K Number
K250389
Device Name
XNY Disposable Gastric Calibration Tube
Date Cleared
2025-04-10

(58 days)

Product Code
Regulation Number
876.5980
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.
Device Description
The XNY Disposable Gastric Calibration Tubes are indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. The device has the advantages of accurately determining residual gastric capacity and easy positioning, and it can be connected to a suction device to extract the gas or liquid in the patient's stomach.
More Information

Not Found

No.

The document describes a medical device, a disposable gastric calibration tube, that performs mechanical functions (e.g., decompressing the stomach, draining fluid, sizing the gastric pouch). The descriptions of its function, design, and performance testing do not mention any computational, analytical, or learning capabilities typical of AI models. There are explicit "Not Found" mentions for AI, DNN, or ML, and for descriptions of training or test sets, which would be crucial for an AI model.

Yes
Explanation: The device is used in gastric and bariatric surgical procedures to decompress the stomach and drain/remove gastric fluid, which are therapeutic interventions.

No

The device is described as assisting in surgical procedures by decompressing the stomach, draining fluids, and sizing the gastric pouch. Its functions are mechanical and procedural, not involving the diagnosis of a disease or condition. The performance studies focus on physical properties and functionality rather than diagnostic accuracy.

No

The device is a physical medical device (gastric calibration tube) and the performance studies focus on mechanical and biocompatibility properties, not software functionality.

No.
The device is a physical tube designed to decompress the stomach, drain fluid, and size the gastric pouch during surgical procedures; it does not perform in vitro examination of specimens derived from the human body.

N/A

Intended Use / Indications for Use

XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.

Product codes

KNT

Device Description

The XNY Disposable Gastric Calibration Tubes are indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. The device has the advantages of accurately determining residual gastric capacity and easy positioning, and it can be connected to a suction device to extract the gas or liquid in the patient's stomach.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

gastric

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Bench studies were performed to evaluate the safety and effectiveness of the device. Tests performed include: Firmness of the balloon, The firmness between the suction port and the catheter, The catheter aspiration function, The firmness between suction port 2 (adapter) and catheter, The firmness of the inflation valve, Balloon leakage test, Biocompatibility Testing, Balloon Inspection, Tube Dimension, Tube scale mark inspection, Tube resistance to bend and break, Firmness of insertion tip and tube, The firmness between suction port 1(Adapter 1) and suction port 2(Adapter 2), and Pipe Clip Sealability Test. All tests passed their respective acceptance criteria.

Key Metrics

Not Found

Predicate Device(s)

K221898

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

FDA 510(k) Clearance Letter - XNY Disposable Gastric Calibration Tube

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 10, 2025

Changzhou Xin Neng Yuan Medical Stapler Co., Ltd.
℅ Jenny Xia
Director
LSI International Inc
504E Diamond Ave., Suite H
Gaithersburg, Maryland 20877

Re: K250389
Trade/Device Name: XNY Disposable Gastric Calibration Tube
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube And Accessories
Regulatory Class: Class II
Product Code: KNT
Dated: February 9, 2025
Received: February 11, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250389 - Jenny Xia Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250389 - Jenny Xia Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anthony Lee -S

Anthony C. Lee Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250389

Device Name: XNY Disposable Gastric Calibration Tube

Indications for Use (Describe)

XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510k Summary

K250389

1. General Information

| Applicant Information | Changzhou Xin Neng Yuan Medical Stapler Co., Ltd.
No.51 Shuishan Rd, Zhonglou Economic Development District
Changzhou, Jiangsu, China
www.xnywhq.com
whq@xnywhq.com |
|---------------------------|---|
| Contact Person | Jenny Xia
LSI International Inc
504E Diamond Ave., Suite H
Gaithersburg, MD 20877
jxia@lsi-consulting.org
301-525-6856 |
| Date Summary Prepared | February 9, 2025 |
| Device Common Name | Disposable Gastric Calibration Tubes |
| Device Trade Name | XNY Disposable Gastric Calibration Tubes |
| FDA Product Code | KNT |
| Classification Name | 876.5980 Gastrointestinal tube and accessories |
| Classification | Class II |
| Predicate Device | Gastrointestinal Boundary Identifier (GIBI HD™), K221898 |

2. Device Description

The XNY Disposable Gastric Calibration Tubes are indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. The device has the advantages of accurately determining residual gastric capacity and easy positioning, and it can be connected to a suction device to extract the gas or liquid in the patient's stomach.

3. Indication for Use

XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.

4. Comparison of Technological Characteristics with the Predicate Device

| Feature | Predicate Device
Calibration Tube,
K221898
| Subject Device
Calibration Tubes
K250389
| Effect on Substantial
Equivalence
|
|-------------|-----------------------------------------------------|---------------------------------------------------|------------------------------------------|
| Indications for Use | The GIBI HD™ is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide. | Product is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. | None |
| Outer Diameter | B-2032: 10.5mm +0/-0.3mm | JZGA/B-32FR: 10.5mm±0.5mm | Outer diameters are similar and no impact to |

