K Number
K250310
Date Cleared
2025-06-27

(144 days)

Product Code
Regulation Number
870.1220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VIKING Fixed Curve Diagnostic Catheter is indicated for use to diagnose cardiac arrhythmias.

Device Description

The Viking Fixed Curve Diagnostic Catheters (also referred to as Viking Catheters) are intended for use in electrophysiology procedures to record, stimulate and pace in the diagnosis of cardiac arrhythmias.

The Viking Catheters receive electrical signals from the cardiac tissue, which are transmitted via the electrode wiring through the connector and cable to an amplifier, where the signals are displayed on an EP recording system for the physician to view and interpret in diagnosing cardiac arrhythmias. The physician inserts the diagnostic catheter into the vasculature percutaneously. To reach the right side of the heart, either the internal jugular vein, the subclavian veins, or the femoral veins can be accessed. To reach the left ventricle, the retrograde approach via the femoral artery can be used. In some cases, the radial artery may be used to access the targeted cardiac chamber. Using fluoroscopic guidance, the physician then navigates the diagnostic catheter to the target site using the catheter steering mechanism, as confirmed by the intracardiac electrograms and/or pacing threshold, confirming electrode to endocardium interface.

Viking Catheters are radiopaque, flexible, insulated catheters. The catheters have a polymer shaft that houses the electrical wiring that connects the electrodes to a connector at the proximal end of the catheter; each electrode wire is soldered into this connector. When the catheter is used, the connector is attached to a sterile cable that connects to the hospital recording and pacing equipment.

The Viking Catheters are offered pre-packaged sterile and are a single use product not to be reused, reprocessed, or resterilized. The method of sterilization is ethylene oxide (EO).

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the VIKING™ Fixed Curve Diagnostic Catheter do not contain any information regarding acceptance criteria and studies demonstrating device performance in the context of an AI/algorithm-based diagnostic device.

The document describes a medical device, specifically a diagnostic catheter, and the FDA's determination of substantial equivalence to a predicate device. The performance data discussed relates to:

  • Bioburden Testing: To confirm bioburden levels were not adversely affected.
  • Design Verification Testing: To demonstrate adherence to product specifications at baseline and after accelerated aging (25 months/761 days).
  • Packaging Verification Testing: To ensure packaging meets design input and maintains sterility.

These tests are standard for physical medical devices to ensure safety, sterility, functionality, and shelf-life, not for evaluating the diagnostic accuracy or effectiveness of an AI algorithm.

Therefore, I cannot provide the requested information for acceptance criteria and studies related to an AI diagnostic device because the provided text is for a physical diagnostic catheter and does not involve AI or algorithmic performance evaluation.

To directly answer your numbered points based only on the provided text, without making assumptions:

  1. A table of acceptance criteria and the reported device performance: Not applicable for AI performance from this document. The document presents performance data for physical device characteristics (bioburden, design verification, packaging verification) to support substantial equivalence, but no specific acceptance criteria or reported values are detailed in the summary itself.
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for AI performance from this document. The document refers to "Design Verification Testing" and "Packaging Design Verification testing" but does not specify sample sizes or data provenance in the context of diagnostic accuracy.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable for AI performance from this document.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for AI performance from this document.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical diagnostic catheter, not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable for AI performance from this document. The "ground truth" for the catheter's physical performance would be its adherence to engineering specifications and international standards.
  8. The sample size for the training set: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.
  9. How the ground truth for the training set was established: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.

In summary, the provided FDA document pertains to the clearance of a physical medical device (a catheter) and does not involve the evaluation of an AI algorithm or its diagnostic performance.

FDA 510(k) Clearance Letter - VIKING™ Fixed Curve Diagnostic Catheter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 27, 2025

Boston Scientific Corporation
Oliver Buttleman
Regulatory Affairs Specialist
4100 Hamline Ave North
Saint Paul, Minnesota 55112

Re: K250310
Trade/Device Name: VIKING™ Fixed Curve Diagnostic Catheter
Regulation Number: 21 CFR 870.1220
Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe
Regulatory Class: Class II
Product Code: DRF
Dated: February 3, 2025
Received: February 3, 2025

Dear Oliver Buttleman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

FDA U.S. FOOD & DRUG ADMINISTRATION

June 27, 2025

Boston Scientific Corporation
Oliver Buttleman
Regulatory Affairs Specialist
4100 Hamline Ave North
Saint Paul, Minnesota 55112

Re: K250310
Trade/Device Name: VIKING™ Fixed Curve Diagnostic Catheter
Regulation Number: 21 CFR 870.1220
Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe
Regulatory Class: Class II
Product Code: DRF
Dated: February 3, 2025
Received: February 3, 2025

Dear Oliver Buttleman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

K250310 - Oliver Buttleman Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250310 - Oliver Buttleman Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250310

Device Name
VIKING™ Fixed Curve Diagnostic Catheter

Indications for Use (Describe)
The VIKING Fixed Curve Diagnostic Catheter is indicated for use to diagnose cardiac arrhythmias.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Boston Scientific
300 Boston Scientific Way
Marlborough, MA 01752
www.bostonscientific.com

