(98 days)
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
IPL Hair Removal Device (Model: KCA511, KCA516,KCA522), is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable treatment window and the spot size is 3.3cm² that is suitable for multiple hair removal areas. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.
IPL Hair Removal Device, model: KCA511, KCA516 and KCA522 have the same indication for use, performance, structure design and operation, the only deference is their appearance size and weight.
IPL Hair Removal is used for removing excess facial hair and body hair and can also reduce hair regrowth. The products can be used on body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs.
It is designed so that the flash window requires full contact with the skin to work (flash),it will no works as long as the flash window does not contact the skin.
This FDA 510(k) clearance letter and summary is for an IPL Hair Removal Device. It details the device's characteristics and its comparison to predicate devices to establish substantial equivalence. However, it does not contain the specific acceptance criteria or the study details proving the device meets those criteria, especially not in the format requested for AI/ML performance evaluations.
The document focuses on:
- Regulatory Classification: Class II, OHT product code, 21 CFR 878.4810 (Laser surgical instrument for use in general and plastic surgery and in dermatology).
- Intended Use: Removal of unwanted hair and permanent reduction in hair regrowth (measured at 6, 9, and 12 months post-treatment).
- Technological Comparison: Comparing the subject device (IPL Hair Removal Device, KCA511/KCA516/KCA522) to a predicate device (K240282) and a reference device (K230122) in terms of intended use, design, specifications, and performance.
- Performance Data: This section only covers non-clinical testing for:
- Biocompatibility: Tests conducted according to ISO 10993 series (cytotoxicity, irritation, skin sensitization).
- Electrical Safety and EMC: Compliance with various IEC 60601 standards.
- Eye Safety: Compliance with IEC 62471.
- Software Verification and Validation: Stating that software requirements were met and hazards mitigated.
Therefore, I cannot provide the requested information because it is not present in the provided text. The document does not describe:
- A table of acceptance criteria and reported device performance (in the context of AI/ML evaluation).
- Sample size, data provenance, number of experts, qualifications, or adjudication methods for a test set. This type of information is typically associated with clinical studies or performance evaluations where a "ground truth" is established, which is entirely absent here.
- Whether a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done or the effect size of human readers with/without AI assistance. This is relevant for AI-assisted diagnostic or decision-making devices, not a direct energy hair removal device.
- Whether a standalone (algorithm only) performance was done. Again, pertaining to AI/ML devices.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.). This is not applicable given the type of device and the information provided.
- The sample size for the training set or how its ground truth was established. This refers to AI/ML model development, which is not described.
In summary, the provided FDA 510(k) clearance letter and summary pertains to a physical IPL hair removal device, not a software or AI/ML-driven diagnostic or decision-support system. As such, the requested details regarding acceptance criteria, study methodologies, and ground truth establishment, which are typical for AI/ML performance evaluation, are not present in this document.
FDA 510(k) Clearance Letter - IPL Hair Removal Device
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov
April 6, 2025
Dongguan Boyuan Intelligent Technology Co.,Ltd
℅ Reanny Wang
General Manager
Shenzhen Reanny Medical Devices Management Consulting Co.Ltd
Room 1509, Jingting Building, Dongzhou Community,
Guangming Street, Guangming
Shenzhen, Guangdong 818000
China
Re: K250253
Trade/Device Name: IPL Hair Removal Device (KCA511/KCA516/KCA522)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: January 26, 2025
Received: January 28, 2025
Dear Reanny Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
2K250253 - Reanny Wang Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
3K250253 - Reanny Wang Page
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2025.05.06 17:23:28 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K250253
Device Name: IPL Hair Removal Device (KCA511/KCA516/KCA522)
Indications for Use (Describe):
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable):
- ☐ Prescription Use (Part 21 CFR 801 Subpart D)
- ☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
K250253 - 510(k) Summary
I. SUBMITTER
Dongguan Boyuan Intelligent Technology Co.,Ltd.
Room 801, Building 3, No.3 Kuiqiao Road, Puxin Lake , Tangxia Town ,Dongguan ,
Guangdong, China.
Phone: 86-13422889315
Fax: 978953150@qq.com
Contact Person: Li Jinlong
Date Prepared: Sep 29, 2024
II. DEVICE
Name of Device: IPL Hair Removal Device
Model(s): KCA511,KCA516,KCA522.
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810
III. PREDICATE DEVICE
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
|---|---|---|---|
| Dongguan Boyuan Intelligent Technology Co.,Ltd. | IPL Hair Removal Device | K240282 | April 3, 2024 |
This predicate has not been subject to a design-related recall.
Reference device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
|---|---|---|---|
| Shenzhen Ulike Smart Electronics Co.,Ltd. | IPL Hair Removal Device | K230122 | April 10, 2023 |
Page 6
This predicate has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
IPL Hair Removal Device (Model: KCA511, KCA516,KCA522), is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable treatment window and the spot size is 3.3cm² that is suitable for multiple hair removal areas. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.
IPL Hair Removal Device, model: KCA511, KCA516 and KCA522 have the same indication for use, performance, structure design and operation, the only deference is their appearance size and weight.
IPL Hair Removal is used for removing excess facial hair and body hair and can also reduce hair regrowth. The products can be used on body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs.
It is designed so that the flash window requires full contact with the skin to work (flash),it will no works as long as the flash window does not contact the skin.
