K Number
K250431
Date Cleared
2025-05-12

(87 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

Medical Diode Laser Hair Removal System adopts 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. The Medical Diode Laser Hair Removal System utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch.

AI/ML Overview

Based on the provided FDA 510(k) clearance letter for the "Medical Diode Laser Hair Removal System (MNLT-D1)", it is important to note that no in-vivo clinical studies were conducted or presented in this submission to demonstrate device performance against acceptance criteria in a clinical setting. The clearance is based on substantial equivalence to a predicate device, supported by non-clinical (bench) testing and compliance with relevant safety and performance standards.

Therefore, the following answers describe the absence of a study proving the device meets acceptance criteria in a clinical sense, and instead focus on what was presented:


Acceptance Criteria and Device Performance (Non-Clinical)

Since no clinical study was conducted as part of this 510(k) submission to demonstrate device performance in terms of hair removal efficacy or permanent hair reduction against clinical acceptance criteria, a table showing "reported device performance" in that context cannot be generated.

Instead, the acceptance criteria and performance were focused on non-clinical tests to demonstrate compliance with recognized standards and substantial equivalence to a predicate device in terms of safety and technical specifications.

Table: Non-Clinical Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (based on standards/comparison)Reported Device Performance (Compliance)
Electrical SafetyCompliance with IEC 60601-1:2020 and IEC 60601-2-22:2012Complied with IEC 60601-1:2020, IEC 60601-2-22:2012
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2020Complied with IEC 60601-1-2:2020
Laser SafetyCompliance with IEC 60601-2-22:2012 and IEC 60825-1:2007 (for Class IV lasers)Complied with IEC 60601-2-22, IEC 60825
Biocompatibility - CytotoxicityNo evidence of cytotoxicity (ISO 10993-5)No Cytotoxicity
Biocompatibility - SensitizationNo evidence of sensitization (ISO 10993-10)No evidence of sensitization
Biocompatibility - IrritationNo evidence of irritation (ISO 10993-23)No evidence of irritation
Functional Specifications (comparison to predicate)Spot Size, Fluence, Frequency, Pulse Duration within acceptable differences compared to predicate device K221312Reported differences were analyzed and deemed not to affect safety/effectiveness (e.g., spot size 12mm x 21mm vs 12.6mm x 20.6mm; fluence 1-65J/cm² vs 1-70J/cm²). All deviations rationale accepted by FDA.
Physical Specifications (comparison to predicate)Power Supply, Dimension, Weight differences acceptable and compliant with IEC 60601-1 mechanical performanceDifferences deemed not to affect safety/effectiveness (e.g., compliant with IEC 60601-1).

Study Details (Relevant to Non-Clinical Testing and Equivalence)

  1. Sample size used for the test set and the data provenance:

    • No clinical test set was used.
    • For the non-clinical tests (e.g., electrical safety, EMC, laser safety, biocompatibility), the device itself (Medical Diode Laser Hair Removal System MNLT-D1) served as the "sample." These tests are typically conducted in a laboratory setting on one or a few units to ensure design compliance.
    • Data provenance is from non-clinical laboratory testing reports conducted by the manufacturer (Shandong Moonlight Electronics Tech Co.,Ltd.) or their designated testing facilities in China. The documentation implicitly indicates these were prospective tests performed for regulatory submission.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Since no clinical study or test set requiring human assessment of performance (like image interpretation or clinical outcomes) was conducted, there was no need for experts to establish "ground truth" in the typical sense for clinical performance.
    • The "ground truth" for non-clinical tests is established by objective measurements and adherence to international standards (e.g., IEC, ISO). The expertise involved would be that of test engineers, metrologists, and quality assurance personnel performing and verifying the tests.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set requiring human rater adjudication was involved. Non-clinical compliance testing relies on quantitative measurements and pass/fail criteria from international standards.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is a laser hair removal system, not an AI-assisted diagnostic tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a hardware medical device; it is not an AI algorithm. Its performance is inherent to its physical and functional specifications.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests, the "ground truth" was based on objective, verifiable measurements against the specified requirements of international consensus standards (e.g., maximum allowable leakage current, specific electromagnetic emission limits, defined biocompatibility thresholds).
  7. The sample size for the training set:

    • Not applicable. There was no "training set" as this is not an AI/machine learning device.
  8. How the ground truth for the training set was established:

    • Not applicable. As there was no training set, no ground truth needed to be established for it.

