(90 days)
Not Found
No
The provided text describes a single-use cannula used for injecting contrast medium, and there is no mention of AI, DNN, or ML in the document.
No
The device is used to inject contrast medium for diagnostic imaging, not to treat a disease or condition.
No
This device is used to inject contrast medium, which aids in imaging for diagnostic purposes, but the device itself does not perform the diagnostic function or interpretation of results. It is an accessory to the diagnostic process.
No
The device description clearly states it is comprised of physical cannulas and describes physical components and testing related to materials, sterilization, and mechanical performance. There is no mention of software functionality being the primary regulated component.
No
The device is used to inject contrast directly into the body, not for the examination of specimens derived from the human body.
N/A
Intended Use / Indications for Use
The Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q are intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
Product codes
ODD
Device Description
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series is comprised of nine (9) sterile, single-use, cannulas designed to inject contrast medium in the biliary or pancreatic duct when used in conjunction with a compatible endoscope.
Each device has two sections: the handle (proximal portion) and the insertion portion. The insertion portion is introduced into the biliary or pancreatic ducts through an endoscope. The distal end of the insertion portion is designed for smooth cannulation of the papilla of Vater or the minor papilla. All models are visible under fluoroscopy and feature a distal marking system.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series models are to be used with compatible endoscopes.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Biliary or pancreatic duct
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following performance data were provided to demonstrate substantial equivalence:
• Biocompatibility testing per ISO 10993-1:2018 including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP )
• Sterilization validation per ISO 11135:2014
• Ethylene oxide residuals per ISO 10993-7:2008
• Packaging validation and shelf life testing in accordance with ISO 11607-1:2019 and ASTM F1980-21
• Mechanical testing and comparative testing to verify device performance:
o Insertion force/ Withdrawal force
o Insertion w/ Stylet
o Attachment and detachment of the hook
o Contrast medium infusion
o Connection strength
o Visibility
o Human Factors Testing
Non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
510(k) Premarket Notification
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
May 27, 2025
Olympus Medical Systems Corp.
℅ Roshana Ahmed
Program Manager, Regulatory Affairs
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581
Re: K250573
Trade/Device Name: Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/
V427Q/ V434Q/V435Q;
Single Use 2-Lumen Cannula V PR-V614M
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: ODD
Dated: February 26, 2025
Received: February 26, 2025
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above
and have determined the device is substantially equivalent (for the indications for use stated in the enclosure)
to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment
date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the
provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket
approval application (PMA). You may, therefore, market the device, subject to the general controls
provisions of the Act. Although this letter refers to your product as a device, please be aware that some
cleared products may instead be combination products. The 510(k) Premarket Notification Database
available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination
product submissions. The general controls provisions of the Act include requirements for annual registration,
listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We
remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be
subject to additional controls. Existing major regulations affecting your device can be found in the Code of
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements
concerning your device in the Federal Register.
Page 2
K250573 - Roshana Ahmed Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA
guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part
820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a
change requires premarket review, the QS regulation requires device manufacturers to review and approve
changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and
approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal
statutes and regulations administered by other Federal agencies. You must comply with all the Act's
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for
devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart
A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections
531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent
parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule").
The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label
and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the
dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR
830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device
Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including
information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K250573 - Roshana Ahmed Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250573
Device Name
Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q;
Single Use 2-Lumen Cannula V PR-V614M
Indications for Use (Describe)
The Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q are intended to be used to
inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or
pancreatic duct in combination with an endoscope.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Page 5
Traditional 510(k)
Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series
Page 1 of 6
510(k) Summary
I. Submitter
Olympus Medical Systems Corporation
2951 Ishikawa-cho
Hachiochi-shi
Tokyo 192-8507
Japan
Contact Person: Seiko Yunoki
Phone: +81 42-642-2111
Date Prepared: February 10, 2025
II. Device
| Device Proprietary Name | Single Use Cannula V PR-V223Q/ V414Q/ V416Q/ V418Q/ V420Q/ V427Q/ V434Q/ V435Q
Single Use 2-Lumen Cannula V PR-V614M |
|------------------------|-------------------------------------------------|
| Common or Usual Name | ERCP Cannula |
| Classification Name | Endoscope and accessories |
| Regulation Number | 876.1500 |
| Product Code | ODD |
| Device Classification | Class II |
III. Predicate Device
Substantial equivalence is claimed to the following device:
• ERCP Catheters (ERCP, GT, FD-GT, HRC, and SUC), K171989, Wilson-Cook Medical Inc.
