(209 days)
The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0)., which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
The Evolution Spine Navigation Instruments are reusable surgical instruments for use with the Medtronic® StealthStation® S8 Navigation System. The system is designed to assist surgeons in the precise localization of anatomical structures, preparation and placement of pedicle screw implants during spinal procedures.
The Evolution Spine navigation Instruments include awls, probes, taps and drivers. The navigated instruments are to be used with Whistler Modular Pedicle Screw System (K182478) and Stowe Pedicle Screw System (K181554).
All instruments are made of stainless steel per ASTM F899. The Evolution Spine navigation Instruments are not compatible with implants from other manufacturers and are designed for use only with Medtronic® StealthStation® S8 Navigation System hardware and software.
The provided FDA 510(k) clearance letter for the Evolution Spine Navigation Instruments focuses on establishing substantial equivalence to predicate devices, primarily through non-clinical performance testing. It indicates that clinical testing was not performed. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted to assess human reader improvement with AI assistance, nor was a standalone algorithm performance study performed as part of this submission for establishing substantial equivalence.
The submission relies on a non-clinical test to demonstrate performance, specifically focusing on positional accuracy of the instruments when used with a specific navigation system.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (from ASTM F2554-18 and Rampimura et al.) | Reported Device Performance |
|---|---|---|
| Positional Accuracy | Meets performance criteria identified in ASTM F2554-18, "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems" | The worst-case construct meets and exceeds all the performance criteria identified in Rampimura et al. and ASTM F2554-18. |
| Positional Accuracy | Meets performance criteria identified in Rampimura et al: "Accurate pedicle screw insertion under the control of a computer-assisted guiding system: Laboratory test and clinical study" | The worst-case construct meets and exceeds all the performance criteria identified in Rampimura et al. and ASTM F2554-18. |
Note: The exact numerical values for positional accuracy criteria from ASTM F2554-18 and Rampimura et al. are not explicitly stated in the provided text. The document only confirms that the device "meets and exceeds" these established benchmarks.
Study Details
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not specify a numerical "sample size" in terms of cases or patients. Instead, it refers to "the worst-case instruments constructs" being tested. This implies a limited number of physical instrument configurations were chosen for rigorous testing, representing the most challenging scenarios for accuracy.
- Data Provenance: The study described is a non-clinical, laboratory-based performance test. This means the "data" is derived from controlled measurements in a lab setting, not from human subjects or clinical data. Therefore, concepts like "country of origin" or "retrospective/prospective" do not apply in the typical sense; it's an engineering test.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:
- This is a non-clinical performance test of a physical device's accuracy. "Ground truth" in this context refers to the engineering specifications and established test methods (ASTM F2554-18 and Rampimura et al. reference values) for positional accuracy. No human experts (e.g., radiologists) were used to establish the ground truth for this performance test. The "truth" is an objective, measurable accuracy.
-
Adjudication Method for the Test Set:
- Not applicable as this was a non-clinical instrument performance test, not a subjective assessment requiring human adjudication. The results are quantitative measurements against established engineering standards.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No. The document explicitly states: "Clinical testing was not necessary to demonstrate the substantial equivalence of the subject devices." This implies that an MRMC study involving human readers and AI assistance was not performed or submitted for this specific clearance.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No. The device is a set of "navigation instruments" used with a navigation system (Medtronic® StealthStation® S8). It's a physical instrument, not an AI algorithm. Its performance is evaluated in conjunction with the navigation system, but the focus is on the instrument's mechanical and tracking accuracy.
-
The Type of Ground Truth Used:
- Engineering Specifications / Established Test Standards: The ground truth for this non-clinical study was defined by the performance criteria established in:
- ASTM F2554-18, "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems"
- Rampimura et al: "Accurate pedicle screw insertion under the control of a computer-assisted guiding system: Laboratory test and clinical study"
- These are recognized industry standards and published research that define acceptable accuracy limits for such devices.
- Engineering Specifications / Established Test Standards: The ground truth for this non-clinical study was defined by the performance criteria established in:
-
The Sample Size for the Training Set:
- Not applicable. This is a physical medical device (surgical instruments), not an AI/machine learning algorithm that requires a "training set" of data.