Summary of Reliance for Clearance:

The FDA clearance for the Medical Diode Laser Hair Removal System (MNLT-D1) was based on:

  • Demonstration of substantial equivalence to an existing legally marketed predicate device (K221312) in terms of indications for use, fundamental technology (diode laser), and principle of operation.
  • Non-clinical testing to prove the device's compliance with electrical safety, EMC, laser safety, and biocompatibility standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10, ISO 10993-23).
  • Engineering analysis confirming that any differences in technical specifications (e.g., spot size, fluence, pulse duration) compared to the predicate device did not raise new questions of safety or effectiveness.

The submission explicitly states: "No clinical study is included in this submission." This indicates that the clearance for this specific device relied entirely on non-clinical data and the regulatory pathway of substantial equivalence.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 12, 2025

Shandong Moonlight Electronics Tech Co.,Ltd.
℅ Ray Wang
General Manager
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd.,
FangShan District
Beijing, 102401
China

Re: K250431
Trade/Device Name: Medical Diode Laser Hair Removal System (MNLT-D1)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: February 14, 2025
Received: February 14, 2025

Dear Ray Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250431 - Ray Wang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250431 - Ray Wang
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

YAN FU -S Digitally signed by YAN FU -S
Date: 2025.05.12 16:40:48 -04'00'

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

K250431

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250431

Device Name
Medical Diode Laser Hair Removal System (MNLT-D1)

Indications for Use (Describe)
The Medical Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Premarket Notification 510(k) Submission - 510(k) Summary

Page 1 of 5

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

The assigned 510(k) Number: K250431

1. Date of Preparation: 04/03/2025

2. Sponsor

Shandong Moonlight Electronics Tech Co.,Ltd.
NO.30, LECHUAN, XINHUA ROAD, ECONOMIC DEVELOPMENT ZONE, WEIFANG, SHANDONG, CHINA.

Contact Person: Jvqiang Xin
Position: General Manager
Tel: +86-15689898388
Email: ng585858@126.com

3. Submission Correspondent

Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China, 102401

Contact Person: Ray Wang
Position: General Manager
Tel: +86-18910677558
Fax: +86-10-56335780
Email: information@believe-med.com

4. Proposed Device Identification

Trade Name: Medical Diode Laser Hair Removal System
Common Name: Powered Laser Surgical Instrument
Model(s): MNLT-D1

Regulatory Information:
Classification Name: Powered Laser Surgical Instrument
Classification: II
Product Code: GEX

Page 6

Premarket Notification 510(k) Submission - 510(k) Summary
Page 2 of 5

Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Review Panel: General & Plastic Surgery

Indication For Use Statement:
The Medical Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

5. Device Description

Medical Diode Laser Hair Removal System adopts 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. The Medical Diode Laser Hair Removal System utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch.

6. Predicate Device Identification

Predicate Device:
510(k) Number: K221312
Product Name: Diode Laser Hair Removal Device
Model: M-I-X MEDICAL
Manufacturer: Guangzhou CHUANG ZAO MEI Technology Co., Ltd

7. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1: 2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

Page 7

Premarket Notification 510(k) Submission - 510(k) Summary
Page 3 of 5

  • IEC 60601-1-2: 2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
  • IEC 60825-1: 2007 Safety of laser products - Part 1: Equipment classification, and requirements
  • ISO 10993-5: 2009 Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10: 2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
  • ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation

8. Clinical Test Conclusion

No clinical study is included in this submission.

