K Number
K250219
Date Cleared
2025-06-17

(144 days)

Product Code
Regulation Number
870.1250
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dorado™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable and self expanding stents in the peripheral vasculature. This catheter is not for use in coronary arteries.

Device Description

The Dorado™ PTA Balloon Dilatation Catheter is a high-performance balloon catheter consisting of an over the wire catheter with a balloon fixed at the distal tip. The proprietary non-compliant, low-profile balloon is designed to provide consistent balloon diameters and lengths even at high pressures. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the stenosis. The novel catheter consists of a distal triple lumen and a proximal coaxial lumen and is designed to optimize the balance between pushabilty and trackability. The over the wire catheter is compatible with 0.035" guidewire and is available in 40, 80, 120, and 135 cm working lengths. The proximal portion of the catheter includes a female luer lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen. Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A re-wrapping tool is also provided on the catheter shaft. A stylet is placed into the tip of the catheter to aid in rewrap/refolding of the balloon. These products are not made with natural rubber latex.

AI/ML Overview

This 510(k) clearance letter is for a medical device (Dorado™ PTA Balloon Dilatation Catheter), not an AI/Software as a Medical Device (SaMD).

Therefore, the information requested in your prompt (e.g., acceptance criteria for AI algorithm performance, sample sizes for test/training sets, expert adjudication, MRMC studies, ground truth establishment) is not applicable to this document.

The document discusses the performance evaluation of a physical medical device, focusing on in vitro pre-clinical testing to demonstrate substantial equivalence to a predicate device. This includes:

  • Performance Data: Trackability, Balloon Burst Strength, Balloon Fatigue, Sheath Compatibility, Dimensional Verification, Simulated Use, etc.
  • Biocompatibility Testing: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility.

The acceptance criteria for this device would be engineering specifications and safety standards for balloon catheters, not statistical metrics for AI algorithm performance.

In summary, based on the provided document, I cannot answer the questions related to AI/SaMD acceptance criteria and study design. The document describes the clearance of a traditional physical medical device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).