IV. Device Description
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series is comprised of nine (9) sterile, single-use, cannulas designed to inject contrast medium in the biliary or pancreatic duct when used in conjunction with a compatible endoscope.
Page 6
Each device has two sections: the handle (proximal portion) and the insertion portion. The insertion portion is introduced into the biliary or pancreatic ducts through an endoscope. The distal end of the insertion portion is designed for smooth cannulation of the papilla of Vater or the minor papilla. All models are visible under fluoroscopy and feature a distal marking system.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series models are to be used with compatible endoscopes.
V. Indications for Use
The Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q are intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
VI. Comparison of Technological Characteristics
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series models have the same intended use, and principles of operation as the predicate devices. The subject device and predicate devices are sterile (via EtO), single use, cannulas which are used with compatible endoscopes. The cannulas are offered with and without stylets and fluoro tips and feature the same distal tip diameters as the predicate devices.
The subject and predicate devices differ as follows:
• the subject device cannulas have a larger maximum insertion portion diameter than the predicate devices.
• the working length of the subject device cannulas is shorter than the working length of the predicate devices.
The differences in insertion portion diameter and working length do not raise different questions of safety or effectiveness when comparing the subject device to the predicate device. Non-clinical testing demonstrates that the slight differences in device design do not alter the safety, efficacy, or performance of the subject devices when compared to the predicate devices.
Page 7
A comparison of the subject and predicate device technological characteristics is provided below.
| | Single Use Cannula V and Single Use 2-Lumen Cannula V PR Series
(PR-V223Q, PR-V414Q, PR-V416Q, PR-V418Q, PR-V420Q, PR-V427Q, PR-V434Q, PR-V435Q, PR-V614M) | ERCP Catheters
(Classic ERCP Catheter, Glo-Tip® ERCP Catheter, Fusion® ERCP Catheter with DomeTip®, Haber RAMP™ Catheter, and Soehendra Universal Catheter) | Analysis |
|-----|-----|-----|-----|
| 510(k) Number | TBD | K171989 | N/A |
| Indications for Use Statement | The Single Use Cannula V PR-V223Q/V414Q/V416Q/ V418Q/ V420Q/V427Q/V434Q/V435Q are intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope. | This device is used for endoscopic cannulation of the ductal system. This device is indicated for adult use only. | Substantially equivalent |
| Models | PR-V223Q
PR-V414Q
PR-V416Q
PR-V418Q
PR-V420Q
PR-V427Q
PR-V434Q
PR-V435Q
PR-V614M | Huibregtse-Katon® ERCP Catheter
Glo-Tip® ERCP Catheter
Glo-Tip II® Double Lumen ERCP Catheter with Radiopaque Bands | Substantially equivalent |
| Distal tip diameter (Fr) | PR-V223Q: 6 Fr
PR-V414Q: 4.5 Fr
PR-V416Q: 4 Fr
PR-V418Q: 3.5 Fr
PR-V420Q: 3.5 Fr
PR-V427Q: 2.5 Fr
PR-V434Q: 4 Fr
PR-V435Q: 3.5 Fr
PR-V614M: 4.5 Fr | Huibregtse-Katon® ERCP Catheter:
• G57289: 5.4Fr
