K Number
K250290

Validate with FDA (Live)

Device Name
SurgiTwin
Manufacturer
Date Cleared
2025-08-29

(210 days)

Product Code
Regulation Number
892.2050
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SurgiTwin is a web-based platform designed to help healthcare professionals carry out pre-operative planning for knee reconstruction procedures, based on their patients' imported imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The system works with a database of digital representations related to surgical materials supplied by their manufacturers. SurgiTwin generates a PDF report as an output. End users of the generated SurgiTwin reports are trained healthcare professionals. SurgiTwin does not provide a diagnosis or surgical recommendation.

Device Description

SurgiTwin is a semi-automated Software as a Medical Device (SaMD) that assists health care professionals in the pre-operative planning of total knee replacement surgery. Using a series of algorithms, the software creates 2D segmented images, a 3D model, and relevant measurements derived from the patient's pre-dimensioned medical images. The software interface allows the user to adjust the plan manually to verify the accuracy of the model and achieve the desired clinical targets. SurgiTwin generates a PDF report as an output. SurgiTwin does not provide a diagnosis or surgical recommendation.

The intended patient population is patients over 22 undergoing total knee replacement surgery without any existing material in the operated lower limb.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for SurgiTwin:

1. Acceptance Criteria and Reported Device Performance

The provided document specifically details acceptance criteria for the segmentation ML model. Other functions (automatic landmark function, metric generation, implant placement, osteophyte removal) are mentioned as having "predefined clinical acceptance criteria" and "all acceptance criteria were met," but the specific numeric criteria are not listed.

Table of Acceptance Criteria (for the Segmentation ML Model) and Reported Device Performance:

MetricAcceptance CriteriaReported Device Performance
Mean DSC (Dice Similarity Coefficient)> 0.95Met (> 0.95, implied by "met the acceptance criteria")
Mean voxel based AHD (Average Hausdorff Distance)< 1.0mmMet (< 1.0mm, implied by "met the acceptance criteria")
5th percentile of the DSC> 0.9Met (> 0.9, implied by "met the acceptance criteria")
95th percentile of the boundary based HD 95 (Hausdorff Distance 95th percentile)< 2.5mmMet (< 2.5mm, implied by "met the acceptance criteria")

2. Sample Size and Data Provenance

The document states:

  • Test Set (Validation Dataset): Not explicitly stated, but it's part of a dataset where the ML model was "tested with the remaining 19%." The total training and testing dataset size is also not explicitly stated in numerical terms (only "datasets from multiple sites").
  • Data Provenance: "Datasets from multiple sites." Institution Name and Institution Location were subgroup definitions, implying a variety of sources, but no specific countries or retrospective/prospective nature are mentioned for the test set. However, the "ML Model Development and Testing Information" sections generally imply retrospective data collection for development and testing.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: The ground truth for the segmentation ML model reference standard was established through labeling, but the document only mentions that the "validation dataset was labeled by different individuals from the training dataset." No specific qualifications (e.g., radiologist with X years of experience) are provided for these individuals.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (such as 2+1 or 3+1) for the establishment of ground truth for the test set. It only mentions labeling was done by "different individuals."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The performance studies described are focused on the device's standalone performance compared to ground truth or automatic vs. manual landmarking, not human reader improvement with AI assistance.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone performance study was clearly done for the ML model. The document states:

  • "The machine learning (ML) model incorporated into SurgiTwin was developed, trained, tested, and validated for its performance."
  • "Comparison of the performance of the segmentation ML model against the predefined ground truth met the acceptance criteria for the model performance."

This indicates that the algorithm's performance was evaluated independently against a ground truth.

Furthermore, studies for "automatic landmark function," "metrics generated by SurgiTwin," "default implant placement algorithm," and "osteophyte removal function" all imply standalone validation of these algorithmic components against predefined criteria or manual annotations by experts.

7. Type of Ground Truth Used

The ground truth used for the segmentation ML model was expert consensus (implied by "labeled by different individuals") or expert annotation. It's referred to as "predefined ground truth."