-
How the Ground Truth for the Training Set was Established:
- Not applicable, as there is no "training set" for this type of device.
FDA 510(k) Clearance Letter - Evolution Spine Navigation Instruments
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov
August 18, 2025
Evolution Spine
James Mirda
Principal Engineer
2300 N. Haskell Ave
Dallas, Texas 75204
Re: K250167
Trade/Device Name: Evolution Spine Navigation Instruments
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: July 21, 2025
Received: July 21, 2025
Dear James Mirda:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K250167 - James Mirda Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
Page 3
K250167 - James Mirda Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K250167
Device Name
Evolution Spine Navigation Instruments
Indications for Use (Describe)
The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0)., which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) Summary
510(k)#: K250167 Prepared on: 2025-08-13
Contact Details
21 CFR 807.92(a)(1)
| Field | Value |
|---|---|
| Applicant Name | Evolution Spine |
| Applicant Address | 2300 N. Haskell Ave Dallas TX 75204 United States |
| Applicant Contact Telephone | 901-550-3377 |
| Applicant Contact | Mr. James Mirda |
| Applicant Contact Email | jmirda@evolutionspine.com |
| Correspondent Name | Evolution Spine |
| Correspondent Address | 2300 N. Haskell Ave Dallas TX 75204 United States |
| Correspondent Contact Telephone | 9015503377 |
| Correspondent Contact | Mr. James Mirda |
| Correspondent Contact Email | jmirda@evolutionspine.com |
Device Name
21 CFR 807.92(a)(2)
| Field | Value |
|---|---|
| Device Trade Name | Evolution Spine Navigation Instruments |
| Common Name | Stereotaxic instrument |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Regulation Number | 882.4560 |
| Product Code(s) | OLO |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|---|---|---|
| K223649 | VERTICALE® Navigation Instruments | OLO |
| K16120 | Medtronic Navigated Instruments | HWE |
| K140454 | Navigated CD HORIZON® SOLERA® Screwdrivers and Taps | OLO |
Device Description Summary
21 CFR 807.92(a)(4)
The Evolution Spine Navigation Instruments are reusable surgical instruments for use with the Medtronic® StealthStation® S8 Navigation System. The system is designed to assist surgeons in the precise localization of anatomical structures, preparation and placement of pedicle screw implants during spinal procedures.
The Evolution Spine navigation Instruments include awls, probes, taps and drivers. The navigated instruments are to be used with Whistler Modular Pedicle Screw System (K182478) and Stowe Pedicle Screw System (K181554).
All instruments are made of stainless steel per ASTM F899. The Evolution Spine navigation Instruments are not compatible with implants
Page 1 of 2
Page 6
from other manufacturers and are designed for use only with Medtronic® StealthStation® S8 Navigation System hardware and software.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0)., which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject "Indications for Use" statement is nearly identical to those of the predicate devices with only minor changes that do not impact the indicated use of the devices.
Technological Comparison
21 CFR 807.92(a)(6)
Evolution Spine Navigation Instruments is substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have substantially equivalent technological characteristics to the predicate devices through comparison of design, intended use, material composition, function, and range of sizes.
Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The worst-case instruments constructs were tested per ASTM F2554-18, "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems". The worst-case construct meets and exceeds all the performance criteria identified in Rampimura et al: Accurate pedicle screw insertion under the control of a computer-assisted guiding system: Laboratory test and clinical study and ASTM F2554-18. Based upon the test data presented and comparison of features, Evolution Spine has determined that the proposed System is substantially equivalent to previously cleared systems.
Clinical testing was not necessary to demonstrate the substantial equivalence of the subject devices.
A comparison of critical dimensions was conducted to support the substantial equivalence of the subject devices.
Biocompatibility was addressed through use of identical manufacturing materials and processes to the devices cleared in K182478 and K181554.
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject Evolution Spine Navigation Instruments has been shown to be substantially equivalent to legally marketed predicate devices.
Page 2 of 2
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).