9. Substantially Equivalent (SE) Comparison

Table 1 General Comparison

ITEMProposed Device Model: MNLT-D1Predicate Device (K221312) Model: M-I-X MEDICALRemark
Product CodeGEXGEXSAME
Regulation No.21 CFR 878.481021 CFR 878.4810SAME
Class22SAME
Indications for useThe Medical Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.SAME
ConfigurationMain UnitMain UnitSAME
HandpieceHandpieceSAME
Foot ControlFoot ControlSAME
Principle of OperationDiode LaserDiode LaserSAME

Page 8

Premarket Notification 510(k) Submission - 510(k) Summary
Page 4 of 5

Laser TypeDiode LaserDiode LaserSAME
Laser ClassificationClass IVClass IVSAME
Laser wavelength808 nm808 nmSAME
Spot Size12mm × 21mm12.6mm × 20.6mmAnalysis 1
Fluence1-65J/cm²1-70J/cm²Analysis 2
Frequency1-10 Hz1-10HzSAME
Pulse Duration10-320ms3-320msAnalysis 3
Power SupplyAC 110V, 50Hz/60HzAC 120V/60HzAnalysis 4
Dimension1180×465×465mm393mm×430mm×1130mmAnalysis 4
Weight74kg48kgAnalysis 4

Analysis 1:
The proposed device is different in Spot Size from the predicate, Spot size only affects the area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness.

Analysis 2:
The proposed device only has slight difference in Fluence with the predicate device. The fluence of the proposed device is within the allowable error range of the predicate device, which can justify that the difference in the parameter of fluence will not raise new safety issues of the proposed device. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device. So the slight difference is considered to have no effect on effectiveness and safety.

Analysis 3:
The proposed device is difference in Pulse Duration with the predicate device. The pulse duration of the proposed device is within the range of that of the predicate device, which is similar to the proposed device's pulse duration, so this difference will not affect safety and effectiveness of the proposed device.

Analysis 4:
The proposed device is different in Power Supply, Dimension and Weight from the predicate device. However, the power supply, dimension and weight difference are just in physical specification and this difference will not raise any issues in safety and effectiveness. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted. Therefore, this difference will not affect safety and effectiveness of the proposed device.

Table 2 Test Comparison

ITEMProposed DevicePredicate Device (K221312) Model: M-I-X MEDICALRemark
EMC, Electrical and Laser Safety

Page 9

Premarket Notification 510(k) Submission - 510(k) Summary
Page 5 of 5

Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22Comply with ANSI/AAMI ES60601-1, IEC 60601-2-22Analysis 5
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SAME
Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SAME

Patient Direct/Indirect Contact Materials and Biocompatibility

Patient Direct/Indirect Contact MaterialsAluminum and GemstoneTip of Handle (6063 Aluminum & S1 Quartz)Analysis 6
CytotoxicityNo CytotoxicityNo CytotoxicitySAME
SensitizationNo evidence of sensitizationNo evidence of sensitizationSAME
IrritationNo evidence of irritationNo evidence of irritationSAME
Biocompatibility testing standardsComply with ISO 10993-5, ISO 10993-10, ISO 10993-23Comply with ISO 10993-5, ISO 10993-10Analysis 5

Analysis 5:
The proposed device is different in Electrical Safety Testing Standard and Biocompatibility Testing Standard from the predicate device. The proposed device was tested according to IEC 60601-1:2020, ISO 10993-5:2009, ISO 10993-10:2021 and ISO 10993-23:2021, which are FDA recognized standard. Therefore, this different will not affect safety and effectiveness of the proposed device.

Analysis 6:
The subject and predicate device have different Patient Direct/Indirect Contact Materials, the contact material of the subject device is aluminum and gemstone. By complying with ISO 10993-5, ISO 10993-10, ISO10993-23, the material of proposed device is determined to be accepted. Therefore, this different will not affect safety and effectiveness of the proposed device.

10. Substantially Equivalent (SE) Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device (K221312).

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.