For other functions:

  • Automatic landmark function and generated metrics were compared to "manual landmark placement by expert annotators" and "manual annotations by expert annotators," respectively, which also points to expert annotation as ground truth.
  • The clinical acceptability of implant placement and osteophyte removal functions was also validated against "predefined clinical acceptance criteria," likely based on expert consensus or established clinical standards.

8. Sample Size for the Training Set

The ML model was "trained with 81% of the dataset." The total size of this dataset is not explicitly stated in numerical terms.

9. How the Ground Truth for the Training Set Was Established

The document states: "The validation dataset was labeled by different individuals from the training dataset." This implies that the training dataset was also labeled, likely by similar "individuals" (presumed experts or annotators). However, the specific methodology for establishing this ground truth for the training set (e.g., number of annotators, adjudication) is not detailed.

FDA 510(k) Clearance Letter - SurgiTwin

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 28, 2025

Twinsight
Meghan McFadden
Quality and Regulatory Affairs Engineer
Biopolis
5 avenue du Grand Sablon
La Tronche, 38700
France

Re: K250290
Trade/Device Name: SurgiTwin
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ, QIH
Dated: January 31, 2025
Received: January 31, 2025

Dear Meghan McFadden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250290 - Meghan Mcfadden Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250290 - Meghan Mcfadden Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250290

Device Name
SurgiTwin

Indications for Use (Describe)

SurgiTwin is a web-based platform designed to help healthcare professionals carry out pre-operative planning for knee reconstruction procedures, based on their patients' imported imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The system works with a database of digital representations related to surgical materials supplied by their manufacturers. SurgiTwin generates a PDF report as an output. End users of the generated SurgiTwin reports are trained healthcare professionals. SurgiTwin does not provide a diagnosis or surgical recommendation.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

510(k) Summary

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Submitter

Twinsight
5 avenue du Grand Sablon
38700 La Tronche
France

Contact Person: Meghan McFadden
Quality and Regulatory Affairs Engineer
Email: qara@twinsight-medical.com
Phone: +33 6 68 02 05 61

Date Summary Prepared: 29 August, 2025

2. Device

2.1. SurgiTwin

Trade Name/Common Name: SurgiTwin
Classification Name: Medical image management and processing system (21 C.F.R. § 892.2050)
Regulatory Class: Class II
Product Code: LLZ, QIH
510(k) Number: K250290

3. Legally Marketed Predicate Device

3.1. Peekmed web

510(k)Product NameClearance Date
K240926PeekMed webDecember 2024

510(k) Summary: SurgiTwin
1

Page 6

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

4. Device Description Summary

SurgiTwin is a semi-automated Software as a Medical Device (SaMD) that assists health care professionals in the pre-operative planning of total knee replacement surgery. Using a series of algorithms, the software creates 2D segmented images, a 3D model, and relevant measurements derived from the patient's pre-dimensioned medical images. The software interface allows the user to adjust the plan manually to verify the accuracy of the model and achieve the desired clinical targets. SurgiTwin generates a PDF report as an output. SurgiTwin does not provide a diagnosis or surgical recommendation.

The intended patient population is patients over 22 undergoing total knee replacement surgery without any existing material in the operated lower limb.

5. Intended Use/Indications for Use

SurgiTwin is a web-based platform designed to help healthcare professionals carry out pre-operative planning for knee reconstruction procedures, based on their patients' imported imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The system works with a database of digital representations related to surgical materials supplied by their manufacturers. SurgiTwin generates a PDF report as an output. End users of the generated SurgiTwin reports are trained healthcare professionals. SurgiTwin does not provide a diagnosis or surgical recommendation.

5.1. Contraindications

SurgiTwin is contraindicated as follows:

  1. Patients under 22 years of age
  2. Presence of orthopedic hardware in the limb to be operated
  3. Pathological anomaly in the limb to be operated including fracture, tumor, or significant bone loss
  4. Patients with severe limb deformities
  5. Contraindications for each implant as indicated by the implant manufacturer

5.2. Indications for Use Comparison

SurgiTwin assists the surgeon in pre-surgical planning of the knee, while the predicate allows planning in the knee, hip, upper limb, and foot. The anatomical region and surgical procedure covered in SurgiTwin are included in its predicate. Comparison of the indications for use therefore supports the substantial equivalence of SurgiTwin and its predicate.