V. Indications for Use
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
VI. Comparison of Technological Characteristics with the Predicate Device
The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device and reference device. Any minor differences between the subject device and the listed predicate device and reference
Page 7
device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference device for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference device. IPL Hair Removal Device is compared with the following Predicate Device and Reference Device in terms of intended use, design, specifications and performance:
| Comparison Elements | Subject device | Predicate device | Reference device | Remark |
|---|---|---|---|---|
| 510(k) Number | Pending | K240282 | K230122 | / |
| Trade name | IPL Hair Removal Device | IPL Hair Removal Device | IPL Hair Removal Device | / |
| Manufacturer | Dongguan Boyuan Intelligent Technology Co.,Ltd. | Dongguan Boyuan Intelligent Technology Co.,Ltd. | Shenzhen Ulike Smart Electronics Co.,Ltd. | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHT | OHT | OHT | Same |
| Device classification | Class II | Class II | Class II | Same |
| Indication for use/ Intended use | IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when | IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when | IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing | Same |
Page 8
| measured at 6, 9 and 12 months after the completion of a treatment regime. | measured at 6, 9 and 12 months after the completion of a treatment regime. | when measured at 6, 9 and 12 months after the completion of a treatment regime. |
|---|---|---|
| Prescription or OTC | OTC | OTC |
| Applicable skin | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V |
| Treatment area | Body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs. | Body parts such as underarm, bikini line, arms, facial hair below the chin line, back and legs. |
Device Design
| Source energy | Supplied by Power Adapter | Supplied by Power Adapter | Supplied by external adapter |
|---|---|---|---|
| Power source | External power supply1)Model:FX65U-24030CInput:100-240V AC 50/60Hz 2AOutput:24.0V DC 3.0A2)Model:DZ072CHL240300UInput:100-240V AC 50/60Hz 1.5AOutput:24.0V DC 3.0A | External power supply;KCA450, KCA446, KCA504, KCA505, KCA506, KCA507, KCA508, KCA509:1) Model: FX48U-120400CInput: 100-240V AC; 50/60Hz; 1.0A maxOutput: 12,0 V d.c.; 4,0 A2) Model: DZ048CHL120400UInput: 100-240V AC; 50/60Hz; 1.5A max | External power supply:100-240 VAC, 50/60Hz |
Page 9
| Output: 12,0 V d.c.; 4,0 AAn external power supply;KCA501, KCA423, KCA448, KCA449, KCA502, KCA521:1) Model: FX48U-120300CInput: 100-240V AC; 50/60Hz; 1.0A maxOutput: 12,0 V d.c.; 3,0 A2) Model: DZ048CHL120300UInput: 100-240V AC; 50/60Hz; 1.5A maxOutput: 12,0 V d.c.; 3,0 A |
|---|
| Sterilization | Not required | Not required | Not required | Same |
|---|---|---|---|---|
| Dimensions | KCA511:178.2 61.9 36mmKCA516: 177.365.437.4mmKCA522: 17770.243.1mm | KCA450: 170.8 *80.4 *41.3mmKCA446: 170.0 *80.5 *41.3mmKCA504: 172.1 *69.0 *37.8mmKCA505: 172.1 *69.0 *37.8mmKCA506: 171.0 *74.8 *41.9mmKCA507: 171.5 *74.5 *37.3mm | 60mm x 38mm x 170mm | Different |
Page 10
| KCA508: 171.0 74.8 41.9mmKCA509: 171.5 74.5 37.3mmKCA501: 163.579.142.8mmKCA423: 164.873.741.1mmKCA448: 162.076.442.2mmKCA449: 162.278.042.7mmKCA502: 164.275.838.3mmKCA521: 162.580.839.1mm |
|---|
Output Specification
| Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
|---|---|---|---|---|
| Wavelength range | (610nm-1200nm) ±15nm | (600nm~1200nm) ±15nm | 560-1200nm | Similar |
| Energy density | 2.18~7.27J/cm² | 1.6~6J/cm2 | 2.4~7.2J/cm2 | Similar |
| Output energy | (9J~20J) ± 20% | (7.11J~17.41J) ± 20% | 9.9~19.8J | Similar |
| Spot size | 3.3cm² | 3.5 cm² | 3.3cm² | Same |
| Pulse duration | 0.45ms±0.2ms | (0.67-4.5)ms ±2ms 8~12ms | 1.15-6.2ms | Different |
| Output intensity level | 5 | 9 | 5 | Same |
| Software/ Firmware/ Microprocess or Control? | Software control | Software control | Software control | Same |
Page 11
Additional Features
| Electrical safety | ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83 | ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83 | ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83 | Same |
|---|---|---|---|---|
| Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | Same |
| Biocompatibility | ISO 10993-5ISO 10993-10ISO 10993-23 | ISO 10993-5ISO 10993-10ISO 10993-23 | ISO 10993-5ISO 10993-10 | Same |
VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
1) Biocompatibility testing
The biocompatibility evaluation for the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices − Part 1: Evaluation and Testing Within a Risk Management Process' " as recognized by FDA. The battery of testing included the following tests:
- ISO 10993-5:2009, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2010, Biological Evaluation of Medical Devices –Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices --Part 23: Tests for irritation
The IPL Hair Removal Device is considered tissue contacting for a duration of less than 24 hours.
2) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, as per the following standards:
Page 12
- IEC 60601-1:2005/AMD1:2012/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-1-11:2015/AMD1:2020 Medical Electrical Equipment –Part 1-11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- IEC 60601-2-83:2019/AMD1:2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
3) Eye Safety
- IEC 62471:2006 Photobiological safety of lamps and lamp systems
4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
Summary
Based on the above performance as documented in this application, IPL Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device and reference device.
VIII. Conclusions
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.