Page 6

K250389 Page 2 of 4

| | B-2036: 11.9mm +0/-0.3mm
B-2040: 13.2mm +0/-0.3mm | JZGA/B-34FR: 11.2mm±0.5mm
JZGA/B-36FR: 11.9mm±0.5mm
JZGA/B-38FR: 12.5mm±0.5mm
JZGA/B-40FR: 13.2mm±0.5mm
JZGA/B-42FR: 13.9mm±0.5mm
JZGA/B-44FR: 14.5mm±0.5mm | substantial equivalence |
|---|---|---|---|
| Working Length | 823.7mm±13.2mm | JZGA type: 700mm±10mm
JZGB type: 900mm±20mm | No impact to substantial equivalence with the similar lengths. |
| Tubing | Dual lumen | JZGA: Dual lumen
JZGB: single lumen | Dual lumen and single lumen only choose different models according to different clinical needs, and no impact to substantial equivalence. |
| Distal Tip | Includes a tip distal to the balloon with 12 aspiration holes. | Includes a tip distal to the balloon with 3 aspiration holes. | The number of suction holes has no substantial effect on the actual use. No impact to substantial equivalence |
| Connector for Suction | The catheter includes an adapter for room suction | There is a matching adapter, which can be adapted to various connectors | None |
| Balloon + inflation Valve | The balloon has an inflation capacity ≥ 100 cc min. The catheter is attached to a 6-inch tubing with a stopcock for filling the balloon. | JZGA type balloon with inflation capacity ≥100cc/min.
JZGB has no balloon | None |
| Tubing Material | Silicone | Medical grade silicone | None |
| Markings | Indication marks at 30, 35, 40, 45, and 50 centimeters. | A type indication marks at 52, 53, 54, 55, 56, 57, 58, 59, 60 centimeters;
B type indication marks at 30, 31, 32, 33…89, 90 centimeters | Marks are included and no impact to Substantial equivalence |
| Connector | Luer taper, tape adapter | Normal joint, tape joint | Similar adapters and no impact to substantial equivalence |
| Sterility | Non-sterile, disposable, single patient use | Sterile, single-use, single patient use | None |

Page 7

K250389 Page 3 of 4

5. Performance Data

Test PerformedPurposeAcceptance CriteriaResults
Firmness of the balloonThe purpose of this test is to evaluate the firmness of the balloon of Disposable gastric calibration tube.The firmness of the balloon is greater than 60N.Pass
The firmness between the suction port and the catheterThe purpose of this test is to evaluate the firmness between the suction port and the catheter of Disposable gastric calibration tube.The firmness between the suction port and the catheter is greater than 5N.Pass
The catheter aspiration functionThe purpose of this test is to evaluate catheter aspiration function of Disposable gastric calibration tube.The catheter should be able to draw out 500ml of liquid within 1 minute.Pass
The firmness between suction port 2 (adapter) and catheterThe purpose of this test is to evaluate the firmness between suction port 2 and catheter of Disposable gastric calibration tube.The firmness between suction port 2 and catheter is greater than 30N.Pass
The firmness of the inflation valveThe purpose of this test is to evaluate the firmness of the inflation valve of Disposable gastric calibration tube.The firmness of the inflation valve is greater than 20N.Pass
Balloon leakage testThe purpose of this test is to evaluate balloon leakage.Can tolerate double the working pressure of 80 ml gas without leakage or damage.Pass
Biocompatibility TestingThe purpose of this test is to evaluate the device to meet ISO 10993 on biocompatibility for Biological evaluation of medical devices.No cytotoxicity
No sensitization
No irritation.Pass
Balloon InspectionThe purpose of this test is to evaluate the appearance of the balloon of Disposable gastric calibration tube.The balloon should be colorless and transparent.
The balloon should be tightly connected to the tube and the connection between the balloon and the tube should be homogeneous and without twisting.Pass
Tube DimensionThe purpose of this test is to evaluate the tube dimension of Disposable gastric calibration tube.The tube length and diameter meet the required measurements.Pass
Tube scale mark inspectionThe purpose of this test is to evaluate the appearance of the tube scale marks on Disposable gastric calibration tube.The scale marks on the tube should be clear and not faded.Pass

Page 8

K250389 Page 4 of 4

Test PerformedPurposeAcceptance CriteriaResults
Tube resistance to bend and breakThe purpose of this test is to evaluate the Disposable gastric calibration tube's resistance to bending and breaking.Hold the tube at both ends and bend the tube so that the 1cm of long tube at both ends are parallel and in contact, hold for 15 seconds, then release, ensuring that no creases form at the bend of the catheter.Pass
Firmness of insertion tip and tubeThe purpose of this test is to evaluate the firmness of the insertion tip and tube of Disposable gastric calibration tube.The connection between the Insertion tip and the tube should be able to withstand a static axial pull of 15N for 15S without breaking or falling off.Pass
The firmness between suction port 1(Adapter 1) and suction port 2(Adapter 2) is greater than 5N.The purpose of this test is to evaluate the firmness between adapter 1 and adapter 2The firmness between adapters is greater than 5N.Pass
Pipe Clip Sealability Test.The purpose of this test is to evaluate the Pipe Clip sealability of disposable gastric calibration tubes.When the Pipe Clip is closed, no bubbles are generated at the catheter port.Pass

6. Basis for Substantial Equivalence

Substantial equivalence of the XNY Disposable Gastric Calibration Tube to the predicate device was established through an evaluation of the indications for use, principle of operation, device design, materials of construction, and an assessment of safety, and effectiveness via bench studies.

The data presented in this summary demonstrates the technological similarity and equivalency of the XNY Disposable Gastric Calibration Tube to the predicate device (GIBI HD, K221898).

The devices:

  • have the same intended use,
  • use the same principle of operation,
  • incorporate the same basic design,
  • use similar construction and material

7. Conclusion

Based on the non-clinical assessments, the comparison of the device classification, indications for use, operating principle, technological characteristics, and biocompatibility that the subject device, the XNY Disposable Gastric Calibration Tube, is substantially equivalent to the predicate device, the GIBI HD, K221898. Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.