510(k) Summary: SurgiTwin
2

Page 7

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

6. Technological Comparison to Predicate

SurgiTwin has been evaluated in comparison to its predicate device regarding intended use, indications for use, design, function, and technology. Both the subject device and its predicate are medical software designed to assist healthcare professionals in orthopedic pre-surgical planning for the adult musculoskeletal system within a clinical setting. Appropriate clinical judgment and experience are mandatory for the use of both devices.

The two devices follow the same workflow, share similar use requirements (such as internet connectivity and user verification and approval of outputs), and offer comparable planning functionalities, including model representation, digital overlap of prosthetic material, and support for both 2D and 3D environments. Additionally, both devices generate a final planning report that incorporates selected images with templates, measurements, and textual descriptions of the patient and/or the planned surgical procedure.

The analysis of technological differences between SurgiTwin and the predicate device performed in the context of this 510(k) submission supports substantial equivalence. A summary of this comparison is presented in the table below.

510(k) Summary: SurgiTwin
3

Page 8

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

Table 1: Summary of predicate and subject device characteristics and rationale for substantial equivalence

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
Product CodeLLZ, QIHLLZ, QIHYes
Regulation Number21 CFR 892.205021 CFR 892.2050Yes
Regulation NameMedical Image Management And Processing SystemMedical Image Management And Processing SystemYes
Intended UsePeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.SurgiTwin is a web-based platform designed to help healthcare professionals carry out pre-operative planning for knee reconstruction procedures, based on their patients' imported imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning. The system works with a database of digital representations related to surgical materials supplied by their manufacturers. SurgiTwin generates a PDF report as an output. End users of the generated SurgiTwin reports are trained healthcare professionals. SurgiTwin does not provide a diagnosis or surgical recommendation.Yes See justificationThe last three sentences have been added for clarification but do not alter the intended use compared to the predicate.

510(k) Summary: SurgiTwin
4

Page 9

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
Indications for UseThis medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical specialties within surgery and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population.This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for total knee arthroplasty procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for medical specialties within surgery and no specific use environment is mandatory. The typical use environment is a room with a computer. The intended patient group is patients over 22 years old identified to require knee reconstruction surgery without any existing material in the operated lower limb.Yes See justificationSurgiTwin allows the surgeon to perform the pre-surgical planning efficiently in the knee, while the predicate allows planning in the hip, knee, upper limb, and foot. The anatomical region and surgical procedure covered in SurgiTwin are included in its predicate. Comparison of the indications for use therefore supports substantial equivalence.

510(k) Summary: SurgiTwin
5

Page 10

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
ContraindicationsNo contraindications specific to this device.SurgiTwin is contraindicated as follows: - Patients under 22 years of age - Presence of orthopedic hardware in the joint to be operated - Pathological anomaly in the limb to be operated including fracture, tumor, or significant bone loss - Patients with severe limb deformities - Contraindications for each implant as given by the implant manufacturerYes See justificationContraindications are added as a precautionary measure to enhance user awareness and ensure safe and appropriate use of the device within its intended population. They do not indicate a fundamental difference in SurgiTwin's technological characteristics, intended use, or performance compared to the predicate.
Clinical PurposePeekMed web allows the surgeon to efficiently perform orthopedic pre-surgical planning in the musculoskeletal systemSurgiTwin allows the surgeon to efficiently perform orthopedic pre-surgical planning in the kneeYes See justificationSurgiTwin allows the surgeon to perform the pre-surgical planning efficiently in the knee, while PeekMed® web allows planning in the hip, knee, upper limb, and foot. The anatomical region and surgical procedure covered in SurgiTwin are included in its predicate. Comparison of the general purpose therefore supports substantial equivalence.
Anatomical RegionsPeekMed web allows the surgeon to perform the pre-surgical planning in the following anatomical regions: - Hip - Knee - Upper limb - FootSurgiTwin allows the surgeon to efficiently perform orthopedic pre-surgical planning in the kneeYes See justificationSee justification above.

