95 W/m²95 W/m²--

Page 3 of 4

Page 9

K250728

Conclusion: All verification activities were successfully completed to confirm the subject device meets product requirements and design specifications established for the device, including compliance with FDA-recognized standards for EMT safety, EMC, biological safety, and recommendations per FDA guidance document for software testing.

The Arthrex Synergy Endoscopic Imaging System did not require animal testing or human clinical studies to support the determination of substantial equivalence.

Based on the same intended use, the same or similar indications for use and technological characteristics, and successful completion of non-clinical testing, the Arthrex Synergy Vision Endoscopic Imaging System is as safe and effective as the legally marketed predicate devices. Any differences between the subject device and predicate devices are considered minor and do not raise different questions concerning safety and effectiveness.

Page 4 of 4

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.