510(k) Summary for Viking Fixed Curve Diagnostic Catheters

1. Submitter

Boston Scientific Corporation
Electrophysiology Division
4100 Hamline Ave North
St. Paul, MN 55112

Contact:
Oliver Buttleman
Regulatory Affairs Specialist
Phone: 651-581-0894
E-mail: oliver.buttleman@bsci.com

Date Prepared: February-03-2025

2. Device

Name of Device(s): VIKING™ Fixed Curve Diagnostic Catheter
Common Name: Viking Fixed Curve Diagnostic Catheter
Classification Name: Electrode Recording Catheter or Electrode Recording Probe
Product Code: DRF
Device Class and Panel: Class II, Cardiovascular
Classification Regulation: 21 CFR Part 870.1220
510(k) Number: K250310

3. Predicate Device

Subject DevicePredicate DevicePredicate 510(k)
VIKING™ Fixed Curve Diagnostic CatheterVIKING DIAGNOSTIC ELECTRODE CATHETERK971265

K250310 - 510(k) Summary for Viking Fixed Curve Diagnostic Catheters 1

Page 6

Boston Scientific
300 Boston Scientific Way
Marlborough, MA 01752
www.bostonscientific.com

4. Device Description

The Viking Fixed Curve Diagnostic Catheters (also referred to as Viking Catheters) are intended for use in electrophysiology procedures to record, stimulate and pace in the diagnosis of cardiac arrhythmias.

The Viking Catheters receive electrical signals from the cardiac tissue, which are transmitted via the electrode wiring through the connector and cable to an amplifier, where the signals are displayed on an EP recording system for the physician to view and interpret in diagnosing cardiac arrhythmias. The physician inserts the diagnostic catheter into the vasculature percutaneously. To reach the right side of the heart, either the internal jugular vein, the subclavian veins, or the femoral veins can be accessed. To reach the left ventricle, the retrograde approach via the femoral artery can be used. In some cases, the radial artery may be used to access the targeted cardiac chamber. Using fluoroscopic guidance, the physician then navigates the diagnostic catheter to the target site using the catheter steering mechanism, as confirmed by the intracardiac electrograms and/or pacing threshold, confirming electrode to endocardium interface.

Viking Catheters are radiopaque, flexible, insulated catheters. The catheters have a polymer shaft that houses the electrical wiring that connects the electrodes to a connector at the proximal end of the catheter; each electrode wire is soldered into this connector. When the catheter is used, the connector is attached to a sterile cable that connects to the hospital recording and pacing equipment.

The Viking Catheters are offered pre-packaged sterile and are a single use product not to be reused, reprocessed, or resterilized. The method of sterilization is ethylene oxide (EO).

5. Intended Use & Indications for Use

Intended Use

The Viking Fixed Curve Diagnostic Catheter is intended for temporary intracardiac sensing, recording, stimulation, and temporary pacing during the evaluation of cardiac arrhythmias. The device is intended for use in adult (not pediatric) patients, with the exclusion of pregnant and/or nursing patients.

Indications for Use

The VIKING Fixed Curve Diagnostic Catheter is indicated for use to diagnose cardiac arrhythmias.

6. Comparison of Technological Characteristics with the Predicate Device

The Viking Catheters incorporate substantially equivalent design, packaging, fundamental

K250310 - 510(k) Summary for Viking Fixed Curve Diagnostic Catheters 2

Page 7

Boston Scientific
300 Boston Scientific Way
Marlborough, MA 01752
www.bostonscientific.com

technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device. The difference between the subject and predicate devices is minor modifications to certain design requirements and test methods to align with current standards and Boston Scientific's quality system.

7. Performance Data

The proposed labeling modifications of the Viking Catheters do not impact the device's substantial equivalence to the previously cleared version of this device. The device is as safe, as effective, and performs as well as the predicate device. The indications for use, intended use, classification, product functions, materials, configuration, and sterility are substantially equivalent to the predicate device.

The following performance data were provided in support of the substantial equivalence:

Bioburden Testing

Bioburden testing, per ISO 11737-1, Microbiological Methods – Part 1: Determination of a Population of Microorganisms on Products, was performed to confirm bioburden levels of the devices were not adversely affected.

Design Verification Testing

Design verification testing was performed on the Viking Catheters to demonstrate that the devices, as manufactured, packaged and 2x EO sterilized, adhere to the requirements set forth in accordance with the applicable product specifications.

Testing was conducted at baseline (T=0) and 25 months (761 days) accelerated aging to support the labeled shelf life.

Packaging Verification Testing

The Packaging Design Verification testing provides evidence that the catheter packaging configurations meet the predetermined design input specification and maintain sterility in conformance with ISO 11607-1.

8. Conclusion

The subject Viking Fixed Curve Diagnostic Catheters are substantially equivalent in indications for use, design, product function, materials, and sterility to the predicate Viking Diagnostic Electrode Catheters. The performance data supports that the subject catheters are substantially equivalent to the predicate device.

K250310 - 510(k) Summary for Viking Fixed Curve Diagnostic Catheters 3

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).