510(k) Summary: SurgiTwin
6

Page 11

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
ContraindicationsNo contraindications specific to this device.SurgiTwin is contraindicated as follows: - Patients under 22 years of age - Presence of orthopedic hardware in the joint to be operated - Pathological anomaly in the limb to be operated including fracture, tumor, or significant bone loss - Patients with severe limb deformities - Contraindications for each implant as given by the implant manufacturerYes See justificationContraindications are added as a precautionary measure to enhance user awareness and ensure safe and appropriate use of the device within its intended population. They do not indicate a fundamental difference in SurgiTwin's technological characteristics, intended use, or performance compared to the predicate.
Clinical PurposePeekMed web allows the surgeon to efficiently perform orthopedic pre-surgical planning in the musculoskeletal systemSurgiTwin allows the surgeon to efficiently perform orthopedic pre-surgical planning in the kneeYes See justificationSurgiTwin allows the surgeon to perform the pre-surgical planning efficiently in the knee, while PeekMed® web allows planning in the hip, knee, upper limb, and foot. The anatomical region and surgical procedure covered in SurgiTwin are included in its predicate. Comparison of the general purpose therefore supports substantial equivalence.
Anatomical RegionsPeekMed web allows the surgeon to perform the pre-surgical planning in the following anatomical regions: - Hip - Knee - Upper limb - FootSurgiTwin allows the surgeon to efficiently perform orthopedic pre-surgical planning in the kneeYes See justificationSee justification above.

510(k) Summary: SurgiTwin
7

Page 12

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
Patient PopulationAdultsAdultsYes
End UsersHealthcare ProfessionalsHealthcare ProfessionalsYes
Device AvailabilitySoftware is cloud-based (not installable) and can be displayed on any personal device or workstation that can run on a web browserSoftware is cloud-based (not installable) and can be displayed on any personal device or workstation that can run on a web browserYes
Software ArchitectureDistributed system (cloud-based). This distributed system is a combination of software modules placed on servers that are able to communicate with each other.Distributed system (cloud-based). This distributed system is a combination of software modules placed on servers that are able to communicate with each other.Yes
WorkflowThe workflow is as follows: Import case images, configure images, identify the case, pre-surgical planning, and export the case.The workflow is as follows: Import case images, configure images, identify the case, pre-surgical planning, and export the case.Yes
Internet ConnectionRequiredRequiredYes
Image SourceReceives medical images from various sourcesReceives medical images from various sourcesYes
Data ProcessingThe software processes data to provide an overlap and dimensioning of digital representations of the prosthetic materialThe software processes data to provide an overlap and dimensioning of digital representations of the prosthetic materialYes
Digital overlap of templatesAllows the overlap of models and the intersection of the modelsAllows the overlap of models and the intersection of the modelsYes
Interactive model positioningYesYesYes

510(k) Summary: SurgiTwin
8

Page 13

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
Interactive model dimensioningYesYesYes
Model RotationYesYesYes
Support for digital prosthetic materials provided by the manufacturersYesYesYes
Contact with the patientNoNoYes
Control of life supporting devicesNoNoYes
Human intervention for image interpretationYesYesYes
Tools for surgical simulation and planningYesYesYes
Preoperative annotation and analysisYesYesYes
Provides values for measurementYes, including distance and angle measurementYes, including distance and angle measurementYes
Automatic bone segmentationYesYesYes
Machine learning models for image segmentationYesYesYes

510(k) Summary: SurgiTwin
9

Page 14

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

CharacteristicPeekmed web K240926SurgiTwin Subject Device K250290Substantially Equivalent?Justification and rationale
MPR ViewYesYesYes
Automatic placement of anatomical landmarksYesYesYes
Approval of anatomical landmarks by userYesYes
Modification of anatomical landmarksYesNoYes See justificationSee discussion of technological differences below.

510(k) Summary: SurgiTwin
10

Page 15

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

6.1. Discussion of Technological Differences

The device and its predicate have the following differences:

  1. SurgiTwin supports automatic landmark placement while PeekMed web supports both manual and automatic landmark placement. Landmark placement is required to be validated by the qualified orthopedic surgeon using the software in order to generate the planning results.
  2. SurgiTwin allows the surgeon to perform the pre-surgical planning in the knee, while PeekMed® web allows planning in the hip, knee, upper limb, and foot.

