K Number
K250270

Validate with FDA (Live)

Device Name
Luja Set
Manufacturer
Date Cleared
2025-04-17

(77 days)

Product Code
Regulation Number
876.5130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Luja Set is indicated for use by patients with urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain.

The product is indicated for male patients only (adults and pediatric above the age of 1 years).

Device Description

The medical device is an intermittent, ready-to-use, sterile, hydrophilic-coated catheter with integrated urine bag. It contains a catheter with a flexible tip that facilitates passage through the sphincter in the urethra. The catheter is shielded by a sleeve, which serves as protection from the user's hands during insertion. Several small holes (micro-holes) by the straight tip, create a draining zone, which allow the urine to flow from the bladder out through the catheter to a urine bag. The drainage end of the catheter is attached to a urine bag with a tear off section to open the bag for emptying. The catheter is sterilized by radiation. The Luja Set is available in sizes: CH10, CH12, CH14, and CH16.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the Luja Set indicates that it is a Special 510(k) Notification. This type of submission is used when a modification is made to a manufacturer's own legally marketed device (the predicate device) where the modification does not require a new intended use and the methods for evaluating the changes are well-established.

Crucially, the provided document describes the device itself (a urological catheter), its intended use, and its similarities/differences to a predicate device (Luja Coudé) and a reference device (SpeediCath Flex Set). It details the non-clinical performance testing performed (flow rate and kink stability), but it DOES NOT contain the actual acceptance criteria or the specific results proving those criteria were met.

The document states: "Bench performance testing was conducted to verify the proposed subject devices met the pre-determined acceptance criteria per specified requirements. Testing was performed on final, finished, and sterilized devices as described in the applicable submission sessions." However, the "applicable submission sessions" are not included in this document.

Therefore,Based on the provided document, I cannot fulfill most of the detailed requests regarding acceptance criteria and study particulars because the document does not contain the acceptance criteria or the specific performance data.

Here's what I can extract and state based only on the provided text, along with what cannot be determined:

Acceptance Criteria and Device Performance

  • Acceptance Criteria: Not explicitly stated in the provided document. The document only mentions that "the proposed subject devices met the pre-determined acceptance criteria per specified requirements."
  • Reported Device Performance: Not explicitly stated in the provided document. The document mentions that "Risk assessment was conducted, and two performance risks were identified: flow rate and kink stability. To support the flow rate and kink stability testing, the SpeediCath Flex Set (K222059) was used as a reference device." However, the actual performance data (e.g., specific flow rates achieved, quantitative kink stability metrics) are not provided.

Study Details (Based on available information and inferences from a 510(k) for a physical medical device)

The information provided is for a physical medical device (urological catheter), not an AI/Software as a Medical Device (SaMD). Therefore, many of the typical questions regarding AI/ML clinical validation studies (like multi-reader multi-case studies, expert adjudication, ground truth establishment for AI models, training set details) are not applicable to this type of device and submission.

Here's a breakdown based on the context of the device:

  1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not listed in the provided document.
    • Reported Device Performance: Not listed in the provided document. The document only states that testing was performed for "flow rate and kink stability."
  2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified.
    • Data Provenance: Not applicable in the context of a physical device's non-clinical bench testing. This typically refers to clinical data. The testing mentioned (flow rate, kink stability) is bench testing on the device itself.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This question is relevant for AI/ML or imaging-based devices where human expert readings establish ground truth. For a physical device like a catheter, "ground truth" is established through standardized physical measurements and engineering specifications, not expert consensus interpreting data.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a physical medical device, not an AI/SaMD. MRMC studies are not relevant here.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For this device, "ground truth" for non-clinical testing is based on engineering specifications, physical measurements, and adherence to international standards (e.g., ISO 20696:2018 for urethral catheters, ASTM F623-19 for Foley Catheters, as cited). It's not a diagnostic ground truth from patient data like pathology or expert consensus.
  8. The sample size for the training set

    • Not Applicable. This device does not involve a "training set" in the context of machine learning.
  9. How the ground truth for the training set was established

    • Not Applicable. This device does not involve a "training set" in the context of machine learning.

FDA 510(k) Clearance Letter - Luja Set

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 17, 2025

Coloplast Corp
Vallabha Tantry
Sr. Regulatory Affairs Specialist
1601 West River Road North
Minneapolis, Minnesota 55411

Re: K250270
Trade/Device Name: Luja Set
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: EZD
Dated: March 18, 2025
Received: March 18, 2025

Dear Vallabha Tantry:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250270 - Vallabha Tantry
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 3

K250270 - Vallabha Tantry
Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250270

Device Name
Luja Set

Indications for Use (Describe)

Luja Set is indicated for use by patients with urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain.