Glo-Tip® ERCP Catheter:
• G22302 (GT-5-4-3): 3 Fr
• G21452(GT1-1-LT): 3.5 Fr
• G21500(GT-1-ST): 3.5 Fr
• G22093(GT-1): 5.5 Fr
• G22621(GT-1-T): 4.5Fr
• G22719(GT-1-TE): 4.5 Fr
• G21965(GT-1-T-ANG): 4.5 Fr
• G22305(GT-1-UT): 4 Fr | Substantially equivalent |
Page 8
| | | Glo-Tip II® Double Lumen ERCP Catheter with Radiopaque Bands:
• G24694: 6Fr | |
|-----|-----|-----|-----|
| Maximum insertion portion diameter (mm) | PR-V223Q: ø 2
PR-V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q: ø 2.1
PR-V614M: ø 2.5 | Huibregtse-Katon® ERCP Catheter:
• G57289: ø 1.8
Glo-Tip® ERCP Catheter:
• G22302 (GT-5-4-3): ø 1.83
• G21452 (GT1-1-LT): ø 1.83
• G21500 (GT-1-ST): ø 1.83
• G22093 (GT-1): ø 1.83
• G22621 (GT-1-T): ø 1.83
• G22719 (GT-1-TE): ø 1.83
• G21965 (GT-1-T-ANG): ø 1.83
• G22305 (GT-1-UT): ø 1.83
Glo-Tip II® Double Lumen ERCP Catheter with Radiopaque Bands:
• G24694: ø 2.0 | Substantially equivalent |
| Working length (mm) | PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q:
• 1950
PR-V614M
• 1700 | Huibregtse-Katon® ERCP Catheter:
• G57289: 1950
Glo-Tip® ERCP Catheter:
• G22302: 2000
• G21452 (long taper): 2000
• G21500 (short taper): 2000
• G22093 (standard): 2000
• G22621 (taper): 2000
• G22719 (taper): 3200
• G21965: 2000
• G22305: 2000
Glo-Tip II® Double Lumen ERCP Catheter with Radiopaque Bands:
• G24694: 2000 | Substantially equivalent |
| Stylet and tip design | PR-V223Q
• Stylet with ball tip | Huibregtse-Katon® ERCP Catheter:
• G57289: With Stylet and Ball tip | Identical |
Page 9
| | PR-V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q:
• Stylet with fluoro tip
PR-V614M
• Fluoro tip | Glo-Tip® ERCP Catheter:
• G22302 (GT-5-4-3): With Stylet and Fluoro tip
• G21452(GT1-1-LT): With Stylet and Fluoro tip
• G21500(GT-1-ST): With Stylet and Fluoro tip
• G22093(GT-1): With Stylet and Fluoro tip
• G22621(GT-1-T): With Stylet and Fluoro tip
• G22719(GT-1-TE): With Stylet and Fluoro tip
• G21965(GT-1-T-ANG): With Stylet and Fluoro tip
• G22305(GT-1-UT): With Stylet and Fluoro tip
Glo-Tip II® Double Lumen ERCP Catheter with Radiopaque Bands:
• G24694: With Stylet and Fluoro tip | |
|-----|-----|-----|-----|
| Sterilization | Ethylene Oxide | Ethylene Oxide | Identical |
| Single-Use/ Reusable | Single Use | Single Use | Identical |
VII. Performance Data
The following performance data were provided to demonstrate substantial equivalence:
• Biocompatibility testing per ISO 10993-1:2018 including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP )
• Sterilization validation per ISO 11135:2014
• Ethylene oxide residuals per ISO 10993-7:2008
• Packaging validation and shelf life testing in accordance with ISO 11607-1:2019 and ASTM F1980-21
• Mechanical testing and comparative testing to verify device performance:
o Insertion force/ Withdrawal force
o Insertion w/ Stylet
o Attachment and detachment of the hook
o Contrast medium infusion
Page 10
o Connection strength
o Visibility
o Human Factors Testing
Clinical data is not required to demonstrate substantial equivalence.
VIII. Conclusion
The non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device. Therefore, it is concluded that the Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series are substantially equivalent to the identified predicate device.