The assessment of these technological differences supports substantial equivalence of SurgiTwin to its predicate because:

  1. The automatic landmarking feature is designed to streamline the placement of landmarks on each bone before surgical planning, but is not intended to provide medical advice on their positioning. The qualified user must review and validate the landmark placement before proceeding with surgical planning. In addition, performance testing and validation confirmed that the automatic landmarking feature is more accurate than that of the predicate. The requirement for user approval ensures that clinicians review landmark placement before continuing.
  2. The anatomical region and surgical procedure covered in SurgiTwin are included in its predicate.

7. Performance Data

Nonclinical performance testing performed on SurgiTwin supports substantial equivalence to the predicate device. Testing was performed in accordance with the following FDA guidance documents:

  • Content of Premarket Submissions for Device Software Functions
  • Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
  • Applying Human Factors and Usability Engineering to Medical Devices
  • Technical Performance Assessment of Quantitative Imaging in Radiological Device Premarket Submissions
  • Off-The-Shelf Software Use in Medical Devices

The following testing was performed:
A. Verification activities to ensure that all features were correctly implemented to meet system requirements and fulfill acceptance criteria.
B. The machine learning (ML) model incorporated into SurgiTwin was developed, trained, tested, and validated for its performance.
a. ML Model Development and Testing Information

510(k) Summary: SurgiTwin
11

Page 16

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

ML models were developed with datasets from multiple sites. The ML model was trained with 81% of the dataset and tested with the remaining 19%. This dataset was designed to represent the intended use population.

b. Subgroup Definition
Datasets were divided according to the following subgroups:

  • Demographics
    • Patient Sex
    • Patient Age
  • Equipment and protocols for image collection
    • Institution Name
    • Institution Location
    • Manufacturer
    • Image slice thickness

The CT image parameters in the final test dataset fall within the acceptable input specifications for the SurgiTwin system:

  • 0.3 mm to 5 mm pixel spacing and slice interval for hip-only and ankle-only scans
  • 0.3 mm to 1.0 mm for scans including the knee region

c. Acceptance Criteria
The acceptance criteria for the segmentation ML model are shown in the following table:

MetricAcceptance Criteria
Mean DSC> 0.95
Mean voxel based AHD< 1.0mm
5th percentile of the DSC> 0.9
95th percentile of the boundary based HD 95< 2.5mm

d. Reference Standard
Comparison of the performance of the segmentation ML model against the predefined ground truth met the acceptance criteria for the model performance.

e. Independence of Training and Validation Data
To prevent data leakage, external validation datasets were collected separately from the development data. The validation dataset was labeled by different individuals

510(k) Summary: SurgiTwin
12

Page 17

K250290
Twinsight
Traditional 510(k) Premarket Notification
SurgiTwin

from the training dataset.

The results from testing confirm that the ML model demonstrates acceptable performance for its intended population.

C. The automatic landmark function was validated using a study to compare automatic landmarking performance to manual landmark placement by expert annotators. The system met predefined clinical acceptance criteria.

D. A validation study was conducted to compare metrics generated by SurgiTwin to those from manual annotations by expert annotators. The system met predefined clinical acceptance criteria for all measurements.

E. A validation study was performed to confirm the accuracy and clinical acceptability of the default implant placement algorithm. All acceptance criteria were met.

F. A validation study was performed to verify the clinical acceptability of the osteophyte removal function. All acceptance criteria were met.

G. External validation tests were performed by qualified personnel in an environment simulating the real end-user environment using a pre-defined test protocol. All acceptance criteria were met.

Nonclinical performance testing demonstrated that the subject device fulfills its intended use with an acceptable performance according to clinical acceptance criteria for all functions. These tests will be repeated and updated when appropriate to ensure that the software continues to meet performance criteria. Performance testing of SurgiTwin therefore supports substantial equivalence to its predicate.

8. Conclusion

Based on the information presented in this 510(k) submission, SurgiTwin has been determined to be substantially equivalent to the legally marketed predicate device in terms of indications for use, intended use, design, technology, and performance.

510(k) Summary: SurgiTwin
13

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).