The product is indicated for male patients only (adults and pediatric above the age of 1 years).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Luja Set Special 510(k) notification

K250705
Page 1 of 6

SPECIAL 510(K) SUMMARY

Submitted by: Coloplast A/S
Holtedam1
3050 Humlebaek
Denmark

Contact Person: Vallabha Tantry
Sr. Regulatory Affairs Specialist
Coloplast Corp.
1601 West River Road North
Minneapolis MN 55411
Phone: +1 (612)-806-1798
Email: usvtan@coloplast.com

Date of Summary: April 16, 2025

Subject Device:

| Trade or Proprietary Name: | Luja Set |
| Item/Model Numbers: | 20071/ 20072/ 20074/ 20076 |
| Common Name: | Urological catheter and accessories |
| Regulation/Classification Name: | Gastroenterology and Urology |
| Regulation Number: | 21 CFR 876.5130 |
| Product Code: | EZD |

Page 6

K250705
Page 2 of 6

| Review Panel: | Gastroenterology/Urology |
| Predicate Device: | K241210, Luja Coudé (Sizes CH 8-CH18) |
| | The predicate device has not been subject of a design-related recall. |
| Reference Device: | K222059, SpeediCath Flex Set |
| | The reference device has not been subject to a design-related recall. |
| Device Description: | The medical device is an intermittent, ready-to-use, sterile, hydrophilic-coated catheter with integrated urine bag. It contains a catheter with a flexible tip that facilitates passage through the sphincter in the urethra. The catheter is shielded by a sleeve, which serves as protection from the user's hands during insertion. Several small holes (micro-holes) by the straight tip, create a draining zone, which allow the urine to flow from the bladder out through the catheter to a urine bag. The drainage end of the catheter is attached to a urine bag with a tear off section to open the bag for emptying. The catheter is sterilized by radiation. The Luja Set is available in sizes: CH10, CH12, CH14, and CH16. |
| Indications for Use: | Luja Set is indicated for use by patients with urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain. The product is indicated for male patients only (adults and pediatric above the age of 1 years). |

Page 7

K250705
Page 3 of 6

Technological Characteristics Comparison

The table below summarizes the technological characteristics of Luja Set as compared to the predicate device.

ParameterSubject devicePredicate device
Luja Set maleLuja Coudé
510(k) NumberK250270K241210
Regulation NameUrological catheter and accessoriesSame
Regulation Number21 CFR 876.5130Same
Product CodeEZDSame
ClassificationIISame
Prescription DeviceYesSame
Intended UseIntermittent catheterization through the urethra.Same
Indications for UseLuja Set is indicated for use by patients with urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain. The product is indicated for male patients only (adults and pediatric above the age of 1 year).Same

Page 8

K250705
Page 4 of 6

ParameterSubject devicePredicate device
Luja Set maleLuja Coudé
Condition of UseIntermittent use and single useSame
DrainageMicro holesSame
Device Categorization per ISO 10993Surface contacting device in contact with mucosal membrane for a prolonged duration of time (24 h ˂ t˂ 30 days)Same
SterilitySAL 10⁻⁶Same
Sterilization Methode-beamSame
Shelf Life2 yearsSame
Available SizesCH/ FR 10, CH/ FR 12, CH/ FR 14, CH/ FR 16CH/ FR 8, CH/ FR 10, CH/ FR 12, CH/ FR 14, CH/ FR 16, CH/ FR 18
Catheter MaterialsPolyurethaneSame
Hydrophilic CoatingPolyvinylpyrrolidone (PVP) basedSame
Swelling media (Wetting Agent)Saline solution with PEGSame
Tip ConfigurationStraight tipFlexible curved tip (bended)
Primary Packaging DescriptionSingle pouch package, dark greySingle and double-loop pouch packages, dark grey
PackagingInner layer: PE-peelSame

Page 9

K250705
Page 5 of 6

ParameterSubject devicePredicate device
Luja Set maleLuja Coudé
MaterialsOuter layer: Printed PETPSame
Effective Catheter LengthEffective length (according to ISO 20696:2018): 33cm (13 inches)Same
Inner ConnectorPolyurethane WhiteSame
Outer Connector2 shot components: 1st shot: Polypropylene 2nd shot: Thermoplastic elastomerSame
Handle2 shot components: 1st shot: Polypropylene 2nd shot: Thermoplastic elastomerSame
Cap2 shot components: 1st shot: Polypropylene 2nd shot: Thermoplastic elastomerPolypropylene
Bag ConnectorLLDPE - Linear Low Density PolyethyleneN/A
Collection Bag(LDPE) LDPE/CO-PEPP/LDPE – Low Density Polyethylene/ Polyethylene-Polypropylene copolymer/ Low Density PolyethyleneN/A

Page 10

K250705
Page 6 of 6

Summary of Non-Clinical Performance Testing

| Non-clinical test summary: | The subject device when compared to the predicate (K241210) device shares similar technological characteristics as summarized in the technological characteristics table. The subject device differs from the predicate device in the straight tip configuration and the urine bag. Testing conducted with the predicate and reference device was deemed applicable to the subject device. Risk assessment was conducted, and two performance risks were identified: flow rate and kink stability. To support the flow rate and kink stability testing, the SpeediCath Flex Set (K222059) was used as a reference device. |
| Catheter performance: | ISO 20696: 2018, Sterile urethral catheters for single use ASTM F623-19, Standard performance specification for Foley Catheter |

Bench performance testing was conducted to verify the proposed subject devices met the pre-determined acceptance criteria per specified requirements. Testing was performed on final, finished, and sterilized devices as described in the applicable submission sessions.

Conclusion

The performance testing demonstrates the subject device is as safe and effective as the predicate device.

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.