K Number
K250189
Device Name
AdvaTx
Manufacturer
Date Cleared
2025-04-01

(69 days)

Product Code
Regulation Number
878.4810
Panel
SU
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVATx at 589 nm is indicated for use for dermatologic treatment of benign cutaneous vascular lesions including but not limited to: treatment of wrinkles, periocular wrinkles, periorbital wrinkles, facial and leg telangiectasia, rosacea, cherry angiomas, port wine stains, hemangiomas and venous lakes, angioma, spider angioma, Poikiloderma of Civatte, inflammatory acne vulgaris, verrucae/warts, scars, striae, psoriasis and the treatment of benign pigmented lesions.

The ADVATx at 1319 nm is indicated for treatment of fine lines and wrinkles. Treatment of atrophic acne scars. Treatment of mild and moderate inflammatory acne vulgaris.

Device Description

The AdvaTx is non-ablative laser with two wavelengths: 1319nm and 589nm. There are multiple indications for use that are specific to each wavelength. The AdvaTx consists of a console, a handpiece that has a scanner and a distance guide. The laser is activated using a trigger that is located on the handpiece.

The AdvaTx produces light pulses with a specific duration, intensity, and spectral distribution allowing for a controlled and confined energy delivery into tissue. Laser use in dermatology relies on the basis that certain targets for energy absorption (chromophores) are capable of absorbing energy from the specific light wavelength (absorptive band) without exclusively being targeted by their highest absorption peak. The working basis of the laser rests on the principle of selective photothermolysis, in which thermally mediated radiation damage is confined to chosen epidermal and/or dermal pigmented targets at the cellular or tissue structural levels.

AI/ML Overview

The provided text is a 510(k) Clearance Letter for the AdvaTx Laser System. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a clinical study with performance metrics in the way a diagnostic AI device would.

Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" are not applicable to this type of regulatory submission for a therapeutic laser system. There is no information provided regarding:

  • A table of acceptance criteria and reported device performance (in terms of specific clinical outcomes or AI metrics).
  • Sample size used for a test set, data provenance, number of experts, adjudication methods, MRMC studies, or standalone algorithm performance.
  • Ground truth used or how it was established for training or test sets.

The relevant information from the document is summarized below based on the nature of the device and the submission.


Summary of AdvaTx Laser System Regulatory Submission (K250189)

The AdvaTx Laser System (K250189) received 510(k) clearance based on demonstrating substantial equivalence to predicate and reference devices, rather than a clinical study establishing performance against pre-defined acceptance criteria for a new and unique clinical outcome. The clearance specifically addresses the addition of an indication for "treatment of benign pigmented lesions" at 589nm, and an increased fluence (up to 60 J/cm2) for the 1319nm wavelength.

1. Table of Acceptance Criteria and Reported Device Performance

As this is a therapeutic laser device seeking 510(k) clearance by demonstrating substantial equivalence, specific "acceptance criteria" in terms of clinical performance metrics, as would be seen for a diagnostic device with an AI component, are not provided in the document. Instead, the "performance" demonstrated is through engineering and safety testing, and comparison of technical characteristics to legally marketed predicate devices.

Characteristic / RequirementAcceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (AdvaTx K250189)
Indications for Use (589nm)Substantially equivalent to predicate/reference devices for dermatologic treatment of benign cutaneous vascular lesions. Addition of "benign pigmented lesions" must be supported.Met: Indicated for: wrinkles, periocular/periorbital wrinkles, facial/leg telangiectasia, rosacea, cherry angiomas, port wine stains, hemangiomas, venous lakes, angioma, spider angioma, Poikiloderma of Civatte, inflammatory acne vulgaris, verrucae/warts, scars, striae, psoriasis, AND treatment of benign pigmented lesions.
Indications for Use (1319nm)Substantially equivalent to predicate/reference devices.Met: Indicated for: treatment of fine lines and wrinkles, atrophic acne scars, mild and moderate inflammatory acne vulgaris.
WavelengthsMatch or be substantially equivalent to predicate/reference.Met: 589nm and 1319nm. (Compared to 589nm predicate, 595nm reference, 1320nm reference, stating 589/595nm and 1319/1320nm are substantially equivalent).
Spot SizeSafe and effective, potentially smaller than predicate.Met: 1.0 mm (Subject device spot size is smaller than predicates, but deemed safe with no additional risks).
Pulse Width (589nm)Match or be within acceptable range of predicate/reference.Met: 20-950 ms
Pulse Width (1319nm)Match or be within acceptable range of predicate/reference.Met: 20-200 ms
Fluence (589nm)Match or be within acceptable range of predicate/reference.Met: Up to 100 J/cm2
Fluence (1319nm)Increased fluence must be safe and effective, and within range of reference devices.Met: Up to 60 J/cm2 (This is an increase from the predicate's 40 J/cm2 but is the same as a reference device, implying safety/efficacy is established by the reference).
Repetition RateMatch or be within acceptable range of predicate/reference.Met: 0.67 - 1.0 Hz
Electrical Safety (IEC 60601-1 Ed 3.2)Compliance with standard.Met: Test performed and deemed compliant.
EMC (IEC 60601-1-2:2014)Compliance with standard.Met: Test performed and deemed compliant.
Laser Specific Safety (IEC 60601-2-22:2012)Compliance with standard.Met: Test performed and deemed compliant.
Biocompatibility (ISO 10993-1:2018)Compliance with standard.Met: Biological Evaluation performed.
Software V&VDocumentation provided as recommended by FDA guidance.Met: Verification and validation testing conducted and documentation provided.
Safety and Effectiveness ConclusionSafe and effective as the predicate device, with no new concerns from changes.Met: Concluded AdvaTx is "as safe and effective as the predicate device" and that "The addition of the higher fluence does not present any new concerns for safety or efficacy."

2. Sample Size Used for the Test Set and the Data Provenance

This information is not applicable/not provided in the context of this 510(k) submission for a therapeutic laser system. The submission relies on technical comparisons and established safety standards. There is no mention of a "test set" in the sense of a dataset for validating an algorithm.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable/not provided. There is no "test set" for which ground truth would be established by experts.

4. Adjudication Method for the Test Set

This information is not applicable/not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Based on the similarities in the safety and effectiveness profiles of the subject and the predicate, no clinical studies were deemed needed to support this submission."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This information is not applicable/not provided. The AdvaTx is a laser system, not a software algorithm.

7. The Type of Ground Truth Used

This information is not applicable/not provided. Clinical performance for specific indications is demonstrated implicitly by the substantial equivalence to predicate devices, which would have had their own evidence for clearance. The safety and technical performance are demonstrated via engineering standards and comparisons.

8. The Sample Size for the Training Set

This information is not applicable/not provided. The AdvaTx is a physical laser device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.